Maxima: Touch Probe Calibration

The touch probe calibration of the Maxima will calibrate and account
for any wear in the touch probe tips. This will allow for a more
consistent bevel and/or groove. AIT recommends a calibration once
every morning before the machine is used for the day. Low volume
labs may be fine with a calibration once a week at bare minimum,
though this is not a recommended practice.

The following are step-by-step instructions for a touch probe


calibration of the Maxima. Prior to beginning the calibration, please
have your Maxima Accessory Kit ready. You will need your steel
calibration disc, highlighted in the image below.
The selection buttons depicted below will be referred to as shown.

1) From the normal operating screen push and hold the MENU button
(H7).
2) A new window will appear asking for an access code. This code is
V1, V2, V3, V4. You only have a few seconds to enter this code
correctly.

3) Insert the steel calibration disc from the accessory kit as shown to
the left. Press the blue OPEN/CLOSE button to secure the disc.
select calibration

4) From the technical menu, press the Right arrow button (H4) to
select calibration.

5) Touch Probe will be highlighted in the drop-down menu. Press the


ENTER button (H7).
6) When the Touch Probe Calibration menu appears, press the START
button to begin the calibration.

7) This picture shows acceptable results. The image should show two
parallel black lines breaking off into two red, angled lines. The left and
right halves of the image should appear mirrored. If the calibration
failed you will receive an error message. If this occurs please call AIT
Technical Support.
Press the ENTER button (H7) twice to save the calibration. Press the
blue OPEN/CLOSE button and remove the steel calibration disc. Press
the ESC button (V4) to return to the normal operating screen.

Maxima: Automatic Wheel Calibration


The automatic wheel calibration of the Maxima will calibrate and
account for any wear in the diamond wheels. This will ensure more
consistent sizing. AIT recommends a calibration once every morning
before the machine is used for the day. Low volume labs may be fine
with a calibration once a week at bare minimum, though this is not a
recommended practice.

1) From the normal operating screen push and hold the MENU button
(H7).
2) A new window will appear asking for an access code. This code is
V1, V2, V3, V4. You only have a few seconds to enter this code
correctly.

3) Insert the steel calibration disc from the accessory kit as shown to
the left. Press the blue OPEN/CLOSE button to secure the disc.
4) From the technical menu, press the Right arrow button (H4) to
select calibration.

5) Use the Down arrow button (H3) and highlight Automatic. Press
the ENTER (H7) button.
6) Press the START button the begin the calibration. After the
calibration has finished press the ENTER button (H7) to save the
calibration.

Press the ESC button (V4) to return to the normal operating screen.

Maxima: Touch Probe Tip Replacement


1) Remove the chucking pad as shown in the image to the left.

From the normal operating screen press and hold the MENU button
(H7). A window will appear asking for an access code. This code is V3,
V3, START, OPEN/CLOSE.

2) Use the Right arrow button (H4) to highlight T.Service and press
the ENTER button (H7).

Use the Down arrow button (H3) to highlight Check optoswitches and
then press the ENTER button (H7).
3) Use the Down arrow button (H3) to highlight Opto Size assembly.
Press and hold the + button (H5) to move the chucking arms down
slightly.

4) Replace the probe tips now. This is easily done by hand but may
require a pair of pliers to loosen them initially.

There are two types of touch probe tips, one type screws on and
another type snaps on.

NOTE: AIT recommends that if one of the touch probe tips needs to be
replaced, the other one be changed at the same time.
5) After replacing the touch probe tips use the Down arrow to
highlight Opto SW sensor Motor and press the + button (H5). The tips
should line up perfectly.

6) Use a philips screw driver to remove the Maxima's cover. There are
four screws, two located on the lower half of each side.

Once open, remove the silver shield seen in the picture to the left. For
this you will need a flat-head screw driver.

The arrows show the locations of each arm's fastener. Loosen the arm
that is off, make the adjustments, and tighten it again.

When the tips meet perfectly, replace the silver shield and the
Maxima's cover.

NOTE: It is important to keep the inside of the machine as dust-free


as possible. AIT recommends the use of a compressed air can,
especially for the mechanics underneath the silver shield.
moving the chucking arms
7) Press the ESC button (V4) to return to the normal operating
screen. When the touch probes and chucking arms are in normal
working positioning, the space shown to the left should be 4mm. If
not, you may have to go back and make further adjustments.
Maxima: Changing the Chucking Adapter

1) To remove the chucking adapter use a 2mm allen wrench to


unscrew the fastener as shown to the left.

2) Remove the adapter and replace it with the Indo adapter that came
standard on your mahcine.

3) Use the allen wrench to secure the adapter.


standard chucking adapter