Вы находитесь на странице: 1из 40

SERVICE-MANUAL HM5011/5010

ENGLISH 4S-5010-0010 09 JUN 1998

Instruments

Spectrum analyzer HM5011 HM5010

Service Manual Adjustment Procedure Circuit Diagrams Alignment Procedure for HM5010/HM 5011 ......... A Control and Adjustment of Supply Voltages ............................................... B Final Alignment - Tuner ..................................... C Alignment - IF-Unit ........................................... D Linearity of Frequency Display ......................... HM5010 XY-board ........................................... Tuner ............................................................... Tracking-Generator .......................................... E Alignment of HM5011 Tracking Generator ...... Tracking-Generator ................................................. Tuner RA-Board ...................................................... Tuner RB-Board ...................................................... IF-Amplifier ............................................................. Main board ............................................................. X-Y Board ............................................................... FC-Board ................................................................ PA-Board ................................................................ CRT-Board .............................................................. Power supply Board ............................................... RA-Board ................................................................ Tracking-Generator ................................................. RB-Board ................................................................ Main board ............................................................. Main board ............................................................. XY-Board ................................................................. XY-Board ................................................................. FC-Board ................................................................ FC-Board ................................................................ PA-Board ................................................................ CRT-Board .............................................................. PS-Board ................................................................ PS-Board ................................................................ PS-Board ................................................................ Block Diagram HM5010/HM5011 ..........................
Subject to change without notice

5 5 6 6 7 10 11 12 13 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39

HM5010/5011

SPECTRUM ANALYZER
Specifications
Frequency
Frequency range: 0.15MHz to 1050MHz (-3dB) Center frequency display accuracy: 100kHz Marker accuracy: (0.1% span + 100kHz) Frequency display res.: 100kHz (4 digit LED) Frequency scanwidth: 100kHz/div. to 100MHz/div. in 1-2-5 steps and 0Hz/div. (Zero Scan) Frequency scanwidth accuracy: 10% Frequency stability: better than 150kHz / hour IF Bandwidth (-3dB): Resolution: 400kHz and 20kHz; Video-Filter on: 4kHz Sweep rate: 43Hz

Amplitude
Amplitude range: -100dBm to +13dBm Screen display range: 80dB (10dB / div.) Reference level: -27dBm to +13dBm (in 10dB steps) Reference level accuracy: 2dB Average noise level: -99dBm (20kHz BW) Distortion: <-75dBc; 2nd and 3rd harmonic 3rd order intermod.: -70dBc (two signals >3MHz apart) Sensitivity: <5dB above average noise level Log scale fidelity: 2dB (without attn.) Ref.: 250MHz IF gain: 10dB adjustment range

Spectrum Analyzer HM5010 / HM5011


Frequency Range 0.15MHz - 1050MHz. 4 Digit Display (Center & Marker Frequency, 0.1MHz resolution) 100 to +13dBm Amplitude Range, 20kHz, 400kHz and Video-Filter Tracking-Generator (HM5011 only): Frequency range: 0.15MHz - 1050MHz. Output Voltage: +1dBm to 50dBm (50). Evolution of the original HM5005/HM5006 has led to the new HM5010/ HM5011 Spectrum Analyzer/Tracking Generator which now extends operation over 1 GHz (frequency range 0.15 to 1050 MHz). Both fine and coarse center frequency controls, combined with a scanwidth selector provide simple frequency domain measurements from 100 kHz/div. to 100 MHz/Div.. Both models include a 4digit numeric LED readout that can selectively display either the center or marker frequency. The HM5011 includes a tracking generator. The HM5010/5011 offer the same operation modes as the HM5005/5006. The instruments are suitable for pre-compliance testing during development prior to third party testing. A near-field sniffer probe set, HZ530, can be used to locate cable and PC board emission "hot spots" and evaluate EMC problems at the breadboard and prototype level. The combination of HM5010 / 5011 with the HZ530 is an excellent solution for RF leakage/radiation detection, CATV/ MATV system troubleshooting, cellular telephone/pocket pager test and EMC diagnostics. There is an optional measurement output for a PC which makes documentation of results easy and affordable with the HO500 Interface.

