AMETEK Drexelbrook makes no warranty of any kind with regard to the material contained in this manual, including, but not limited to, implied warranties or fitness for a particular purpose. Drexelbrook shall not be liable for errors contained herein or for incidental or consequential damages in connection with the performance or use of material. Copyright 2004 AMETEK Drexelbrook
Series RCT Universal III Mark II Transmitter with HART Protocol using 409-T Electronics
DREXELBROOK
An ISO 9001 Certified Company
205 Keith Valley Road Horsham, PA 19044 US Sales 1-800-553-9092 24 Hour Service 1-800-527-6297 International + 215-674-1234 Fax + 215-674-2731 E-mail drexelbrook.info@ametek.com Web www.drexelbrook.com
Table Of Contents
Table of Figures .............................................................................................................................. vii Introduction ............................................................................................................... 1 Section 1: 1.1 System Description........................................................................................................... 1 1.2 Technology........................................................................................................................ 1 1.3 Models Available ............................................................................................................... 2 1.4 Sensing Element Model Numbering ................................................................................. 3 1.5 Area Classifications .......................................................................................................... 3 Installation ................................................................................................................. 5 Section 2: 2.1 Unpacking ......................................................................................................................... 5 2.2 Hazardous Location Installations...................................................................................... 5 2.3 Mounting the Electronic Unit............................................................................................. 5 2.4 Wiring the Electronic Unit................................................................................................ 10 2.5 Wiring the Sensing Element............................................................................................ 11 2.6 Spark (Static Electricity) Protection ................................................................................ 12 2.7 Surge Voltage (Lightning) Protection .............................................................................. 13 2.8 RFI (Radio Frequency Interference) Filters..................................................................... 14 2.9 Electrostatic Filters ......................................................................................................... 16 2.10 Digital Integral Meter....................................................................................................... 16 Section 3: Configuration and Calibration with Drexelbrook PC Software ........................... 19 3.1 General Description ........................................................................................................ 19 3.2 Model Number ................................................................................................................ 19 3.3 System Requirements .................................................................................................... 19 3.4 Installing The RS232 Modem.......................................................................................... 20 3.4.1 Installing The USB Modem ............................................................................................. 21 3.5 Install the Windows Version HARTWin 2.1 Software on Hard Drive ............................... 22 3.6 Description of Function Keys .......................................................................................... 22 3.7 Configuration .................................................................................................................. 24 3.7.1 Level Configuration ......................................................................................................... 24 3.7.2 Vessel Configuration ....................................................................................................... 25 3.7.3 Lower and Upper Range Values (LRV and URV) ........................................................... 25 3.8 Calibration....................................................................................................................... 26 3.8.1 Point Calibration.............................................................................................................. 27 3.8.2 Level Calibration ............................................................................................................. 28 3.8.3 Application Example ....................................................................................................... 28 3.9 PC Status Messages ...................................................................................................... 30 3.10 Set D/A Trim.................................................................................................................... 31 3.11 Strapping Table ............................................................................................................... 31 3.12 Digital Integral Meter Configuration ................................................................................ 33 3.13 Save/Print Entries ........................................................................................................... 33 3.14 Validation ........................................................................................................................ 34 3.14.1 Validation Design Concept.............................................................................................. 34 3.14.2 Validation Procedures..................................................................................................... 34 3.14.3 Validation Results ........................................................................................................... 36 3.15 Calibration & Configuration via 401-44-3 Display/Keypad .............................................. 37 Section 4: 4.1 4.2 Configuration and Calibration with Model 275 with Drexelbrook Device Description............................................................................ 39 Drexelbrook Device Description...................................................................................... 39 Start-up ........................................................................................................................... 40
Configuration .................................................................................................................. 41 Calibration....................................................................................................................... 42 Point Calibration.............................................................................................................. 43 Capacitance Calibration.................................................................................................. 44 D/A Trim .......................................................................................................................... 44 Strapping Table ............................................................................................................... 46
Configuration & Calibration with Model 275 Communicator without a Device Description............................................................................................... 47 5.1 Range/Span Control ....................................................................................................... 47 5.2 Basic Rules & Conventions of HART Configuration Software (Model 275) .................... 48 5.3 Tag ID.............................................................................................................................. 48 5.4 Set Up Procedures ......................................................................................................... 49 5.5 Reading Input and Output............................................................................................... 52 5.6 Calibration Using Actual Tank Level................................................................................ 53 5.7 Bench Calibration (if needed) ......................................................................................... 54 5.7.1 Bench Calibration Information Sheet .............................................................................. 55 5.8 Point Calibration.............................................................................................................. 56 5.8.1 Fine Tuning Calibration ................................................................................................... 56 5.8.2 Selecting Engineering Units............................................................................................ 56 5.9 Handheld Calibrator Error Messages.............................................................................. 57
Section 6: Troubleshooting ...................................................................................................... 59 6.1 Identifying a Problem/Symptom ...................................................................................... 60 6.2 Troubleshooting Loop Connection................................................................................... 60 6.3 Rosemount Model 268 or 275 Calibrator Cannot Identify or Find Device....................... 61 6.4 Rosemount Model 275 Calibrator with Device Description Cannot Identify or Find Device......................................................... 61 6.5 Universal III Transmitter Does not Communicate with Drexelbrook PC software ................................................................................................ 62 6.6 Troubleshooting Transmitter ............................................................................................ 63 6.6.1 Transmitter Drift Test ....................................................................................................... 64 6.7 Troubleshooting Sensing Element .................................................................................. 65 6.8 Troubleshooting Coaxial Cable ....................................................................................... 66 6.9 Static Electricity .............................................................................................................. 68 6.10 Radio Frequency Interference......................................................................................... 68 6.11 Factory Assistance.......................................................................................................... 68 6.12 Field Service ................................................................................................................... 69 6.13 Customer Training ........................................................................................................... 69 6.14 Return Equipment........................................................................................................... 69 6.15 Universal III Mark II Troubleshooting Guide .................................................................... 71 Section 7: Specifications.......................................................................................................... 73 7.1 Transmitter Specifications ............................................................................................... 73 7.2 Coaxial Cable Specifications .......................................................................................... 74 7.3 Enclosure Specifications................................................................................................. 74 7.4 Approvals Available......................................................................................................... 75 7.4.1 ATEX Approvals - Install Per 420-0004-203-ID............................................................... 75 7.4.2 CSA Approvals - Install Per 420-0004-225-CD............................................................... 76 Section 8: Normal Maintenance............................................................................................... 77 8.1 Viewport Cleaning........................................................................................................... 77 Appendix A: Approval Drawings .................................................................................................A1 A.1 ATEX APPROVAL DRAWINGS ...................................................................................... A1 A.2 CSA APPROVAL DRAWINGS ........................................................................................ A7
Table of Figures
Figure 1-2 Figure 1-1 Figure 2-1 Figure 2-2 Figure 2-3 Figure 2-4 Figure 2-5 Figure 2-6 Figure 2-7 Figure 2-8 Figure 2-9 Figure 2-10 Figure 2-11 Figure 2-12 Figure 2-13 Figure 2-14 Figure 3-1 Figure 3-1a Figure 3-2 Figure 3-3 Figure 3-4 Figure 3-5 Figure 3-6 Figure 3-7 Figure 3-9 Figure 3-10 Figure 3-11 Figure 3-12 Figure 3-13 Figure 3-13a Figure 3-14 Figure 3-15 Figure 3-16 Figure 3-17 Figure 3-18 Figure 3-19 Figure 3-20 Figure 4-1 Figure 6-1 Figure 6-2 Figure 6-3 Figure 6-4 Figure 6-5 Figure 6-6 RF Admittance Sensing Element with Cote-Shield..................................................... 1 Capacitance Sensing Element.................................................................................... 1 Recommended Conduit Connection ........................................................................... 6 Mounting Dimensions ................................................................................................. 6 Installing Sensing Element.......................................................................................... 8 Common Installation Mistakes .................................................................................... 8 Installing Sensing Element in Agitated Vessel ............................................................ 9 Providing Ground Reference....................................................................................... 9 Universal III Mark II Wiring Connections Integral Mounting ...................................... 10 Universal III Mark II Wiring Connections, Remote Mounting.................................... 10 Spark Protection for Integral Sensing Elements ....................................................... 12 Spark Protection for Remote Sensing Elements....................................................... 13 Surge Voltage Protection .......................................................................................... 14 Radio Frequency Interference (RFI) Filters............................................................... 15 Electrostatic Filter ..................................................................................................... 16 Digital Meter in Housing with Viewport ..................................................................... 17 RS232 Modem Assembly & Loop Connection.......................................................... 20 USB Modem Assembly & Loop Connection ............................................................. 21 Selecting COM ports during software installation ..................................................... 22 PC Software Menu Screen ....................................................................................... 23 Configure Transmitter from Menu screen .................................................................. 24 Level Configuration from Menu screen ..................................................................... 24 Vessel Configuration from Menu screen ................................................................... 25 LRV & URV Configuration from Menu screen ........................................................... 25 Range Span Jumpers ............................................................................................... 26 Point Calibration from Menu screen.......................................................................... 27 Level Calibration area of Menu screen ..................................................................... 28 Application Example Diagram................................................................................... 29 PC Software Menu Screen View of Application Example ........................................ 29 PC Software Menu Screen View of Main Menu ....................................................... 30 Setting D/A Trim Menu Screen "Pop-Ups" ............................................................... 31 Menu Screen Transforms to Strapping Table ............................................................ 32 Configure Meter Pop-up from Menu Screen; values relate to LRV & URV ............... 33 Print Pop-up from Menu ............................................................................................ 33 Capacitance on Menu Screen................................................................................... 34 Real Time View Pop-up from Menu Screen .............................................................. 35 Typical Printout of Transmitter Data .......................................................................... 36 Typical Transmitter Loop............................................................................................ 40 Place capacitor on transmitter from probe to ground terminals ................................ 63 C-Box Adjustment ..................................................................................................... 64 Span Range Capacitance Values ............................................................................. 64 Transmitter Drift Test ................................................................................................. 65 Sensing Element Testing, Material Below the Sensing Element............................... 66 Sensing Element Testing, Material Covering the Sensing Element .......................... 66
vii
Introduction
SECTION 1: INTRODUCTION
1.1 System Description
The instructions in this manual are for the AMETEK Drexelbrook RCT Series Universal III Mark II for level measurement in liquids, slurries, interfaces, and granulars.
k air
d C A d
k air
k media
k media
Each AMETEK Drexelbrook RCT Series system consists of a Universal III Mark II series two-wire, 4-20 mA electronic unit and a 700 series sensing element. A 380 series connecting cable is also supplied for connection of the sensing element to the electronic unit. The RCT Series system is an admittance-to-current transducer. A change in level produces a change in admittance which results in a change of current. It is termed a two-wire transmitter because the same two wires that are used to power the unit also indicate the change in level (4-20 mA).
kA C= d
kA C = d
1.2
Technology
In a simple capacitance sensing element, when the level rises and material covers the sensing element, the capacitance within the circuit between the sensing element and the medium (conductive applications) or the sensing element and the vessel wall (insulating applications) increases. This is due to the dielectric constant (k) of the material which causes a bridge misbalance. The signal is demodulated (rectified), amplified, and the output is increased. There are drawbacks, however, especially when there is coating of the sensing element.
R Xc
An RF Admittance level transmitter is the next generation. Although similar to the capacitance concept, Universal III Mark II employs a radio frequency signal and adds the Cote-Shield circuitry within the Electronics Unit. Built-in oscillator buffer and chopper drive circuits permit separate measurement of resistance and capacitance. Since the resistance and the capacitance of any coating are of equal magnitude (by physical laws), the error generated by a coating can be measured and subtracted from the total output. This patented Cote-Shield circuitry is designed into the Universal III Mark II series and enables the instrument to ignore the effect of buildup or material coating on the sensing element. The sensing element is mounted in the vessel and provides a change in RF admittance indicating presence of material. The Cote-Shield circuitry prevents the transmission of RF current through the coating on the sensing element. The only path to ground available for the RF current is through the material being measured.
Oscillator circuitry through phase shift cancels small amounts of RF current flow (both Resistive and Capacitive) caused by coating
High Resistance
(More Coating is Easier to Ignore)
Xc R
Coating
Tank Wall
Little to No Resistance
The result is an accurate measurement regardless of the amount of coating on the sensing element. By far the most versatile technology available it works with all types of materials in a vast array of conditions, from cryogenics to high temperature, from vacuum to 10,000psi pressure.
