Вы находитесь на странице: 1из 11

The Effects of Gray Scale Image Processing on

Digital Mammography Interpretation Performance1


Elodia B. Cole, MS, Etta D. Pisano, MD, Donglin Zeng, PhD, Keith Muller, PhD, Stephen R. Aylward, PhD,
Sungwook Park, Cherie Kuzmiak, DO, Marcia Koomen, MD, Dag Pavic, MD, Ruth Walsh, MD, Jay Baker, MD,
Edgardo I. Gimenez, MD, Rita Freimanis, MD

Rationale and Objectives. To determine the effects of three image-processing algorithms on diagnostic accuracy of digi-
tal mammography in comparison with conventional screen-film mammography.
Materials and Methods. A total of 201 cases consisting of nonprocessed soft copy versions of the digital mammograms
acquired from GE, Fischer, and Trex digital mammography systems (1997–1999) and conventional screen-film mammo-
grams of the same patients were interpreted by nine radiologists. The raw digital data were processed with each of three
different image-processing algorithms creating three presentations—manufacturer’s default (applied and laser printed to
film by each of the manufacturers), MUSICA, and PLAHE—were presented in soft copy display. There were three radiol-
ogists per presentation.
Results. Area under the receiver operating characteristic curve for GE digital mass cases was worse than screen-film for all
digital presentations. The area under the receiver operating characteristic for Trex digital mass cases was better, but only with
images processed with the manufacturer’s default algorithm. Sensitivity for GE digital mass cases was worse than screen film
for all digital presentations. Specificity for Fischer digital calcifications cases was worse than screen film for images processed
in default and PLAHE algorithms. Specificity for Trex digital calcifications cases was worse than screen film for images pro-
cessed with MUSICA.
Conclusion. Specific image-processing algorithms may be necessary for optimal presentation for interpretation based on
machine and lesion type.
Key Words. Digital mammography; ROC curve; image processing.
© AUR, 2005

The separation of acquisition from display, in full-field display at each point in the image formation chain. Be-
digital mammography, allows for optimization of image fore display of the final image, some type of image pro-
cessing is applied to the raw digital mammographic im-
age. Ideally, the application of an image processing algo-
Acad Radiol 2005; 12:585–595 rithm will function to improve the visibility of lesions
1 From the Department of Radiology and Lineberger Comprehensive Can- rather than just improving the aesthetic appeal of images.
cer Center (E.B.C., E.D.P.) and Department of Radiology (S.R.A.), University Several studies have specifically evaluated the effect of
of North Carolina, CB#7515, Radiology Research Labs, 106 Mason Farm
Road, Chapel Hill, NC 27599; University of North Carolina, Department of various types of image processing algorithms on radiolo-
Biostatistics, Chapel Hill, NC (D.Z., K.M.); WiBro Terminals Labs, Telecom- gist performance (1–7).
munication R&D Center, Samsung Electronics Co., Gyenoggi-do, Korea
(S.P.); Department of Radiology, University of North Carolina, Chapel Hill, Image processing can take place in several places in
NC (C.K., M.K., D.P.); Department of Radiology, Duke University Medical the imaging chain: at the acquisition station, at the review
Center, Durham, NC (R.W., J.B., E.I.G.); Department of Radiology, Wake
Forest University, Winston-Salem, NC (R.F.). Received October 28, 2004; station, or in between. Generally, the end user presses a
revision received and accepted January 6, 2005. Supported by Susan G. predefined button that will display a processed image.
Komen Foundation Grant. Address correspondence to: E.B.C. e-mail:
ecole@unc.edu
Beyond the individuals who actually developed the algo-
© AUR, 2005 rithms or were responsible for implementing them in the
doi:10.1016/j.acra.2005.01.017 various digital mammography systems, very little is

