Peak airway pressure is measured at the airway opening (Pao) and is routinely
displayed by mechanical ventilators. It represents the total pressure needed to
push a volume of gas into the lung and is composed of pressures resulting from
inspiratory flow resistance (resistive pressure), the elastic recoil of the lung and
chest wall (elastic pressure), and the alveolar pressure present at the beginning of
the breath (positive end-expiratory pressure [PEEP]—see also Fig. 1: Respiratory
Failure and Mechanical Ventilation: Components of airway pressure during
mechanical ventilation, illustrated by an inspiratory-hold maneuver. ). Thus:
Fig. 1
Any elevation in peak airway pressure (eg, > 25 cm H2O) should prompt
measurement of the end-inspiratory pressure (plateau pressure) by an end-
inspiratory hold maneuver to determine the relative contributions of resistive and
elastic pressures. The maneuver keeps the exhalation valve closed for an
additional 0.3 to 0.5 sec after inspiration, delaying exhalation. During this time,
airway pressure falls from its peak value as airflow ceases. The resulting end-
inspiratory pressure represents the elastic pressure once PEEP is subtracted
(assuming the patient is not making active inspiratory or expiratory muscle
contractions at the time of measurement). The difference between peak and
plateau pressure is the resistive pressure.
Elevated resistive pressure (eg, > 10 cm H2O) suggests the endotracheal tube has
been kinked or plugged with secretions or the presence of an intraluminal mass,
increased intraluminal secretions, or bronchospasm. An increase in elastic
pressure (eg, > 10 cm H2O) suggests decreased lung compliance from edema,
fibrosis, or lobar atelectasis; large pleural effusions, pneumothorax or
fibrothorax; extrapulmonary restriction as may arise from circumferential burns
or other chest wall deformity, ascites, pregnancy, or massive obesity; or a tidal
volume too large for the amount of lung being ventilated (eg, a normal tidal
volume being delivered to a single lung because of malpositioning of the
endotracheal tube).
Adjustable ventilator settings differ with mode but include respiratory rate, tidal
volume, trigger sensitivity, flow rate, waveform, and inspiratory/expiratory (I/E)
ratio.
A/C ventilation is the simplest and most effective means of providing full
mechanical ventilation. In this mode, each inspiratory effort beyond the set
sensitivity threshold triggers delivery of the fixed tidal volume. If the patient
does not trigger the ventilator frequently enough, the ventilator initiates a breath,
ensuring the desired minimum respiratory rate.
SIMV also delivers breaths at a set rate and volume that is synchronized to the
patient's efforts. In contrast to A/C, however, patient efforts above the set
respiratory rate are unassisted, although the intake valve opens to allow the
breath. This mode remains popular, despite the fact that it neither provides full
ventilator support as does A/C nor facilitates liberating the patient from
mechanical ventilation.
Pressure control ventilation is similar to A/C; each inspiratory effort beyond the
set sensitivity threshold delivers full pressure support maintained for a fixed
inspiratory time. A minimum respiratory rate is maintained.
In pressure support ventilation, a minimum rate is not set; all breaths are
triggered by the patient. Pressure is typically cut off when back-pressure causes
flow to drop below a certain point. Thus, a longer or deeper inspiratory effort by
the patient results in a larger tidal volume. This mode is commonly used to
liberate patients from mechanical ventilation by letting them assume more of the
work of breathing. However, no studies indicate that this approach is more
successful.
NIPPV can be given in the form of continuous positive airway pressure (CPAP)
or bilevel positive airway pressure (BIPAP). In CPAP, constant pressure is
maintained throughout the respiratory cycle with no additional inspiratory
support. With BIPAP, the physician sets both the expiratory positive airway
pressure (EPAP) and the inspiratory positive airway pressure (IPAP), with
respirations triggered by the patient. Because the airway is unprotected,
aspiration is possible, so patients must have adequate mentation and airway
protective reflexes and no imminent indication for surgery or transport off the
floor for prolonged procedures. NIPPV should be avoided in patients who are
hemodynamically unstable or those with evidence of impaired gastric emptying,
such as occurs with ileus, bowel obstruction, or pregnancy. In such
circumstances, the swallowing of large quantities of air may result in vomiting
and life-threatening aspiration. Indications for conversion to endotracheal
intubation and conventional mechanical ventilation include the development of
shock or frequent arrhythmias, myocardial ischemia, and transport to a cardiac
catheterization laboratory or surgical suite where control of the airway and full
ventilatory support are desired. Obtunded patients and those with copious
secretions are not good candidates. Also, IPAP must be set below esophageal
opening pressure (20 cm H2O) to avoid gastric insufflation.
NIPPV can be used in the outpatient setting. For example, CPAP is often used for
patients with obstructive sleep apnea (see Sleep Apnea: Obstructive Sleep
Apnea), whereas BIPAP can be used for those with concomitant central sleep
apnea (see Sleep Apnea: Central Sleep Apnea) and obstructive sleep apnea or for
chronic ventilation in patients with progressive neuromuscular diseases.
Ventilator settings: Ventilator settings are tailored to the underlying condition, but
the basic principles are as follows.