700-0002-227 C Tri-Clamp 7 Remote - 100 feet cable GP 6 0 700-0204-038 (remote only) 7 1 700-0005-028 316 SS 9 700-0005-385 Q 8" 300# 700-0204-038 (remote only) D ANSI 4 700-0001-052 600# 6 Flange 150#, 300# and 700-0005-029 Hastelloy C 8 Remote - 25 feet Tri-ax 7 A 700-0005-385 2 R 1" 600# 6 4 700-0001-052 6 4 700-0001-052 E Din Flange 9 9 Remote - 50 feet Tri-ax 7 0 700-0001-052 6 5 S 1 1/2" 600# 700-0001-062 3 700-0005-048 6 5 700-0001-062 6 5 700-0001-062 F ANSI Flange 900#, 1500#, 2500# Mounting Options A Remote - 75 feet Tri-Ax 7 6 T 2" 600# 6 700-0002-042 4 700-0005-054 6 6 700-0002-042 6 6 700-0002-042 1 3/4" A NPT B Remote - 100 feet Tri-Ax 7 6 7 U 2 1/2" 600# 700-0002-052 5 700-0005-085 316 SS Technology 6 7 700-0002-052 6 7 700-0002-052 2 1" C Tri-Clamp C Remote - 10 feet Hi-temp GP composite 7 V 3" 600# 8 700-0002-062 6 700-0005-085 Hastelloy C R RF 6 6 8 700-0002-062 6 8 700-0002-062 3 1 1/2" D ANSI Flange 150#, 300# and 600# D Remote - 25 feet Hi-temp GP composite 7 W 4" 600# 0 700-0005-018 7 700-0005-285 316 SS Measurement Type 7 0 700-0005-018 7 0 700-0005-018 47 2" E Din Flange E Remote - 35 feet Hi-temp GP composite 7 Technology X 6" 600# 1 700-0005-028 8 700-0005-285 Hastelloy C C Continuous 7 1 700-0005-028 7 1 700-0005-028 57 1 1/4" F ANSI Flange 900#, 1500#, 2500# F Remote - 50 feet Hi-temp GP composite 7 9 700-0005-385 316 SS R RF Y 8" 600# Input 7 2 700-0005-029 7 2 700-0005-029 2 700-0005-029 67 3" 1 3/4" G Remote - 75 feet Hi-temp GP composite Technology Measurement Type 7 3 700-0005-048 1 316/316LSS Flange 7 A 700-0005-385 Hastelloy C T Two Wire 7 3 700-0005-048 77 4" 3 700-0005-048 0 700-0001-052 9 2 1" H Remote - 100 feet Hi-temp GP composite R RF C Continuous Technology 2 CS Flange 7Output Type 4 700-0005-054 7 4 700-0005-054 A 1" 150# 7 4 700-0005-054 Technology Mounting1/2" 3 1 Options Sensing Element R RF Measurement Type Input 3 Hastelloy C 276 Flange 70 5Analog Only 700-0005-085 316 SS 7 5 700-0005-085 316 SS B 1 1/2" 150# R RF C ContinuousType 4 2" Z Z Special 7 5 700-0005-085 316 SS A NPT Measurement Wire 1 4-20 mA HART T Two 4 CS Flange with Fitted TFE Face 7 6 700-0005-085 Hastelloy C 7 6 700-0005-085 Hastelloy C C 2" TFE Measurement Type 1 0 700-0001-018 150# Probes C Tri-Clamp Input C Continuous 6 700-0005-085 Hastelloy C 5 1 1/4" Output Type 5 7316/316L SS Flange with Fitted TFE Face 72 7 Foundation Fieldbus SS 700-0005-285 316 7 7 700-0005-285 316 SS D 2 1/2" 150# 1 6 700-0002-360 C Continuous Wire ANSI T Two Input 7 700-0005-285 316 SS D 6 3"Flange 150#, 300# and 600# 6 7Hastelloy C 276 Flange with Fitted TFE Face Application Hastelloy C 70 8Analog Only 700-0005-285 Type 7 8 700-0005-285 Hastelloy C E 1" 1 8 700-0001-022 300# Input Wire Output Type E Din Flange 7 4" T Two HART A 7CS 8 NPT or Tri-Clamp 71 94-200mA15 Khz 0 Degree 700-0005-385 316 SS 7 9 700-0005-385 316 SS 9 700-0001-0161/2" 700-0005-285 Hastelloy C F 1 300# 1 T Two Wire 0 Analog Only F ANSI Flange 900#, 1500#, 2500# A 1" 150# Output 1 15 Fieldbus Type B 7316SS NPT or Tri-Clamp 9 700-0005-385 316 SS 72 A FoundationKhz 45 Degree C 700-0005-385 Hastelloy 7 A 700-0005-385 Hastelloy C G 2" (remote only) 2 0 700-0209-002 300# Output TypeHART 1 3/4" 150# B 1 1/2" 1 4-20Application mA 0 Analog Only Khz 0 Degree C 7Hastelloy C276 NPT or Tri-Clamp C 2 100 9 0 700-0001-052Type 9 0 700-0001-052 A 700-0005-385 Hastelloy H 2 2 1 700-0009-0021/2" 300# 0 Analog Only Fieldbus 2 1" 150# C 2" 2 4-200mA15 Khz 0 Degree 1 Foundation Khz 45 Degree HART P 9PFA NPT 3 Mounting Options 100 Mounting Options 2 700-0001-026 150#SS 700-0001-052 I 3"0316 2 1 4-20Application Type mA HART 3 1 1/2" 150# D 2 2 FoundationKhz 45 Degree 1 True Level 0 Degree 15 Fieldbus Sensing Element Insertion Length 4 A NPT A NPT J 3" Carbon Steel 2 3 700-0001-026 300# Mounting Options 2 FoundationKhz 0Type E 1" 4 2" 300# 0 100 Fieldbus 15 Application Degree 2 C Tri-Clamp Degree (in millimeters) 5 TrueKhz 0 Degree Level 45 C Tri-Clamp K 4" 150#SS A NPT 2 4 700-0001-074 316 Application Type 5 1 1/2" 300# 1/4" F 1 15 Khz 045 Degree 45 Degree 0 CUTKhz Degree 3 100 MONITOR Inactive Length or Cote-Shield Length 6 D ANSI Flange 150#, 300# and 600# D ANSI Flange 150#, 300# and 600#4" 300# L 2 5 700-0001-074 Carbon C Tri-Clamp Steel 0 15 Khz 00 Degree 6 3" G 2" 2 100 Khz Degree 1 15 Khz 45 0 Degree 4 ETrue Level Display (in millimeters)300# ut Type Digital Din Flange E Din Flange 2 6 700-0001-074 Hastelloy C M 6" 150# D ANSI Flange 150#, 7 4"1/2" 300# 300# 1 15 Khz 45 Degree H 2 and 600# 3 100 Level Degree 4545 Degree 2 0 Degree 5 FTrueKhzWithout 900#, 1500#, 2500# nalog Only Cooling Extension 0 ANSI Flange Meter - Standard Housing F ANSI Flange 900#, 1500#, 2500# 6" 300# N 2 7 700-0001-034 2 100 Khz 0 Degree E Din Flange A 1" 150# I 3" 4 True MONITOR 3 100 1Level 0 Meter, 45 Degree 6 CUTKhzWithDegree Keypad, and Viewport Housing1 3/4" -20 mA HART 1 3/4" P 8" 316 4 0 700-0002-024 150#SS 3 100 Khz 4545 Degree Degree J 3" 300# 5 True Level 0 Degree F ANSI Flange 900#, B 1 1/2" 150# 1500#, 2500# 4 oundation Fieldbus 2Digital Display 1" Reserved 2 1" Q 8" 300# 4 1 700-0002-024 Hastelloy C 4 True MONITOR C 2" 150# K 4" 6 CUT Level 0 Degree - Standard Housing 5 True Level 45 Degree 1 3/4" Application Type 30 1Without Meter 1/2" 3 1 1/2" R 1" 600# 4 2 700-0002-224 5 True MONITOR D 2 1/2" 150# L 4" 300# Digital Display Keypad, and Viewport Housing 6 CUT Level 45 Degree With Meter, 2 1" 0 15 Khz 0 Degree Approvals 41 2" 4 2" S 1 4 3 700-0002-0271/2" 600# E 1" 300# 6 CUT MONITOR Meter - Standard Housing M 6" 150# 0 Without Digital Display 1 15 Khz 45 Degree 3 1 1/2" 0 No Approval 5 1 1/4"Reserved 5 1 1/4" T 2" 4 4 700-0002-028 600# Digital Display Keypad, and Viewport Housing F 1 1/2" 300# N 6" 300# 1 With Meter, - Standard Housing 0 Without Meter 2 100 Khz 0 Degree F FM APPROVAL 6 3" 6 3" 4 2" U 2 4 5 700-0002-0571/2" 600# 0 Without Meter - Standard Viewport Housing G 2" 300# P 8" 150# 1 With Meter, Keypad, and Housing Reserved 3 100 Khz 45 Degree Approvals C CSA APPROVAL 7 4" 7 4" 5 1 1/4" V 3" 4 6 700-0002-036 600# 1 With Meter, Keypad, and Viewport Housing H 2 1/2" 300# Q 8" 300# Reserved 4 True Level 0 Degree 0 ATEX APPROVAL A 1" 150# A No Approval A 1" 150# 6 3" W 4" 4 7 700-0002-037 600# I R 3" 150# 1" 600# Reserved Approvals 5 True Level 45 Degree F FM APPROVAL Suppression Surge/Noise B 1 1/2" 150# B 1 1/2" 150# X 6" 4 8 700-0002-056 600# 7 4" J 3"1/2" 600# S 1 300# 0 CSA APPROVAL Approvals No additional filtering required 6 CUT MONITOR C No Approval 0 C 2" 150# C 2" 150# 4 9 700-0002-227 600# Y 8" A 1" 150# K 4" 150# T 2" 600# Approvals F ATEX APPROVAL 0 FM APPROVAL Digital Display A No Approval RFI Filter 2 1/2" 150# 1 Loop D 2 1/2" 150# D 1 316/316LSS 1 1/2" 6 0 700-0204-038 (remote only) Flange 150# B L 4"1/2" 600# U 2 300# 0 FM APPROVAL C No Approval F CSA APPROVAL Suppression 0 Without Meter - Standard Housing Surge/Noise E E 1" 300# 2 Probe RFI Filter1" 300# 2 6 4 700-0001-052 CS Flange M 6" 150# V 3" 600# F CSA APPROVAL C 2" 150# A FM APPROVAL C ATEX APPROVALProbe RFI Filter 1 With Meter, Keypad, and Viewport Housing 0 No additional 1 1/2" 300# 3 Loop and F filtering required F 1 1/2" 300# 3 6 5 700-0001-062 Hastelloy C 276 Flange N 6" 300# W 4" 600# C CSA APPROVAL Suppression Surge/Noise A ATEX APPROVAL Filter D 2 1/2" 150# 1 Loop RFI Filter 2" 300# Reserved 4 SurgeG G 2" 300# 4 6 6 700-0002-042 CS Flange with Fitted TFE FaceP 8" 150# X 6" 600# A ATEX APPROVAL filtering required 0 No additional Surge/Noise Suppression E 1" 300# 2 5 Probe RFI Filter2 1/2" 300# Surge Filter H 2 1/2" 300# H 5 6 7 700-0002-052 316/316L SS Flange with Fitted TFE8" 300# Q Face Y 600# Surge/Noise Suppression 1 Loop RFI Filter RFI Filter 0 No additional filtering required 3 Approvals F 1 1/2" 300# 6 Loop and Probe3" 150# Filter Surge I 3" 150# I 6 6 8 700-0002-062 Hastelloy C 276 Flange with Fitted TFE Face R 1 600# 1" 316/316LSS Flange 0 No additional filtering required 2 ProbeSurge Filter 1 Loop RFI Filter 4 Loop RFI Filter 3" 300# 0 No Approval 7 RFI and Surge Protection J 3" 300# J G 2" 300# A 7 0 700-0005-018 CS NPT or Tri-Clamp S 2 1/2" 600# 1 CS Flange 1 Loop RFI Filter RFI Filter 3 Probe RFI Filter 2 5 Probeand Probe Surge Filter F FM APPROVAL 8 Desalter Filter 4" 150# K 4" 150# K H 2 1/2" 300# B 7 1 700-0005-028 316SS NPT or Tri-Clamp T 3 600# 2" Hastelloy C 276 Flange 2 ProbeSurge Filter RFI Filter 4 Loop and Probe RFI Filter 3 6 Loop and Probe4" 300# Filter Surge C CSA APPROVAL Remote/Integral & Cable L 4" 300# L I C NPT or 7 2 700-0005-029 Hastelloy C276 3" 150#Tri-Clamp 4 1/2" 600# U 2 CS Flange with Fitted TFE Face 3 Loop and Probe RFI Filter 5 Probe SurgeFilter 4 Loop IntegralProtection 7 RFI 0 SurgeMFilter150# and Surge 6" A ATEX APPROVAL M 6" 150# P 7 3 700-0005-048 PFA NPT J 3" 300# V 5 