Вы находитесь на странице: 1из 9

Journal of Pharmaceutical & Biomedical Analysis, Vol. 13. No. 2. pp. 89 97.

1995
Copyright (~) 1995 Elsevier Science Ltd
Pergamon Printed in Great Britain. All rights reserved
0731-7085/95 $9.50 + 0.00
0731-7085(94)00106-5

Application issues in bioanalytical method validation,


sample analysis and data reporting
D A R I O U S H D A D G A R , * PAMELA E. BURNETT, M. G E R R Y CHOC,? KEITH
GALLICANO:~ and JOHN W. H O O P E R

Analytical Department, Phoenix International Life Sciences, St Laurent (Montreal), Quebec, Canada
+Bioanalytical Department, Sandoz Pharmaceuticals Corporation, East Hanover, N J, USA
Drugs Directorate, Health Protection Branch, Bureau of Drug Research, Health Canada, Ottawa, Canada

Abstract: Although some degree of consensus has been reached concerning the requirements for acceptable method
validation, the procedures used to establish them vary significantly between laboratories. Also, issues arising from
application of these requirements during validation and subsequent sample analysis need to be addressed. The purpose of
this paper is to discuss application issues concerning prerequisites to method validation, and all validation criteria for
evaluation of method reliability and overall performance. Other poorly addressed issues such as re-validation, cross-
validation, partial sample volume, multicomponent analysis and reporting will also be discussed. Although many issues
discussed are of a general nature, the scope of this presentation is primarily to address issues arising from the validation
and routine application of chromatographic methods.

Keywords: Bioanalytical; chromatography; validation; stability; biofluids.

Introduction cation of these requirements during validation


and in subsequent sample analysis, need to be
The validation of bioanalytical methods has addressed. A round table discussion at the
been the subject of discussion in recent con- recent AAPS meeting (15-19 November 1992
ferences and papers [1-8]. These have in- San Antonio, Texas) partly addressed some of
tended to provide guiding principles for valid- these issues.
ation of analytical methods used in bioavail- The purpose of this paper is to summarize
ability, bioequivalence and pharmacokinetic broadly based views on application issues in
studies in humans and animals. It has been bioanalytical method validation, sample
generally agreed that the key criteria for analysis and data reporting from scientists
evaluation of method reliability and overall engaged in bioanalytical analysis and its regu-
performance are: lation. This paper will discuss application
(i) analyte stability; problems and potential solutions to these
(ii) method selectivity; problems. Although many issues discussed are
(iii) limit of quantitation; of a general nature, the scope of this paper is
(iv) accuracy; primarily to address issues arising from the
(v) precision; validation and routine application of
(vi) relationship between response and chromatographic methods.
concentration (e.g. linearity);
(vii) recovery; and
(viii) ruggedness.
Prerequisites to Method Validation
Although some degree of consensus has
been reached concerning the requirements for The first prerequisite for method validation
acceptable method validation, the procedures is a developed analytical method. The purpose
used to establish them vary significantly among of method validation is to establish that an
laboratories. Also, issues arising from appli- accurate, precise and rugged method has been
* Author to whom correspondence should be addressed.
Note: The contents of this paper represent the personal views of the authors and do not necessarily reflect the policies
of the institutions with which they are affiliated,

89
90 DARIOUSH DADGARet al.