Tidal volume and respiratory rate set the minute ventilation. Too high a volume
risks overinflation; too low a volume risks atelectasis. Too high a rate risks
hyperventilation and respiratory alkalosis along with inadequate expiratory time
and autoPEEP; too low a rate risks inadequate minute ventilation and respiratory
acidosis. A tidal volume of 8 to 10 mL/kg ideal body weight (see Respiratory
Failure and Mechanical Ventilation: Initial Ventilator Management in ALI/ARDS
) is usually appropriate, although some patients with normal lung mechanics
(particularly those with neuromuscular disease) benefit from tidal volumes on the
high end of this range to prevent atelectasis, whereas patients with ALI/ARDS or
acute exacerbations of COPD or asthma may require lower volumes (see
Respiratory Failure and Mechanical Ventilation: Mechanical ventilation in
ALI/ARDS). Ideal body weight (IBW) rather than actual body weight is used to
determine the appropriate tidal volume for patients with lung disease receiving
mechanical ventilation:
Sensitivity adjusts the level of negative pressure required to trigger the ventilator.
A typical setting is –2 cm H2O. Too high a setting will cause weak patients to be
unable to trigger a breath. Too low a setting may lead to overventilation by
causing the machine to auto-cycle. Patients with high levels of autoPEEP may
have difficulty inhaling deeply enough to achieve a sufficiently negative intra-
airway pressure.
The ratio of the time spent in inhalation versus that spent in exhalation (I:E ratio)
can be adjusted in some modes of ventilation. A normal setting for patients with
normal mechanics is 1:3. Patients with asthma or COPD exacerbations should
have ratios of 1:4 or even more to limit the degree of autoPEEP.
The inspiratory flow rate can be adjusted in some modes of ventilation (ie, either
the flow rate or the I:E ratio can be adjusted, not both). The inspiratory flow
should generally be set at about 60 L/min but can be increased up to 120 L/min
for patients with airflow limitation to facilitate having more time in exhalation,
thereby limiting autoPEEP.
Fio2 is initially set at 1.0 and is subsequently decreased to the lowest level
necessary to maintain adequate oxygenation.
PEEP can be applied in any ventilator mode. PEEP increases end-expired lung
volume and reduces airspace closure at the end of expiration. Most patients
undergoing mechanical ventilation may benefit from the application of PEEP at 5
cm H2O to limit the atelectasis that frequently accompanies endotracheal
intubation, sedation, paralysis, and/or supine positioning. Higher levels of PEEP
improve oxygenation in disorders such as cardiogenic pulmonary edema and
ARDS. PEEP permits use of lower levels of Fio2 while preserving adequate
arterial oxygenation. This may be important in limiting the lung injury that may
result from prolonged exposure to a high Fio2 (≥ 0.6). PEEP increases
intrathoracic pressure and thus may impede venous return, provoking
hypotension in the hypovolemic patient, and may overdistend portions of the
lung, thereby causing ventilator-associated lung injury (VALI). By contrast, if
PEEP is too low, it may result in cyclic airspace opening and closing, which in
turn may also cause VALI from the resultant repetitive shear forces.
Sedation and comfort: Although some patients tolerate mechanical ventilation via
endotracheal tube without sedatives, most require continuous IV administration
of sedatives (eg, propofol Some Trade Names
DIPRIVAN
Click for Drug Monograph
, lorazepam Some Trade Names
ATIVAN
Click for Drug Monograph
, midazolam Some Trade Names
No US trade name
Click for Drug Monograph
) and analgesics (eg, morphine Some Trade Names
DURAMORPH
MS CONTIN
MSIR
ROXANOL
Click for Drug Monograph
, fentanyl Some Trade Names
ACTIQ
DURAGESIC
SUBLIMAZE
Click for Drug Monograph
) to minimize stress and anxiety. These drugs also can reduce energy expenditure
to some extent, thereby reducing CO2 production and O2 consumption. Doses
should be titrated to the desired effect, guided by standard sedation/analgesia
scoring systems. Patients undergoing mechanical ventilation for ALI/ARDS
typically require higher levels of sedation and analgesia. The use of propofol
Some Trade Names
DIPRIVAN
Click for Drug Monograph
for longer than 24 to 48 h requires periodic monitoring of serum triglyceride
levels.
Last full review/revision August 2007 by Brian K. Gehlbach, MD; Jesse Hall,
MD
Mechanical Ventilator
A mechanical ventilator is a machine that makes it easier for patients to breathe until they
are able to breathe completely on their own. Sometimes the machine is called just a
ventilator, respirator or breathing machine. Usually, a patient is connected to the ventilator
through a tube (called an endotracheal tube) that is placed in the windpipe. Sometimes,
patients can use a machine that assists breathing through a mask or mouthpiece but this
may not work with severe respiratory problems. Despite their life-saving benefits,
mechanical ventilators carry many risks. Therefore, the goal is to help patients recover as
quickly as possible to get them off the ventilator at the earliest possible time.
• To deliver oxygen
• To eliminate carbon dioxide
• To ease the work of breathing
The main job of our lungs is to get oxygen into the body and to get rid of carbon dioxide.
