Proteomics/Protein Identification - Mass


Spectrometry/Instrumentation

This Section:

How does a Mass Spectrometer work?


A mass spectrometer is made up of three
components: an ion source, mass analyzer, and a
detector. The unknown sample which may
originate as solid, liquid, solution or vapor, is
presented to the ionization source. After ionizing
the sample, the ions of the sample are passed to
the mass analyzer region where separation based
on the mass-to-charge ratio occurs. Once separated
by the analyzer, the ions then enter the detector
portion of the mass spectrometer. At this point, the
machine calculates the mass-to-charge ratio and the relative abundance of each of the different
ions. From this information, a spectrum graph can be created such as the one to the right.

Most mass spectrometers are maintained under a vacuum to improve the chances of ions
traveling from ionization source to detector without interference by collision with air
molecules.

Ion Source
The ion source is the mass spectrometer component which ionizes the sample to be analyzed.
Ionization mainly serves to present the sample as vaporized ions which can be acted upon by
the mass analyzer and measured by the ion detector. There are many different methods
available to ionize samples, such as positive or negative ion modes. The ionization method
chosen should depend on the type of sample and the type of mass spectrometer.

Ionization Methods:
Page 2 of 10

File:ES-MassSpec.jpeg
There are three types of ionization methods, electron ionization, Electronspray ionization
chemical ionization and photo-induced ionization. Electron
ionization involves application of an electrical current to the
sample to induce ionization. Chemical ionization involves
interaction of the sample with reagent molecules to induce
ionization. Ions produced are often denoted with symbols that
indicate the nature of the ionization, eg. [M+H]+ is used to
represent a molecule which is protonated.

■ Electron ionization
■ Electrospray Ionization
■ Field Desorption / Field Ionization
■ Chemical Ionization
■ Matrix Assisted Laser Desorption Ionization
■ Fast Atom Bombardment
■ Thermal Ionisation

The methods commonly used in proteomics are ‘Matrix MALDI


Assisted Laser Desorption Ionization’ or MALDI and
‘Electrospray Ionization’ also known as ESI. Atmospheric
pressure chemical ionization and Atmospheric pressure photo-ionization are two other forms as
well.

MALDI uses a solid support target plate where a UV active matrix (solid or liquid) is spotted
on the plate followed by the sample over the matrix. The laser hits the spot on the crystallized
matrix and transfers energy from the matrix molecule to the sample. This energy transfer
vaporizes the sample, sending a plume of ions into the MALDI source. This plume of ions is
then collected and held in the source until a pulse sends them all out simultaneously. If the
MALDI is attached to a Time of Flight(TOF) mass analyzer these ions are then sent down the
TOF tube (typically ~2m) and are separated according to their velocity (light ions hitting first).
MALDI is the preferred instrumentation for proteomics due to ease of reading the spectrum;
most ions are found in the +1 charge state [M+H].

Electrospray, on the other hand, is done by injecting the sample dissolved in a slightly acidic
solution through a heated capillary that has a voltage around 10v. This allows for highly
charged particles to be formed at the tip of the capillary. As the particles evaporate, their
charge/volume increases to a point where charge repulsion forces take over and the particle will
explode. A small drop will form which continues the process until individual molecules are in
the gas phase and charged. These ions will then travel into the analyzer, typically a quadrupole,
to be scanned one mass at a time. The molecules in electrospray tend to be multiply charged
and even though the upper mass limit of a quadrupole is 2000 m/z, the multiple charges allow
for high mass ions or proteins to be identified. Proteins will have a charge envelope in which
each peak has a different amount of charges on them. Special software is needed to deconvolve
the multiple charge species peaks into a single mass peak, such as MassLynx from Waters. This
Page 3 of 10

form of ionization is good for most compounds, although is not best with neutral or low polarity
molecules.

Atmospheric pressure chemical ionization involves interaction of the sample with reagent
molecules to induce ionization. As in electrospray ionization, liquid is pumped through a
capillary. At the tip it is nebulized and a corona discharge takes place ionizing the molecules.
The molecules interact with the analyte and transfer their charge. This form of ionization is
good for small thermally stable molecules.

Atmospheric pressure photo-ionization uses photons to excite and ionize the molecules after
they have been nebulized. This form of ionization is good for neutral compounds.

Mass Analyzers
The mass analyzer is the component that separates the ions created from the ion source by their
mass-to-charge ratios. Mass analyzers are based on the principles of charged particles in an
electric or magnetic field. By using Lorentz force law and Newton’s second law of motion you
can generate the following equation:

Where m is the mass, the ionic charge is q , a the acceleration, E is the electric field, and the
vector cross product of the ion velocity and the magnetic field is v x B. This equation says that
two particles with the same mass to charge ratio (m/q) will behave exactly the same.