developed. Criteria for evaluation of validation taken, for use in demonstrating selectivity. In
data should be established a priori. the absence of reference samples for meta-
However, often validation is begun on a bolites or degradation products, either or both
"promising" method. If validation criteria are of the following experiments may sufficiently
not met, further development work ensues and demonstrate selectivity.
the method is then revalidated. In order to (i) Use of biological samples from dosed
break the vicious circle of development work subjects may be the best solution (e.g. from a
during validation, it is advisable to establish pilot study). These samples should be analysed
that the following parameters are acceptable using the usual chromatographic conditions,
prior to initiating the aspects of validation and under "stretched" conditions i.e. changing
which require analysis of large numbers of chromatographic conditions to resolve as many
samples (for example, intra-batch and inter- potentially merged peaks as possible.
batch precision and accuracy). Examination of chromatograms from subject
samples collected at various times following
Selectivity the dose can reveal related peaks due to
It is recommended that the selectivity of the substances that are absent in the pre-dose
method should be established with respect to sample and increase in response following the
endogenous substances, metabolite(s) and dose, and subsequently decrease.
known degradation products. Interferences (ii) If concentrations are high enough, and
which are likely to be present in small quan- the UV spectra of potential metabolites/
tities may adversely affect the quantitation of degradation products are different from that of
unknown samples at concentrations approach- the parent drug, diode array or other multiple
ing the limit of quantitation (LOQ). wavelength detectors may be used to ensure
Selectivity with respect to interferences from peak purity.
endogenous substances in biological fluids can (iii) In the absence of subject samples from
usually be established by processing a a pilot study, just before the start of analysis of
minimum of six independent sources of the study samples, a few samples at the estimated
same blank matrix. Careful examination of tmax s for parent drug, metabolites and degrad-
chromatograms across the time windows of ation products can be analysed, using only a
peaks of interest is required to evaluate selec- portion of the available sample volume, as
tivity. Here it should be emphasized that it is described above for pilot study samples. This
not appropriate to test only one source of will ensure that study samples will not be
blank [7]. Although it would be preferable that wasted should interference from metabolites/
all tested blanks, if obtained under controlled degradation products be found.
conditions, be free from interference, factors (iv) If analysis of samples from dosed sub-
such as subject food and beverage intake, use jects is not feasible, a number of the potential
of vitamin supplements, use of over-the- metabolites could be produced by in vitro
counter and prescription drugs other than the incubation with liver homogenate, and result-
one(s) being assayed, and cigarette smoking ing metabolites analysed using the method's
can affect selectivity. Therefore, even blanks chromatographic conditions.
obtained under "controlled conditions" may (v) To account for potential degradation
not ensure inter- and intra-subject uniformity products produced by acid hydrolysis in the
both during selectivity testing and during the stomach, in vitro incubations in acidic media
study. Due to factors discussed above, and/or may be performed.
method artifacts, zero per cent interference (vi) Anticipating interferences from con-
may not always be achievable, and in such comitant medications and their metabolites
cases, minor interferences may be allowable may not always be feasible. Method modifi-
provided that the pre-defined precision and cations are usually required if such problems
accuracy criteria for quantification at the LOQ arise during analysis of the study samples, and
are respected. may necessitate revalidation, the extent of
Real problems arise when analyte meta- which is described later in this paper.
bolites or known degradation products are not (vii) Last, and by no means least, is to use a
available. Under these circumstances, if poss- more specific detector such as the mass
ible, custom synthesis of metabolites and spectrometer. The new generation of
known degradation products should be under- atmospheric pressure HPLC-MS, MS-MS and
B I O A N A L Y T I C A L M E T H O D VALIDATION, SAMPLE ANALYSIS AND D A T A REPORTING 91

sensitive GC-MS instruments have greatly for injection at time 0, intermediate times and
enhanced the power of analytical chemists to after the expected maximum delay before
produce highly sensitive and specific methods. completion of a batch of samples, e.g. 24, 36,
or 48 h. By pooling the extracted QCs at each
Stability concentration, the starting concentration for
To ensure that compound integrity is main- all aliquots is identical, and the comparison of
tained throughout the workup process, certain determined concentrations following storage to
stability tests should be performed during the the time 0 concentration is independent of the
final stages of method development. extraction reproducibility.