585
COLE ET AL Academic Radiology, Vol 12, No 5, May 2005

Table 1 collected between 1997 and 1999. The Food and Drug
Lesion Type and Cancer Status for All 201 Cases
Administration had not yet clinically approved any of the
Calcifications Masses full-field digital mammography systems at the time these
Benign Malignant Benign Malignant Normals
data were collected. Each was in various stages of the
clinical trials process. However, the machines used were
Fischer 16 8 22 14 14 the same as the ones ultimately approved for clinical use
GE 10 9 14 10 9
by the Food and Drug Administration. The MUSICA and
Trex 22 8 20 13 12
PLAHE algorithms were applied to all 201 cases. Each
manufacturer’s default algorithm was applied to the cases
acquired using that manufacturer’s digital mammography
system.
known about why specific algorithms were chosen or un-
The algorithms were not always applied successfully
der what conditions they are applied (ie, Is the specific
because of incorrect interpretation of image parameters by
image processing algorithm more suitable for mass char-
the image processing software. Image format standardiza-
acterization or calcification characterization? Is the spe-
tion had not yet been implemented at the start of this
cific image processing algorithm better suited for applica-
tion to specific breast densities? Does the specific algo- study. Therefore, across platforms, the image formats var-
rithm improve lesion visibility or radiologist diagnostic ied considerably from one manufacturer to the next and
accuracy?). The purpose of this study was to quantify the even within manufacturer. Cases acquired from one proto-
effect of algorithms applied to digital mammograms of type system differed in format (byte order, image size,
women with dense breasts acquired from three different gray-scale representation) from another provided by the
digital mammography systems on radiologist performance same manufacturer. Table 2 shows the distribution of
when compared with screen-film mammography. cases that were successfully processed and thus used for
the final reader study. The application of an algorithm
was considered successful for an image if the resultant
image was interpretable in the judgment of a board-certi-
METHODS AND MATERIALS
fied Mammography Quality Standards Act (MQSA)-quali-
A total of 201 cases were obtained from the Interna- fied breast imaging radiologist (E.D.P.). The processing
tional Digital Mammography Development Group image success rate was 93.5% (188/201) for the MUSICA algo-
archive. This case set is described in detail in a previous rithm, 89.1% (179/201) for the manufacturer’s default
article (8). Table 1 shows the breakdown of the cases by algorithms, and 67.2% (135/201) for the PLAHE
lesion type and cancer status. Each case consisted of four algorithm.
standard view digital and screen-film mammograms of the The default algorithms were applied by each manufac-
same patients. The patients were enrolled and imaged at turer to the images acquired on their respective machines
one of seven institutions under the protocols of a previous by applications specialists. The default processed digital
study (8). Three digital mammography systems were test- mammograms were subsequently printed to film for radi-
ed: the SenoScan (Fischer Imaging Corp, Denver, CO), ologist interpretation in the reader study. Agfa Corpora-
the Senographe 2000D (General Electric Corp., Wauke- tion provided a stand-alone version (version 2.0.0) of
sha, WI), and the Trex Digital Mammography System their MUSICA software for use in this study. Research
(Hologic Inc., Bedford, MA). Each patient was imaged on personnel at our institution applied PLAHE and MUSICA
only one type of digital mammography system. Available algorithms to all digital mammograms. For radiologist
screen-film mammography systems at each site were used interpretation in the reader study, PLAHE and MUSICA
to obtain the screen-film mammograms during this study. cases were displayed in soft copy format using a mam-
Three algorithms were applied to each of the digital im- mography soft copy review workstation (Sun platform,
ages: MultiScale Image Contrast Amplification two high-resolution high contrast monitors) running Mam-
(MUSICA), a commercially available algorithm (Agfa- moview software, as described elsewhere (9).
Gevaert N.V., Belgium); Power Law Adaptive Histogram Nine readers participated in the reader study. All had
Equalization (PLAHE), an algorithm developed at our mammography experience, with an average experience of
institution; and the manufacturer’s recommended or de- 11 years (range 1–18 years). Eight of nine readers were
fault algorithm, the algorithms recommended and imple- American Board of Radiology (ABR) -certified attending
mented by each manufacturer at the time this data was radiologists. One reader was a breast imaging fellow. The