600# 3" 316/316L SS Flange with Fitted TFE Face 4 Loop Surge Filter 6 Probeand Probe Surge Filter 5 DesalterRemote - no cable 8 Filter 6" Surge/Noise Suppression 1 Surge Filter300# N 6" 300# N Sensing Element Insertion Length 7 4 700-0005-054 K 4" 150# W 6 600# 4" Hastelloy C 276 Flange with Fitted TFE Face 5 Probeand Probe Surge Filter 7 RFI Remote/Integral & Cable and Surge Protection 6 Loop Surge Filter10 feet cable GP 0 No additional filtering required 2 Remote - 150# P 8" 150# P 8" (in 7 5 700-0005-085 316 SS millimeters) L 4" 300# X A 600# NPT or Tri-Clamp 6" CS 6 Loop and Probe Surge Filter 8 DesalterIntegralProtection Filter 7 RFI 0 Remote - 25 feet cable GP and Surge 1 Loop RFI Filter 3 Q 8" 300# Q 8" 300# 7 6 700-0005-085 Hastelloy C Inactive Length or Cote-Shield Length Y B 600# 8" 316SS NPT or Tri-Clamp 7 RFI Remote/Integral & Cable and Surge Protection M 6" 150# 8 DesalterRemote - 35 feet cable GP Filter 1 2 Probe RFI Filter 4 R 1" 600# R 1" no cable 600# (in millimeters) 7 7 700-0005-285 316 SS 1 316/316LSS Flange or Tri-Clamp C Hastelloy C276 NPT 8 DesalterIntegral Filter 0 Remote/Integral & Cable N 6" 300# 2 Remote -1/2" feet cable GP 10 600# 3 Loop and Probe RFI Filter 5 50 S 1 1/2" 600# S 1 7 8 700-0005-285 Hastelloy C 2 CS Flange Extension P PFA Cooling NPT Remote/Integralcable 1 Remote - 25 feet cable GP 0 Integral - 75 & Cable P 8" 150# 3 Remote no 4 Loop Surge Filter 6 T 2" 600# T 2" 600# 7 9 700-0005-385 316 SS 3 Hastelloy C 276 FlangeInsertion Length Sensing Element 0 Integral - 10 feet cable GP 2 Remote - 35 feet cable GP 1 Remote no cable cable GP 4 5 Probe Surge Filter 7 100 feet U 2 1/2" 600# U 2 1/2" 600# A 700-0005-385 Hastelloy C Q 8" 300# 7 (in millimeters) 4 CS Flange with Fitted TFE Face 1 Remote - 10 feet cable GP - no cable 3 2 Remote - 25 feet Tri-ax GP 5 cable 6 Loop and Probe Surge Filter 8 Remote 25 V 3" 600# R 1" 600# V 3" 50 600# 9 0 700-0001-052 5 316/316L SS Flange with Fitted TFE Face Inactive Length or Cote-Shield Length 2 Remote - 10 feet cable GP 4 35 3 25 cable 6 Remote - 75 feet cable GP 7 RFI and Surge Protection 9 Remote - 50 feet Tri-ax GP Mounting Options W 4" 600# W 4" 600# S 1 1/2" 600# (in millimeters) 6 Hastelloy C 276 Flange with Fitted TFE Face 3 Remote - 25 feet cable GP 5 50 4 7 Remote - 35 feet cable GP 8 Desalter Filter A Remote - 75 feet Tri-AxGP A NPT X 6" 600# X 6" 100 feet cable 600# T 2" 600# A CS NPT or Tri-Clamp Cooling Extension 4 Remote - 35 feet cable GP 6 75 feet cable 5 50 8 Remote - 25 feet Tri-ax GP Remote/Integral & Cable B Remote - 100 feet Tri-Ax Y 8" 600# Y 8" 600# C Tri-Clamp U 2 1/2" 600# B 316SS NPT or Tri-Clamp 5 Remote - 50 feet cable GP 7 Remote - 50 feet Tri-ax GP 100 6 75 feet cable GP cable 9 Remote - 10 feet Hi-temp GP composite 0 Integral C 1 316/316LSS Flange 1 316/316LSS Flange ANSI Flange 150#, 300# and 600# D V 3" 600# C Hastelloy C276 NPT or Tri-Clamp 6 Remote - 75 feet cable GP 8 25 Tri-ax 7 100 A Remote - 75 feet cable GP composite Tri-Ax 1 Remote - no cable D Remote - 25 feet Hi-temp GPE Din Flange 2 CS Flange 2 CS Flange W 4" 600# P PFA NPT 7 Remote - 50 feet Tri-ax GP 100 9 8 25 feet cable B Remote - 35 feet Hi-temp GP composite 2 Remote - 10 feet cable GP E Remote 100 feet Tri-Ax Flange 3 Hastelloy C 276 Flange 3 Hastelloy C 276 F ANSI Flange 900#, 1500#, 2500# Sensing Element Insertion Length 8 25 Tri-ax X 6" 600# A Remote - 75 feet Tri-Ax 9 50 3 Remote - 25 feet cable GP F Remote - 50 feet Tri-ax 4 CS Flange with Fitted TFE Face C Remote - 10 feet Hi-temp GP composite 4 CS Flange with Fitted 1 3/4" TFE Face (in millimeters) 9 B Remote - 50 feet Tri-ax 100 Y 8" 600# A D Remote 75 feet Tri-Ax 4 Remote - 35 feet cable GP G Remote - 25 feet Tri-Ax Flange with Fitted 5 316/316L SS Flange with Fitted TFE Face 5- 75 feet Hi-temp GP composite TFE Face 316/316L SS 2 1" Inactive Length or Cote-Shield Length A Remote - 10 feet Hi-temp GP composite Tri-Ax C Remote 35 feet Hi-temp GP composite B Remote 75 feet Tri-Ax 100 1 316/316LSS Flange E 5 Remote - 50 feet cable GP HTFE Face6- 100 feet Hi-temp GP composite 6 Hastelloy C 276 Flange with Fitted Hastelloy C 276 Flange with1/2" TFE Face 3 1 Fitted (in millimeters) B Remote - 25 feet Hi-temp GP composite 100 D Remote - 10 feet Hi-temp GP composite C Remote 50 feet Tri-Ax F 2 CS Flange 6 Remote - 75 feet cable GP A CS NPT or Tri-Clamp A Sensing Element CS NPT or Tri-Clamp 4 2" Cooling Extension C Remote - 10 feet Hi-temp GP composite E 35 D Remote - 25 feet Hi-temp GP composite G Remote - 75 feet Hi-temp GP composite 7 Remote - 100 feet cable GP B 316SS NPT or Tri-Clamp 3 Hastelloy C 276 Flange B Z Z NPT or Tri-Clamp 1 1/4" 316SS Special 5 D Remote - 25 feet Hi-temp GP composite F 50 E Remote - 35 feet Hi-temp GP compositeProbes H Remote - 100 feet Hi-temp NPT or Tri-Clamp 8 Remote - 25 feet Tri-ax C Hastelloy C276 NPT or Tri-Clamp C 1 0 700-0001-0186 TFE Hastelloy C276 GP composite 4 CS Flange with Fitted TFE Face 3" E Remote - 75 feet Hi-temp GP composite G Remote - 35 feet Hi-temp GP composite F 50 Sensing Element 9 Remote - 50 feet Tri-ax P PFA NPT P 1 6 NPT PFA 700-0002-3607 4" 5 316/316L SS Flange with Fitted TFE Face F H Remote - 50 feet Hi-temp GP composite 100 feet Hi-temp GP composite 1 Z 700-0001-022 Z 8 SpecialGP composite A Remote - 75 feet Tri-Ax Sensing Element Insertion G Remote - 75 feet Hi-tempElement Insertion Length Length Sensing A 1" 150# 6 Hastelloy C 276 Flange with Fitted TFE Face G Remote - 75 feet Hi-temp GP compositeProbes Sensing Element 1 9 700-0001-016 TFE H Remote - 100 feet Hi-temp GP composite 1 0 millimeters) 700-0001-018 (in millimeters) B Remote - 100 feet Tri-Ax (in B 1 1/2" 150# A CS NPT or Tri-Clamp H Remote - 100 feet Hi-temp GP composite only) Z 6 Special 2 Z 700-0209-002 (remote Sensing Element 1 0 700-0002-360 C Remote - 10 feet Hi-temp GP composite Inactive Length or Cote-Shield Length C 2" 150# Inactive Length or Cote-Shield Length B 316SS NPT or Tri-Clamp Sensing Element 1 1 700-0001-022 0 Special 2 Z 700-0009-002 TFE Probes Z 8 700-0001-018 (in millimeters) D Remote - 25 feet Hi-temp GP composite (in 150# D 2 1/2" millimeters) Z Z Special 1 9 700-0001-016 TFE Probes 6 700-0002-360 C Hastelloy C276 NPT or Tri-Clamp 2 2 700-0001-026 316 SS 0 700-0001-018 E Remote - 35 feet Hi-temp GP composite Cooling Extension Cooling Extension E 1" 300# 1 6 700-0001-022 TFE Probes 1 0 700-0001-018 (remote Steel 2 8 700-0209-002 Carbon only) 3 700-0002-360 700-0001-026 P PFA NPT F Remote - 50 feet Hi-temp GP composite 9 700-0001-016F 1 1/2" 300# 1 6 700-0002-360 316 SS 1 700-0009-002 2 8 700-0001-022 4 700-0001-074 Sensing Element Insertion Length G Remote - 75 feet Hi-temp GP composite 1 8 700-0001-022G 2" 300# 2 0 700-0209-002 (remote only) 1 9 700-0001-016 Carbon Steel 2 700-0001-026 2 5 700-0001-074 316 SS (in millimeters) H Remote - 100 feet Hi-temp GP composite 1 9 700-0001-016H 2 1/2" 300# 2 1 700-0009-002 (remote Steel 0 700-0209-002 3 700-0001-026 Carbon only) 2 6 700-0001-074 Hastelloy C Sensing Element Inactive Length or Cote-Shield Length 3" 150# 2 0 700-0209-002I(remote only) 2 700-0001-026 2 4 700-0001-074 316 SS 1 700-0009-002 2 7 700-0001-034 316 SS Z Z Special (in millimeters) 3" 300# 2 1 700-0009-002JCarbon Steel 3 2 2 700-0001-026 316 SS 2 5 700-0001-074 316 SS 4 0 700-0002-024 Carbon Steel 1 0 700-0001-018 TFE Probes Cooling Extension 2 2 700-0001-026K 4" 150# 316 SS 2 4 700-0001-074 Carbon Steel 2 6 700-0001-074 Hastelloy C 4 3 700-0001-026 1 700-0002-024 1 6 700-0002-360 2 3 700-0001-026L 4" 300# 2 5 700-0001-074 Carbon Steel 4 700-0002-224 316 SS 2 7 1 8 700-0001-022 4 2 700-0001-034 2 4 700-0001-074M 6" 150# 316 SS 2 6 700-0001-074 Hastelloy C 0 1 9 700-0001-016 4 5 700-0002-027 Carbon Steel 3 700-0002-024 316 SS 2 5 700-0001-074N 6" 300# 2 7 700-0001-034 Hastelloy C 6 700-0001-074 Carbon Steel 1 700-0002-024 2 0 700-0209-002 (remote only) 4 4 700-0002-028 Hastelloy C 2 6 700-0001-074P 8" 150# 4 7 700-0001-034 316 SS 0 700-0002-024 2 2 700-0002-224 Hastelloy C 2 1 700-0009-002 4 5 700-0002-057 2 7 700-0001-034Q 8" 300# 4 3 700-0002-027 Hastelloy C 1 700-0002-024 0 700-0002-024 2 2 700-0001-026 316 SS 4 6 700-0002-036 316 SS 4 0 700-0002-024R 1" 600# 4 2 700-0002-224 316 SS 1 700-0002-024 4 4 700-0002-028 Hastelloy C 2 3 700-0001-026 Carbon Steel 7 700-0002-037 4 1 700-0002-024S 1 1/2" 600# 4 3 700-0002-027 Hastelloy C 5 700-0002-057 2 4 700-0001-074 316 SS 4 2 700-0002-224 8 700-0002-056 4 2 700-0002-224T 2" 600# 4 4 700-0002-028 6 700-0002-036 2 5 700-0001-074 Carbon Steel 4 3 700-0002-027 9 700-0002-227 4 3 700-0002-027U 2 1/2" 600# 4 7 700-0002-037 4 700-0002-028 4 5 700-0002-057 2 6 700-0001-074 Hastelloy C 6 0 700-0204-038 (remote only) 4 4 700-0002-028V 3" 600# 4 6 700-0002-036 5 700-0002-057 4 8 700-0002-056 2 7 700-0001-034 6 4 700-0001-052 4 5 700-0002-057W 4" 600# 4 7 700-0002-037 6 700-0002-036 4 9 700-0002-227 4 0 700-0002-024 316 SS 6 5 700-0001-062 4 6 700-0002-036X 6" 600# 4 8 700-0002-056 7 700-0002-037 6 0 700-0204-038 (remote only) 4 1 700-0002-024 Hastelloy C 6 700-0002-042 4 7 700-0002-037Y 8" 600# 4 9 700-0002-227 4 700-0001-052 4 2 700-0002-224 6 8 700-0002-056 7 700-0002-052 1 4 8 700-0002-056 (remote316/316LSS Flange 6 0 700-0204-038 only) 4 9 700-0002-227 5 700-0001-062 4 3 700-0002-027 6 8 700-0002-062 2 CS Flange 4 9 700-0002-227 6 4 700-0001-052