When a patient's lungs are no longer able to do this job completely, we use a ventilator to
help. Most commonly, patients are put on a mechanical ventilator when they are in
respiratory failure. Respiratory failure is the situation when the patient has a low level of
oxgyen in the blood, even while getting oxygen therapy and/or when the level of carbon
dioxide rises too much in the blood. Some patients need help from a ventilator even though
they still have nearly normal levels of oxygen and carbon dioxide in the bloodstream. This
can be true when breathing is very uncomfortable. Sometimes patients are placed on a
ventilator because of other serious injuries that require treatment, which may interfere with
breathing temporarily.
In most cases, mechanical ventilators are used for patients who cannot breathe by
themselves. The only other choice would be to allow the patient to die, while using
medicines to maintain comfort (see sections on Code Status and Withdrawal of Life-
Sustaining Treatments). Mechanical ventilators do not actually fix diseases, but rather keep
the patient alive while the hospital staff finds out why the patient has difficulty breathing
and treats the disease that is causing the difficulty.
Risks:
Some of the risks of mechanical ventilation include:
• Infections - The endotracheal tube in the windpipe makes it easier for bacteria to get into the
lungs. As a result, the lungs develop an infection, which is called pneumonia. The risk of
pneumonia is about 1% for each day spent on the ventilator. Pneumonia can often be treated
with antibiotics. Sometimes the pneumonia can be severe or difficult to treat because of
resistant bacteria (see General Information).
• Collapsed Lung - This is called a pneumothorax. The mechanical ventilator pushes air into the
lungs. It is possible for a part of the lung to get over-expanded which can injure it. Air sacs may
leak air into the chest cavity and cause the lung to collapse. If this air leak happens, doctors can
place a tube in the chest between the ribs to drain out the air leaking from the lung. The tube
allows the lung to re-expand and seal the leak (see related Information Sheet on Chest Tube
Thoracostomy). Rarely, collapse of the lung can cause death.
• Lung damage - When the lungs are diseased and not functioning well, they are at greater risk of
injury. The pressure to put air into the lungs with a ventilator can be hard on the lungs.
• Side Effects of Medications - Patients may be given medications, called sedatives, to make
them more comfortable while the ventilator pushes air in and out of the lungs. These
medications make patients sleepy and help them forget unpleasant experiences. The
medications can build up in the body and the patient may remain in a deep sleep for hours to
days, even after the medicine is stopped. Although the doctors and nurses try hard to get just the
right amount of medication for a patient, it is not easy to get it perfectly right.
• Maintenance of Life - In some very sick patients, trying to keep the patient alive means that
dying actually takes longer. Sometimes the lungs fail because the body is dying, and using the
ventilator in place of the lungs only serves to put off what is inevitable - death. In this way, the
ventilator may increase the length of time that patients are uncomfortable in their final days.
Sometimes, doctors can give a good idea how likely use of mechanical ventilation will lead to a
successful recovery. Very often, however, doctors can only give a rough idea of the likelihood
that a patient will survive and go home after mechanical ventilation. A decision about
continuing mechanical ventilation or not may come up if a patient is not showing any recovery
or is continuing to get worse (see sections on Code Status and Withdrawal of Life-Sustaining
Treatments).
Mechanical Ventilation
By Richard RobinsonThe Gale Group Inc., Gale…. Gale Encyclopedia of Surgery, 2004more »
Definition
Mechanical ventilation is the use of a mechanical device (machine) to inflate and deflate the lungs.
Purpose
Mechanical ventilation provides the force needed to deliver air to the lungs in a patient whose own
ventilatory abilities are diminished or lost.
Description
Breathing requires the movement of air into and out of the lungs. This is normally accomplished by the
diaphragm and chest muscles. A variety of medical conditions can impair the ability of these muscles to
accomplish this task, including:
• muscular dystrophies
• motor neuron disease, including ALS
• damage to the brain's respiratory centers
• polio
• myasthenia gravis
• myopathies affecting the respiratory muscles
• scoliosis
Mechanical ventilation may also be used when the airway is obstructed, especially at night in sleep
apnea.
Mechanical ventilation may be required only at night, during limited daytime hours, or around the
clock, depending on the patient's condition. Some patients require mechanical ventilation only for a
short period, during recovery from traumatic nerve injury, for instance. Others require it chronically,
and may increase the number of hours required over time as their disease progresses.
Mechanical ventilation is not synonymous with the use of an oxygen tank. Supplemental oxygen is
used in patients whose gas exchange capacity has diminished, either through lung damage or
obstruction of a major airway. For these patients, the muscles that deliver air work well, but too little
oxygen can be exchanged in the remaining lung, and so a higher concentration is supplied with each
breath. By the same token, many patients who require mechanical ventilation do not need supplemental
oxygen. Their gas exchange capacity is normal, but they cannot adequately move air into and out of the
lungs. In fact, excess oxygen may be dangerous, since it can suppress the normal increased respiration
response to excess carbon dioxide in the lungs.