Input
Input impedance: 50 Input connector: BNC Input attenuator: 0 to 40 dB (4 x 10dB steps) Input attenuator accuracy: 1dB/10dB step Max. input level: +10dBm, 25VDC (0dB attenuation) +20dBm (40dB attenuation)

Tracking Generator
Output level range: -50dBm to +1dBm (in 10dB steps and var.) Output attenuator: 0 to 40dB (4 x 10dB steps) Output attenuator accuracy: 1dB Output impedance: 50 (BNC) Frequency range: 0.15MHz to 1050MHz Frequency response: 1.5dB Radio Frequency Interference (RFI): <20dBc

Divers
AM-Demodulator output for head-sets. Permissible load impedance >8

General
Display: CRT. 6 inch, 8 x 10 div. intern. graticule Trace rotation: Adjustable on front panel Line voltage: 115 / 230V 10%, 50-60Hz Power consumption: approx. 20W Operating ambient temperature: 0C..+40C Protective system: Safety Class I (IEC 1010-1) Weight: approx. 7kg Cabinet: W 285, H 125, D 380 mm
Subject to change without notice 5/96

Accessories supplied: Line Cord, Operators Manual. Optional accessories, 50-feedthrough termination HZ22 Viewing Hood HZ47, Near Field Probe Set HZ530, Carrying Case HZ96-2, Transient Limiter HZ560

HO500 Computer Interface for HAMEG Spectrum Analyzer


This HO500 computer interface offers the facility to transfer a calibrated frequency spectrum from any HAMEG spectrum analyzer to a PC. The HO500 interface is a 8 bit ISA BUS card installed in the PC, which transfers data via an interface cable. The software supplied allows a hard copy print out (including parameters) of the frequency spectrum, in Windows Format. Signal aquisition occurs 2 to 3 times per second. The picture consists of 10 bit vertical by approx 3600 point horizontal display. The PC monitor display is in SVGA Format with 800 x 600 pixels. For comparison measurements, a previosly stored reference curve can be recalled. The software supplied works under Windows 3.1, 3.11 and WIN95. A simple XY analog output is required to connect the HO500 to the spectrum analyzer.

Subject to change without notice

EMC-MEASUREMENT EQUIPMENT
Specifications
Frequency Frequency range: 0.1MHz to 1000MHz (lower frequency limit depends on probe type) Output impedance: 50 Output connector: BNC-jack Input capacitance: 2pF (high imped. probe) Max. Input Level: +10dBm (without destruction) 1dB-compression point: -2dBm (frequency range dependent) DC-input voltage: 20V max. Supply Voltage: 6V DC 4 AA size batteries Supply-power of HM5010/5011 Supply Current: 8mA (H-Field Probe) 15mA (E-FieldProbe) 24mA(High imp.Probe) Probe Dimensions: 40x19x195mm (WxDxL) Housing: Plastic; (electrically shielded internally) Package contents: Carrying case 1 H-Field Probe 1 E-Field Probe 1 High Impedance Probe 1 BNC cable (1.5m) 1 Power Supply Cable (Batteries or Ni-Cads are not included)

Near Field Sniffer Probes HZ 530


The HZ530 is the ideal toolkit for the investigation of RF electromagnetic fields. It is indispensable for EMI precompliance testing during product development, prior to third party testing. The set includes 3 hand-held probes with a built-in pre-amplifier covering the frequency range from 100kHz to over 1000 MHz. The probes - one magnetic field probe, one electric field probe, and one high impedance probe - are all matched to the 50 inputs of spectrum analyzers or RFreceivers. The power can be supplied either from batteries, Ni-Cads or through a power cord directly connected to an HM5010/HM5011 series spectrum analyzer. Signal feed is via a 1.5m BNC-cable. When used in conjuction with a spectrum analyzer or a measuring receiver, the probes can be used to locate and qualify EMI sources, as well as evaluate EMC problems at the breadboard and prototype level. They enable the user to evaluate radiated fields and perform shield effectiveness comparisons. Mechanical screening performance and immunity tests on cables and components are easily performed. the connected measurement system which is proportional to the magnetic radio frequency (RF) field strength existing at the probe location. With this probe, circuit RF sources may be localized in close proximity of each other. The H-field will decrease as the cube of the distance from the source. A doubling of the distance will reduce the Hfield by a factor of eight (H = 1/d); where d is the distance. In the actual use of the H-field sensor one observes therefore a rapid increase of the probes output voltage as the interference source is approached. While investigating a circuit board, the sources are immediately obvious. It is easily noticed which component (i.e. IC) causes interference and which does not. In addition, by use of a spectrum analyzer, the maximum amplitude as a function of frequency is easily identified. Therefore, one can eliminate early in the development components which are not suitable for EMC purposes. The effectiveness of countermeasures can be judged easily. One can investigate shields for "leaking" areas and cables or wires for conducted interference. (near the insulation resistance of the printed circuit material) and is loading the test point with only 2 pF (80 at 1 GHz). Thereby one can measure directly in a circuit without significantly influencing the relationships in the circuit with the probe. One can, for example, measure the quantitative effectiveness of filters or other blocking measures. Individual pins of ICs can be identified as RFI sources. On printed circuit boards, individual problem tracks can be identified. With this Hi-Z probe individual test points of a circuit can be connected to the 50 impedance of a spectrum analyzer.