4 6
9 0
RCT RF Series 13 Universal III Mark II RCT 10, 11, 12, RF Series Models RF Series RF Series RF Series
Available
R
1
C
2
11
12
0 C
10
13
14
15
R
1
10
11
12
13
14
15
Introduction
1.4
13XXXXX18 Water-like conductive liquids 12XXXXX23 Concentric waterlike insulating liquids 13XXXXX40 Low viscosity conducting liquids
TFE-covered Concentric Shield 100oF @ 1000 PSI rod with carbon 1.66" OD 300oF @ 500 PSI steel concentric 1 1/2" NPT shield TFE-covered rod FEP-covered rod X*-covered rod X*-covered rod X*-covered cable Urethanecovered cable PFA-covered cable Rod 3/4" OD 3/4" NPT Rod .56" OD 3/4" NPT Rod .54" OD 3/4" NPT Rod .84" OD 1" NPT Cable 5/16" OD 3/4" NPT Cable 3/4" OD 2" NPT Cable .093" OD 3/4" NPT Cable 5/16" OD 1" NPT 100oF @ 1000 PSI 450oF @ 500 PSI 100oF @ 1000 PSI 300oF @ 500 PSI 100oF @ 1000 PSI 250oF @ 500 PSI 100oF @ 1000 PSI 250oF @ 500 PSI 100oF @ 1000 PSI 250oF @ 500 PSI 150oF @ 5 PSI 100oF @ 1000 PSI 300oF @ 500 PSI 250oF @ 5 PSI
700-2-24
12XXXXX43 Interface of liquids 700-2-27 containing ketones and esters 13XXXXX47 Thick conducting liquids 13XXXXX45 Conducting liquids and interfaces 13XXXXX70 Agitated conducting liquids and granulars 10XXXXXXX Heavy-duty for abrasive granulars 13XXXXX74 Long lengths of conducting liquids 12XXXXXXX Insulating liquids and granulars 700-2-37 700-2-57 700-5-18 700-5-19 700-5-54
*X is a fluorocarbon-type insulation
1.5
Area Classifications
The standard electronic unit mounted in the durable housing (RCTXXX1, 2) is dual-rated and meets the following conditions: Type NEMA 4X Waterproof / Corrosion Resistant IP 65 See Section 1.4 for detailed specifications of sensing elements that are most often recommended with a RCT Series system. Contact the factory or your local representative if additional information is required. The electronic unit and sensing element are connected by a three-terminal coaxial cable. Drexelbrook cables are available in See Section 7.2 for Specifications: General Purpose: Triax: Composite: (first 10 feet high temperature) 3
Installation
SECTION 2: INSTALLATION
2.1 Unpacking
Carefully remove the contents of the carton and check each item against the packing list before destroying any packing material. If there is any shortage or damage, report it immediately to the factory.
2.2
2.3
2.3
Packing Gland CONDULET Recommended Conduit Assembly Do Not Disturb! Connection Hold here while tightening condulet. Hold here to install or remove sensing element from vessel.
Figure 2-1
WRONG
Allows Moisture Infiltration Use only cable supplied by AMETEK Drexelbrook
CORRECT
CONDUIT BREATHER DRAIN
WRONG
CORRECT
Figure 2-2
Installation
2.3
2.3
LACK OF PROPER GROUND (EARTH) CONNECTION BETWEEN SENSOR MOUNTING AND VESSEL WALL
STILLING WELL LACKS VENT HOLES PROBE CONTACTS SIDE OF STILLING WELL PROBE MAY FLEX INSIDE PIPE
Installation
2.3
If the vessel is non metallic see grounding sketch below. sensors with factory supplied concentric shield or ground rod do not need additional grounding. Measure ground continuity from housing ground screw to metal wall of vessel. A good ground will measure less than 5 ohms.
RUN 14 AWG. WIRE FROM FABRICATED GROUND TO GREEN SENSOR GROUND SCREW
A
GROUND ROD A ground rod can be fabricated out of any metal compatible with the process. Use material that is at least 3/8inch dia. such as pipe, All-Thread, or tubing. The ground rod must be parallel and relatively close to the sensor.
*This is a sensing element ground reference and possibly different from an electrical power ground.
B
CLAMP
GROUND WIRE 1/4 inch or larger dia. stainless steel rope that is anchored or weighted can be used.
C
MOUNT SENSOR IN METAL PIPE Mounting the level sensor inside a metal pipe provides an excellent ground reference.
PLASTIC TANK
D
METAL PIPING Metal piping that connects to the tank bottom can be used as a ground reference. Use only if the process material is greter than 1000 uMHOS/cm such as acids and caustic.
E
WEIGHT
SUBMERGED METAL STRUCTURE Use any constantly submerged metal object such as: pumps, agitators, or thermowells. Use only if the process material is greter than 1000 uMHOS/cm such as acids and caustic.
F
SENSOR
PUMP
METAL
F
METAL FLANGE A submerged metal flange or orifice plate can be used. Use only if the process material is greater thatn 1000 uMHOS/Cm. such as acids and caustic.
GROUND CLAMP
2.4
Universal III - MK II
GND MVV+
Universal III - MK II
Universal III - MK II
+ To Signal
- To Signal
Universal III - MK II
GND
M-
V-
V+
+ To Signal - To Signal
1. All devices must be wired in series. Voltage driven devices require a series voltage dropping resistor. 2. Signal Terminals + & - can operate with minimum of 12.0 VDC with 20mA flowing. 3. If the field wiring is to be in hazardous areas, then suitable safety barriers are required between the power supply and the field to provide for intrinsically safe wiring. 4. Clipped Shield Wire must NOT touch housing!
Figure 2-8
10
Installation
2.4
Integral units are pre-wired at the factory. Figure 2-7 shows the wiring of the integral unit. For remote units, the signal connections are made to the four-terminal block on the front of the chassis. Due to the low power consumption of the instrument, the wiring need only be light gauge (e.g. 20 AWG). Twisted shielded pair cables are recommended for lengths over 200 feet. The cable from the sensing element is connected to the five-terminal strip on the back side of the instrument chassis. The cable connections are probe (PRB) or center wire (CW), ground (GND), and shield (SHD). See Figure 2-8 for wiring connections of the remote unit. Only coaxial cables supplied by Drexelbrook should be used to connect the transmitter to the sensing element. Use of other cables can result in unstable performance.
2.5
11
2.6
Figure 2-9
Ground (Green)
Spark Protection for Remote Sensing Elements If spark protection is supplied for a remote sensing element, use the following instructions for installing the spark protection. A. Attach the mounting link on the spark protector to the sensing element center connection screw. B. Connect the green wire from the spark protector to the ground screw. C. Feed the coax cable into the condulet.
12
Installation
2.6
Make sure the transmitter has a ground attached either on sensing element side or loop side of the unit. Unless the assembly is attached to a metallic vessel, chassis of transmitter is not grounded.
Figure 2-10
2.7
13
2.7
Figure 2-11
2.8
Ground the electronic unit and housing with a minimum of 14 gauge wire to a good earth ground. Make sure that conduits entering and leaving the housing have a good electrical ground connection to the housing
14
Installation
2.7
Universal III - MK II
GND MVV+
PAD GROUND
SHD
PROBE
+ To Signal
SERIAL NO. MODEL NO. YEAR OF MANUFACTURE
- To Signal
Universal III - MK II
SHLD GND CW
15
Figure 2-13
Installation
The meter is not inserted into the 4-20 mA loop. It receives power and data directly from the Drexelbrook smart transmitter via attached mini ribbon cable. See Figure 2-14. If the Ribbon Cable is attached while the transmitter is Powered Up, the transmitter may have to be restarted before the DIM will display correctly. The meter can be configured using the Drexelbrook PC software, the Model 275 Calibrator, or from the DIM itself.
Figure 2-14
17
SECTION 3:
3.1
General Description
The 401-700-20/40 software package allows the use of any Windows 9X/NT/2000/XP-based personal, laptop, or notebook computer to calibrate the HART Protocol transmitter. The PC software can be used in place of the Rosemount 268 or 275 handheld calibrators used for multi-Process Variable (PV) transmitters.
3.2
Model Number
4 0 1 - 0 7 0 0 - 0 2 X/4 X 2X= 1 PC Software Package includes: RS232 Modem Assembly 401-0700-004 (Figure 3.1). 2X=2 PC Software Package includes: Contents in 401-0700-021, HART 6.0 (DOS version) on a 312" Floppy Disk, and HARTWin version 2.1 or greater on a CD-ROM. 4X=1 PC Software Package includes: USB Modem Assembly 401-0700-007(Figure 3.1a). 4X=2 PC Software Package includes: Contents in 401-0700-41, Utilities and Drivers on a CDROM, and HARTWin version 2.3 or greater on a CDROM. 4 0 1 - 0 7 0 0 - 0 0 6 HART 6.0 (DOS version) on a 312" Floppy Disk 4 0 1 - 0 7 0 0 - 0 3 1 HARTWin version 2.X on a CD-ROM.