Ana(vte stability. Published literature should Ruggedness


be investigated, or laboratory tests should be If it is discovered late in the validation
conducted to determine whether pure analyte process that a method is not easily transferred
and/or solutions of the analyte (e.g. drug, between systems, analysts, or analytical
metabolite(s) and internal standard) are stable columns of the same type, further development
under normal laboratory conditions of heat, and revalidation becomes necessary. To
humidity, light and air exposure. minimize the chances of this occurring, the
following should be investigated during the
In process stability. It must be demonstrated latter stages of development.
that drugs remain intact if left for several hours
at room temperature in the biological matrix. Column effect. Analyses should be per-
Certain analytes, e.g. captopril, acetyl salicylic formed on at least two columns containing two
acid, etc. undergo immediate changes/degrad- different lots of the identical packing material,
ation in the biological matrix. In such cases, preferably one column which was used for
appropriate additives, e.g. enzyme inhibitors, method development and the other previously
anti-oxidants and/or derivatizing agent may be unused.
added. Reduction of the collecting device
temperature might improve stability. It may be Mobile phase effect. For HPLC methods, the
necessary to extract samples immediately after effect of small variations in the mobile phase
collection, and store the dried frozen residue solvent ratio (i.e. less than 2% of the amount
for later reconstitution and analysis. If this of each component), and buffer pH were
procedure is used, the calibration standards applicable, should be examined and reported.
and QC samples must be treated in the same For GC methods, other parameters such as
manner. small variations in oven temperature and gas
It should also be noted that some matrices flow rates should be examined.
encountered at intermediate stages of sample Assessment of the above parameters should
workup such as buffered biological matrix and provide an indication of the method's ability to
the back-extraction medium may affect the maintain critical separations when faced with
stability. Again, if instability of analytes is expected day to day mobile phase variations,
observed at any of, these stages, appropriate and column to column variability.
additives should be considered, or processing
temperature reduced. Ruggedness acceptability. Peak shapes and
resolution from other peaks in the matrix must
Processed sample stability. Reconstituted remain visually acceptable. The limit of quan-
extracted samples must remain stable in the titation must demonstrate a reproducible
reconstitution solvent at the temperature of the response readily distinguishable from the noise
auto-injector until they are injected, as well as level.
at other lower temperature(s), e.g. 4°C, to
allow for storage of prepared samples that Internal standard (IS)
cannot be immediately chromatographed Many bioanalytical methods are based on
owing to instrument problems. Processed the internal standard quantitation method.
sample stability should be examined in repli- Internal standard is added at the earliest stage
cate at each QC concentration. After recon- of sample preparation to minimize error by
stitution, replicate samples of each QC concen- compensation. It is commonly believed that a
tration should be pooled, mixed and aliquoted good IS should be structurally similar to the
92 DARIOUSH DADGAR et al.