The E-Field Monopole Probe


The E-field monopole probe has the highest sensitivity of the three probes. It is sensitive enough to be used as an antenna for radio or TV reception. With this probe the entire radiation from a circuit or an equipment can be measured. It is used, to determine the effectiveness of shielding measures. With this probe, the entire effectiveness of filters can be measured by measuring the RFI which is conducted along cables that leave the equipment and may influence the total radiation. In addition, the E-field probe may be used to perform relative measurements for certification tests. This makes it possible to apply remedial suppresion measures so that any requalification results will be positive. In addition, pre-testing for certification tests may be performed so that no surprises are encountered during the certification tests.
SCALE = 10dB/DIV.

The High-Impedance Probe


The high-impedance probe (Hi-Z) permits the determination of the radio frequency interference (RFI) on individual contacts or printed circuit traces. It is a direct-contact probe. The probe is of very high impedance

The H-Field Near-Field Probe


The H-Field probe provides a voltage to

SCALE = 10dB/DIV.

Frequency Response E-Field Probe (typical)

SCALE = 10dB/DIV.

Frequency Response H-Field Probe (typical)

Frequency Response High Impedance Probe (typical)

Subject to change without notice

Alignment Procedure for HM5010/HM 5011


Attention! The opening of covers or removal of parts is likely to expose live parts and accessible terminals which can be dangerous to life. Maintenance, service and alignment should be carried out by qualified personnel only, which is acquainted with the danger involved. When aligning the HM5010/5011 it is assumed that all sub-assemblies of the instrument are completely pretested and working correctly. The tuner, IF-unit, and tracking generator should be pre-aligned. When aligning a HM5010, a separate tracking generator unit must be available and connected to the HM5010 for some specific adjustments. Prior to the alignment procedure, the instrument must warm up for 60 minutes. All adjustments are carried out by means of a plastic screw driver or a ceramic adjustment tool. The alignment is divided into the following steps: A Checking of supply voltages D Linearity alignment B Alignment of the tuner E Tracking generator alignment C Alignment of the IF-unit F Check of overall adjustment The numbering system is related to the respective pictures. Screen shots are designated as PIC 1 HF-Synthesizer 100 kHz to 1000 MHz, i.e. HM 8133 2 BNC cable, BNC T-connector, 2 x 10dB attenuator 50 Ohm 1 Voltmeter i.e. HM 8011-3

Control and Adjustment of Supply Voltages


12V to an accuracy of 0.1V via Trimpot 4 PIC 1 minimum brightness via Trimpot 2 PIC 1 maximum brightness via Trimpot 3 PIC 1 astigmatism via Trimpot 1 PIC 1 12V +12V +5V 5V +38V +138V tolerance 0.2V tolerance 0.1V tolerance 0.2V tolerance 0.2V tolerance 1V tolerance 1V
1

adjust: adjust: adjust: adjust: check: check: check: check: check: check:

The corresponding voltage test points to check direct voltage can be measured at the measuring connector strip (see PIC 2). Basic adjustment: When Y-pos. knob is snapped in on front board, adjust beam approx. 2 mm below bottom graticule line via R801.
4 3

1 2 3 4 5 6

+138V +38V GND +5V +12V -12V

PIC 2. XYF-Board (partial)