So what this equation is basically saying is that the mass to charge ratio acts as a determinant of
acceleration of the ion, which can also be represented as the addition of the electric field plus
the cross product of the ion velocity and magnetic field.

Scanning Mass Analyzers

Scanning mass analyzers need to separate the ions based on their relative abundances and mass
to charge ratios. Electromagnetic fields are used to separate the ions based on their mass to
charge ratios, by using a slit they are able to regulate which mass to charge ratio ions get to the
detector. Once selected for a particular mass to charge ratio, the ion current is then recorded as
a function of time which is analagous to mass.

Sector Mass Spectrometer


Page 4 of 10

Mass spectrometer that uses a mass analyzer using magnetic, electric or static sector in it is
called a sector instrument. This also works in combination of sectors like BEB, magnetic-
electric-magnetic. In the present days, these sector instruments are mostly double focusing i.e,
the ion beams are focused in both direction and velocity. Here is how a magnetic sector mass
spectrometer works, imagine a tube like thing between two electromagnets. When the electrons
are passed through the tube from one end to another the magnetic field bends the electron
stream by exerting turning force on it. Then m/z ratio is determined.

Orbitrap Mass spectrometer

Orbitrap Mass spectrometer is the one in which ions are subjected to an electric field produced
by electrodes tangentially. Also these ions are trapped between outer electrodes, where as ions
are attracted with an electrostatic attraction to the inner electrode which is balanced by
centrifugal force. Therefore these ions move in a circular manner around the inner electrode
also back and forth with a particular m/z ratio. By using this ion oscillation based on the m/z
ratio, the trap can act as a mass analyzer.

Quadrupole Mass Spectrometers

You can think of this method as a filter or funnel which only allows certain ion masses to pass
through. The "funnel" is actually a combination of positively and negatively charged metal rods
which together form a channel through which the ions travel. The theory is that only selected
masses will be able to pass through the channel, as all other ions won't have a stable trajectory
though it and will hit the quadrupole rods, stopping the ion from reaching the detector.

Ion trap Quadrupole Spectrometer

A quadrupole ion trap mass spectrometer consists of hyperbolic electrodes with a ring and two
endcaps, which is the core of this instrument. In this method, ions are trapped and then
sequentially ejected into a conventional electron multiplier detector from the ion trap. That way
all ions can be stored during the process of mass analysis. Recent findings showed that using 1
mtorr of helium gas in the trapping volume substantially improved the resolution of the
instrument with the kinetic energy of the ions reducing and the ion trajectories contracted to the
center of the trap. A packet can be formed with a given m/z ions. This spectrometer is used
widely for commercial purposes because of the high resolution and it is inexpensive.

Linear Quadrupole Mass Spectrometer

In the linear ion trap mass spectrometer the ions are trapped within a set of quadrupole rods to
hold the ions radially and end electrodes to maintain the ions axially with a static electrical
potential. It is simply said that unlike Quadrupole Ion Trap, which uses a 3-D field, Linear
Quadrupole Ion Trap uses a 2-D field. It has a selective mass filter that detects the ions of
particular m/z ratio. This method has advantages like higher ion storage capacity and faster
scanning technique.
Page 5 of 10

Time of Flight Mass Analyzers

The TOF (Time of Flight) is a mass analyzer that allows ions to flow down a field free region;
which allows the ions with a greater velocity, lighter ions, to hit the detector first. This is
especially compatible with MALDI due to the fact that the TOF needs a pulsed instrument for
its source. In this way ions are generated in the MALDI source and held there for a brief time
and all are pulsed into the TOF at the same exact time. In this way, If all ions have the same
kinetic energy, the ions with the lower mass will have a higher velocity and reach the detector
first; whereas the ions with the higher mass will have a lower velocity and hit the detector last.

The kinetic energy of an ion leaving a source if given by:

Where velocity v is defined by the Length of the path L divided by time t.

By substituting this equation into the first and solving for time you arrive at.

From this equation you can easily see how mass directly affects travel time. Mass is directly
proportional to time. Using the m/e portion of the equation you can clearly see how a larger
mass means a longer time, and likewise how a lower mass would mean a smaller m/e and thus
shorter travel time.

Ion Cyclotron Resonance Spectrometer

ICR is a form of trapped Ion mass analyzer, that specifically is defined as a static trap. It is
basically a box with three parallel metal sides. Trapped ion analyzers work by keeping the ions
in the trap and controlling the ions by using positive and negatively charged electrical fields in
a carefully series of timed events.

