Вы находитесь на странице: 1из 4

Clinical Biomechanics 14 (1999) 357±360

Accuracy of centre of pressure measurement using a piezoelectric


force platform
James Middleton a,*, Peter Sinclair b, Ray Patton b
a
Moorong Spinal Unit, Royal Rehabilitation Centre, Ryde, Sydney NSW 1680, Australia
b
School of Exercise and Sport Science, The University of Sydney, Sydney NSW, Australia
Received 29 May 1997; accepted 14 August 1998

Abstract
Objective. To evaluate the accuracy in determining centre of pressure when using a force platform for stabilometry research.
Design. Technical report.
Background. Previous studies have determined the accuracy of locating the centre of pressure by applying force through a single
point. In human stabilometry research however, force is distributed across two feet and the centre of pressure located between the
feet.
Methods. Forces were applied to a Kistler force platform using a force applied through a single point as well as when force was
distributed between two metal blocks.
Results. Errors in calculating centre of pressure were substantially larger when force was applied through a single point than
when the force was distributed between two blocks.
Conclusions. Evaluating centre of pressure accuracy by applying force through a single point overestimates the errors inherent in
human stabilometry research.

Relevance
Evaluating changes in the centre of pressure of human subjects standing on a force platform is a common tool for evaluating
control mechanisms of balance. Accurate information regarding the centre of pressure is essential for the validity of this re-
search. Ó 1999 Elsevier Science Ltd. All rights reserved.

Keywords: Force platform; Stabilometry; Balance; Accuracy; Centre of pressure

1. Introduction When using a force platform for stabilometry, force is


spread over an area of the two feet and the CoP is lo-
Movement of the Centre of Pressure (CoP) with a cated between the feet. Testing CoP accuracy using a
subject standing on a force platform has been used to single point force is convenient, but does not realistically
provide information about postural control in both represent the loading patterns used for stabilometry re-
normal and pathological situations [1,2]. search. The purpose of this study was to evaluate the
Bobbert and Schamhardt [3] evaluated the accuracy accuracy of CoP calculation using a two `foot' model
of determining CoP by applying forces through a known that more closely resembled the clinical situation of an
point of application to a Kistler force platform. Errors individual standing quietly on a force platform.
varied in magnitude up to 20 mm, depending on the
position across the plate. Average errors were 3.5 mm in
the x (short axis) and 6.3 mm in the y (long axis) di-
rection. These errors increased substantially towards the 2. Methods
edge of the force platform, particularly outside the
central region bounded by the four force transducers. A Kistler force platform, type 9281A (Kistler In-
struments AG, Winterthur, Switzerland) was used to
measure ground reaction forces from which CoP data
was calculated using the two di€erent models of force
*
Corresponding author. application. The Kistler force platform was embedded in
0268-0033/99/$ ± see front matter Ó 1999 Elsevier Science Ltd. All rights reserved.
PII: S 0 2 6 8 - 0 0 3 3 ( 9 8 ) 0 0 0 7 9 - 5
358 J. Middleton et al. / Clinical Biomechanics 14 (1999) 357±360

and replaced between each trial. In several positions


along the x and y axes, six trials were collected at each
location. Clear indentations were made by the apparatus
from which Vernier callipers (precision 0.05 mm) were
used after each data collection to measure displacement
of the point of application. Data was collected for 10 s
periods whilst the apparatus applied a static point force
and mean coordinates for the point of application were
calculated from CoP formulas. Coordinates calculated
from the force data (CoPforce ) were then compared with
the measured point of application coordinates
(CoPmeasured ) to assess the accuracy of CoP calculation.