3.3
System Requirements
PC Requirements Windows 95, 98, ME, 2000, XP. The USB modem is not compatible with Windows 95, 98 First Edition, or NT. It is recommended that the software be installed on a hard drive with 20 megabytes or more of space available. Input to Modem RS232 or USB Port, from one of the COM serial ports (COM1, COM2, etc.). The PC provides operating power for the modem but not for the transmitter. Output (to Transmitter being Calibrated) 4-20 mA in HART Protocol. 19
3.4
HART COMPATIBLE
401-0700-004
20
Notes:
HART COMPATIBLE
1. Modem will operate from 32 F to 122 F (0 C to 50 C). It can be stored from -40 F to +185 F (-40 C to +85 C). 0% to 95% relative humidity - non condensing. 2. Servic Department 1-800-527-6297 or 1-215-674-1234
USB
401-0700-007
3.5
Figure 3-2
E. Select communication port [Com 1, Com 2, etc.] and then click OK. See Figure 3-2 F. If you are not sure which communication port you are using (such as when first using a USB modem), select Search Ports, then OK. The software automatically will seek out the correct one. In either case the software begins to communicate with the HART protocol transmitter and returns with a view (below) containing name plate data, Tag ID and all default or existing configuration information. This is the same as if you clicked on the Read Transmitter function button. G. The next view, shown in Figure 3-3, appears automatically, displaying current transmitter database for calibration set-up for your selected Tag ID. The Scratch Pad will automatically show the last message (last user, last calibration, etc.) up to 32 characters. If this is a new transmitter, the Tag ID is user-defined. Serial number, transmitter software version, range, etc. is automatically entered from the name plate data embedded in the transmitter:
3.6
22
3.6
PC Software Menu Screen automatically communicates all "name plate data" from transmitter
Figure 3-3
23
3.7
Configuration
Refer to Figure 3-3 PC Software Menu Screen. Configuration involves downloading information to the HART protocol transmitter that is specific to the application and vessel that is being measured. Calibration requires that application information and two points of level and/or capacitance be supplied to the transmitter from the calibration software. A. Begin configuration by using Tag ID (8 characters) to identify the unit or vessel. Use the Scratchpad (32 characters) to record the date of calibration or other similar notes. Press Tab or Enter on your keyboard. B. Select Level or Vessel in the Analog Loop Assign selection box. Press Tab or Enter on your keyboard. Level configuration sets the output to follow the level of the material being measured. Vessel configuration sets the output to follow the strapped volume or weight in the vessel. For example, gallons in a horizontal vessel. C. Edit Damping Time from 0-90 seconds, if desired. D. Click on Write to Transmitter. E. Move to Level Configuration section of menu.
Figure 3-4
3.7.1
Level Configuration
A. Select Level Units. The default is feet. Choose the units that correspond to the level measurement. Figure 3-5 B. Edit the Maximum Level to agree with the actual tank height (not the length of the sensing element). C. Click on Write to Transmitter and move to the Vessel Configuration section of the menu. 24
Figure 3-6
Figure 3-7
A. Edit LRV (Lower Range Value) to display the output you want to see when the transmitter generates 4 mA current. The default LRV is 0 feet. B. Edit URV (Upper Range Value) to display the output you want to see when the transmitter generates 20 mA current. The default URV is 100 feet for Universal III. C. Click on Write to Transmitter. Configuration is now complete. 25
3.8
Calibration
There are two methods for calibrating the transmitter using the PC software: Point Calibration (menu button selection): Uses the two known level points in the vessel for calibration. The further apart the two points are for the calibration the better the accuracy of the overall measurement. Always initiate the point calibration process by selecting the Point Calibration button on the PC menu screen and following the prompts in the pop-up window. Level Calibration (lower right of window): See Figure 3-3 Uses capacitance values obtained from the AMETEK Drexelbrook Service department (or a previous calibration or identical application) for the zero and span calibration data. Call 1-800527-6297. Please provide your DE purchase order number, transmitter serial number, vessel and application data to the Service Engineer. Level calibration is done using the Level Calibration data fields on the PC menu screen.
It is permissible or sometimes even recommended that both methods be used in order to establish a calibration standard. For example, if the vessel was already filled before the calibration was attempted and it is difficult or impossible to lower the level to establish the second point, it would be best to use a calculated zero capacitance for the low point and actual level for the high point. While this wouldnt be as accurate as two known level points, it will be reasonably accurate until an actual low point calibration can be established. The Service department will help in calculating high or low capacitance values. Because calibration involves determining two known points of capacitance, a span (or range) jumper provides an adjustment for the change in capacitance required to produce full scale current. The Range Span Jumper is located on the back of the transmitter chassis. See Fig. 3-9. Each Range Span position on the Universal III advances the range in inches or feet to approximately five times the previous setting. See Table 3-1
Figure 3-9
26
Configuration & Calibration Universal III Mark II Span Range Setting Chart Sensing Element Length Vs. Span Position Number / Maximum pf Jumper Position = Maximum pF = System # 1 20 2 100 3 450 4 2000 5 6 10000 40000
Maximum Sensing Element Length In Feet N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A 5.4 9.2 1.2 1.4 7.8 N/A 5.4 5.4 N/A 1.2 1.0 N/A 20 40 5.3 6.25 20 N/A 20 20 4.2 5.3 4.8 N/A N/A 200 20 31 N/A 4.1 N/A N/A 21 20 20 4.0 N/A 920 N/A 140 N/A 14 N/A N/A 95 N/A N/A 14
Conducting Liquids: RCTX3XXXXX18 N/A RCTX3XXXXX74 N/A RCTX3XXXXX45 N/A RCTX3XXXXX70 N/A RCTX3XXXXX40 N/A RCTX3XXXXX47 N/A RCTX3XXXXX22 N/A RCTX3XXXXX65 N/A RCTX3XXXXX72 N/A Interface Application: RCTX2XXXXX45 N/A RCTX2XXXXX43 N/A RCTX2XXXXX47 N/A Insulating K = 1.5-5: RCTX2XXXXX22 N/A RCTX2XXXXX18 N/A RCTX2XXXXX74 N/A RCTX2XXXXX45 N/A RCTX2XXXXX70 N/A RCTX2XXXXX40 N/A RCTX2XXXXX43 N/A RCTX2XXXXX47 N/A
2.7 12.5 20 N/A N/A 8.3 20 N/A N/A N/A 8.0 37.5 166 N/A N/A 3.6 16.2 20 N/A N/A 3.8 17.1 76 N/A N/A 6.0 20 N/A N/A N/A 5.0 20 N/A N/A N/A 5.0 14 N/A N/A N/A Table 3-1 Span Range Position for a number of common Sensing Elements
Figure 3-10
3.8.2
Level Calibration
Level calibration uses zero and span capacitance values as the calibration data . These values can be obtained from the AMETEK Drexelbrook Service department (or from a previous calibration or identical application). Please be prepared when you call (1-800-527-6297) with the purchase order number and the serial number of the transmitter. A. Go to Level Calibration area of the menu. B. Enter Lower Level value. Press Tab or Enter. C. Enter Lower Capacitance value. Press Tab or Enter. D. Enter Upper Level value. Press Tab or Enter. E. Enter Upper Capacitance value. Press Tab or Enter. F. Click on Write to Transmitter.
Figure 3-11
Maximum Level 20 feet URV = 1185 gallons 20 mA [19.5 feet] High Point 16 feet Hi Cal Low Point 3 feet Lo Cal LRV Alternate* [0 f eet] LRV 4 mA Point
Figure 3-12
*LRV may either reference the bottom of the vessel, bottom of the sensor, or an elevated p oint on the sensor.
Figure 3-13
29
3.9
PC Status Messages
Status Message: SPAN TOO SMALL Difference between URV and LRV is less than 10% of range. Example: For 0 to 10 foot calibration points: LRV=3.0 feet and URV=3.8 feet. When calibration points are too close together, overall accuracy of calibration is adversely affected. Action: The calibration points should be farther apart. Calibration Status Message: RAISE SPAN JUMPER Based on LRV, URV, and capacitance calibration data, the estimated 100% capacitance exceeds selected range by greater than 10%. Example: A unit in Range 4 (2000 pF) projects maximum capacitance equal to 2500 pF. Error message is displayed. Action: Raise Range jumper (Section 3.8, Table 3-1) to position 5 for this example. Calibration Status Message: LOWER SPAN JUMPER Based on LRV, URV, and capacitance calibration data, the estimated 100% capacitance value is less than 10% of the maximum pF for the range below the selected range. Example: A unit in Range Span 4 (2000 pF) projects maximum capacitance equal to 400 pF. Action: Lower Range jumper (Section 3.8, Table 3-1) to position 3 (Max. pF of 450 pF) for this example. Real-time Status Message: UNDERRANGE Present capacitance/milliampere value is less than -5% of range. Examples: Center wire connection is broken. Sensing element is not operating. An elevated Zero is used and actual level is below 4mA point. Vessel has lost its RF ground reference. Action: Check connections and ground. Recalibrate if this level is in operational range of process. Real-time Status Message: OVERRANGE Present capacitance/ milliamp value over 105% of system span. Examples: Actual level exceeds span point on sensing element. Cut in sensing element insulation or shorted coax. Action: Check sensing element and coax. Re-calibrate if this level is in operational range of process. 30
Figure 3-13a
3.10
Figure 3-14
3.11
Strapping Table
31
Table 3-2
1. 2. 3. 4. 5. 6. 7. 8. 9. 10. 11. 12. 13. 14. 15. 16. 17. 18. 19. 20. 21.
0 5 10 15 20 25 30 35 40 45 50 55 60 65 70 75 80 85 90 95 100
Figure 3-15
By clicking on Last Read Values, this view may also be used to review existing strapping tables previously entered.
32
Configure Meter Pop-up from Menu Screen; values relate to LRV & URV
Figure 3-16
Figure 3-17
3.13
3.14 Validation
More and more industries are requiring formal validation of their processes for their customers as well as for various government regulatory agencies. The Universal III Mark II Series Transmitter has this capability built in.
Figure 3-18
B. At the Main configuration screen observe the Level Calibration, Lower Capacitance, and Upper Capacitance values and the Lower Level and Upper Level values. Select an NPO 34
Figure 3-19
* Every capacitor manufactured will generate a slightly different capacitance value within its specified tolerance. By marking the capacitor and using only this capacitor for testing and validating the AMETEK Drexelbrook Universal III Transmitter, the system should produce repeatable results within transmitter specifications.
35
3.14.3
Validation Results
See Fig. 3-20. If the information that is shown (or printed) matches the initial readings within system specifications, then it can be verified that the calibration and configuration is as originally set. It can also be verified that the transmitters response falls within acceptable tolerances. The system has passed validation tests. Using the Save/Print feature built into the transmitter allows the ability to comply with the record-keeping needed for many processes by regulatory agencies.
AMETEK Drexelbrook
205 Keith Valley Road Horsham, PA 19044 Telephone: 215-674-1234 FAX: 215-674-2731 Service: 800-527-6297
Tag-ID: Scratch PAD: Analog Loop Assign: Damping Time: Level Units: Maximum Level: Level Type:
Serial Number: Software Version: Range Position: Type: gallons 1200.00 gal Vertical 0.00 ft 10.00ft
1172 3.1 4 30
Level Configuration
feet 10.00 ft Standard
Vessel Configuration
Vessel Units: Maximum Capacity: Vessel Type: LRV (4mA): URV (20 mA):
Level Calibration
Lower Level: 0 ft Upper Level: 10 ft Lower Capacitance: 50 pf Upper Capacitance: 2000 pf
Strapping Table
Vessel Type: Number of Points: Level In
0.00 1.00 2.00 3.00 4.00 5.00 6.00 7.00 8.00 9.00 10.00 11.00 12.00 13.00 14.00 15.00 16.00 17.00 18.00 19.00 20.00
Percent
0.00 5.00 10.00 15.00 20.00 25.00 30.00 35.00 40.00 45.00 50.00 55.00 60.00 65.00 70.00 75.00 80.00 85.00 90.00 95.00 100.00
Percent
0.00 5.00 10.00 15.00 20.00 25.00 30.00 35.00 40.00 45.00 50.00 55.00 60.00 65.00 70.00 75.00 80.00 85.00 90.00 95.00 100.00
Figure 3-20 36
Enter
See? Edit
Up/Down Enter Up/Down Enter Up/Down Enter Enter Enter
Up/Down
Min Value
Enter
Edit 0 - 99900
Enter
Min Value
View Min Value
Max Value
Edit 0 - 99900
Up/Down Enter Up/Down Enter Enter
Max Value
View Max Value
Up/Down
Up/Down
Return
Enter
Return
Enter
See?
View LRV
Enter Up/Down Enter Enter No
Value = 4.0?
Up/Down
Edit
Edit 0 - 99900
Up/Down
Enter
Enter
See?
View URV
Up/Down
Edit
Enter Enter
Up/Down
Up/Down
Up/Down
Edit 0 - 90
Up/Down
Up/Down Enter
Edit
Enter Enter
Enter
Edit 0 - 99900
S.S.HH
Enter
CAP (Capacitance)
Enter
View Capacitance
Enter
See?
View Low Point
Enter
37
Up/Down Sc.PO (Scale Percent Out)
Enter
Up/Down See?