586
Academic Radiology, Vol 12, No 5, May 2005 IMAGE PROCESSING AND DIGITAL MAMMOGRAPHY

Table 2
Lesion Type and Cancer Status Case Distributions for Successfully Processed
Cases Used in Reader Study per Image Processing Algorithm

MUSICA PLAHE Default

Calcifications 16 16 16
Benign
Mass 20 19 20
Fischer Calcifications 8 8 8
Malignant
Mass 13 13 13
Normal 10 10 8
Fischer totals 67 66 65

Calcifications 10 8 10
Benign
Mass 13 3 12
GE Calcifications 9 3 9
Malignant
Mass 8 5 9
Normal 9 4 8
GE totals 49 23 48

Calcifications 22 14 22
Benign
Mass 19 10 15
Trex Calcifications 8 4 7
Malignant
Mass 12 10 12
Normal 11 8 10
Trex totals 72 46 66

MUSICA: MultiScale Image Contrast Amplification; PLAHE: Power Law Adaptive His-
togram Equalization.

mograms. Another three readers were assigned to read all


readable cases for PLAHE and to the corresponding
screen film mammograms. Yet another three readers were
assigned to read all readable cases for default and to the
corresponding screen film mammograms. All nine readers
began the reader study by completing the screen film
mammograms interpretations first on a standard radiology
multiviewer appropriately masked for mammography.
After a minimum 4-week washout period, each of the
nine readers then read the digital mammograms to which
they were assigned.
The readers were provided with structured paper forms
to facilitate consistent reporting of mammographic find-
Figure 1. Time spent reading mammograms is between 10 and ings. The readers reported clinically significant lesion lo-
40 hours for our nine readers. On average, 55% of their time is cation (breast, o’clock location, anteroposterior depth),
spent on screening mammography and 45% is spent on diagnos-
lesion-specific Breast Imaging Reporting and Data Sys-
tic mammography.
tems (BI-RADS) (American College of Radiology, Res-
ton, VA) characteristics, and a probability of malignancy
average amount of time spent by the readers as a group based on a 5-point scale (1 definitely not malignant, 2
interpreting mammograms per week was 28.11 hours probably not malignant, 3 possibly malignant, 4 probably
(range 10 – 40 hours/week; 55% screening, 45% diagnos- malignant, and 5 definitely malignant). Responses for le-
tic) (Fig 1). sion location and probability of malignancy were com-
Three readers were assigned to read all readable cases pared with ground truth based on biopsy or 1-year fol-
for MUSICA and to the corresponding screen film mam- low-up using methods described elsewhere (8). If the