analyte under analysis. Although the above light of the above, inclusion/exclusion of points
statement is generally true, there are occasions in the standard curve should be established a
when it is not so. Such was the case in liquid- priori, and the following provides possible
liquid extraction method followed by HPLC guidelines for this.
analysis of the tricyclic antidepressant doxepin, Provided the calibration curve consists of at
and its metabolite, nordoxepin (Dadgar, un- least seven non-zero single standards, up to
published data). The structurally similar drug, two non-zero standrds may be removed from
imipramine did not minimize error by compen- the calibration curve if at least one of the
sation for nordoxepin, whereas it successfully following valid reasons exists and a minimum
did so for doxepin in spiked samples. This is of five non-zero standards remain in the curve.
because parent and metabolite differed from (a) loss of sensivitity;
each other in their extractabilities under the (b) poor chromatography;
extraction conditions used. Under this circum- (c) loss during sample processing;
stance, it was necessary to use two internal (d) if, when included in the calibration curve,
standards, imipramine as IS for doxepin, and it clearly biases the QC results, and the
desipramine as IS for nordoxepin. back-calculated standard concentration
Two other important criteria for choosing an deviates substantially from its nominal
internal standard are: value.
(i) it should elute close to the retention time In many bioanalytical laboratories, the cali-
of the analyte of interest. Therefore, for bration curves and QC samples are prepared
some multicomponent analyses two or simultaneously and frozen for storage,
more internal standards are needed; normally at the same temperature as is in-
(ii) it should normally elute after the parent tended for storage of biological samples
drug peak (where reversed-phase derived from clinical/toxicological samples.
chromatography is used) so that the However, some laboratories prefer to prepare
possibility of interference with faster elut- calibration curves fresh with each batch of
ing more polar metabolites is obviated. samples analysed. The advantage of preparing
and freezing calibration curves is that the effect
of time and possible degradation of analyte(s)
Method Validation
is the same for the calibration samples and the
In this section, commonly encountered study samples.
application issues in bioanalytical method In order to generate an accurate "analytical"
validation are discussed under the heading of calibration curve independent of possible time
each validation criterion. effect, the calibration curve may be prepared
fresh with each batch of samples while QC
Calibration c u r v e samples are prepared and stored frozen with
The joint conference report [1] states that study samples in order to account for the "time
batch acceptance should be based on QC effect". A simple approach to using fresh
acceptance criteria. However, bioanalytical calibration curves is to prepare a series of
laboratories react differently and somewhat working calibration standards at concen-
arbitrarily regarding inclusion or exclusion of a trations 10 or 20 times greater than those
standard point deviating greatly from the cali- intended for biological standards, in a suitable
bration curve. Some leave it in the curve and dilution solution such as i:1 methanol-water.
some drop it. As it happens, there are times These working calibration standards may be
when, for example, the LOQ standard may stored provided stability is previously demon-
deviate by, say 40%, but the next point in the strated over the maximum period over which
calibration curve demonstrates a minor devi- they will be stored. Then, on a daily basis,
ation from nominal. However, dropping the blank biological matrix is spiked with the
LOQ standard causes the next point to deviate working calibration standards in a ratio of, e.g.
greatly from the nominal value. 1:20 working standard:biological blank.
Another example of a commonly encoun- Dilution of biological matrix with working
tered problem is where removal of a given standards will be compensated for by adding an
standard causes a QC to fail to meet its equal volume of the working standard dilution
acceptance criteria when the QC would other- solution (free of analyte(s)) to the study
wise have been within the acceptable range. In samples.
B I O A N A L Y T I C A L M E T H O D VALIDATION, SAMPLE ANALYSIS AND D A T A REPORTING 93