PIC 1. PS-Board 5

Instrument Rear Side

CRT-Board

Subject to change without notice

Final Adjustment - Tuner

The tuner is already aligned by the factory. When changing the 1st mixer, it might be necessary to realign cavity filter: Set signal to 500MHz -27dBm to input, center frequency to 500MHz, Scanwidth to 0.5MHz, turn all three M3 screws to max. amplitude. In case curve is not uniform, coils L1, L2, and L3 can be adjusted to maximum uniformity (fig. 22) by slightly bending them. In case signal in frequency range is shacky, 2nd local oscillator might not be snapped in correctly. The correct snapin of the PLL is visible via the lock detect LED (D2), which has to be lit constantly without flickering if PLL is snappedin correctly. The oscillator signal has to be fixed (crystal dependent) at 1.32GHz, it may not drift. The PLL is snapped-in when the tuning voltage VT (measure at PAD1) is between 0.5 and 4.5V. The alignment will bring the tuning voltage to the center of the tuning range. VT of the 2nd LO has to be between 2V and 2.5V. Constantly check tuning voltage while aligning tuner. Alignment is necessary in two cases: VT<2.0V and VT>2.5V. Case 1 VT<2.0V Correction: a ) Solder all adjustment areas to connect trace from C1 to C2. b ) Remove excess solder at center conductor of resonator pad. Case 2 VT>2.5V Correction: a ) Adjustment areas may not be soldered to connect trace from C1 to C2. b ) Add solder to resonator pad.

Alignment of IF-Unit

fig. 22 for HM5010

Measurement setup: Apply two different signals to the input of the HM5011 via BNC T-connector and 10dB attenuators. Adjust HM5010/5011: Center frequency to 500MHz; all attenuators on (40 dB) Bandwidth 400kHz; video filter off; scanwidth 0.5MHz/div.; marker off. Adjust Tracking Generator: Attenuator 20dB; level max. (HM5011 only) RF-generator: frequency 500MHz; level 5dBm
The output voltage of the tracking generator is visible on the screen by an overlay of the fixed frequency. A "zero" point fig. 23) is visible within the signal to the right or to the left of the 500MHz spectral line (fig. 23 (5011 only). IF Filter Curve at 400kHz Bandwidth

align: align: align:


VR 3

Step 1

Align (with plastic screw driver) Coil L1 (PIC 5) to maximum output amplitude and symmetry to Y-axis. Step 2 Align (with plastic screw driver) Coils 3 + 4 + 13 + 14 (PIC 5) to symmetrie to Y-axis. The "zero" point must be at the maximum (at 500MHz) (fig. 23). fig. Stept 3 If necessary, repeat step 1 and 2. VR 1

L 10 L 13 L 11 L 14 L 12

L7

L3 L1

L8 L4 L9

fig. 23 for HM5011

VR 4

VR 2

PIC 5 IF-Amp Board

Subject to change without notice

IF Filter Curve at 20kHz Bandwidth (HM5010)

align:

Step 1

Align (with plastic screw driver) Coils L7+8+9+10+11+12 until "zero" point is exactly centered at 500MHz (as with 400kHz Bandwidth). Watch for symmetry of Filter Curve.

IF Filter Curve at 20kHz Bandwidth (HM5011) Pre-Alignment: Remove T-connector. Connect tracking generator module directly to the input of the HM5011. Turn off attenuators on tracking generator module and on HM5011 (0 dB). The output voltage of the tracking generator module is now visible as horizontal line (with slight ripple). adjust: Step 1 Step 2 Adjust (with plastic screw driver) Coils 7 + 8 + 9 + 10 + 11 + 12 to maximum screen height of displayed output voltage. Add attenuators of HM5010/5011 as soon as "line" has reached the middle of the screen (30 dB). This alignment has to be performed repeatedly in order to optimize settings.

adjust:

Fine Alignment: Connect BNC-T-connector again as in the beginning of the IF alignment. Set attenuator of tracking generator module to 20dB, scanwidth to 0.5 MHz/div. If necessary, re-adjust center frequency (adjust 500MHz spectral line to screen center). "Zero" point is now visible to the right or to the left of screen center (amplitude maximum) (fig. 24). fig.

adjust: adjust:

Step 1 Step 2

Align (with plastic screw driver) Coils L7+8+9+10+11+12 until "zero" point reaches maximum. Watch for symmetry of Filter Curve. This alignment has to be performed repeatedly in order to optimize settings. fig. 24