Mechanical ventilation systems come in a variety of forms. Almost all systems use a machine called a
ventilator that pushes air through a tube for delivery to the patient's airways. The air may be delivered
through a nasal or face mask, or through an opening in the trachea (windpipe), called a tracheostomy.
Much rarer are systems that rhythmically change the pressure around a patient's chest when the
pressure is low, air flows into the lungs, and when it increases, air flows out.
Ventilators
Ventilators can either deliver a set volume with each cycle, or can be set to a specific pressure regimen.
Both are in common use. Volume ventilator settings are adjustable for total volume delivered, timing of
delivery, and whether the delivery is mandatory or determined by the patient's initial inspiratory effort.
Pressure ventilators deliver one of two major pressure regimens. Continuous positive airway pressure
(CPAP) delivers a steady pressure of air, which assists the patient's inspiration (breathing in) and resists
expiration (breathing out). The pressure of CPAP is not sufficient to completely inflate the lungs;
instead its purpose is to maintain an open airway, and for this reason it is used in sleep apnea, in which
a patient's airway closes frequently during sleep.
Bilevel positive airway pressure (BiPAP) delivers a higher pressure on inspiration, helping the patient
obtain a full breath, and a low pressure on expiration, allowing the patient to exhale easily. BiPAP is a
common choice for neuromuscular disease.
The choice of ventilator type is partly determined by the knowledge and preferences of the treating
physician. Settings are adjusted to maintain patient comfort and appropriate levels of oxygen and
carbon dioxide in the blood.
Masks vs. tracheostomy
Delivery of air from a ventilator may be either through a mask firmly held to the face, or through a tube
inserted into the trachea toward the bottom of the throat. A mask interface is called noninvasive
ventilation, while a tracheostomy tube is called invasive ventilation.
Until the mid-1990s, invasive ventilation was the option used by virtually all patients requiring long-
term mechanical ventilation. For some patients, tracheostomy continues to be a preferred option. It is
commonly used when 24-hour ventilation assistance is required, and may be preferred by patients who
find masks uncomfortable or unsightly. Some patients feel ventilation through a "trach tube" is more
reassuring. Tracheostomy is also the preferred option for most patients with swallowing difficulties.
The potential to choke and suffocate on improperly swallowed food is avoided with a tracheostomy.
Tracheostomies may require more frequent suctioning of airway secretions, produced in response to the
presence of the tube and the inflatable cuff that some patients require to hold it in place. The risk of
infection is higher, and air must be carefully humidified and cleaned, since these functions are not
being served by the nasal passages. Tracheostomies do not prevent speech, despite misinformation to
the contrary that even some doctors believe. Speech requires passage of air around the trach tube,
which can occur either with an uncuffed tube, or with the presence of a special valve that allows air
passage past the cuff.
Noninvasive interfaces come in a variety of forms. A simple mouthpiece may be used, which a patient
bites down on to seal the lips around the tube as the pressure cycle delivers a breath. Most masks are
individually fitted to the patient's face, and held in place with straps. A tight fit is essential, since the
pressure must be delivered to the patient's lungs, and not be allowed to blow out the sides of the mask.
Masks may be used around the clock. Nasal masks do not prevent speech, though the tone may change.
Oral or full-face masks do interfere with speech, and are typically used at night or intermittently
throughout the day, for patients who do not need continuous ventilation assistance.
Other alternatives
The iron lung was an early mechanical ventilation device, and is still in use in some hospitals. The
patient's head remains outside of it, while the interior depressurizes. This allows air to push in to the
lungs. Repressurizing deflates the lungs again.
A device that works on the same principle is the chest shell (something like a turtle's shell swung
around to the front). The pneumobelt applies pressure to deflate, and relaxes it to allow inflation. A
rocking bed is used for nighttime ventilation. Tilting the head of the bed down deflates the lungs by
allowing the abdominal contents to press against the diaphragm. Reversing the angle reverses the
process, allowing inflation.
Preparation
Patients with diseases in which mechanical ventilation may be required are advised to learn as much as
possible about treatment options before they become necessary. In particular, it is important to learn
about and make decisions about invasive vs. noninvasive ventilation before the time comes. Many
patients who begin ventilation with emergency tracheostomy have a difficult time switching to
noninvasive ventilation later on (though it is certainly possible).
It is often a good idea to try out different masks and other interfaces before their need arises, and to
have these fitted in preparation for a planned transition to the ventilator. Patients can find support
groups and other sources of information to learn more about the options and the features of each means
of ventilation. Patients may have to help educate their doctors if they are not familiar with noninvasive
options.
Patients with neuromuscular disease may have as much or more need for a deep cough as they do for
ventilatory assistance, and many patients who undergo emergency tracheostomy do so because their
airways have become clogged with mucus build up. Physical therapy cough assistance and a cough
assist device are important options for full respiratory health.
Normal results
Mechanical ventilation is a life saver, and provides comfort and confidence to patients who require it.
Proper ventilation restores levels of oxygen and carbon dioxide in the blood, improving sleep at night
and increasing the ability to engage in activities during the day. When combined with proper
respiratory hygiene, it can prolong life considerably. Patients with progressive diseases such as ALS
may wish to consider end-of-life decisions before commencing mechanical ventilation, or before the
ability to communicate is lost.
