YR1030 User's Manual (Trial version)

Table of Contents
Outline
Safety Tips
Safe work practices
Instruments Overview
Terminal Wiring
Display Panel Symbols
Power Management
Battery Saver
Battery Charging
Operation and Menus
Button operation
Measuring
Auto range and Manual Range
Data Retention
Resistance gear cleared
Voltage profile cleared
Measuring battery
Resistance profile base number
Sorting measure
Parameter settings
Maintenance
General Maintenance
Manual calibration
Service and Parts
General Specifications
Accuracy specifications
Caution
To avoid electrical shock or personal injury, and cause damage to the instrument, please read "Safety
Information" and "Warnings and Precautions" before use.
YR1030 is a 2,000 counts instrument while displaying resistance and voltage values.
Instrument rechargeable battery users a built in Li-Ion Battery for the power supply, and a DOT matrix
screen display.

Safety Tips
In internal resistance and terminal voltage measurement range does not exceed the maximum input
voltage of 30VDC. Never use this to measure an AC voltage signals, doing this will damage the unit
and could cause a fire. This is for DC voltage only.
Charging is done with the Micro USB charging port. Standard 5VDC USB port with 500mA is fine.
The internal charging circuit is selectable between 200mA or 400mA charging speeds.
Select the 200mA charging speed if the USB port cannot provide 40mA or higher for charging.
Instruments Overview

Terminals Wiring
name Explanation
ui1 Signal input terminal
ui2 Signal input terminal
com Common ground
io Constant current output terminal

Display Panel Symbols


1. Normal mode

Numbering Explanation
Voltage value is displayed
The values shown
Gear indicator
Automatic range state
Clear enabled
Lock status
Power Battery voltage
Charging input voltage
2. Battery sorting Display Mode

Numbering Explanation
14. A number of batteries that have been tested
13. Class B Number of batteries
12: Class C Number of batteries
C fraction of the selected bit, when the measured cell belongs to the present level, the highlighted
symbol cursor appears
Fraction B of the selected bit, when the measured cell belongs to the present level, the highlighted
symbol appears cursor
A fraction of the selected bit, when the measured cell belongs to the present level, the highlighted
symbol appears cursor.
Power Management
Battery Saver
Menu 4 - Auto Off options:
This can be set on or off for automatic shutdown. You can select the shutdown between 5 minutes to 60
minutes. The meter with automatically shut once it reaches the set time if now measurements have been
taken in the defined time period.

Low power mode (Battery Saver)


These settings can be turned on or off. This is selectable between 5 minutes to 30 minutes.
Automatically shut down before reaching the set time when it detects no battery measurement or
buttons pressed. The meter will automatically enter a low power mode. It displays "-" symbols when its
in low power mode. Press any key on the panel to exit the low power mode.

Battery Charging:
Instrument using MicroUSB charging port interface, an input voltage of 4.5V-5.5V and 500mA. The
battery charging current can be set to 200mAh or 400mAh.
Operation and Menus
POWER set / cancel key (hereinafter referred to as P key):
Click switched off state.
Normal mode/sorting mode: Press enter shutdown interface P, and H bonds by complete shutdown RR
auxiliary keys.
Normal mode and sort mode, click on the P button to enter Menus interface.
Menus next screen, click on the P button to enter the selected Menus option.
Menus option, click on the P button to cancel the setting return Menus interface.
HOLD / ZERO R Save / Enter key (hereinafter referred to as key HZ):
Shutdown screen, click HZ confirm the shutdown.
Normal mode and sort mode, click on the button to display the HZ lock, click Unlock again.
Normal mode and sorting mode, long press zero HZ to enable operation button display, press and
released again.
Menus next screen, click on the HZ key to exit the Menus screen.
Menus option, click the Save Settings button HZ and return to Menus interface.
Parameter settings, click HZ RR and RU toggle key function.
Move RANGE R / addend key (hereinafter referred to as Construction RR):
Shutdown screen, click RR key to cancel the shutdown.
Normal mode and sorting mode, click resistance RR handover speed manual and auto ranging.
Under Menus screen, move Menus option on RR button click. Press and hold more than 3S linked.
Menus option, select the shift key parameters click RR, or change the parameter value. Press and hold
more than 3S linked.
RANGE U Down / subtraction key (hereinafter referred to as the RU key):
Normal mode and sorting mode, click on the button to switch voltage RU-speed manual & auto
ranging.
Menus next screen, click on the button RU Down Menus option. Press and hold more than 3S linked.
Menus option, click the down key to select RU parameters, or change the parameter value. Press and
hold more than 3S linked.
Measuring
Automatic and Manual ranging
Voltage function and resistance meter, respectively, two manual and automatic range selection. When
auto range state, the instrument can automatically select the correct range. Or you can select manual
ranges. The meter defaults to AUTO when you first turn it on. When the measurement function is auto
ranging AUTO: after displaying the function symbols R or U.
Enter and exit the manual range:
In normal mode or sorting mode, press (RANGE R) or (RANGE U) enter the manual range. Or U-R
measurements, enter the manual range minimum range. Continue pressing to increase the range.
Press (RANGE R) or (RANGE U) when the maximum range again, back to the automatic range.
Data Retention
To keep the displayed current reading, press HOLD. Press HOLD again to return to normal display.