ICR specifically works on the principle that in a magnetic field, ions move in a circular path
whose frequency is mass dependent. So using the cyclotron frequency you can surmise the
mass. Equating the Lorentz force in a magnetic field to the equation for centripetal force yields.
Page 6 of 10

You can then easily solve the above equation for the frequency F or

Groups of ions with the same mass to charge ratios will have the same cyclotron frequency but
may be moving out of synch with one another, this is why an excitation pulse is needed to bring
the resonant ions into phase with one another and the excitation pulse. Next the ions which will
be passing close to the ICR cell receiver plates cause "image currents" which can be collected
and amplified and analyzed. This signal registered in the receiver plates depends both on the
number of ions and their distances from the receiver plates.

Fourier transformation

This mass spectrometry includes a mass analyzer that uses cyclotron frequency of the ions to
determine their m/z ratio. Here a magnetic field with electric trapping plates known as penning
trap is used to hold the ions. An oscillating electric field with the magnetic field at right angle to
it is used to excite the trapped ions to a larger cyclotron radius. This results in packets of
oscillating ions. The trapping plates detects the signal as a image current and results in a
interferogram with sine waves called free induction decay (FID). By performing a Fourier
transform to this data, a mass spectrum can be generated with useful signal.

Detector
Those ions which pass through analyzer are now separated by the desired methods. This mass
spectrometry component records the charge induced by an ion passing by a surface or current
produced when an ion hits a surface. From these charges or currents, a mass spectrum can be
produced as well as measure the total number of ions at each each m/z which are present. Due
to the fact that the number of ions entering the detector at any given moment is minuscule,
signal amplification is often necessary.

Next section: Types of Mass Spectrometry

References
■ Experimental Techniques: MALDI TOF,
http://www.microbiology.science.ru.nl/tech/malditof/
■ An Introduction to Mass Spectrometry,
http://www.astbury.leeds.ac.uk/facil/MStut/mstutorial.htm
Page 7 of 10

■ Electron Ionization Sources: The Basics,


http://www.spectroscopymag.com/spectroscopy/article/articleDetail.jsp?id=358670
■ What other techniques are used to produce ions?, http://www.asms.org/whatisms/p11.html
■ Magnetic Sector Mass Spectrometry,
http://www.wooster.edu/Chemistry/analytical/ms/default.html
■ http://en.wikipedia.org/wiki/Orbitrap
■ The Whys and Wherefores of Quadrupole Ion Trap Mass Spectrometry,
http://www.abrf.org/ABRFNews/1996/September1996/sep96iontrap.html
■ http://en.wikipedia.org/wiki/Quadrupole_ion_trap
■ Fourier Transform Ion Cyclotron Resonance theory: a brief review,
http://www.emsl.pnl.gov/docs/msd/mass_spec/home/fticrtut.html
■ Fourier Transform Mass Spectrometry, http://www.bumc.bu.edu/Dept/Content.aspx?
DepartmentID=385&PageID=7205

Articles Summarized
Principles and Applications of Liquid Chromatography-Mass Spectrometry
in Clinical Biochemistry

Pitt JJ. The Clinical Biochemist Reviews Volume:3019-34 (2009)


(http://www.pubmedcentral.nih.gov/articlerender.fcgi?artid=2643089)

Main Focus

One benefit of Liquid Chromatography - Mass Spectrometry is there are different


configurations to allow for results more geared toward the experiment.

Summary

LC-MS was slow to take off due to


inefficient technologies. However, in
the mid 1990’s, after new technologies
were developed, it became more
popular. This was in part due to its
high specificity and ability to handle
more complex mixtures than the other
options available at the time, such as
GC-MS. Mass spectrometry starts by
ionizing samples to generate charged
molecule fragments then analyzing
their mass to charge-ratio. There are different technologies used to perform the mass
spectrometry part of LC-MS to obtain different results, some being more versatile than others,
such as changing the ion source, changing the mass analyzer, changing the ion suppression as
well as changing the direct injection method.
Page 8 of 10

Different ion sources that can be used clinically would be an electrospray ionization source, an
atmospheric pressure chemical ionization source or an atmospheric pressure photo-ionization
source. Electrospray ionization, also known as ESI, works well on polar molecules;
metabolites, xenobiotics and peptides are some examples. Atmospheric pressure chemical
ionization source, also known as APCI, is great analyzer for small thermally stable molecules
and neutral non-polar molecules such as free steroids. Atmospheric pressure photo-ionization,
also known as APPI, works well with neutral compounds such as steroids.

There are four different types of mass analyzers that can be used; these are the quadrupole
analyzers, time-of-flight analyzers, ion trap analyzers, and hybrid analyzers. Quadrupole
analyzers are are widely used because their ease of scanning and good quality quantitative data.
Time-of-flight analyzers are used for small molecules because of its high sensitivity. Ion trap
analyzers have the ability to fragment an ionize ions several times giving so-called MSn
capabilities. Hybrid analyzers can be switched between ion trap mode and conventional
quadrupole mode.

