Академический Документы
Профессиональный Документы
Культура Документы
Analyzer Measurements
Operating a network analyzer is not always
a straightforward task.
Rudy Gloria, Agilent Technologies
Agilent Technologies
Types of Network Analyzers components on the production tions similar to actual opera-
line. Like VNAs, SNAs require tion. However, these realistic
A vector network analyzer
an external or internal sweep- operating conditions also cause
(VNA) is the most powerful
ing signal source and signal-sep- the output power of the amplifi-
kind of network analyzer. It is
aration hardware, but they just er to exceed the compression or
used for design verification and
need simple amplitude-only burn-out level of the analyzer’s
final test in manufacturing
detectors, rather than complex receiver.
when the complete amplitude
(and more expensive) phase-
and phase characteristics of an
coherent detectors. When the amplifier under test
electrical network must be dis-
(AUT) needs an input level high-
played. These characteristics
A network/spectrum analyzer er than the network analyzer’s
include S-parameters, magni-
eliminates the circuit duplica- source can provide, you can put
tude and phase, standing wave
tion of a network and a spec- a preamplifier in front of the
ratios (SWRs), insertion loss or
trum analyzer in a testbench AUT to boost the power level. If
gain, attenuation, group delay,
setup. These combination you add a coupler on the output
return loss, reflection coeffi-
instruments can be economical of the preamplifier, a portion of
cient, and gain compression.
alternatives in the design and the boosted input signal can be
VNAs can measure frequencies
test of active components such used for the analyzer’s refer-
from 5 Hz to 110 GHz. VNA
as amplifiers and mixers, where ence channel (figure 2). This con-
hardware consists of a sweep-
analysis of signal performance figuration removes the pream-
ing signal source (usually inter-
is also needed. Frequency plifier’s frequency response and
nal), a test set to separate for-
coverage ranges from 10 Hz drift errors (by ratioing), yield-
ward and reverse test signals,
to 1.8 GHz. ing an accurate measurement of
and a multichannel, phase-
the AUT alone.
coherent, highly sensitive
Using a network analyzer is not
receiver.
always a straightforward task. When the output power of the
The following techniques can AUT exceeds the input compres-
A scalar network analyzer
help you improve the speed and sion level of the analyzer’s
(SNA) measures only the ampli-
accuracy of network analyzer receiver, some type of attenua-
tude portion of the S-parame-
measurements. tion is needed to reduce the
ters, and so obtains measure-
output level. This can be accom-
ments such as transmission
Boosting and Attenuating Signal plished with couplers, attenua-
gain and loss, return loss, and
Levels When Measuring High- tors, or a combination of both.
SWR. After a passive or active
Power Amplifiers You must choose components
component has been designed
that can absorb the high power
using the more extensive mea- Testing high-power amplifiers
from the AUT without sustain-
surement capability of a VNA, can be challenging when the
ing damage, as most loads
an SNA may be a more cost- signal levels needed for test are
designed for small-signal use
effective measurement tool for beyond the stimulus/response
can only handle up to about
revealing out-of-specification range of the network analyzer.
one watt of power.
High-power amplifiers often
require high input levels for
characterization under condi-
2
The frequency-response effects
of the attenuators and couplers
can be removed or minimized
by using the appropriate type of
error correction. One concern
when calibrating with extra
attenuation is that the input
levels to the receiver may be
low during the calibration cycle.
The power levels must be signif-
icantly above the noise floor
of the receiver for accurate
measurements. Therefore,
network analyzers that have
narrowband, tuned receivers
are typically used for high-
power applications because
their noise floor is typically less
than or equal to -90 dBm, and
they exhibit excellent receiver
Figure 2. Interfacing a network analyzer to an amplifier that requires an input level higher than the
linearity over a wide range network analyzer’s source can provide
of power levels.
3
reverse direction to reduce
1:Transmission &M Log Mag 0.5 dB/ Ref 0.00 dB
noise effects in the measure- 2:Off
ment of S22 and S12. Many dB 8714 1SEC VS 0.129SEC
1:
Compensating for Time Delay for
More-Accurate Cable -.5
Measurements -1
M1
source frequency and tuned -2
1
receiver at the same time to
Start 10.000 MHz Stop 3 000.000 MHz
make stimulus-response mea-
surements. Because the fre-
quency of a signal coming from Figure 4a. Erroneous transmission response of a 12-foot cable (lower trace) compared to the true
a device under test (DUT) may response (upper trace).