587
COLE ET AL Academic Radiology, Vol 12, No 5, May 2005

reader determined there were no significant findings, no cept for the Trex mass cases. The variance of the AUC
probability of malignancy was recorded. difference of digital and screen film mammography was
Three primary outcomes were analyzed separately: 1) relatively large between Fischer digital mammograms and
area under the receiver operating characteristic (ROC) screen film mammograms. Likewise, the variance of the
curve (AUC) from a nonparametric ROC analysis of can- AUC difference was relatively large between GE digital
cer/no cancer; 2) sensitivity, with no findings and 1–2 mammograms and the screen-film mammograms (Fig 6).
defined as “benign” and 3–5 as “malignant”; and 3) speci- Tests based on the multivariate analysis of AUC data
ficity, with no findings and 1–2 defined as “benign” and show that, on average, the digital mammograms produced
3–5 as “malignant.” For each outcome, a general linear by the GE system showed a significantly worse AUC than
multivariate model analysis was used to fit data and re- the screen film mammograms (P ⫽ .009). In addition, the
peated measures tests based on the Geisser-Greenhouse digital mammograms across all machine types showed a
test were used to test the significance of main effects as worse AUC for both mass (P ⫽ .007) and calcifications
well as interactions among the factors of machines, lesion (P ⫽ .042) lesion types (Table 3).
types, and digital display methods. To study which of the machine type, lesion type, and
In the multivariate linear model, the outcome was the processing method variables produce the difference between
repeated measurements of outcomes (AUC difference, digital mammography and screen film mammography, multi-
sensitivity of specificity differences) of a particular reader variate analysis was conducted of AUC data including the
in different conditions of three machines and two lesion interactions among these factors. Trex digital mammography
types; the design matrix included a three-level categorical produced a larger AUC difference with screen film mam-
predictor of three display methods and the baseline per- mography than Fischer (P ⫽ .019), but no significant differ-
formance with film screen was also adjusted in the model. ence was observed between GE and Fischer (P ⫽ .193).
“Main effects” refer to the overall effects of each factor.
There was no significant difference seen among the three
The interaction factors were put in the model at the same
processing methods (MUSICA, PLAHE, default) (P ⫽
time. All the results from the residual analysis indicated a
.0.168, P ⫽ .485) or between the two lesion types (P ⫽
goodness of fit and showed evidence of validity of nor-
.0.822). The resulting 95% confidence intervals for AUC
mality assumption. Thus the identity link function we
difference are shown in Fig 6. GE mass cases processed
used in the model fits the data well. Data analysis was
with each of the three image processing methods produced a
performed using SAS Software, Version 8.0 (SAS Insti-
significantly worse AUC than screen film. Trex mass cases
tute, Cary, NC).
processed with default produced significantly better AUC
than screen film. All other digital combinations were statisti-
cally indistinguishable from screen film.
RESULTS
ROC curve analysis was performed. The outcomes
measured were area under the ROC curve, sensitivity, and Sensitivity
specificity. Figure 2 shows the ROC curves for digital The sensitivity difference between digital mammogra-
and screen film mammography by machine type and le- phy and screen-film mammography was calculated. Table
sion type across all image processing algorithms. The 3 reports the average of this difference by machine type
ROC curves for digital and screen film mammography by and lesion type and shows that the sensitivity from GE
machine type and lesion type per image processing algo- digital mammography was lower than the sensitivity from
rithm are displayed in Fig 3–5. screen film mammography. However, the opposite was
seen when digital mammograms were generated from the
Trex system for both lesion types and for the Fischer sys-
AUC tem for calcifications. The variation of the sensitivity dif-
ference was relatively large for the digital mammograms
The primary outcome of interest was the area under generated by Fischer and GE systems and screen film
the ROC curve difference between digital mammography mammograms.
and screen film mammography. Table 3 reports the aver- Tests based on multivariate analysis of sensitivity data
age of this difference by machine type and lesion type. shows that, on average, GE digital mammography had
The AUC for digital mammography was worse than significantly less sensitivity than screen film mammogra-
screen film for all machine and lesion combinations ex- phy (P ⫽ .011); and, on average, the digital mammo-

588
Academic Radiology, Vol 12, No 5, May 2005 IMAGE PROCESSING AND DIGITAL MAMMOGRAPHY

Figure 2. Area under the receiver operating characteristic curves by machine and lesion type
across all image processing algorithms; value presented in parentheses for each modality.

grams across all machine types had less sensitivity for observed between GE and Fischer. There was no signifi-
mass lesion type (P ⫽ .014) (Table 1). cant difference seen between the three processing meth-
To study which combinations of the machine type, ods (P ⫽ .837 and P ⫽ .745) and between the two lesion
lesion type, and image processing algorithm factors pro- types (P ⫽ .229). The resulting 95% confidence intervals
duce the difference between digital mammography and for sensitivity difference are shown in Fig 7. GE mass
screen film mammography, multivariate analysis was con- cases processed with each of the image processing algo-
ducted of sensitivity data including the interactions among rithms had significantly lower sensitivity than screen film.
these factors. Trex produced a larger variation in sensitiv- There was no statistically significant sensitivity difference
ity difference with screen film mammography than Fi- between all other remaining digital combinations and
scher digital (P ⫽ .005), but no significant difference was screen film.

589
COLE ET AL Academic Radiology, Vol 12, No 5, May 2005

Figure 3. Area under the receiver operating characteristic curves by machine and lesion type av-
eraged across the three default readers; value presented in parentheses for each modality.