BOOKS
Bach, John R. Noninvasive Mechanical Ventilation. Hanley and Belfus, 2002.
Kinnear, W. J. M. Assisted Ventilation at Home: A practical Guide. Oxford: Oxford Medical Publications, 1994
The mechanics of inspiratory support are more complex than previously considered. It has
been established that cyclical inflation and deflation injures lung parenchyma and worsens
outcome (1). Large tidal volume ventilation, to “normalize” blood gases has been shown to
worsen outcome in lung injury (2), presumably due to excessive pressure induced stretch
injury of the parenchyma. Modern ventilation strategy involves attempting to achieve an
adequate minute volume with the lowest possible airway pressure (as this relates to the
degree of alveolar distension). The pressure that we are interested in minimizing is at the
level of the alveolus, the plateau pressure.
The rate, pattern and duration of gas flow control the interplay between volume and
pressure. In volume controlled modes, a desired tidal volume is delivered at a specific flow
(peak flow) rate, using constant, decelerating or sinusoidal flow patterns: the airway pressure
generated may be higher than is desirable. In pressure controlled modes, flow occurs until a
preset peak pressure is met over a specified inspiratory period, the flow pattern is always
decelerating: the tidal volume may be lower than that desired. Moreover, as pulmonary
mechanics change, so too does the delivered tidal volume.
Ventilator “cycling” refers to the mechanism by which the phase of the breath switches
from inspiration to expiration. Modes of ventilation are time cycled, volume cycled or flow
cycled. Time cycling refers to the application of a set “controlled” breath rate. In “controlled
ventilation” a number of mandatory breaths are delivered to the patient at a predetermined
interval.
The respiratory rate may be controlled by the operator or the patient. The patient may
breathe spontaneously, and with modern ventilators these breaths are supported either by
delivering facsimiles of the controlled breaths synchronously with the patient’s effort or by
allowing the patient more subjective control. Pressure support is a form of flow cycled
ventilation in which the patient triggers the ventilator and a pressure limited flow of gas is
delivered. The patient determines the duration of the breath and the tidal volume, which may
vary from breath to breath.
References
(1) Marini JJ. New options for the ventilatory management of acute lung injury. New Horiz 1993; 1(4):489-503.
(2) Ventilation with lower tidal volumes as compared with traditional tidal volumes for acute lung injury and the acute respiratory distress syndrome. The Acute
Respiratory Distress Syndrome Network. N Engl J Med 2000; 342(18):1301-1308.
Mechanical ventilation
In medicine, mechanical ventilation is a method to mechanically assist or replace spontaneous
breathing.
This may involve a machine called a ventilator or the breathing may be assisted by a physician or other
suitable person compressing a bag or set of bellows. Traditionally divided into negative-pressure
ventilation, where air is essentially sucked into the lungs, or positive pressure ventilation, where air (or
another gas mix) is pushed into the trachea.
It can be used as a short term measure, for example during an operation or critical illness (often in the
setting of an intensive care unit). It may be used at home or in a nursing or rehabilitation institution if
patients have chronic illnesses that require long-term ventilatory assistance.
Owing to the anatomy of the human pharynx, larynx, and esophagus and the circumstances for which
ventilation is required then additional measures are often required to "secure" the airway during
positive pressure ventilation to allow unimpeded passage of air into the trachea and avoid air passing
into the esophagus and stomach. Commonly this is by insertion of a tube into the trachea which
provides a clear route for the air. This can be either an endotracheal tube, inserted through the natural
openings of mouth or nose or a tracheostomy inserted through an artificial opening in the neck. In other
circumstances simple airway maneuvres, an oropharyngeal airway or laryngeal mask airway may be
employed. If the patient is able to protect their own airway such as in non-invasive ventilation or
negative-pressure ventilation then no airway adjunct may be needed.
Mechanical ventilation is often a life-saving intervention, but carries many potential complications
including pneumothorax, airway injury, alveolar damage, and ventilator-associated pneumonia.
[citation needed].
In many healthcare systems prolonged ventilation as part of intensive care is a limited resource (in that
there are only so many patients that can receive care at any given moment). It is used to support a
single failing organ system (the lungs) and cannot reverse any underlying disease process (such as
terminal cancer). For this reason there can be (occasionally difficult) decisions to be made about
whether it is suitable to commence someone on mechanical ventilation. Equally many ethical issues
surround the decision to discontinue mechanical ventilation.
Types of ventilators
Ventilation can be delivered via:
• Hand-controlled ventilation such as:
Breath termination
Modes of ventilation are classified by the means that they determine the inspired breath is complete.
This is sensed by either pressure or volume.
• Volume ventilation - A predetermined tidal volume (Vt) is set for the patient and is delivered
with each inspiration. The amount of pressure necessary to deliver this volume will fluctuate
from breath to breath based on the resistance and compliance of the patient and ventilator
circuit. If the tidal volume is set at 500ml, the ventilator will continue to inspire gas until it
reaches its goal. Upon completion of the inspired volume, the ventilator will open a valve
allowing the patient to passively exhale.