Number of calibration curves accuracy and precision are met with the larger
The question of how many calibration curves sample volume.
to run with each batch should be answered by QC samples can be prepared at, e.g. 1/2 and
consideration of whether or not study samples 1/4 the L O Q for replicate extraction of double
are run, for example, singly, in duplicate or in and quadruple the usual extraction volume,
triplicate. If study samples are run singly, then respectively. These L O Q QCs are analysed
a single calibration curve should normally be against the calibration curve extracted using
used. If replicates of study samples are the usual volume. The usual precision and
analysed, then identical replication of standard accuracy criteria should be met. In addition,
curves is desirable. selectivity tests should be conducted using the
In order not to mask the accuracy of the increased blank urine volumes.
method, calibration curves should be pro-
cessed identically during validation and during Limit of quantimtion
study sample analysis. The limit of quantitation (LOQ) must be
differentiated from the limit of detection
(LOD). The value of the L O D is the smallest
Validation of partial or increased sample concentration that can be distinguished from
volume the background noise. The limit of detection
A common difficulty arises during biological can be defined in different ways. However, no
sample analysis when less than the validated matter how the limit of detection is defined,
sample volume is available, and a partial the limit of quantitation should be at least
volume must be used for analysis. This pro- twice the response of the L O D , and within the
cedure needs to be validated. In addition, pre-defined accuracy and precision boun-
samples analysed and found to be above the daries, normally within _+20% of nominal with
calibration curve range require dilution for a RSD -<20%.
reanalysis. It is important that the accuracy and pre-
A simple form of validation for partial cision of the L O Q be obtained using L O Q QC
sample volume is to use 1/4, 1/3 and 1/2 of samples, i.e. independently from the cali-
replicate volumes of QC samples of appro- bration curve, because the L O Q standard
priate concentrations. QC concentrations are which is used to construct the calibration curve
chosen both above the calibration curve range influences the regression equation. In this
to demonstrate accurate dilution to within the manner, the precision and accuracy of the
range, and at concentrations such that once L O Q can be defined in both inter- and intra-
diluted, they are near, but not below the LOQ. batch tests. The practice of pushing the L O Q
The partial sample volume is brought up to the to the limits of the L O D in order to compete
validated volume by addition of blank matrix, with other more sensitive methods should be
and analysed. The calculated concentrations of avoided.
diluted QCs, when multiplied by their
respective dilution factors must fall within the Stability
defined precision and accuracy criteria for that Data obtained from analysis of study
QC. samples are suspect if not accompanied by
Occasionally, it may be desirable to increase supporting data assuring the stability of the
the volume of sample matrix, compared with drug and metabolites in the matrix. It is
that of the calibration standards, so that the important to emphasize that stability tests must
instrument response is within the standard be designed in such a way as to detect any
curve range. For example, when analysing degradation, over the maximum period of time
urine, it may be necessary to use a larger that study samples will be stored prior to
volume to obtain reliable concentrations owing analysis. Often, stability data are based on
to a large void volume. The interpolated duplicate or triplicate determinations of the
concentration, corrected for volume, can be concentrations of high and low QC samples, at
below the L O Q concentration established for multiple time points after the start of storage to
the assay, provided that the response is above allow "trends" to be detected. However, the
that of the L O Q standard. Validation is issue is not whether there is a trend in
necessary to show that selectivity is not com- degradation, but whether the study samples
promised and the pre-defined criteria for with the longest storage time are adequately
94 DARIOUSH DADGARet al.

preserved at the time of analysis. For example, in storing study samples (typically -20°C;
in order to demonstrate 12 month stability at the stability samples), and the second in
-20°C, the stored samples should be analysed liquid nitrogen (nominally at -<-196°C),
when freshly collected or prepared, at inter- or in another suitable freezer (colder than
mediate times (e.g. 3 and 6 months following -130°C). The latter are the reference
preparation) and after 12 months of storage. samples. After storage for an appropriate
To show processed sample stability, for time, replicates of both the stability and
example over 36 h (as discussed previously), the reference samples are analysed on the
samples should be analysed at time 0, at same standard curve and the results
intermediate times such as 12 and 24 h, and at compared.
36 h. At very cold temperatures (--130°C)
If stability testing is conducted on QC stability is assured even for unstable drugs.
samples (as opposed to samples from dosed This has been demonstrated using theor-
subjects) it is recommended to directly com- etical calculations based on the Arrhenius
pare responses of stored and fresh QCs equation, which show that at -130°C, (the
analysed at the same time, or to compare temperature in the vapour above liquid
interpolated values provided the values are nitrogen was used to be conservative)
interpolated from the same calibration curve. reaction rates are a factor of approximately
Two types of essential stability study have 103 lower than at -20°C for reactions with
already been discussed. Here, stability of the very low activation energies (5 kcal
analytes under prolonged storage conditions mole l), and factors of 106-1013 lower for
and freeze-thaw will be discussed. a.ctivation energies in the usual range (10-
20 kcal mole-I). We have confirmed this
L o n g term stability. The stability and by evaluating the degradation of acetylsali-
tendency for adsorption to the storage con- cylic acid (ASA) in human plasma at
tainer should be assessed for all analytes in the -20°C and in liquid nitrogen. A S A is
biological matrix using the exact type of con- known to be one of the most unstable
tainer (e.g. glass, polypropylene) to be used drugs in plasma. While the drug was
for study sample storage. Stability must be degraded approximately 60% over 4
proven over at least the maximum period of months at -20°C, no degradation was
storage of study samples, under the tempera- detected after storage in liquid nitrogen for
ture conditions to be used for study samples. the same period. Further, calculations
Ideally, the control samples used for evalu- based on the ASA reaction rate constant
ation of long term stability, should be those obtained from the -20°C data, and assum-
obtained from dosed patients/volunteers, ing an improbably low activation energy of
collected, pooled, aliquoted and stored at the 5 kcal mole -1 (worst case), showed that
same time and in the same way as study calculated ASA degradation at -130°C
samples. This provides maximum assurance of was <0.05% after 4 months. These and
the integrity of all analytes in study sample for other evaluations of this approach will be
a given study. However, for most drugs where published separately.
metabolite reversion to the drug, or ongoing Our studies suggest that the "second"
metabolism in the frozen matrix are not prob- method of assessing long term frozen stability
lems, blank matrix samples spiked at different has superior precision and accuracy to that of
concentrations are an acceptable alternative the traditional method. This is probably
and are usually used. because it eliminates errors resulting from
Two procedures for the assessment of long preparation of different fresh batches at each
term frozen stability are discussed below, the stability time point. It appears that such errors,
first traditional, and the second developed when combined with assay variation, can lead
recently by some of us [8]. to erroneous conclusions concerning stability
(1) Replicates of stored and freshly prepared or lack thereof. However, many laboratories
QC samples are analysed and their do not have access to liquid nitrogen equip-
responses compared. ment, and this is a limiting factor in the
(2) A batch of stablility samples is prepared application of this method.
and divided into two sub-batches, the first
stored at the temperature intended for use Freeze-thaw stability. The influence of three
BIOANALYTICAL METHOD VALIDATION, SAMPLE ANALYSIS AND DATA REPORTING 95