Specificity average, the digital mammograms across all digital sys-


The specificity difference between digital mammogra- tems gave less specificity for calcifications (P ⫽ .003)
phy and screen film mammography was calculated. Table (Fig 8).
3 reports the average of this difference by machine type To study which combinations among the machine
and lesion type. It shows that the specificity from digital type, lesion type, and processing method factors pro-
mammography was less than the specificity from screen duced the difference between digital mammography
film mammography across all machine types for and screen film mammography, multivariate analysis
calcifications. was conducted of specificity data including the interac-
Tests based on multivariate analysis of specificity data tions among these factors. The mass cases produced
show that on average, the digital mammograms produced larger variation in specificity difference between digital
by the Fischer digital system gave significantly less speci- mammography and screen film mammography than the
ficity than screen film mammograms (P ⫽ .044); and, on calcifications cases (P ⫽ .002), but no significant dif-

590
Academic Radiology, Vol 12, No 5, May 2005 IMAGE PROCESSING AND DIGITAL MAMMOGRAPHY

Figure 4. Area under the receiver operating characteristic curves by machine and lesion type av-
eraged across the three MultiScale Image Contrast Amplification readers; value presented in pa-
rentheses for each modality.

ference was observed among the three machine types. DISCUSSION


No difference was seen among three processing meth-
ods and between the two lesion types. The resulting Radiologist sensitivity and specificity are dependent not
95% confidence intervals for specificity difference are only on the interpretation skill of the radiologist, but also to
shown in Fig 8. The Fischer calcifications cases pro- a certain extent on just how visible lesions actually are.
cessed with default and PLAHE, and Trex calcifica- First, the lesion must be distinguishable from the surround-
tions cases processed with MUSICA produced signifi- ing background (normal breast tissue) to be detected mam-
cantly lower specificity than did screen film. All re- mographically. Similarly, diagnosis of a lesion is only made
maining digital combinations showed no difference in when the lesion’s features can be classified. Image-process-
specificity from screen film. ing algorithms are applied to digital mammograms to alter

591
COLE ET AL Academic Radiology, Vol 12, No 5, May 2005

Figure 5. Area under the receiver operating characteristic curves by machine and lesion type av-
eraged across the three Power Law Adaptive Histogram Equalization readers; value presented in
parentheses for each modality.

the visual presentation of the digital mammograms through density of the patient, and the lesion type if there are
manipulation of the gray-scale values of the pixels making findings.
up the mammographic image. Ideally, this manipulation, be There are several studies that have addressed the effects
it point (individual pixel), area (clustered groups of pixels), of spatial resolution on calcification detection in digital
or global (whole image) would lead to an image where le- mammography (10 –12) for 100-micron digitized images.
sions are more distinguishable from normal tissue to allow None showed improved sensitivity for digital mammography
the radiologists to use their interpretation skill. The lesion over screen film. With increased spatial resolution, microcal-
has to be seen before it can be interpreted. An image-pro- cifications should be better visualized. However, this resolu-
cessing algorithm’s effect on an input image will be depen- tion does not alone provide the ability to see smaller features
dent on the spatial resolution of the input image, the contrast such as calcifications; the available contrast resolution is also
resolution available in the input image, the overall breast important. We would expect calcifications to be better visu-

592
Academic Radiology, Vol 12, No 5, May 2005 IMAGE PROCESSING AND DIGITAL MAMMOGRAPHY

Table 3
Main Effects by Machine Type and Lesion Type of AUC difference, Sensitivity Difference, and Specificity Difference Between
Digital and Screen-Film Mammography

AUC difference Sensitivity difference Specificity difference

Factor Type Mean (SD) P value Mean (SD) P value Mean (SD) P value

Machine Fischer ⫺0.042 (0.037) .321 ⫺0.029 (0.048) .581 ⫺0.061 (0.023) .044*
GE ⫺0.205 (0.043) .009* ⫺0.255 (0.056) .011* ⫺0.037 (0.039) .398
Trex 0.026 (0.012) .091 0.082 (0.032) .061 ⫺0.073 (0.038) .113
Lesion Calc ⫺0.077 (0.026) .042* ⫺0.032 (0.044) .508 ⫺0.129 (0.024) .003*
Mass ⫺0.070 (0.014) .007* ⫺0.102 (0.024) .014* 0.015 (0.040) .725

AUC: area under the receiver operating characteristic curve.