Up/Down
Cal?
Enter Enter
Up/Down
Edit 0 - 99900
Edit
Up/Down
Enter
Up/Down
Enter
Enter
Enter
See?
View High Point
Enter C.CAL (Cap Cal) Up/Down Enter Enter
Enter Up/Down
See?
Up/Down Enter
Enter
Up/Down
Enter
Edit 0 - 99900
Enter
Lower Level
Up/Down
Enter
Cal?
Edit 0 - 99900
tyPE (Type)
Up/Down
Up/Down
Enter Up/Down
Enter
Lower Cap
Up/Down
Enter
Enter
Up/Down
Upper Level
Up/Down
Enter
Enter
Upper Cap
Up/Down
Enter
Edit 0 - 99900
Enter
Upper Cap
Up/Down
Enter
Up/Down
Enter Up/Down
See?
Up/Down
Enter
Enter
Return
Enter
Return Edit
Enter Up/Down
Enter
Enter
Enter
Escape Menu
L.FAC (Load Defaults)
Enter
Sure?
Enter Up/Down
Up/Down
SECTION 4:
4.1
CONFIGURATION AND CALIBRATION WITH MODEL 275 WITH DREXELBROOK DEVICE DESCRIPTION
The Drexelbrook Device Description for the Rosemount 275 handheld calibrator makes it easy to calibrate a Universal III Mark II. The device description software is stored in the memory module (located in the back portion) of the calibrator. To determine if your Model 275 Handheld calibrator has the Drexelbrook Device Description loaded, do the following steps: Turn on the calibrator. From the top screen, push 1. Offline. Push 1. New Configuration A Table of Contents is shown that lists all the Manufacturers in alphabetical order. Drexelbrook is third on the list in the current release. Select the Manufacturer (Drexelbrook) and a list of supported devices is displayed (Universal III). Return to top screen, by backing up from arrow keys. The Memory Module with the device description can be purchased as a direct replacement either from Drexelbrook (401-700-25) or from the local FisherRosemount Service Center. Appendix A describes configuration of the Universal III Mark II transmitter with a Rosemount 275 handheld calibrator without a device description.
39
4.2
Start-up
After the Universal III Mark II transmitter is installed and loop power is applied, per Section 2, do the following: 1. Connect the Rosemount Model 275 as shown in Figure 4-1. 2. Turn on the Calibrator and look for the ONLINE screen to appear. ONLINE means that the 275 Calibrator has recognized the Universal III and is ready for Configuration and Calibration. 3. You must start the process by doing the Configuration first-- followed by Calibration. There are also instructions for configuring the Strapping Tables and for doing a D/A Trim to make the loop output agree with a calibration standard for loop current.
Figure 4-1
+ Nominal
24Vdc Supply 17-30 Vdc
7 ABC
8 DEF
9 GHI
Calibrator or PC Modem may be connected anywhere on the transmitter side of the 250 ohm resistance. Voltage at the transmitter terminals must be at least 12 volts with 20 mA of loop current.
7 ABC
8 DEF
9 GHI
Resistance added if necessary to make total loop resistance at least 250 ohms during the communication process.
9 GHI
7 ABC
8 DEF
9 GHI
7 ABC
8 DEF
9 GHI
7 ABC
8 DEF
9 GHI
7 ABC
8 DEF
7 ABC
8 DEF
9 GHI
40
Configuration
UNIV. III: <TAG> ONLINE 1. LEVEL 2. VESSEL 3. CAP 4. REF. CAP. 5. AO 6. % OF RNG NNNu NNNu NNNpF NNNpF NNNmA NNN% Required only if weight or volume readings are desired.
UNIV. III: <TAG> MAX LEVEL NNNNN XXXXX UNIV. III: <TAG> LRV NNNNN XXXXX UNIV. III: <TAG> URV NNNNN XXXXX
UNIV. III: <TAG> LEVEL CONFIG 1. 2. 3. 4. 5. 6. 7. 8. LEVEL TYPE AAA LEVEL UNITS AAA MAX LEVEL NNNu LRV NNNu URV NNNu DAMP TIME NNNs RANGE POSITION N CHG ANLG LOOP ASGN
UNIV. III: <TAG> LEVEL UNITS AAAAA <select from list>
4.3 Configuration
Start
UNIV. III: <TAG> VESSEL CONFIG PROCESS VARIABLES CONFIGURATION MENU BASIC SETUP DETAILED SETUP REVIEW UNIV. III: <TAG> ENTER MINIMUM METER VALUE NNNNNNNNNN XXXXXXXXXX ENTER METER SPAN VALUE NNNNNNNNNN XXXXXXXXXX UNIV. III: <TAG> BASIC SETUP 1. TAG 2. LEVEL UNITS 3. VESSEL UNITS 4. DEVICE INFORMATION 5. VESSEL TYPE 6. DAMP TIME 1. 2. 3. 4. 5. 6. 7. VESSEL UNITS AAA VESSEL TYPE AAA MAX CAPACITY NNNu DAMP TIME NNNs LRV NNNu URV NNNu CHG ANLG LOOP ASGN
UNIV. III: <TAG> CONFIGURATION 1. LEVEL CONFIG 2. VESSEL CONFIG 3. METER CONFIG 4. STRAPPING TABLE 5. CALIBRATION
UNIV. III: <TAG> DAMP TIME NNNNNs XXXXX UNIV. III: <TAG> CURRENT ANALOG LOOP ASSIGN IS AAAAA
UNIV. III: <TAG> ONLINE 1. DEVICE SETUP 2. LEVEL 3. VESSEL 4. CAP 5. REF. CAP. 6. AO 7. % OF RNG 8. LRV 9. URV UNIV. III: <TAG> DETAILED SETUP 1. HART OUTPUT 2. SIGNAL CONDITION 3. CALIBRATION
1. 2. 3. 4. 5.
41
AAAAA AAAAA AAAAA AAAAA AAAAA UNIV. III: <TAG> REVIEW MANUFACTURER LEVEL UNITS VESSEL UNITS VESSEL TYPE LRU URV DAMP TIME RANGE POSITION ANLG LOOP POLL ADDR TYPE DISTRIBUTOR MODEL SERIAL # TAG DATE DESCRIPTOR SCRATCH PAD DEV ID UNIVERSAL REV FLD DEV REV SOFTWARE REV HARDWARE REV
UNIV. III: <TAG> SELECT ANALOG LOOP ASSIGN 1. LEVEL 2. VESSEL 3. EXIT
LEGEND
UNIV. III: <TAG> DEVICE INFORMATION 1. DISTRIBUTOR 2. MODEL 3. SERIAL # 4. TAG 5. DATE 6. DESCRIPTOR 7. SCRATCHPAD 8. DEV ID 9. REVISION #'S
UNIV. III: <TAG> HART OUTPUT 1. POLL ADDR 2. NUM REQ PREAMS
AAA - ALPHABETICAL DISPLAY NNN - NUMERIC DISPLAY XXX - FIELD MAY BE EDITTED u - UNITS pF - PICOFARADS mA - MILLIAMPS s - SECONDS % - PERCENT
OPTIONAL FUNCTION
OTHER FUNCTION
1. CHG ANLG LOOP ASGN 2. LRV NNNu 3. URV NNNu 4. DAMP TIME NNNs 5. % OF RNG NNN%
4.3
Configuration (Continued)
Following is the keystroke sequence for Configuration using the Model 275 Calibrator. Select Device Setup. Select Configuration Menu. Select Level Config. Select Level Typeedit Level Typereturn to Level Config screen. Select Level Unitsedit Level Unitsreturn to Level Config screen. Select Max Leveledit Max Levelreturn to Level Config screen. Select LRVedit LRVreturn to Level Config screen. Select URVedit URVreturn to Level Config screen. Select Damp Timeedit Damp Timereturn to Level Config screen. Select Chg Anlg Loop Assignedit Current Loop Assign. If current loop assign is Level and is correct, go to next screen and select Exit. Proceed to 4.3 Calibration. If Vessel configuration is to be selected, choose Vessel, go back to Config screen and select Vessel Config. Edit all values as done for Level Config. Select Exit and proceed to 4.4 Calibration.
4.4
Calibration
There are two methods of calibrating the Universal III transmitter: Point Calibration or Capacitance Calibration. Point calibration uses the actual level in your vessel for calibration. The further apart the two points are for the calibration, then the better the accuracy of the overall measurement. Capacitance calibration uses values obtained from the Drexelbrook Service department (or a previous calibration or identical application) for the zero and span calibration data. Call 1-800-527-6297 for assistance. Please provide the purchase order number, transmitter serial number and application information to the Service Engineer. It is permissible or sometimes even recommended that both methods be used in order to establish a calibration standard. For example, if the vessel was already filled before the calibration was attempted and it is difficult or impossible to lower the level to establish the second point, 42
4.4
Calibration (Continued)
it would be best to use a calculated zero capacitance for the low point and actual level for the high point. While this wouldnt be as accurate as two known level points, it will be reasonably accurate until an actual low point can be established. The Service department will help in calculating high or low capacitance values.
UNIV.III: <Tag> POINT CAL 1. LOW POINT CAL 2. HIGH POINT CAL
UNIV.III: <Tag> CALIBRATION 1. STRAPPING TABLE 2. D/A TRIM 3. POINT CAL 4. CAPACITANCE CAL 5. TRUE LEVEL 6. CHNG ANLG LOOP ASGN
UNIV.III: <Tag> CAPACITANCE CAL 1. LOWER LEV NNNu 2. LOWER CAP NNNpF 3. UPPER LEV NNNu 4. UPPER CAP NNNpF
43
44
Strapping Table
UNIV.III: <Tag> ENTER POINT NUMBER X X UNIV.III: <Tag> ENTER NUMBER OF POINTS FOR TABLE X X UNIV.III: <Tag> POINT NUMBER X ENTER INPUT LEVEL VALUE: XXX XXX UNIV.III: <Tag> SELECT EDIT METHOD 1. EDIT 1 POINT 2. EDIT ALL POINTS 3. EXIT UNIV.III: <Tag> POINT NUMBER: X ENTER OUTPUT CAPACITY VALUE: XXX XXX UNIV.III: <Tag> POINT NUMBER: X ENTER INPUT LEVEL VALUE: XXX XXX UNIV.III: <Tag> POINT NUMBER: X ENTER OUTPUT CAPACITY VALUE: XXX XXX
D/A Trim is NOT a calibration! This is a pre-calibrated alignment to precision factory settings and is rarely in need of change. The procedure is intended only as a slight "meter" adjustment to a known external reference.
Strapping Table
UNIV.III: <Tag> POINT NUMBER: X ENTER INPUT AS A % OF MAXIMUM LEVEL: XXX XXX
UNIV.III: <Tag> POINT NUMBER: X ENTER OUTPUT AS A % OF MAXIMUM CAPACITY: XXX XXX
UNIV.III: <Tag> POINT NUMBER: X ENTER OUTPUT AS A PERCENT OF MAXIMUM CAPACITY: XXX XXX
45
UNIV.III: <Tag> TABLE EDIT/REVIEW: 1.CHANGE # POINTS 2. EDIT POINTS 3. REVIEW POINTS 4. SEND POINTS 5. EXIT
UNIV.III: <Tag> POINT #X: INPUT: XXX (UNIT) % INPUT: XXX% OUTPUT: XXX (UNITS) % OUTPUT: XXX%
UNIV.III: <Tag> CALIBRATION 1. STRAPPING TABLE 2. D/A TRIM 3. POINT CAL 4. CAPACITANCE CAL 5. TRUE LEVEL 6. CHNG ANLG LOOP ASGN
UNIV.III: <Tag> SELECT EDIT METHOD 1. EDIT 1 POINT 2. EDIT ALL POINTS
UNIV.III: <Tag> POINT NUMBER: X ENTER INTPUT AS A % OF MAXIMUM LEVEL: XXX XXX
UNIV.III: <Tag> POINT NUMBER X HAS BEEN MODIFIED. WAIT AS POINT IS WRITTEN TO TRANSMITTER.