IF Gain - Adjustment of different Bandwidths Set scanwidth to 0.2MHz/div. Switch bandwidth repeatedly between 400kHz and 20kHz. The distance between the output voltage (line) of the tracking generator and tge 400MHz amplitude may not vary. If the amplitude values do not match exactly, adjust by means of R-trimmer VR2 (PIC5) (PIC5). Linearity of IF-Amplifier Gain The linearity of the IF-amplifier gain has to be checked through the entire display range. Apply a 400MHz (-27dBm) signal directly to the input of the HM5010/5011. Adjust scanwidth to 5MHz/div, release attenuator switched (0 dB), select filter bandwidth of 400kHz. The spectral line should reach the upper screen edge. Use attenuators to reduce signal in 10db steps, whereby each individual attenuation step has to reduce the level by 10dB 1dB. In case deviation is as follows: - A the drop of the individual attenuation steps is larger than 10dB 1dB, or - B the drop of the individual attenuation steps is smaller than 10dB 1dB, the linearity of the attenuators has to be adjusted as follows: Set attenuator to -40dB. Adjust spectral line exactly to -40dB (center line) via VR1, VR3, VR4. Then set attenuators back to 0dB und adjust spectral line by means of trimpots VR802A and VR801 (XY board) to zero point and base line to bottom graticule line. This procedure has to be repeated until the settings at -40dB and 0dB are correct.

Linearity of Frequency Display

Settings for HM5010/5011: Center frequency at 500MHz; all attenuator switches released; filter bandwidth 400kHz; video filter off; scanwidth to 100MHz/div.; marker off. Measurement Setup: Apply signal of 500MHz 27dBm to input of HM 5010/5011.

Check: Check:

Basic setting: The HM5010/5011 has to be adjusted that the noise level touches the bottom graticule line. The spectral line of 500MHz -27dBm reaches the top graticule line and is situated in the center of the screen. Adjust 500MHz spectral line exactly to screen center via X-pos knob (on front of the unit). Upper frequency limit: Check if frequency of minimum 1050 can be set as center frequency.

Subject to change without notice

Adjust HM5010/5011 to basic setting (see above). Apply input signal of 100MHz, level +7dBm Overriding the input allows for the harmonics of the input signal to become visible (fig. 7) This fig. 7). simplifies the adjustment of individual spectral lines in horizontal direction.

adjust:

Spectral line at 400MHz Turn X-ampl. knob on front of instrument to have the 4 spectral lines matched with the corresponding graticule lines on the screen(fig. 8 fig. 8).
fig. 7

adjust:

Spectral line at 100MHz With Trimpot RV171 align to match this spectral line with the corresponding graticule line on the screen (fig. 9 fig. 9). Zero Peak With Trimpot RV173 align to match this spectral line with the corresponding graticule line on the screen (fig. 9 fig. 9).
fig. 8

adjust:

adjust: adjust:

Spectral line at 600-1000MHz With Trimpot RV186 adjust to match 700MHz to the correct graticule line. Spectral line at 800 MHz; With Trimpots RV181 + RV183 for 900MHz and RV197 for 1000MHz (PIC 3) adjust to match this spectral line with the corresponding graticule line on the screen.
fig. 9

adjust:

Beam length With Trimpot RV101 adjust beam to end approx. 1mm beyond the right CRT boundary (graticule) (Fig. 11 - base line too short) Fig. (Fig. 10 - 300 to 1000MHz spectral lines not correct ) Fig.

fig. 10

RV 112

RV113

fig. 11

RV 149 RV 101 RV 183 RV 181 RV 186 RV 197 RV 171 RV 173 PIC 3 - MB-Board Set Trimpot RV113 that frequency will not display below 990MHz when center frequency is set to lowest frequency
Subject to change without notice

Adjust Marker. Settings on HM5010/5011: Center frequency to 500MHz, all attenuators in off position, filter bandwidth to 400kHz, video filter off. Scanwidth to 100MHz, Marker on. Signal 100MHz +7dBm. Adjust Marker to 500MHz display and set to 500MHz mark on screen via RV149 (PIC 4). Turn Marker left to max. and adjust to 990MHz via RV112. VR 3 VR 2