*Indicates statistically significant difference between digital and screen film mammography.

Figure 6. 95% confidence intervals for area under the receiver Figure 7. 95% confidence intervals for sensitivity difference be-
operating characteristic difference between digital and screen film tween digital mammography and screen film mammography.
mammography. Each vertical interval represents the machine Each vertical interval represents the machine type, image pro-
type, image processing type, and lesion type combination seen cessing type, and lesion type combination seen by three readers.
by three readers. The intervals marked with asterisks were statis- The intervals marked with asterisks were statistically significant.
tically significant.

best calcification sensitivity performance relative to screen


alized on screen film compared with digital mammography film (Fig 7). The Fischer system was next with its 50 mi-
based on the higher spatial resolution for screen film, but the crons per pixel size and contrast resolution of 12 bits (4,096
increased contrast difference inherent with digital mammog- distinct gray levels), and GE was third with spatial resolu-
raphy is thought to compensate (13). Based on that assump- tion of 100 microns per pixel size (lowest spatial resolution
tion, we would also expect the sensitivity difference between system tested) and contrast resolution of 15 (32,768 distinct
digital and screen film mammography to decrease as the gray levels) for one GE prototype and 16 bits (65,536 dis-
pixel size of the digital image gets smaller and smaller (in- tinct gray levels) for another. An interesting finding in our
creasing spatial resolution). These trends were found in our study was calcifications sensitivity performance in certain
study, in which calcification visualization on these digital instances was slightly better than the mass sensitivity perfor-
mammograms was worse than screen film overall, with a mance, for which we believe lesion subtlety (the visibility of
narrowing of the difference between digital mammography the lesion in relation to the surrounding breast tissue) was a
and screen film mammography for sensitivity with increas- factor.
ing spatial resolution. The overall breast density for each patient whose images
The Trex digital system, which had 40 microns per pixel were included in this study was at least heterogeneously
size (highest spatial resolution system tested) and contrast dense (8), making this a difficult case set to interpret. The
resolution of 14 bits (16,384 distinct gray levels), had the overall breast density has been shown to affect radiologist

593
COLE ET AL Academic Radiology, Vol 12, No 5, May 2005

mammography and screen film mammography. This speci-


ficity difference improvement is probably due to the fact that
the digital mammograms for this study were obtained with
the location of the lesion known by the technologist and
presumably optimally positioned and compressed to show
the characteristics of the lesion more clearly.
Digital mammography in general is considered more spe-
cific than screen film mammography based on previous stud-
ies (8,15,16). However, none of those studies, or the study
reported here, looked at randomly selecting the order of the
imaging modalities. However, there is a study comparing
digital mammography to screen film mammography that did
include randomization of the order of imaging modalities—
the Digital Mammography Imaging Screening Trial
Figure 8. 95% confidence interval for specificity difference be-
tween digital mammography and screen film mammography. (DMIST), which was open to accrual between 2001 and
Each vertical interval represents the machine type, image pro- 2003 and is currently in analysis—should resolve whether
cessing type, and lesion type combination seen by three readers.
The intervals marked with asterisks were statistically significant.
the increased specificity seen with digital is simply temporal.