D/A Trim
UNIV.III: <Tag> SETTING FLD DEV OUTOUT TO 4mA UNIV.III: <Tag> ENTER METER VALUE XXXX UNIV.III: <Tag> FLD DEV OUTPUT 4mA EQUAL TO REFERENCE METER? 1. YES 2. NO UNIV.III: <Tag> SETTING FLD DEV OUTPUT TO 20mA UNIV.III: <Tag> ENTER METER VALUE XXXX
UNIV.III: <Tag> FLD DEV OUTPUT 20mA EQUAL TO REFERENCE METER? 1. YES 2. NO
1. 2. 3. 4. 5. 6. 7. 8. 9. 10. 11. 12. 13. 14. 15. 16. 17. 18. 19. 20. 21.
0 5 10 15 20 25 30 35 40 45 50 55 60 65 70 75 80 85 90 95 100
46
SECTION 5:
Configuration & Calibration Of Universal III With Model 275 Communicator Without A Device Description
The Model 275 Communicator can be used to enter and download configuration/ calibration values to the Universal III transmitter.
5.1
Range/Span Control
Definitions: LRV - Lower Range Value of control loop. LSL - Lower Sensor Limit corresponding to minimum level point. PV - Process Variable; level value from transmitter. SV - Secondary Variable; could be volume or weight-based on PV. URV - Upper Range Value of control loop. USL - Upper Sensor Limit corresponding to maximum level point. The first step in configuration is to identify the span range position of the instrument. There is a jumper on the back of the electronics chassis: Range Span See Fig. 3-9. The Range Span provides continuous adjustment of the change in capacitance required to produce full scale current. Each Range Span position advances the range in inches or feet to approximately five times the previous setting. Table A-1 gives the range span position for a number of common sensing elements and range of measurement. Probe Length Vs. Span Position Number / Maximum Pf Jumper Position: Max Pf System 1 20 2 100 3 450 4 2000 5 6 10000 40000
Maximum Probe Length In Feet N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A 2.7 8.3 8.0 3.6 3.8 6.0 5.0 5.0 5.4 9.2 1.2 1.4 7.8 N/A 5.4 5.4 N/A 1.2 1.0 N/A 12.5 20 37.5 16.2 17.1 20 20 14 20 40 5.3 6.25 20 N/A 20 20 4.2 5.3 4.8 N/A 20 N/A 16 20 76 N/A N/A N/A N/A 200 20 31 N/A 4.1 N/A N/A 21 20 20 4.0 N/A N/A N/A N/A N/A N/A N/A N/A N/A 920 N/A 140 N/A 14 N/A N/A 95 N/A N/A 14 N/A N/A N/A N/A N/A N/A N/A N/A
Table 5-1
Conducting Liquids RCTX3XXXXX18 N/A RCTX3XXXXX74 N/A RCTX3XXXXX45 N/A RCTX3XXXXX70 N/A RCTX3XXXXX40 N/A RCTX3XXXXX47 N/A RCTX3XXXXX22 N/A RCTX3XXXXX65 N/A RCTX3XXXXX72 N/A Interface Application RCTX2XXXXX45 N/A RCTX2XXXXX43 N/A RCTX2XXXXX47 N/A Insulating K=1.5-5 RCTX2XXXXX22 RCTX2XXXXX18 RCTX2XXXXX74 RCTX2XXXXX45 RCTX2XXXXX70 RCTX2XXXXX40 RCTX2XXXXX43 RCTX2XXXXX47 N/A N/A N/A N/A N/A N/A N/A N/A 47
5.2
5.3
Tag ID
To name or identify the transmitter, use the 8-character Tag ID. The Tag ID is entered or changed in the following menus. a) b) c) d) e) f) g) h) i) a) b) c) d) e) f) g) h) i) j) k) select the ONLINE MAIN menu 1 DEVICE SETUP 3 BASIC SETUP 1 TAG change (use alphanumeric keys) Tag ID F4 ENTER F2 SEND F4 OK F4 OK or: select the ONLINE MAIN menu 1 DEVICE SETUP 3 BASIC SETUP 4 DEVICE INFORMATION 4 TAG change (use alphanumeric keys) Tag ID F4 ENTER F2 SEND F4 OK F4 OK F3 HOME 48
5.4
Set Up Procedures
49
5.4
50
5.4
51
5.5
52
5.6
Observe on the home screen that your PV URV or PV LRV value has been changed by the calibration procedure. You must manually change these values back. Re-enter your desired PV URV and PV LRV at this time. This completes the first calibration point of the two-step process. For the second point, change the level in your tank and repeat the entire above procedure. 53
5.7
Observe on the home screen that your PV URV or PV LRV value has been changed by the calibration procedure. You must manually change these values back. Re-enter your desired PV URV and PV LRV at this time. This completes the first calibration point of the two-step process. Change the value on the C-box to correspond to the Picofarads of a full tank and repeat the entire above procedure for a high point calibration. 54
Company City Customer P.O. Number Filled out by: Date State Item Phone DE# Tag No. Fax
Material being Measured - Fill out any known information Name of Material: Dielectric Constant: (K) Conductivity: (g) Other: Level Measurement Vessel Shape Interface Measurement Vertical Cylinder Horizontal Cylinder Other
Installation Details NPT Thread Mount Flange Mount- if flange mount B= inches E= A= inches H= inches D= inches H= inches IL = inches Unless specified otherwise, calibration values of zero and span capacitances will be based on 4-20 mA being over the entire range of 'H'. If other values are desired specify LRV and URV in inches. LRV = URV = inches inches
H
B E
IL
URV
LRV D
Calculations by Drexelbrook LRV ________ Calculated Zero Capacitance ________ pF URV ________ Calculated Span Capacitance ________ pF Calculated by: ____________________ Date _________
Phone Fax
800-527-6297 215-443-5117
55
5.8
Point Calibration
Equipment Required: UNIVERSAL III HART Smart Transmitter Model 275 HART Calibrator 24V Power source Approximately 250 ohm total loop resistance Two known process levels applied to the probe This procedure uses an example of a point calibration for full-scale (20mA) = 35 ft. and zero (4mA) = 1.5 ft.
at different times.
[ENTER] Basic Setup Select highlighted unit Home Online (Generic) Desired units are now selected.
56
5.9
Error Message: Upper Range Value Too Low The Full scale projected from the entered URV and LRV is less than 5% of the present capacitance range jumper setting. Example: For 0 to 50 ft. calibration: span select jumper = 1000 pF, LRV = 0 ft. @ 10 pF, URV = 40 ft. @ 80 pF generates this error. Action: The Span Select Jumper Must Be Lowered. This allows the URV-LRV span to be greater than 5% of the capacitance range jumper setting.
NOTE: Respond to the prompt which is displayed immediately after the error message that reads Restore Device Value? with <N>.
57
Troubleshooting
SECTION 6:
Problem/Symptom
TROUBLESHOOTING
Tests in order of probability Check calibrator connections Check for 250 resistance (min.) Check voltage at transmitter Check transmitter Check calibrator connections Check for 250 resistance (min.) Check voltage at transmitter Check transmitter Check calibrator connections Check for 250 resistance (min.) Check voltage at transmitter Check transmitter Try another modem Check voltage at transmitter Check polarity of loop Test Transmitter Check for moisture in head of sensor Test Sensing Element Test Transmitter Check Calibration Test transmitter without sensing element (drift test) Verify proper sensing element ground reference Test Transmitter Check process level Check for Static Discharge Check for radio interference Check Signal Loop Connections Reference Section(s) 6.2 and 6.3 6.6 6.2 and 6.3 6.6 6.2 and 6.5 6.6 6.2 (6.3, 6.4, or 6.5) 6.6 Probable loop problem Faulty connection in loop Often a result of loop connection problems Often a result of loop connection problems Comments Often a result of loop connection problems
Rosemount 268 or 275 Calibrator gives error message that no device was found Rosemount 275 Calibrator gives error message that device could not be identified Cant communicate with transmitter using Drexelbrook PC Software
0 mA output all the time (no measurable output current at any time) More than 20 mA output all the time (output current always exceeds 20 mA) Output drifts (output accuracy varies slowly over timee.g. hours or days) Output erratic - (output jumps around noticeably in terms of seconds or minutes) Output intermittent (output jumps quickly usually between >0mA and some "on scale" value Inaccurate readings (Level readings are incorrect compared to actual known level) Reading does not change with level
6.7 6.6 Section(s) 3, 4, and 5 6.6 Fig. 2-6 6.6 6.9 6.10 6.7 6.8 Erratic readings often show actual process conditions. Look for bubbles or stratification, etc. Intermittent Loop Connection
Have you verified actual level? (At times even sight gauges can be misleading.) Be sure that level is really changing. Possible plugged or unvented stilling well. Probable high point cal/low point cal reversal or inverted interface application.
Comments When process material fills what was originally airspace between the sensor and a nozzle or the vessel, it no longer behaves like a coating. It measures like actual level. Contact Drexelbrook. TFE and stainless steel in the sensor's pressure seal have widely different coefficients of expansion that sometimes permit pressure leaks to occur. Re-torqueing the packing assembly can usually fix the problem. Contact the Service department for the proper torque values and procedure.
6.1
Identifying a Problem/Symptom
Use Table 6-1 as a guide to find and correct a problem when it occurs. Most problems are not related to transmitter failure. It is important to be methodical when tracking down a problem. If you experience a problem that you cannot solve using this guide, call the Drexelbrook 24-hour Service Hot line at 1-800-527-6297 or 215-674-1234. You may also E-mail us at the Internet address: drexelbrook.service@ametek.com. Further service information may be found at our World Wide Web address www.drexelbrook.com. When you contact us, be prepared to give the service person as much information as you can about the model numbers, application requirements, and the materials being measured. At the end of this section, a form is available to organize the information that will help us resolve the problem. Prior to your call, a copy of the completed form can be faxed directly to the Service department at (215) 443-5117.
6.2
60
Troubleshooting
6.3
Start
Calibrator does not communicate with transmitter
2. If loop current is 0 mA, check polarity of wiring at transmitter. If OK check for open loop.
No
3. If loop current is between 0 and 3 mA transmitter is likely bad. Test with procedure in Section 5.6. 4. Check voltage at transmitter, it should be at least 12 volts when 20 mA is flowing in loop.
Yes
5. Disconnect the three probe cable connections at transmitter and retry. If it now communicates, check cable and probe as described in Sections 5.7 & 5.8.
No
1. Check Power Supply source resistance. The 250 ohms may be built in as with the Drexelbrook 401500 series or most DCS inputs. 2. If not, add enough resistance for loop to be at least 250 ohms.
Yes
1. Check calibrator connections to loop per loop drawing Figure 4-1. 2. Check for "noise" and ripple on loop with oscilloscope. Maximum noise level per HART Foundation is 1.2 mV rms (500 - 10 kHz). Maximum ripple (47 - 125 hZ) specification is .2V p-p. Line noise can sometimes be traced to things like motor speed controller wiring in close proximity with transmitter. Noise can sometimes be overcome by increasing loop resistance thereby increasing calibrator signal. Noise effects can also be reduced by connecting calibrator directly at transmitter. Generally noise is only a problem when calibrator is communicating with transmitter and does not affect normal operation of transmitter. 3. Check voltage at transmitter it should be at least 12 V when 20 mA is flowing in loop. 4. If wiring is correct, go to Section 6.6 to test transmitter.
6.4
Rosemount Model 275 Calibrator with device description cannot identify or find device
This condition may be the result of trying to calibrate a transmitter with software less than version 3.0 (transmitters shipped prior to approximately January 1997). There are two options you can use to identify the device and delete the error message. Option 1 - Configure the transmitter as a "generic" device per the instructions in this manual. Option 2 - Contact the Drexelbrook Service department (1-800-527-6297) about upgrading your transmitter to the latest software.
61
6.5
Start
Modem does not communicate with transmitter.
Yes
No
No
Yes
1. Check Power Supply and loop source resistance (the 250 ohms may be built in as with the Drexelbrook 401-500 series or most DCS inputs). 2. If not there add enough resistance for loop to be at least 250 ohms.
No
No
Yes
Is It Polling?
No
Yes
Check: 1. Is modem non-isolated from ground and/or is your laptop plugged into AC power.? If so you may have ground problem. Contact Service department. 2. Some laptops don't provide enough voltage to drive modem correctly from COM Port. Check with modem supplier or try a different laptop. 3. Checkout "Yes" response tests listed below.
Yes
Make it stop polling or power transmitter from a separate source - then re-check operation.