L 10 L 13 L 11 L 14 VR 4 L 12

L7

L3 L1

VR 1

L8 L4 L9

RV 112

RV 113

RV 149

RV 183 RV 181 RV 186 RV 197 RV 171 RV 173

RV 101

PIC 3 - Main-Board 9

Subject to change without notice

XY-Board

RV 802A
1

1 2 3 4 5 6

+138V +38V GND +5V +12V -12V

RV 801

Subject to change without notice

10

11 Tuner M3 Screws Cavity - Filter M3 Screws Cavity - Filter C1 C2 MPAD1

Subject to change without notice

1st Mixer

Tracking-Generator

RV4

RV3

C68 C69

MPAD2

Subject to change without notice

12

E 1. 1.1 1.2 1.3 1.4 1.5 2. 2.1 2.2 2.3 2.4 2.5

Alignment of HM5011 Tracking Generator Required Instruments: Spectrum Analyzer - minimum 1000MHz Completely assembled HM5011 Multimeter to measure DC voltage Oscilloscope, i.e. HM303 Coax cabel Preparation: Assemble Spectrum Analyzer completely (without case). All sub-assemblies have to be pre-checked. HM5011 has to be warmed up. Connect VCO output of RA board to VCO input (MCX connector). Connect 12MHz reference clock of TG board to RCA connector ST1 on RB board.

3.

Check Signalling Lines Cable connection W1 Pin Nr. 1 2 3 Description -12V +12V +5V Signal Supply OP Supply Supply PLL

4.

Check Supply and Bias Voltages Pin Nr. 1 1 3 5 3 3 Description 1. VCO Amp. 2. VCO Amp. fix LO Amp. Var. Amp. 1. Power Amp. 2. Power Amp. Voltage +5.8 V +5.8 V +5.9 V +5.6 V +3.6 V +4.8 V

Comp.Nr. C1 C2 C11 C6 C9 C12

5.

Alignment of fixed LO

5.1 The 2nd LO oscillates at a frequency of 1.35 GHz. The level at the mixer input IC3 Pin2 is -17dBm. Oscillation of the 2nd LO can also be measured with the H or E-Field Probe (HZ530). 5.2 Lock-Detect LED (D4) will confirm the correct PLL lock when lighting up without flickering or going out. The oscillator signal has to remain fixed crystal dependent) at 1.35GHz and may not drift. PLL is locked if tuning voltage VT (measure at PAD2) is between 0.5 and 4.5V. Align tuning voltage to medium tuning range. VT of fixed LO should be between 2 and 2.5V. The tuning volage has to be check constantly during alignment. Place coax resonator CR2 bottom side to the edge of soldermask next to center conductor, and solder on the left and the right side. 5.3 Alignment is necessary in two cases: VT<2.0V and VT>2.5V. Case 1 VT<2.0V Correction: a ) Solder all adjustment areas to connect trace from C1 to C2. b ) Remove excess solder at center conductor of resonator pad. Case 2 VT>2.5V Correction: a ) Adjustment areas may not be soldered to connect trace from C1 to C2. b ) Add solder to resonator pad.

13

Subject to change without notice

6. 6.1 6.2 6.3

Control of fixed LO-Level To check 2nd LO-level, cap C20 (22pF) has to be removed and 50Ohm Coax-cable has to be connected to 13dB attenuator of output. Connect other end of coax cable with adequate test analyzer. Adjust test analyzer to 1.35 GHz center frequency; the LO-level to be measured has to be -17dBm (1 dB) (consider attenuation of cable). If cap is not removed and coax cable is connected parallel to mixer to the VCO-branch, a level of approx. -24 dBm (1dB) should be measured due to decreased load resistance. Check of VCO Adjust HM5011 to Zero Scan, center frequency to 500MHz. Remove Cap C14 (22pF), connect one side to attenuator output, other side stand up in the air. Connect Cap side sticking up in the air to center conductor of 50 Ohm cable. Solder shielding of cable directly next to Cap to ground. Connect other end of Cap to Analyzer input (i.e. Advantest R3361A). Start 1.35GHz Stop 2.35GHz reference level: 0dBm. VCO signal which can be observed on the test analyzer has to shift according to center frequency (tuning voltage). Tuning range has to be at least from 1350 to 2350 Mhz. continuously and without interruption. It should not go under absolute level of +7dBm (coax cable of 1 mtr. has 1-2 dB of attenuation at 2GHz). Level has to be between +7 dBm and +10 dBm. Connect coax cable parallel to 1st mixer to VCO-branch. A level of +5 dBm to +8 dBm due to decreased load resistance over frequency band. Check of Attenuators: Set -30dB attenuation to Tuner and to 0dB attenuation to TG. Set Center Frequency to 500MHz, 100MHz/Div. span. Connect tracking generator to tuner input. TG-line has to be visible. Switch on additional attenuators. Attenuation for each attenuator switch has to be 10dB (1 dB) TG Level Adjustment: Connect TG to reference analyzer. Set HM5011 to Zero Scan and 500MHz Center Frequency. Release all attenuator buttons of TG (0-dBm) and turn TG-level to maximum. Adjust amplitude to +1dBm with Pot RV4 on TG. turn TG-level to minimum. Adjust amplitude to -10dBm with Pot RV3 on TG. Amplitude has to be adjustable now between +1dB and -10dB. Final Check