Limitations
ROC performance, showing decreased sensitivity and speci- Although there were 201 cases included in this study,
ficity (14) with increased breast density. This effect is true image-processing failures resulted in a lower number of
of screen film mammography and this would be true of digi- cases being usable for the reader study. We suspect that the
tal mammography as well, unless significant visualization use of more stable digital mammography systems now com-
were possible through the larger contrast range. But although mercially available and the widespread use of image format
the maximum range of contrast values is set per machine standards in medical imaging, such as Digital Imaging and
type, the actual breast tissue composition and dose dictates Communications in Medicine (10), which was not available
the range of contrast values for each individual patient. for all machines at the time of this study, would have led to
The relative difference between each digital mammogra- better input image consistency in this study and fewer fail-
phy system and screen film mammography for specificity is ures of the algorithms, especially PLAHE. More than 96%
about the same. Specificity difference between all the digital (52/54) of the PLAHE-processed cases were not available
machine types and screen film do not vary much with each because of image-formatting variations within machine types
of the image processing algorithms applied, regardless of that were not anticipated at the time of algorithm develop-
lesion type. Both PLAHE and manufacturer’s default seem ment.
to be specifically optimized for improved characterization of The three image-processing algorithms tested in this
masses across each of the machine types. A small relative study— default, PLAHE, and MUSICA— had all performed
difference was noticed in specificity, regardless of machine well in previous studies where they were compared with
type for calcifications (Fig 8). This difference would be ex- other image-processing algorithms (3,8), although not neces-
pected, given that the images used in this study were screen- sarily better than screen film. There certainly could be better
ing mammograms and not diagnostic magnification or spot algorithms, even improved versions of the ones tested here,
compression views, in which better characterization of calci- that could lead to different results. The robustness of algo-
fications is possible. rithms and the parameter settings used can also be the cause
In assessment of the specificity (the ability to distinguish of image display differences.
lesion features that are suspicious for cancer from lesion All of the algorithms tested had some flexibility in regard
features not suspicious for cancer), it was expected that digi- to optimizing parameter settings. The possibilities in some
tal mammography would perform better for specificity by instances were limitless. We conducted a small preference
benefit of the digital mammograms being acquired after the study by one expert radiologist to determine the optimal
lesion was found with screen film mammography. There parameter settings for the MUSICA and PLAHE algorithms
was a smaller mean difference between digital mammogra- to be used for the cases included in this study. These opti-
phy and screen film and there was less variability among mal parameter settings were selected based on the radiolo-
readers in regard to specificity difference between digital gist’s impression of the quality of the image data in side-by-