• Pressure ventilation - A predetermined peak inspiratory pressure (PIP) is determined based on
the patient's condition and pathophysiology. The ventilator will flow gas into the patient until
this set pressure is reached. Upon reaching the preset PIP, the ventilator allows for passive
exhalation. Caution and close observation must be given in this mode due to potential for either
hypoventilation or hyperventilation because the tidal volume is variable.
Several manufactures have incorporated features from both of theses modes in an attempt to
accommodate patients needs.
These modes are flow-variable, volume-targeted, pressure-regulated, time-limited modes (for example,
pressure regulated volume control - PRVC). This means that instead of providing an exact tidal volume
each breath, a target volume is set and the ventilator will vary the inspiratory flow at each breath to
achieve the target volume at the lowest possible peak pressure. The inspiratory time (Ti) limits the
length of the inspiratory cycle and therefore the I:E ratio. Pressure regulated modes such as PRVC,
Auto-flow (Draeger)or Average Volume Assured Pressure Support (AVAPS) from Philips can most
easily be thought of as turning a volume mode into a pressure mode with the added benefit of
maintaining more control over tidal volume than with strictly pressure-control.
Breath initiation
The other method of classifying mechanical ventilation is based on how to determine when to start
giving a breath. Similar to the termination classification noted above, microprocessor control has
resulted in a myriad of hybrid modes that combine features of the traditional classifications. Note that
most of the timing initiation classifications below can be combined with any of the termination
classifications listed below.
• Assist Control (AC). In this mode the ventilator provides a mechanical breath with either a pre-
set tidal volume or peak pressure every time the patient initiates a breath. Traditional assist-
control used only a pre-set tidal volume—when a preset peak pressure is used this is also
sometimes termed Intermittent Positive Pressure Ventilation or IPPV. However, the initiation
timing is the same—both provide a ventilator breath with every patient effort. In most
ventilators a back-up minimum breath rate can be set in the event that the patient becomes
apnoeic. Although a maximum rate is not usually set, an alarm can be set if the ventilator cycles
too frequently. This can alert that the patient is tachypneic or that the ventilator may be auto-
cycling (a problem that results when the ventilator interprets fluctuations in the circuit due to
the last breath termination as a new breath initiation attempt).
• Synchronized Intermittent Mandatory Ventilation (SIMV). In this mode the ventilator provides a
pre-set mechanical breath (pressure or volume limited) every specified number of seconds
(determined by dividing the respiratory rate into 60 seconds - thus a respiratory rate of 12
results in a 5 second cycle time). Within that cycle time the ventilator waits for the patient to
initiate a breath using either a pressure or flow sensor. When the ventilator senses the first
patient breathing attempt within the cycle, it delivers the preset ventilator breath. If the patient
fails to initiate a breath, the ventilator delivers a mechanical breath at the end of the breath
cycle. Additional spontaneous breaths after the first one within the breath cycle do not trigger
another SIMV breath. However, SIMV may be combined with pressure support (see below).
SIMV is frequently employed as a method of decreasing ventilatory support (weaning) by
turning down the rate, which requires the patient to take additional breaths beyond the SIMV
triggered breath.
• Controlled Mechanical Ventilation (CMV). In this mode the ventilator provides a mechanical
breath on a preset timing. Patient respiratory efforts are ignored. This is generally
uncomfortable for children and adults who are conscious and is usually only used in an
unconscious patient. It may also be used in infants who often quickly adapt their breathing
pattern to the ventilator timing.
• Pressure Support Ventilation (PSV). When a patient attempts to breathe spontaneously through
an endotracheal tube, the narrowed diameter of the airway results in higher resistance to airflow,
and thus a higher work of breathing. PSV was developed as a method to decrease the work of
breathing in-between ventilator mandated breaths by providing an elevated pressure triggered
by spontaneous breathing that "supports" ventilation during inspiration. Thus, for example,
SIMV might be combined with PSV so that additional breaths beyond the SIMV programmed
breaths are supported. However, while the SIMV mandated breaths have a preset volume or
peak pressure, the PSV breaths are designed to cut short when the inspiratory flow reaches a
percentage of the peak inspiratory flow (e.g. 10-25%). New generation of ventilators provides
user-adjustable inspiration cycling off threshold, and some even are equipped with automatic
inspiration cycling off threshold function. This helps the patient ventilator synchrony.[4] The
peak pressure set for the PSV breaths is usually a lower pressure than that set for the full
ventilator mandated breath. PSV can be also be used as an independent mode.
• Continuous Positive Airway Pressure (CPAP). A continuous level of elevated pressure is
provided through the patient circuit to maintain adequate oxygenation, decrease the work of
breathing, and decrease the work of the heart (such as in left-sided heart failure — CHF). Note
that no cycling of ventilator pressures occurs and the patient must initiate all breaths. In
addition, no additional pressure above the CPAP pressure is provided during those breaths.
CPAP may be used invasively through an endotracheal tube or tracheostomy or non-invasively
with a face mask or nasal prongs.