freeze-thaw cycles at concentrations represen- effect of the matrix on, for example, peak
tative of the high and low analyte concen- shape does not complicate interpretation of
trations in the matrix should be examined in recovery data. Recovery may be determined
replicates. Data for freeze-thaw stability may by comparison of interpolated concentrations
be obtained in two ways: of extracted vs unextracted QCs provided they
(i) stability samples are frozen and thawed are from the same calibration curve.
three times and then analysed. This
method is less time consuming, however, if
samples are proven unstable, then exper- Revalidation
iment design, (ii) should be performed.
When it is necessary to make changes to
(ii) Quality control samples are analysed with-
chromatographic conditions, or the sample
out being frozen at first, and then after
processing procedure of an analytical method,
each cycle of freeze-thaw so that a trend,
revatidation may be necessary. The decision
if it exists can be seen. If it was proven, for
regarding which parameters require revalid-
instance, that QC samples are only stable
ation should be based on logical consideration
after two cycles of freeze-thaw, then study
of the specific validation parameters likely to
samples should not be frozen and thawed
be affected by the change.
more than twice.
For example, changes to extraction or back-
extraction media may be expected to affect
Recovery
selectivity, recovery, precision and accuracy,
High recovery of analyte(s) from the matrix
without affecting freeze/thaw stability in the
is desirable. However, sometimes it may be
biological matrix. A change to the analytical
necessary to intentionally sacrifice high
column or mobile phase may be expected to
recovery in order to achieve better selectivity,
affect linearity and selectivity without affecting
and this is acceptable provided that adequate
recovery. A guideline to revalidation is
sensitivity, precision and accuracy are
presented in Table 1.
achieved. Solvents such as ethyl acetate
normally give rise to high recovery of analyte,
however this solvent simultaneously extracts
Cross-validation
many interfering compounds, therefore,
provided that an adequately sensitive detection Linearity, intra-batch precision and accuracy
limit is attained with good precision and Cross validation refers to applying a valid-
accuracy, the extent of recovery should not be ated method in a given biological matrix to the
considered an issue in bioanalytical method same type of matrix from another species, or to
development and validation. a similar matrix (e.g. plasma and serum) from
Recovery is best tested by directly compar- the same species, or to the same matrix with a
ing responses of replicates of extracted QC change in anticoagulant.
samples with replicates of extracted blank Cross validation may be carried out as
matrix to which analyte has been added at the follows: a calibration curve is prepared in the
same nominal concentration. In this way, any validated matrix; replicates at all QC concen-