Check: 1. Is correct COM Port selected at startup? 2. Is there any software running that would re-direct COM Port such as Windows, mouse drivers, terminal emulation software, or TSR's. (This can be tested by booting from Drexelbrook Calibration software in the A: drive) 3. Possible bad RS-232 cable or defective modem. 4. Check modem connections to loop per loop drawing on Figure 3-1. 5. Go to Section 5.6 to test transmitter.
1. Check calibrator connections to loop per loop drawing Figure 4-1. 2. Check for "noise" and ripple on loop with oscilloscope. Maximum noise level per HART Foundation is 1.2 mV rms (500 - 10 kHz). Maximum ripple (47 - 125 Hz) specification is .2V p-p. Line noise can sometimes be traced to things like motor speed controller wiring in close proximity with transmitter. Noise can sometimes be overcome by increasing loop resistance thereby increasing calibrator signal. Noise effects can also be reduced by connecting calibrator directly at transmitter. Generally noise is only a problem when calibrator is communicating with transmitter and does not affect normal operation of transmitter. 3. Check voltage at transmitter. It should be at least 12 volts when 20 mA is flowing in the loop. 4. If wiring is OK, go to Section 6.6 to test transmitter.
62
Troubleshooting
6.6
Troubleshooting Transmitter
To troubleshoot the transmitter, use one of the following tests, depending on the device used with your calibration. Some of the following tests require the use of high quality fixed capacitors in the picofarad ranges or a Drexelbrook calibrator box (C-box 401-6-81). Contact the Drexelbrook Service department if you need a C-box. Fixed temperature stable capacitors (NPO types) can often be found at many electronic supply houses. TRANSMITTER TEST - Using a PC or 275 Handheld with device Description 1. Determine the span range currently selected on the electronic unit. See Figure 3-9 2. Using the MAX pF values listed in Figure 6-3, select a capacitance value near the midpoint of the MAX pF range. For example, position #4 has a MAX pF range of 2000 pF. Select a 1000 pF capacitance for this test. 3. Remove all three connections of the coaxial cable at the transmitter's probe terminals. 4. Place capacitor on transmitter from PROBE to GND (ground) terminals See Figure 6-1.
CAP
Always Install to Local Codes / Requirements / Directives as Mandated by the Authority Having Jurisdiction Install Per 420-0004-203-ID
WARNING: To Prevent Ignition of Flammable or Combustible Atmospheres, disconnect Power Before Servicing.
1 2 3 4 RANGE SPAN
6 SPARE
5a. Using a PC - Using the real-time view (F4), verify that the displayed capacitance value is within the value and tolerance printed on the test capacitor. If using a Drexelbrook C-box, be sure to add the standing capacitance of the box (10pF low range, 20 pF normal range). 5b. Using a Rosemount 275 handheld with device description - Viewing the HOME screen, verify that the displayed capacitance value is within the value and tolerance printed on the test capacitor. If using a Drexelbrook C-box, be sure to add the standing capacitance of the box (10pF low range, 20pF normal range).
GROUND PAD
SHD
PROBE
Figure 6-1
6. If the displayed capacitance value is within the stated tolerance, the unit is working. If the displayed capacitance value is not accurate, call 1-800-527-6297. 63
6.6
Transmitter
1 2 3 4 RANGE SPAN
2. Using the C-box, adjust the capacitance until 4 mA is achieved. Record value. See Figure 6-2 3. Adjust C-box capacitance until 20 mA is achieved. Record value. 4. Using Figure 6-3, verify that the capacitance value recorded at 20 mA is less than the MAX pF value for the Span Range Position of the electronic unit.
GROUND PAD
SHD
PROBE
C-Box
5. Verify that the loop configuration is Level Config (signal output linear to level vs. Vessel Config). 6. Subtract pF value recorded for 4 mA from the pF value recorded for 20 mA. Divide this number in half. Example: 4 mA = 120 pF 20 mA = 800 pF 800 pF 120 pF = 680 pF 680 pF 2 = 340 pF 7. Adjust the capacitance box (C-box) to the number figured in step 6. The signal should read approximately 50%.
Figure 6-2
8. If the display reads 50%, the unit is working properly. If the display is not accurate, contact the Service department at 1-800-527-6297 for further assistance. SPAN RANGE MAX pF 1 20 JUMPER POSITION 2 3 4 5 100 450 6
Figure 6-3
Troubleshooting
Transmitter
Universal III - MK II
GND MVV+
mA Meter on Record
24 VDC
Figure 6-4
65
Figure 6-5
Testing the Sensing Element - Step 2 (Figure 6-6) Raise the level in the vessel to cover as much of the sensor as possible. Repeat the measurement made in step 1. Readings of 1 meg-ohm or less indicate a pinhole or crack in the sensing element insulation. Failed insulation is not field repairable. Consult the Service department for further assistance.
Figure 6-6
6.8
Troubleshooting
6.8
OHMMETER
0
8 8
8 8 8
"0" "0" "0"
NO CONNECTION
4. Check for continuity of each conductor. Short out two of the coaxial cable conductors and measure these two conductors at the other end. A reading close to 0 ohms should be shown.
67
6.9
Static Electricity
Static electricity can cause the 4-20 mA output to appear to jump around in an erratic fashion with a time period of few seconds. Applications that are prone to static electricity include insulating liquids that may be agitated or pumped and granulars that may be air-conveyed at high rates of speed. Conductive liquids and conductive granulars tend not to generate static electricity. In addition to causing erratic readings, static electricity can cause instrument failure. If you ever get a static shock from a sensing element, you need spark protection. (See section 2.5) Drexelbrook normally supplies static electricity discharge devices (spark protectors) with its sensing elements that are going to be used in these types of applications. If you need a spark protector, contact the Drexelbrook Service department.
Troubleshooting
Trained field servicemen are available on a time-plusexpense basis to assist in start-ups, diagnosing difficult application problems, or in-plant training of personnel. Contact the service department for further details. Periodically, AMETEK Drexelbrook instrument training seminars for customers are held at the factory. These sessions are guided by Drexelbrook engineers and specialists, and provide detailed information on all aspects of level measurement, including theory and practice of instrument operation. For more information about these valuable workshops, write to AMETEK Drexelbrook, attention: Communications/ Training Group, or call direct + 215-674-1234.
69
Please include a purchase order even if the repair is under warranty. If repair is covered under warranty, you will not be charged.
Ship equipment freight prepaid to: AMETEK Drexelbrook 205 Keith Valley Road Horsham, PA 19044-1499 COD shipments will not be accepted.
70
Troubleshooting
AMETEK Drexelbrook AMETEK Drexelbrook Universal III Mark II Troubleshooting Guide Universal III Mark II Troubleshooting Guide Service Department (800) 527-6297 FAX(215) 443-5117
Company
Fax #
DE #
City/State
Serial #
Serial #
Temp.
Mounting
Other
Provide as much of the following information as possible. All of the information is available from the Drexelbrook Calibration Software, or from a Rosemount Model 275 with Drexelbrook Device Description (DD) installed. Information with an asterisk is available from a Rosemount 268 or 275 in the Generic mode.
AMETEK Drexelbrook HART Protocol Software Version
Vessel Sketch
*Tag ID
*Scratch Pad
Analog Loop Assign
*Damping Time
Level Configuration
*Serial Number
Software Version
Range Position
Type (00/30)
Vessel Configuration
Level Units
Maximum Level
Level Type
*LRV (4mA)
*URV (20mA)
True Level Reference
Vessel Units
Maximum Capacity
Vessel Type
Capacitance Calibration
Lower Level
Lower Capacitance
Upper Level
Upper Capacitance
Show principal tank dimensions, including vessel construction, mounting location, nozzle, LRV, URV, present level, etc.
Level
Vessel
Capacitance
Reference
Loop Current
Percentage
Status
71
Specifications
SECTION 7: SPECIFICATIONS
7.1 Transmitter Specifications
Power Requirement 12 to 30 VDC Minimum of 12 VDC at 20 mA Input Range 409-T: 1.0 to 40,000 pF Output Range 4-20 mA Accuracy 0.25% of range. Accuracy includes the combined effects of linearity, hysteresis, and repeatability. It refers to the transmitter only and is measured at reference conditions of 25 degrees C 1, 1055% R.H. and 24 1.2 Vdc, using an admittance standard (applied to the transmitter sensor terminals) in place of the sensor. Load Resistance Maximum Load Resistance = 750 ohms Minimum Load Resistance With HART Communications = 250 ohms Without HART Communications = 0 ohms Temperature Effect 1% of range per 50F (30C). Supply Voltage Effect < 0.1% from 12 to 30 VDC. Effect of Load Resistance < 0.1% for full resistance range at 24 VDC supply. Response to Step Change < 1 second standard (to 90% of final value); 0-90 seconds available with delay. Fail-Safe Low-Level Fail-Safe (LLFS) standard. Also called direct-acting because current increases as the level increases. High-level Fail-Safe (HLFS). Also called reverse-acting because current decreases as level increases. There Are No Devices That Are Absolutely Fail-Safe".
Fail-safe means that in the event of the most probable failures, the instruments will fail safely. Probable failures include things such as loss of power and transistor and component failures. If your application needs absolute fail-safe, a backup instrument should be installed.
73
Nonincendive Equipment Class I Div. 2, Groups A, B, C, and D; Class II, Div 2, Groups E, F, and G, Class III Div 2 Sensing Element Cable Length 100 feet maximum.
7.2
7.3
Enclosure Specifications
The RCT transmitter and remote sensor enclosures are constructed of aluminum and are rated both NEMA 4X and IP65. Conduit entries are 3/4 NPT with M20 adapters available The enclosures are powder coated to provide additional corrosion resistance.
74
Specifications
270-0101-812
205 KEITH VALLEY ROAD Level Control HORSHAM, PA 19044 Amb Temp -40C...+75C Ui = 30V 0344 Ii = 140 mA II 1 GD EEx ia IIC T4 T135C Pi = 1W KEMA NO. 03 ATEX1249 X Ci = 0.001 uF Li = 0mH SERIAL NO. ACA-XXXXXX 12-30 VDC ENCLOSURE TYPE IP65 4-20mA INSTALL PER 420-004-203-ID T135C
RCT Series
II 1 GD EEx ia IIC T4
270-0101-804
0344
Chassis Only
For indoor locations enclosure must be minimum IP20 rated. For outdoor locations enclosure must be minimum IP44 rated. A higher degree may be required for harsh environments. See Note 1
II 1 G EEx ia IIC T4 0344 KEMA NO. 03 ATEX1249 X Ui : 30V Ii : 140 mA Pi : 1W Ci : 0.001uF Li : 0mH Amb Temp: -40C...+75C 12-30VDC 4-20mA
II 1 G EEx ia IIC T4
Note 1: Note 2:
Note 3:
Always Install to Local Codes / Requirements / Directives as Mandated by the Authority Having Jurisdiction For applications in explosive atmospheres caused by air / dust mixtures, a degree of ingress protection of ot least IP 60 according to EN 60529 will only be achieved when certified cable and conduit entries are used providing a degree of ingress protection of at least IP65 according to EN60529 The aluminum enclosure must be protected from mechanical friction and impact that could cause ignition capable sparks. WARNING: Substitution of Components May Impair Intrinsic Safety. WARNING: To Prevent Ignition of Flammable or Combustible Atmospheres, disconnect Power Before Servicing
270-0101-847
Remote Sensor
75
Normal Maintenance
SECTION 8:
8.1
NORMAL MAINTENANCE
The viewport (if supplied) is made of Borosilicate glass and can be cleaned with any common glass cleaning product (e.g.: Windex, Isopropyl alcohol, etc.) that is suitable for the Class and Division rating of the specific system installation.
Viewport Cleaning
77
Appendix A
A1
A.1
A2
Appendix A
A.1
A3
A.1
A4
Appendix A
A.1
A5
A.1
A6
Appendix A
A.2
A7
A.2
A8
Appendix A
A.2
A9
A.2
A10
Appendix A
A.2
A11
A.2
A12
Appendix A
A.2
A13
A.2
A14
205 Keith Valley Road, Horsham, PA 19044 US and Canada: 1-800-553-9092 International: +1 215-674-1234 24-Hour Service: +1 215-527-6297 Fax: +1 215-674-2731 E-mail: drexelbrook.info@ametek.com Website: www.drexelbrook.com