7. 7.1 7.2 7.3

7.4

7.5

8. 8.1 8.2 8.3 8.4 8.5 9. 9.1 9.2 9.3 9.4 9.5 9.6 9.7 10.

10.1 Position of TG-level has to align in both bandwidths (400kHz and 20kHz). If necessary, align with IF-Amp. 10.2 TG-level knob has to be adjustable from +1 to -10dBm.

Subject to change without notice

14

Tracking-Generator

15

Subject to change without notice

Tuner RA-Board

Subject to change without notice

16

Tuner RB-Board

17

Subject to change without notice

IF AMP IF PCB

+12V R8 221R R5 220R L1101 +12V L1102 R9 3K32 R6 10K0 4.7V= R7 10K0 KA1102 T1 BC547 C8 10n TO XYF PCB P803 S.3 VID BW F3 RF2 C11 82p +5V 100u 6 5 4 3 2 1 6 5 4 3 2 1 +12V GND +5V GND VIDEO SIGNAL BW 100u KA1101

C9 0.1u

Subject to change without notice


C12 33p 82p T2 BF961 F4 R2 47R5 L2 0.82u IF GAIN +12V +5V 1 2 3 4 GND IF GAIN +12V +5V C5 39p C6 RF2 R3 2K21 C7 33p L5 15u XT1 27.125MHz R4 470R C13 220p C14 220p TO AL (RF UNIT) KA901 S.1 R33 180R +8.8V R15 221R R18 470k F10 RF2 RF2 C27 10n C28 120p C31 270p C25 10n R17 47R5 11V= 0.7Vpp F11 RF2 C33 120p C34 10n R14 220R NTC R23 22K1 R22 4K75 T5 BC547 VR4 100K R20 47K5 C16 0.1u CD3 1nF CD4 1nF HAMEG INSTRUMENTS GERMANY Title IF AMP HM5005/6 IF GAIN FROM ATTEN. RF UNIT S.1 VID VIDEO SIGNAL Size B Date: May 5, 1994 Document Number IF5005.SCH Sheet 2 of 8 REV IF GAIN RANGE D3 1N4148 D4 400mVpp C36 330p 7.6V= 8 R26 100K 1N4148 R27 150k R25 180R C42 10n C44 22p C35 10n R38 22K NTC R39 2K21 R37 3K92 C43 0.1u R24 100R U F 1 4 C22 1p8 C32 1p8 C30 270p R30 3K32 C39 10n F12 C29 1p8 C26 0.1u R32 100R C40 180p R35 2K21 R34 2K2 NTC F13 RF2 C38 180p F14 RF2 C37 0.1u R28 470K R29 47R5 >2Vpp 9 RF2 F7 RF2 R13 100k T4 BF961 C23 120p C20 270p C21 270p C24 10n R16 181R T6 BF961 R36 332R R31 4K75 GAIN OVERALL 200mVpp VR3 500R 6 C41 22u VR2 50K IC1 MC3356P VCC 7 G F9 1 2 3 4 5 11 12 13 15 16 17 18 19 20

C10 10n

IF IN

FROM TUNER 2nd IF OUT

30.0 MHz

IF

C4 8p2

C1 8p2

C3

22p

R1 100k

250mV

C2 47p

F1 RF3

VR1 200R TUNER GAIN COMP.