4
CH1 S21 &M Log MAG 0.5 dB/ Ref 0.00 dB
Cor
Hld
Figure 4b. Measured data can jump as the sweep rate changes in different bands.
ment does not have the frequen- passbands and cables. T/R- However, if port 2 of the device
cy-shift error. The upper trace based analyzers offer only one- is connected directly to the
of figure 4b shows a measurement port calibration for reflection network analyzer’s test port,
of the DUT using the same measurements, which corrects you cannot assume a good load
100-ms sweep time, but with the for errors caused by directivity, termination. In this case, mea-
matching cable in R channel. source match, and frequency surement accuracy can be
response, but not load match. improved considerably by plac-
Terminating the Unmeasured ing an attenuator (6 to 10 dB,
Port for Accurate Reflection One-port calibration assumes a for example) between port 2
Measurements with good termination at port 2 of of the device and the test port
Two-Port Devices the DUT (the port not being of the analyzer. This improves
measured), as load match is not the effective load match of the
Making accurate reflection mea-
corrected. One way to achieve analyzer by twice the value of
surements on two-port devices
this is by connecting a high- the attenuator.
using a transmission/reflection
quality load (such as a load
(T/R)-based analyzer requires a
from a calibration kit) to port 2
good termination on the unmea-
of the device. This technique
sured port. This is particularly
yields measurement accuracy
important for low-loss, bidirec-
on par with more expensive
tional devices such as filter
S-parameter-based analyzers
that use full two-port calibration.
5
Figure 5 shows an example of
how this works. When measur-
ing a filter with 1 dB of inser-
tion loss and 16 dB of return
Analyzer port 2 match:
loss, using an analyzer with an 18 dB (0.126)
18-dB load match and 40-dB Directivity:
directivity would yield a worst-
40 dB (0.010) DUT
16 dB return loss (0.158)
case measurement uncertainty 1 dB loss (0.891)
for return loss of -4.6 dB,
0.158
+10.4 dB (figure 5a). This is a
large variation that might cause 0.891*0.126*0.891 = 0.100
a filter that didn’t meet its
specifications to pass, or a Measurement uncertainty:
good filter to fail. Figure 5b –20 * log (.158 + 0.100 + 0.010)
= 11.4 dB (–4.6dB)
shows how adding a high-quality –20 * log (0.158 – 0.100 – 0.010)
10-dB attenuator (for example, = 26.4 dB (+10.4 dB)
VSWR = 1.05, or 32-dB match)
improves the load match of the
analyzer to 29 dB ((2 x 10 + 18 dB) Figure 5a. Proper termination is important to keep measurement uncertainty reasonable.
combined with 32 dB). Now
the worst-case measurement
uncertainty is reduced to
+2.5 dB, -1.9 dB, which is
much more reasonable. Load match:
18 dB (.126)
6
Using Frequency-Offset Mode to
Accurately Measure Mixers,
Converters, and Tuners
FREQ
Frequency-translating devices ON off
7
CH1 CONV MEAS log MAG 10 dB/ REF 10 dB
ACTIVE CHANNEL ENTRY
RESPONSE
HP-IB STATUS
S
Reference Mixer
H 8753D 30 KHz-3GHz
NETWORK ANALYZER
PROBE POWER
FUSED
Ref Out
PORT 1 PORT 2
RF IF
10 dB LO
10 dB Lowpass
Filter
10 dB
LO
DUT 3 dB
Signal Generator
For high-dynamic-range ampli- Increasing Accuracy When There are five basic ways to
tude measurements, a reference Measuring Noninsertible Devices overcome the potential errors
mixer must be used (figure 6b). when making a through connec-
Full two-port error correction
This mixer provides a signal to tion for a noninsertible device:
provides the best accuracy
the R channel for proper phase
when measuring RF and
lock, but does not affect mea- 1. Use electronic calibration
microwave components. But if
surements of the DUT as it is (ECal) modules. Using an ECal
you have a noninsertible device
not in the measurement path. module with the connectors
(for example, one with female
A reference mixer must also be that match the DUT is the sim-
connectors on both ports), then
used for phase or delay mea- plest and fastest noninsertible
its test ports cannot be directly
surements. The reference mixer calibration method. Full two-
connected during calibration.
and the DUT must share a port error correction, defined
Extra care is needed when mak-
common LO to guarantee phase at the test ports, is achievable.
ing this through connection,
coherency. When testing mixers,
especially when measuring a
either technique requires an 2. Use a very short through.
device that has poor output
IF filter to remove the mixer’s This allows you to disregard
match, such as an amplifier
undesired mixing products as the potential errors. When you
or a low-loss device.
well as the RF and LO leakage connect port 1 to port 2 during
signals. a calibration, the analyzer cal-
8
culates the return loss of the 3. Use the “swap equal swap in the male-to-female
second port (the load match) as adapters” method. For this adapter (you now have two
well as the transmission term. method you use two matched male test ports), and do the
When the calibration kit defini- adapters of the same electrical reflection portion of the calibra-
tion does not contain the cor- length, one with male/female tion. Now you are ready to mea-
rect length of the through, an connectors and one that match- sure your device. All the
error occurs in the measure- es the DUT. adapters in the calibration kits
ment of the load match. If a are of equal electrical length
barrel is used to connect port 1 For example, suppose your (even if their physical lengths
to port 2, the measurement of instrument test ports are both are different). This process is
the port 2 match will not have male, such as the ends of a pair shown in figure 7.