Table 1
Guidelines for revalidation

Method parameter changed Parameters to revalidate

Extraction solvent, buffer, back extraction matrix Linearity, recovery, selectivity, LOQ, intra-batch precision and
or injection solvent accuracy, in-process stability. Additionally, if injection solvent is
changed, processed sample stability, but no recovery or in-process
stability.

Chromatographic column, mobile phase compo- Linearity, selectivity, intra-batch precision and accuracy for the
sition (e.g. significant change in retention times), LOQ and other QCs
detector type or wavelength

Extending the upper end, or reducing the lower end Linearity, LOQ (if reduced), intra-batch precision and accuracy
of the calibration curve range at revised upper and lower levels

Internal standard Selectivity, intra-batch precision and accuracy, and recovery


96 DARIOUSH DADGAR et al.

trations including the LOQ are prepared in 0.8


both the matrix from the validated method, 0.7
and in the matrix to be validated; all QC
0.6
samples are back-calculated from the same
0.5
calibration curve in the validated matrix; the
method is considered cross validated if the e~ 0.4
p
determined concentration of QCs in the matrix 0.3
to be validated satisfy the acceptance criteria. 0..2

0.1
Stability
All types of stability studies previously dis- 10 11 12 13 14 15 16 17 18 19 20
cussed (long term, freeze-thaw, in-process, Assay R S D (%)
processed sample) should be evaluated and o One + Two O Three
reported under cross-validation. Figure 1
P r o b a b i l i t y of r e j e c t i n g an a n a l y t i c a l b a t c h w h e n m e a s u r -
Selectivity ing o n e , two and t h r e e a n a l y t e s .

Selectivity should be evaluated in the revised


matrix as interferences may differ between
matrices.

methods. For a three analyte assay having a


Analysis of Study Samples and Reporting of
RSD of 15% for each analyte, greater than
Study Data
30% of the runs will have to be repeated on the
Multianalyte methods basis of chance alone. This analysis assumes
A recent conference report [1] recommends that variability is constant across analyte con-
batch acceptance criteria defined as follows: centration levels; any increase in variability at
duplicate QCs at three concentration levels the concentration of the low control would
(low, intermediate, high) with four of six further increase the probability of run
calculated to be within +20% of their rejection.
respective nominal values (no two at the same In multicomponent methods, it is recom-
concentration level may be outside +20%). mended that each component be treated
This approach is now generally accepted. In independently with regard to acceptance or
practice, using an analytical method character- rejection, and reanalysis. Data for previously
ized by acceptable inter-batch precision accepted components that is generated simul-
(-<15% RSD) results in minimal rejection of taneously with a repeat analysis for a failed
analytical runs on the basis of chance alone. component should not be evaluated or
However, rejection of a substantial proportion reported.
of runs is possible when using some multi-
analyte procedures. Acceptance of a truncated calibration curve
In a multianalyte procedure where all Sometimes, during analysis of study
analytes are measured in the same processed samples, the full concentration range of the
sample chromatogram, failure to meet QC calibration curve may not be available. If the
acceptance criteria for any one analyte triggers low end of the curve is not available, an
a repeat analysis of the subject samples in "elevated" LOQ of concentration equal to the
order to generate reportable data for that lowest acceptable calibration standard or QC
analyte. The repeat analysis also potentially may be used. All study samples with analyte
generates additional reportable data for the concentrations falling below the elevated LOQ
other analyte(s). This "compounding" of the must be reanalysed.
acceptance criteria increases the probability of When the upper portion of the calibration
overall run rejection. Treating each control curve is not available, a truncated range should
value independently, the probability of run be used, by setting the revised upper end of the
rejection can be calculated for single and calibration range equal to the highest accept-
multianalyte methods. Figure 1 shows the able standard or QC sample. All study samples
resulting probability of run rejection vs assay with concentrations determined to be above
%RSD for one, two and three analyte the revised range must be reanalysed.
BIOANALYTICAL METHOD VALIDATION, SAMPLE ANALYSIS AND DATA REPORTING 97