• Positive end-expiratory pressure (PEEP) is functionally the same as CPAP, but refers to the use
of an elevated pressure during the expiratory phase of the ventilatory cycle. After delivery of the
set amount of breath by the ventilator, the patient then exhales passively. The volume of gas
remaining in the lung after a normal expiration is termed the functional residual capacity
(FRC). The FRC is primarily determined by the elastic qualities of the lung and the chest wall.
In many lung diseases, the FRC is reduced due to collapse of the unstable alveoli, leading to a
decreased surface area for gas exchange and intrapulmonary shunting (see above), with wasted
oxygen inspired. Adding PEEP can reduce the work of breathing (at low levels) and help
preserve FRC.
Sighs
An adult patient breathing spontaneously will usually sigh about 6–8 times per hour to prevent
microatelectasis, and this has led some to propose that ventilators should deliver 1½–2 times the
amount of the preset tidal volume 6–8 times per hour to account for the sighs. However, such high
quantity of volume delivery requires very high peak pressure that predisposes to barotrauma. Currently,
accounting for sighs is not recommended if the patient is receiving 10-12 mL/kg or is on PEEP. If the
tidal volume used is lower, the sigh adjustment can be used, as long as the peak and plateau pressures
are acceptable.
Sighs are not generally used with ventilation of infants and young children.
Initial FiO2
Because the mechanical ventilator is responsible for assisting in a patient's breathing, it must then also
be able to deliver an adequate amount of oxygen in each breath. The FiO2 stands for fraction of
inspired oxygen, which means the percent of oxygen in each breath that is inspired. (Note that normal
room air has ~21% oxygen content). In adult patients who can tolerate higher levels of oxygen for a
period of time, the initial FiO2 may be set at 100% until arterial blood gases can document adequate
oxygenation. An FiO2 of 100% for an extended period of time can be dangerous, but it can protect
against hypoxemia from unexpected intubation problems. For infants, and especially in premature
infants, avoiding high levels of FiO2 (>60%) is important.
Positioning
Prone (face down) positioning has been used in patients with ARDS and severe hypoxemia. It improves
FRC, drainage of secretions, and ventilation-perfusion matching (efficiency of gas exchange). It may
improve oxygenation in > 50% of patients, but no survival benefit has been documented.
Prophylaxis
• To protect against ventilator-associated pneumonia, patients' beds are often elevated to about
30°.
• Deep vein thrombosis prophylaxis with heparin or sequential compression device is important
in older children and adults.
• A histamine receptor (H2) blocker or proton-pump inhibitor may be used to prevent
gastrointestinal bleeding, which has been associated with mechanical ventilation
Modification of settings
In adults when 100% FiO2 is used initially, it is easy to calculate the next FiO2 to be used and easy to
estimate the shunt fraction. The estimated shunt fraction refers to the amount of oxygen not being
absorbed into the circulation. In normal physiology, gas exchange (oxygen/carbon dioxide) occurs at
the level of the alveoli in the lungs. The existence of a shunt refers to any process that hinders this gas
exchange, leading to wasted oxygen inspired and the flow of un-oxygenated blood back to the left heart
(which ultimately supplies the rest of the body with unoxygenated blood).
When using 100% FiO2, the degree of shunting is estimated by subtracting the measured PaO2 (from
an arterial blood gas) from 700 mmHg. For each difference of 100 mmHg, the shunt is 5%. A shunt of
more than 25% should prompt a search for the cause of this hypoxemia, such as mainstem intubation or
pneumothorax, and should be treated accordingly. If such complications are not present, other causes
must be sought after, and PEEP should be used to treat this intrapulmonary shunt. Other such causes of
a shunt include:
• Alveolar collapse from major atelectasis
• Alveolar collection of material other than gas, such as pus from pneumonia, water and protein
from acute respiratory distress syndrome, water from congestive heart failure, or blood from
haemorrhage
Connection to ventilators
There are various procedures and mechanical devices that provide protection against airway collapse,
air leakage, and aspiration:
• Face mask - In resuscitation and for minor procedures under anaesthesia, a face mask is often
sufficient to achieve a seal against air leakage. Airway patency of the unconscious patient is
maintained either by manipulation of the jaw or by the use of nasopharyngeal or oropharyngeal
airway. These are designed to provide a passage of air to the pharynx through the nose or
mouth, respectively. Poorly fitted masks often cause nasal bridge ulcers, a problem for some
patients. Face masks are also used for non-invasive ventilation in conscious patients. A full face
mask does not, however, provide protection against aspiration.
• Laryngeal mask airway - The laryngeal mask airway (LMA) causes less pain and coughing than
a tracheal tube. However, unlike tracheal tubes it does not seal against aspiration, making
careful individualised evaluation and patient selection mandatory.
• Tracheal intubation is often performed for mechanical ventilation of hours to weeks duration. A
tube is inserted through the nose (nasotracheal intubation) or mouth (orotracheal intubation) and
advanced into the trachea. In most cases tubes with inflatable cuffs are used for protection
against leakage and aspiration. Intubation with a cuffed tube is thought to provide the best
protection against aspiration. Tracheal tubes inevitably cause pain and coughing. Therefore,
unless a patient is unconscious or anaesthetized for other reasons, sedative drugs are usually
given to provide tolerance of the tube. Other disadvantages of tracheal intubation include
damage to the mucosal lining of the nasopharynx or oropharynx and subglottic stenosis.