the correct phase, and the of test-port cables, and your
error-correction algorithm will device has two female ports. 4. Modify the through-line stan-
not remove the effects of an Put a female-to-female through dard. If your application is
imperfect port 2 impedance. adapter, usually on port 2, and manufacturing test, the require-
do the transmission portion of ment of the swap-equal-
This approach works reason- the calibration. After the four adapters method for additional
ably well if the through connec- transmission measurements, adapters may be a drawback.
tion is quite short. However,
for a typical network analyzer,
“short” means less than one
hundredth of a wavelength. If Port 1 DUT Port 2 Non-insertible device
the through connection is one
tenth of a wavelength at the
frequency of interest, the 1. Transmission cal using
Port 1 Adapter Port 2
corrected load match will A adapter A.
be no better than the raw load
match. As the through length 2. Reflection cal using
approaches one-quarter of a Port 1 Adapter Port 2 adapter B. Length of
B
wavelength, the residual load adapters must be equal.
match can actually get as high
as 6 dB worse than the raw load
3. Measure DUT using
match. For a 1-GHz measure- Port 1 DUT Adapter Port 2
B adapter B.
ment, one hundredth of a
wavelength means less than
3 mm (about 0.12 inches).
Figure 7. Swap-equal-adapters method for measuring noninsertible devices
9
Instead, it is possible to modify Selecting the Appropriate Furthermore, the VNA calcu-
the calibration kit definition Measurement Parameters to lates group delay data from
to include the length of the Prevent Aliasing Problems phase data. If the slope of
through line. If the calibration the phase is reversed, the
When measuring a DUT that
kit has been modified to take group delay will change sign.
has a long electrical length, you
into account the loss and delay A surface acoustic wave (SAW)
need to select the appropriate
of the through, then the correct filter may appear to have nega-
measurement parameters care-
value for load match will be tive group delay–clearly not a
fully. A VNA samples its data at
measured. It’s easy to find correct answer. If you suspect
discrete frequency points, then
these values for the male-to- aliasing in your measurement,
“connects the dots” on the dis-
male through and the female- try decreasing the spacing
play to make it more visually
to-female through. First, do a between frequency points
appealing. If the phase shift of
swap-equal-adapters calibra- (by increasing the number of
the DUT changes by more than
tion, ending up with test ports points or reducing the frequency
180 degrees between adjacent
that are either both female or span) and see if the data on the
frequency points, the display
both male. Then simply mea- VNA’s display changes.
can look like the phase slope is
sure the “noninsertible”
reversed. The data is undersam-
through and look at S21 delay Figure 8a shows a measurement
pled and aliasing occurs. This
(use the midband value) and of a SAW bandpass filter on a
is analogous to filming a wagon
loss at 1 GHz. Use this value network analyzer with 51 points
wheel in motion where the
to modify the calibration kit. in the display. The indicated
wheel appears to spin backward
group delay is negative–a physi-
because the number of frames
5. Use the adapter-removal cal impossibility. But if the
per second is too low to accu-
technique. Many VNA models
rately portray the motion.
offer an adapter-removal tech-
nique to eliminate all effects
of through adapters. This 1: Transmission Delay 500 ns/ Ref 0s
2: Transmission Phase 100 / Ref 0.00
technique yields the most 51 POINT TRACE Meas1:Mkr1 140.000 MHz
s
–1.1185
2:
Figure 8a. Decreasing the spacing between frequency points can create an aliasing problem.
In this case, the display used 51 points.
10
number of display points is
1: Transmission Delay 500 ns/ Ref 0s
increased to 201, as in figure 8b, 2: Transmission Phase 100 / Ref 0.00
201 POINT TRACE Meas1:Mkr1 140.000 MHz
it becomes clear that the VNA 1 –1.3814 s
Delay
Known
settings created an aliasing Positive
problem.
1: Ref = 0 seconds
11
Using a Segmented Sweep
to Increase Accuracy at Fast
Sweep Speeds
12
of the port 2 coupler, but simply
reverses the direction of signal
travel (figure 10). By reversing the Source
Simplifying Multiport R1 R1 A A B B R2 R2
OUT IN Source Coupler OUT IN IN OUT Coupler Source IN OUT
Measurements OUT IN IN OUT
13
Visit Agilent’s library of application notes,
traiing courses, FAQs, tutorials, and more at
www.agilent.com/find/test
By internet, phone, or fax, get assistance
with all your test & measurement needs
Online assistance:
www.agilent.com/find/assist
Agilent Technologies