2.2. Stationary `feet' model

Two rectangular model `feet' (illustrated in Fig. 1),


made of steel to the same dimensions (25 cm long and
10 cm wide) and ®lled with lead to approximately
Fig. 1. Model `feet' used to simulate standing and test the accuracy of 300 N, were used to simulate standing and test the ac-
CoP measurement under static loading conditions.
curacy of CoP measurement.
The weight and CoP of each model `foot' were de-
a concrete foundation according to the manufacturer's termined independently by positioning each `foot' as
speci®cations and measured the orthogonal components precisely as possible in the geometric centre of the force
of force along the x, y and z axes as shown in Fig. 1. platform. Six separate trials were collected for each
Raw signals from each transducer were recorded by model `foot' and mean coordinates of the CoP and the
personal computer at a sampling frequency of 50 sam- weight were calculated individually for each `foot'.
ples per second via a 12 bit analogue-to-digital converter With two model `feet' positioned symmetrically about
(Data Transition DT 2801A). the geometric centre of the force platform, the CoP
The point of application of the resultant of the would not be expected to coincide with the force plat-
ground reaction forces was calculated from the orthog- form centre owing to the inevitable di€erence in weight
onal force components using equations from the Kistler and CoP for each model `foot'. Predicted CoP between
instruction manual. the two `feet' was calculated using the following equa-
tion:
2.1. Point of force application model Wtfoot1  CoPfoot1 ‡ Wtfoot2  CoPfoot2
CoPpredicted ˆ
Wtfoot1 ‡ Wtfoot2
An apparatus was constructed to apply a vertical
force of approximately 300 N through a single point of (where Wt represents the weight of each `foot' in New-
application. The pin-point tip of the apparatus was tons and CoP coordinates are in millimetres).
carefully positioned on the force platform at predeter- Trials were collected with the medial edges of the
mined positions. Three trials were collected around each `feet' symmetrically positioned parallel to the x-axis in
position shown in Table 1, with the apparatus removed positions illustrated by Fig. 1. These positions were

Table 1
Mean and standard deviation of the di€erences between CoPmeasured and CoPforce in the x and y directions using the single point of force application
model

Point of application (y-axis) Point of application (x-axis)

0 mm 60 mm 120 mm 150 mm

xerror yerror xerror yerror xerror yerror xerror yerror

0 mm 0.6 (0.6) 2.5 (1.2) 1.7 (0.6) 2.9 (1.7) 4.4 (1.0) 2.8 (2.6) 6.2 (1.0) 5.1 (2.8)
50 mm 0.4 (0.3) 3.3 (1.8) 1.9 (0.8) 4.0 (1.7)
100 mm 0.9 (0.4) 3.1 (1.8) 0.9 (0.6) 3.8 (2.3) 3.9 (1.2) 4.8 (2.7)
150 mm 0.5 (0.2) 2.3 (1.1) 1.3 (0.7) 2.2 (1.6)
200 mm 0.9 (0.6) 3.9 (1.6) 3.1 (1.0) 4.2 (1.7)
250 mm 0.9 (0.7) 11.6 (1.1)
J. Middleton et al. / Clinical Biomechanics 14 (1999) 357±360 359