• Esophageal obturator airway - sometimes used by emergency medical technicians and basic
EMS providers not trained to intubate. It is a tube which is inserted into the esophagus, past the
epiglottis. Once it is inserted, a bladder at the tip of the airway is inflated, to block ("obturate")
the esophagus, and oxygen is delivered through a series of holes in the side of the tube which is
then forced into the lungs.
• Cricothyrotomy - Patients who require emergency airway management, in whom tracheal
intubation has been unsuccessful, may require an airway inserted through a surgical opening in
the cricothyroid membrane. This is similar to a tracheostomy but a cricothyrotomy is reserved
for emergency access.[1]
• Tracheostomy - When patients require mechanical ventilation for several weeks, a tracheostomy
may provide the most suitable access to the trachea. A tracheostomy is a surgically created
passage into the trachea. Tracheostomy tubes are well tolerated and often do not necessitate any
use of sedative drugs. Tracheostomy tubes may be inserted early during treatment in patients
with pre-existing severe respiratory disease, or in any patient who is expected to be difficult to
wean from mechanical ventilation, i.e., patients who have little muscular reserve.
• Mouthpiece - Less common interface, does not provide protection against aspiration. There are
lipseal mouthpieces with flanges to help hold them in place if patient is unable.
Terminology
Terminology used in the field of mechanical ventilation and respiratory support:
• APRV Airway pressure release ventilation
• ASB Assisted spontaneous breathing—also ASV = assisted spontaneous ventilation
• ASV Adaptive support ventilation—a patented technology—closed-loop mechanical
respiration, a further development of MMV. Can also stand for assisted spontaneous ventilation.
• ATC Automatic tube compensation
• Automode Automode
• BCV Biphasic Cuirass Ventilation
• BIPAP Bilevel Positive Airway Pressure
• CMV Continuous mandatory ventilation
• CPAP Continuous positive airway pressure
• CPPV Continuous positive pressure ventilation
• EPAP Expiratory positive airway pressure
• HFV High frequency ventilation
• HFFI High frequency flow interruption
• HFJV High frequency jet ventilation
• HFOV High frequency oscillatory ventilation
• HFPPV High frequency positive pressure ventilation
• ILV Independent lung ventilation—separate sides positive pressure ventilation.
• IPAP Inspiratory positive airway pressure
• IPPV Intermittent positive pressure ventilation
• IRV Inversed ratio ventilation— mechanical ventilation with switched respiration phases/time
rate.
• LFPPV Low frequency positive pressure ventilation
• MMV Mandatory minute volume
• NAVA Neurally Adjusted Ventilatory Assist
• NIF Negative inspiratory force—amount of force generated by a patient against a closed valve;
greater than 20 cmH2O indicates an adequately strong diaphragm.
• NIV Non-invasive ventilation
• PAP Positive airway pressure
• PAV and PAV+ Proportional assist ventilation and proportional assist ventilation plus
• P/F ratio Ratio of PaO2 off an ABG and FiO2 off the ventilator. P/F < 200 indicates ARDS,
P/F < 300 indicates ALI
• PCMV (P-CMV) Pressure controlled mandatory ventilation
• PCV Pressure controlled ventilation or PC Pressure control—pressure-controlled, fully
mechanical ventilation.
• PEEP Positive end-expiratory pressure
• PNPV Positive negative pressure ventilation—switching pressure mechanical ventilation
• PPS Proportional pressure support
• PRVC Pressure regulated volume controlled ventilation
• PSV Pressure Support Ventilation or PS—supported spontaneous respiration, see also ASB.
• RSBI Rapid shallow breathing index—ratio of breath rate divided by the tidal volume.
RSBI<105 declares a patient can be extubated and maintain themselves. Also indicates patient
has a good chance of staying extubated.[12]
• (S) IMV (Synchronized) intermittent mandatory ventilation
• S-CPPV Synchronized continuous positive pressure ventilation
• S-IPPV Synchronized intermittent positive pressure ventilation
• TNI Therapy with nasal insufflation—nasal high-flow mechanical ventilation for respiration
support.
• VCMV (V-CMV) Volume controlled mandatory ventilation
• VCV Volume controlled ventilation or VC, Volume Control—volume-controlled, fully
mechanical ventilation.
• VS Volume Support
• ZAP Zero airway pressure—spontaneous respiration under atmospheric pressure.
References
1. ^ Colice, Gene L (2006). "Historical Perspective on the Development of Mechanical
Ventilation". In Martin J Tobin. Principles & Practice of Mechanical Ventilation (2 ed.). New
York: McGraw-Hill. ISBN 978-0071447676.
2. ^ Chamberlain D (2003) "Never quite there: A tale of resuscitation medicine" Clinical
Medicine, Journal of the Royal College of Physicians' 3 6:573-577