Reassay o f samples between, for example, the mean of duplicate


In most studies, some samples will require reanalyses, and the median value of the
reassay. Criteria for identifying these samples, original and reanalyses. It is helpful to use a
their analysis and reporting should be estab- flow chart to define the decision making
lished a priori. Reasons for reanalysing bio- process used to report results from multiple
analytical samples can be summarized as determinations, given different scenarios of
follows: original and repeat sample concentrations
found.
Results are analytically unacceptable. These
refer to samples with unreportable concen-
Conclusion
trations due to:
(i) equipment failure; It appears that bioanalytical validation pro-
(ii) p o o r chromatography; cedures and acceptance criteria will continue to
(iii) loss during sample processing; evolve. It is likely, and desirable, that future
(iv) samples with concentrations below an changes will be based more on statistical
elevated L O Q ; considerations and the calculated impact of
(v) samples with concentrations above the various criteria on the end use of the data,
accepted range of the calibration rather than on "consensus" of leaders in the
curve; field. The authors of this p a p e r are conducting
(vi) pre-dose samples with observed con- research in this area, to better link the impact
centrations needing confirmation; of validation requirements and acceptance
(vii) processing errors, such as incorrect criteria on the determination of bioequival-
addition of internal standard or other ence and other pharmacokinetic evaluations.
reagents;
(viii) rejected batches;
(ix) low injection volume (and thus low References
response). [1] V.P. Shah, K.K. Midha, S. Dighe, l.J. McGilveray,
Samples in these categories should be re- J.P. Skelly, A. Yacobi, T. Layloff, C.T. Viswanathan,
analysed singly, and provided that the batch C.E. Cook, R.D. McDowall, K.A. Pittman and S.
Spector, J. Pharm. Sci. 81,309-312 (1992).
meets acceptance criteria, the value from the [2] D. Dadgar and M.R. Smith, Trend Anal. ('hem. g,
reanalysis in each case is reported. 115-117 (1986).
[3] V.P. Shah, Clin. Res. Pract. Drug. Reg. Affairs 5, 51-
60 (1987).
Pharmacokinetic outliers. It is debatable [4] A.R. Buick, M.V. Doig, S.C. Jeal, G.S. Land and
whether a sample with a concentration in- R.D. McDowall, J. Pharm. Biomed. Anal. 8,629-637
congruous with the pharmacokinetic profile (199(I).
[,5] A.G. Causey, H.M. Hill and L.J. Phillips, J. Pharm.
should be reanalysed. However, if a decision is Biomed. Anal. 8, 625-628 (1990).
made to repeat the analysis, the reanalysis [6] G.P. Carr and J.C. Wahlich, J. Pharm. Biomed. Anal.
should be conducted in duplicate. 8, 613-618 (1990).
[7] H.T. Karnes, G. Shiu and V.P. Shah, Pharm. Res. 8,
Reporting a single result from multiple 421-426 (1991).
determinations requires pre-defined criteria [8] J.W. Hooper, D. Lessard, K. Selinger and D. Dadgar,
for data selection. Such criteria are normally A new, more reliable method for the determination of
the long term frozen stability of drugs in biological
based on the percentage difference of the mean fluids. Eighth International AAPS Conference,
of repeat values from the original value. Based Orlando, Florida (1993).
on the criteria, the decision is made to report [Received for review 28 January 1994;
either the original sample value, or a choice revised manuscript received 20 May 1994]

Вам также может понравиться