Taking a Battery Measurement


Normal mode
Connect the test leads and turn it on. Do not connect the meter to any battery with a voltage over 28V
The two leads (two lines per probe) are in contact with the positive and negative battery measured.
Stable value to be displayed is read out

Resistance Cleared
Red and black wire test line twisted pair line, a white line and a white line can be twisted to enhance
immunity, improve the measurement accuracy. When measuring a very low resistance or interference
will cause a little deviation measurements, then use the ZR (resistance profile clear) function.
Measuring the resistance has extremely low resistance profile interfering signal introduced through the
test line, the HOLD button is enabled by pressing the function is cleared, an internal storage by the zero
point value, the low end of the range measurement using improved accuracy.
Side Note:
1. After clearing the test line placement is subject to change, etc., may be a little change in the
measured value.
2.Enabling ZR function meter with a steady increase in the time required than the usual wait for a few
seconds.
3.Resistance measurements, voltage measurements are inaccurate value leads or floating interfaces, no
reference value.
4.Each test line test leads have good contact with the battery to be tested, and can not have a connection
between each test line.
Sorting Measurement Feature
Sorting capabilities help improve battery efficiency rating. By setting Ra, Ua, Rb, Ub, Rc, Uc
parameters, and sorted into the interface, the measured battery parameters, to classify the battery.
Battery condition is classified as follows, Rx <Ra and Ux> Ua on the A-class battery, so obtained class
B and C cells. So before the battery is fully charged battery Make sure the rating, or the same power,
and standing over a period of time (1h or more). Classification parameters, please refer to
"Preferences."
Battery sorting mode
Connect the test leads and turn it on. Confirmed that the instrument does not exceed the measured
battery voltage limit 28V.
The two leads (two lines per pen and the probe) are in contact with the positive and negative battery
measured.
Stable value to be displayed is read out.
Sorting parameters set in line with the battery, will count is incremented at the appropriate level.

Parameter settings
Click the Set button to enter meter Menus mode:
Normal mode: Set to normal mode
Sorting mode: Set the sorting mode parameters
Backlight setting: a backlight parameter setting, brightness 10% -99%, triggers a shutdown Steady
delay 5S-60S.
Automatic shutdown: automatic shutdown function setting, open and close, the delay 5min-60min.
Low-power setting, open and close, the delay 5min-30min.
Set sub-file: sorting mode A, B, C and the resistance profile of the battery voltage condition value.
A battery needs to comply with Class <Ua Rxa (set value)> Uxa (set value) Ra of the
The battery needs to comply with Class B <Ub Rxb (set value)> Uxb (set value) Rb
The battery needs to comply with Class C <Uc Rxc (set value)> Uxc (set value) Rc
Counting sub-file: sorting mode A, B, C the number of file cell, and zero setting
Calibration:
Resistance calibration file, set to zero the voltage profile
It may be individually calibrated for each gear position.
Factory settings: Restore factory calibration settings
Charge setting: built-in battery charging current is provided, the charging current 200mA 400mA
Maintenance
General Maintenance
1. Avoid hot, humid environments. This can cause interface circuit to prevent moisture and oxidation.
2. Avoid direct strong sunlight for a long time the LCD screen to prevent screen damage over time.
3. Keep the battery at around 40% power (or battery voltage between 3.7V-3.9V) to extend the life.

External test line:


1. Always check test lines for each signal line off the situation, to avoid breakage caused by
measurement errors.
2. Avoid hot and humid environments to prevent oxidation and moisture probe on the probes.

Manual calibration
Enter Menus 7- Calibration interface

Resistance calibration
Resistance Cal Range Resistor connected to the corresponding range
20mΩ file 10mΩ - 20mΩ
200mΩ file 100mΩ - 200mΩ
2Ω file 1Ω - 2Ω
20Ω file 10Ω - 20Ω
200Ω file 100Ω - 200Ω
Select the range, and then press the key or RU RR key to adjust the measured values of the display to
the nearest value of the resistor connected. Save the settings and exit.
Note: The connection resistance does not calibrated, the display is a random value.

Zero voltage settings


When the test leads are shorted and the voltage display value cannot return to zero, zero adjustment
needs to be done. Short the test leads gear is selected, then display the selected voltage value, press the
zero key RR. Save and exit.
Service and Parts
If the instrument fails, first check the battery voltage and probe lines. Review this manual for proper
use of the instrument. Still having issues, please contact the seller for additional support.

General Specifications
Maximum voltage between any terminal: 30VDC
Resolution: 2,000 count
Update Rate: 4 samples per second
Working Temperature: 0 °C -40 °C
Storage Temperature: -20 °C -60 °C
Temperature Coefficient: 0.1 * (a predetermined accuracy) /°C (<18 Or °C> 28°C)
Battery: 3.7V nominal 900mAh Li-ion
Power consumption: <60mA at 3.7Vin the 200mΩ range and above *(backlight off)
<120mA at 3.7Vin the 20mΩ to 200mΩ range *(backlight off)
<10mA when it’s in low power mode
0mA (shutdown)
Accuracy Standards
Precision format specifications: ± ([percentage degree] + [least significant digit])
DC voltage Range Resolution Accuracy Display Mode
2V 1mV 0.8% + 51.999V
20V 10mV 0.8% + 519.99V
28V 0.1V 0.8% + 5 29.9V
Internal resistance measurement (1) Resolution Accuracy Display Mode
20mΩ0.01mΩ0.7% + 722.00mΩ (ZR function is enabled)
200mΩ0.1mΩ 0.5% + 5220.0mΩ
2Ω1mΩ 0.5% + 52.200Ω
20Ω10mΩ 0.5% + 522.00Ω
200Ω0.1Ω 0.6% + 5220.0Ω
(1)In the range of 5% -95% range, it is not in this range (less than 5% of the measuring range gear, gear range or
the range of greater than 95%) error may be greater than the accuracy.