chosen to represent typical foot positions during stab- Mean error for the two `foot' model was signi®cantly
ilometry research. Alignment was achieved using ®ne less than that for the single point of force application
felt tip markings parallel to the x-axis at the desired (p < 0:001), even though only single point force results
displacements. Six trials were performed at each dis- from the more accurate central region were included in
placement to test reliability. Mean CoPforce coordinates the comparison. If measurements from the whole force
between the `feet' for each individual trial were com- platform were included, these di€erences would be even
pared to the coordinates of the CoPpredicted for each more apparent.
displacement. Comparison between means in di€erent
positions across the force platform was undertaken to
determine whether any systematic error existed. 4. Discussion
Student's t-test was used to compare mean absolute
CoP errors between forces applied by the model `feet' Single point of force results were found to be consis-
with the mean for errors applied through a single point tent with the ®ndings previously described by Bobbert
of force. Only single point trials within the region and Schamhardt [3] with accuracy doubled in the x-axis
bounded by the force transducers were used for this compared with the y-axis direction. The distance be-
comparison for reasons that will be discussed later. tween the force transducer positions on the force plat-
forms used in both experiments were proportionately
greater in the y-axis direction by a ratio of approximately
3. Results 2:1. Calculation of moments therefore results in multi-
plication of errors in force measurements by the same
Measurement of the weight and CoP of each model factor. Consistent with this result was the ®nding that
`foot' was repeatable within 3 N and 1 mm. mean absolute errors in this study were approximately
Force platform CoP accuracy for the single point of half those demonstrated by Bobbert and Schamhardt [3],
force application was greatest along the mid-lines of the where a force platform of almost double the dimensions
force platform and deteriorated as forces were applied was used. The forces applied by Bobbert and Scham-
further from the centre. Table 1 shows the absolute hardt [3] were larger than those for the present study,
di€erence between coordinates measured directly with however their results demonstrate that larger forces do
Vernier callipers and calculated from force platform not result in greater magnitudes of error.
data. Also consistent with Bobbert and Schamhardt [3] was
Accuracy of CoP prediction using the single point of the ®nding that the accuracy of CoP deteriorated mark-
force model was better for x than y coordinates, with the edly towards the edge of the force platform when using a
magnitude of the mean absolute di€erence (or error) for single point of force application. As their force platforms
all data points collected within the central region were larger, this e€ect was greater for their study.
(n ˆ 144) being nearly twice as great in the y-direction Despite the apparent potential for error when man-
(3.4 mm) as in the x-direction (1.8 mm). Accuracy de- ually placing the `feet', this model demonstrated much
teriorated towards the edge of the force platform, as better accuracy than when using a single point of force
demonstrated by Table 1. application. Manually re-positioning the `feet' between
Mean di€erences between CoPpredicted and CoPforce trials with alignment of the `feet' using lines drawn with
from trials with the model `feet' are shown in Table 2. a ®ne felt-tip pen would be expected to introduce errors
Di€erences never exceeded 1.25 mm and 2.00 mm for of at least 0.5 mm. Despite this, errors were generally
xCoP and yCoP respectively, regardless of `foot' position. found to be less than 1 mm, even when the feet were
In contrast to the single point of application, there was positioned on the edge of the force platform.
no trend for errors to increase as the `feet' moved apart. Bobbert and Schamhardt [3] hypothesized that errors
in determination of the point of application using CoP
calculation could be explained by force plate bending
Table 2 and the fact that the transducers were therefore not
Absolute value of the di€erence between CoPpredicted and CoPforce be- loaded exactly at their centres. Point application of force
tween the two `feet' loading would be expected to result in greater force
platform deformation than standing with two feet and
Distance between model `feet' xerror yerror
hence may overestimate the signi®cance of potential
(mm) Mean (SD) Mean (SD)
error during stabilometry tasks. Furthermore, as the
10 0.4 (0.3) 1.0 (0.6) `feet' were positioned symmetrically about the centre of
100 0.4 (0.2) 1.0 (0.8) the platform, errors resulting from each `foot' deforming
200 1.0 (0.3) 0.5 (0.6) the plate would be likely to cancel out. Further inves-
300 0.7 (0.2) 1.0 (0.7)
400 1.1 (0.7) 0.5 (0.2)
tigation is warranted to determine the e€ect of placing
the `feet' asymmetrically on the platform.
360 J. Middleton et al. / Clinical Biomechanics 14 (1999) 357±360

The present study suggests that the accuracy expected References


during two legged stabilometry tasks would be greater
than that previously indicated by single point of appli- [1] Hufschmidt A, Dichgans J, Mauritz KH, Hufschmidt M. Some
methods and parameters of body sway quanti®cation and their
cation loading techniques. Errors in determining the
neurological applications. Arch Psychiat Nervenkr 1980;228:135±
location of the centre of pressure are likely to be less 50.
than 2 mm when the feet are positioned symmetrically [2] Ekdahl C, Jarnlo GB, Andesson SI. Standing balance in healthy
about the centre of the force platform. subjects. Scand J Rehab Med 1989; 21:187±95.
[3] Bobbert MF, Schamhardt HC. Accuracy of determining the point
of force application with piezoelectric force plates. J Biomech
1990;23:705±10.
Acknowledgements

This study was supported by a Northern Sydney Area


Health Service grant (No. 94/21).

Вам также может понравиться