There are also different steps that can be changed in the Liquid Chromatography part of LC-MS
to make the procedure more versatile such as changing the flow rate, the mobile phase, the
resolution and through-put and the quanitation (calibration). The parameters and conditions
involved in LC-MS can be changed to optimize the assay. However, these conditions are
specific to the particular analyte and the LC separation. Therefore there are no general
conditions. Some of these conditions include using a dilute solution of the analyte and using
single MS or tandem MS. It is also important to make sure the LC-MS system is working
properly with protocols to detect deviations from normal performance. LC-MS is used in the
Biochemical screening for genetic disorders as well as therapeutic drug monitoring and steroid
hormones.

New Terms

GC-MS
Gas chromatography-mass spectrometry (GC-MS) is a method that combines the features
of gas-liquid chromatography and mass spectrometry to identify different substances
within a test sample. ( http://en.wikipedia.org/wiki/GC-MS)
Ion source
A device in which gas ions are produced, focused, accelerated, and emitted as a narrow
beam. Also known as ion gun; ionization source. ( http://www.answers.com/topic/ion-
source )
Calibration
a set of graduations to indicate values or positions —usually used in plural
( http://www.merriam-webster.com/dictionary/calibration )
Deviation
noticeable or marked departure from accepted norms of behavior ( http://www.merriam-
webster.com/dictionary/deviation )
Xenobiotics
Page 9 of 10

a chemical compound (as a drug, pesticide, or carcinogen) that is foreign to a living


organism ( http://www.merriam-webster.com/dictionary/xenobiotics )

Course Relevance

How is this applicable to the proteomics class?

It is important to be able to analyze what makes up a protein so that we can understand the
affects of that protein on an organism. Being able to see the differences between a given protein
and its normal counterpart could help in the understanding of genetic diseases. LC-MS gives
scientists the ability do protein separation.

Websites Summarized
What is Mass Spectrometry?

Chiu CM, Muddiman DC. http://www.asms.org/whatisms/index.html (3/28/09)

Main Focus

This website discusses what mass spectrometry is, the history behind it, and how it works.

Summary

Mass spectrometry is a powerful analytical technique that is used to identify unknown


compounds, to quantify known compounds, and to elucidate the structure and chemical
properties of molecules. Compounds can be identified at very low concentrations. Mass
spectrometry is used by a wide range of professionals such as physicians, astonomers, and
biologists. For example it can monitor the breath of patients by anesthesiologists during surgery
and determine the composition of molecular species found in space. It can be used to identify
structures of biomolecules as well sequence them. It is also used to identify and quantitate
compounds of complex organic mixtures. Mass spectrometry was invented by J.J. Thomson in
a vacuum tube. His invention was used to discover a number of isotopes, to determine the
relative abundance of the isotopes, and to measure their "exact masses" which are important in
the foundation for later developments in diverse fields ranging from geochronology to
biochemical research.

A mass spectrometer is an instrument that measures the masses of individual molecules that
have been converted into ions by becoming electrically charged. A mass spectrometer does not
actually measure the molecular mass directly, but rather the mass-to-charge ratio of the ions
formed from the molecules. The charge on an ion is denoted by the integer number of the
fundamental unit of charge, and the mass-to-charge ratio. It represents daltons per fundamental
unit of charge. Results are in a generated mass spectrum. A mass spectrum is a graph of ion
intensity as a function of mass-to-charge ratio.
Page 10 of 10

New Terms

dalton
a unit of mass for expressing masses of atoms, molecules, or nuclear particles equal to
1⁄12 of the atomic mass of the most abundant carbon isotope ( http://www.merriam-
webster.com/medical/dalton )
corona
a faint glow adjacent to the surface of an electrical conductor at high voltage
( http://www.merriam-webster.com/dictionary/corona )
mass spectrum
the spectrum of a stream of gaseous ions separated according to their differing mass and
charge ( http://www.merriam-webster.com/dictionary/mass%20spectrum )
elucidate
to make lucid especially by explanation or analysis ( http://www.merriam-
webster.com/dictionary/elucidate )
isotopes
any of two or more species of atoms of a chemical element with the same atomic number
and nearly identical chemical behavior but with differing atomic mass or mass number
and different physical properties ( http://www.merriam-webster.com/dictionary/isotopes )

Course Relevance

How is this applicable to the proteomics class?

Proteomics is the study of the protein and is defined as the qualitative and quantitative
comparison of proteomes under different conditions to further unravel biological processes.
Mass spectrometry allows for the study of the components of an organism and how they create
their individual functions.

Retrieved from "http://en.wikibooks.org/wiki/Proteomics/Protein_Identification_-


_Mass_Spectrometry/Instrumentation"

■ This page was last modified on 20 July 2009, at 09:32.


■ Text is available under the Creative Commons Attribution-ShareAlike License; additional
terms may apply. See Terms of Use for details.