594
Academic Radiology, Vol 12, No 5, May 2005 IMAGE PROCESSING AND DIGITAL MAMMOGRAPHY

side comparison to screen film. The most consistently pre- City, Iowa; Stephen Feig, MD, Mt. Sinai School of Medi-
ferred parameters across both lesion types (masses and cine, New York, NY; Brad Hemminger, PhD, University of
calcifications) were selected for each of the MUSICA and North Carolina, Chapel Hill, NC; Roberta Jong, MD,
PLAHE algorithms and applied to all cases in the reader FRCPC-Sunnybrook & Women’s Health Science Center,
study. Toronto, Ontario, Canada; Daniel Kopans, MD, Massachu-
There have been changes in the various digital mam- setts General Hospital, Boston, MA; Andrew Maidment,
mography systems tested here since this study was com- PhD, University of Pennsylvania Medical Center, Philadel-
pleted including: detector changes, hardware changes, and phia, PA; Bahjat Qaqish, PhD, University of North Carolina,
software changes that have led to more consistent image Chapel Hill, NC; Rene Shumak, MD, Ontario Breast Screen-
quality over earlier system versions. The Trex system ing Program, Toronto, Ontario, Canada; Melinda Staiger,
tested here was Food and Drug Administration–approved MD, Monmouth Medical Center, Long Branch, NJ; Mark
in 2001 under the name Lorad by Hologic, Inc. (Bedford, Williams, PhD, University of Virginia, Charlottesville, VA;
MA), but was never marketed. The Fischer SenoScan and Martin Yaffe, PhD, Sunnybrook & Women’s Health
tested here has undergone a number of software changes Science Center, Toronto, Ontario, Canada.
throughout the years. In November 2003, Fischer intro-
duced a new image-processing algorithm that was de- REFERENCES
signed not only to improve the overall image presentation
1. Shah AJ, Wang J, Yamada T, et al. Digital mammography: a review of
quality compared with older Fischer default algorithms, technical development and clinical applications. Clin Breast Cancer
but also to improve the speed of loading images into the 2003; 4:63–70.
2. Pisano ED, Cole EB, Hemminger BM, et al. Image processing algo-
review station for review. The GE Senographe 2000D has rithms for digital mammography: a pictorial essay. Radiographics 2000;
also undergone changes in display software and now of- 20:1479 –1491.
3. Pisano ED, Cole EB, Major S, et al. Radiologists’ preferences for digital
fers a second image-processing algorithm, Premium View, mammographic display. The International Digital Mammography Devel-
in addition to the default algorithm at the review worksta- opment Group. Radiology 2000; 216:820 – 830.
tion. Whether any of these changes results in earlier de- 4. Baydush AH, Floyd CE, Jr. Improved image quality in digital mammog-
raphy with image processing. Med Phys 2000; 27:1503–1508.
tection of breast cancer by radiologists over previous ver- 5. Chakrabarti K, Thomas JA, Kaczmarek RV, et al. Optimization of view-
sions of a given manufacturer’s digital mammography ing conditions and phantom image quality evaluations on GE DMR and
full-field digital mammography system. J Digit Imaging 2000; 13(Suppl
system or over screen film mammography would require 1):226 –227.
additional reader studies. 6. Qian W, Li L, Clarke L, et al. Digital mammography: comparison of
adaptive and nonadaptive CAD methods for mass detection. Acad Ra-
Although image processing alone was not a main effect diol 1999; 6:471– 480.
of digital mammography versus screen film mammography 7. Nawano S, Murakami K, Moriyama N, et al. Computer-aided diagnosis
in our study, it does seem to have an effect when combined in full digital mammography. Invest Radiol 1999; 34:310 –316.
8. Cole EB, Pisano ED, Kistner EO, et al. Diagnostic accuracy of digital
with lesion and machine type. Training in the visual presen- mammography in patients with dense breasts who underwent problem-
tation of lesion characteristics when a given image process- solving mammography: effects of image processing and lesion type.
Radiology 2003; 226:153–160.
ing algorithm is applied to a particular digital mammography 9. Hemminger BM. Soft copy display requirements for digital mammogra-
system should be provided during applications training for phy. J Digit Imaging 2003; 16:292–305.
10. Chan HP, Vyborny CJ, MacMahon H, et al. Digital mammography.
digital mammography review workstations. Based on our ROC studies of the effects of pixel size and unsharp-mask filtering on
results, multiple image-processing algorithms may need to the detection of subtle microcalcifications. Invest Radiol 1987; 22:581–
589.
be incorporated into future soft copy workstations to allow 11. De Maeseneer M, Beeckman P, Osteaux M, et al. Detecting clustered
for optimal viewing of specific lesion types for each of the microcalcifications in the female breast: secondary digitized images
digital mammography systems. versus mammograms. J Belge Radiol 1992; 75:173–178.
12. Karssemeijer N, Frieling JT, Hendriks JH. Spatial resolution in digital
mammography. Invest Radiol 1993; 28:413– 419.
13. Feig SA, Yaffe MJ. Digital mammography. Radiographics 1998;
ACKNOWLEDGMENTS
18:893–901.
The authors would like to acknowledge the contributions 14. Barlow WE, Lehman CD, Zheng Y, et. al. Performance of diagnostic
mammography for women with signs or symptoms of breast cancer.
of the International Digital Mammography Development J Natl Cancer Inst 2002; 94:1151–1159.
Group, especially the principal investigators of the project 15. Kuzmiak CM, Millnamow GA, Qaqish B, et al. Comparison of full-field
digital mammography to screen-film mammography with respect to
(3) that led to the creation of the image archive from which diagnostic accuracy of lesion characterization in breast tissue biopsy
the raw images used in this study were obtained: Emily Co- specimens. Acad Radiol 2002; 9:1378 –1382.
16. Lewin JM, Hendrick RE, D’Orsi CJ, et al. Comparison of full-field digital
nant, MD, University of Pennsylvania Medical Center, Phila- mammography with screen-film mammography for cancer detection:
delphia, PA; Laurie Fajardo, MD, University of Iowa, Iowa results of 4,945 paired examinations. Radiology 2001; 218:873– 880.

595

Вам также может понравиться