R10

2.2k

500mVpp

T3 BC557

D1 1N4148

C15 22p

C19 1p8

D2

R11 4K75

1N4148

F19

R19 4K75

R21 100k

RF2

C18 120p

R40 2K21

C17 0.1u

CD1 1nF

CD2 1nF

0V(250 kHz)

IF-Amplifier

+12V

BW

+8V(12.5 kHz)

31

32
+VREF +28V L-D IC106B LM324 R135 P109 5 47k5 -6V2 1 C140 * R168 2 10 8 9 R137 1M5 R136 8k25 1 C125 14 (VT) +5V 12 IC106D LM324 1 3 +VREF 100R R141 9k09 +28V R143 100R R133 56k2 4 C115 P110 Z Y X1 X0 X 14 15 4 1 2 3 1 2 3 GND BLANK GND 2 0.1u 1 R124 1k50 SAWTH IC101A TL074 C105 22nF R114 5k11 -6V2 R142 15k0 13 12 1 1 IC106A LM324 0.1u R170 * Y1 Y0 1 2 R127 150k 3 Z1 Z0 C116 0.1u 1 C109 3 5 R145 * C143 470p TO CRT W701 S.7 R147 * R146 * PCB 10k0 9 10 11 6 C B A INH +5V +28V R144 * +VREF IC105 HEF4053 C110 10nF C131 0.1u D144 * D110 BAS16 C126 1000u /16V * 3 IC107 * I N ADJ R100 0R O U 2 R166 * R165 * 1 C134 * D128 BAV99 3 C132 * D180 * R198 56k2 * TO PS P1006 S.8 C139 0.1u 7 6 R134 C128 1000u /16V R138 1M5 2 G OU N IN D -12V (+12V) GND (-5V) +5V +32V 3 R162 100R IC128 LM79L05 RV197 2k5 R179 * R196 10k0 RV180 * 49k9 * R181 0R 30k1 R187 3k32 R185 * L-A1 301k 4k99 RV171 5K0 R192 20k0 R172 2k74 R173 47k5 R176 * D171 BAV70 2 R194 * 13 D197 BAS16 RV173 500K L-A2 R175 * R182 4k75 3k01 R189 1k00 RT102 * R186 221k D181 BAW56 2 RT101 2k2 3 1 R193 61k9 301R R140 562R 499R 0R R183 10k0 RV183 100K L-C2 R197 47k5 IC106C LM324 R180 * R171 1k50 RV176 * R174 * R191 30k0 R195 0R R184 0R R188 L-B RV186 10K R178 * 10k0 L-C1 RV181 5K0 6 5 4 3 2 1 6 5 4 3 2 1

ANALOG SECTION

R105

IC103 HEF4051

R106

TO FC PCB W304 S.5

R107

G D C B A T N V D C

Main board

P107

X0 X1 X2 X3 X4 X5 X6 X7

13 14 15 12 1 5 2 4

R104

MOLEX6

1 2 3 4 5 6

INH

1 2 3 4 5 6

A B C

6 11 10 9

R103

R102

R101

IC102 HEF4051

R110

R108

+VREF

X0 X1 X2 X3 X4 X5 X6 X7

13 14 15 12 1 5 2 4

R109

CF

RV111 10K /10turns 2

INH

A B C

6 11 10 9

R130 39R

C127 0.1u

R112 10k0

+28V

+5V

R139 39R R131 39R C124 22u /35V R132 39R

R177 2k74

RV113

1k0

R111 33k2

CF-MIN +5V R156 221k R157 8k25 IC104B LM393/M 6 7 LM393/M 8 -5V +5V R158 1M00 R167 22k1 R160 100R -6V2 R159 1k50 C129 0.1u C130 4n7 5 D103 BAS16 R161 51R1 -6V2

-6V2

1 1

2 C120 +VREF SW-AMPL +28V R150 47k5 4 2 1 3 IC104A 0.1u RV101 100K R120 22k1 R117 22k1 D109 BAS16 R123 274k R126 10k0 T101 BC860C R153 * R113 6M8 C107 R125 8k25 9 8 10 IC101C TL074 R190 6M8 3 1 R148 10k0 RV149 10K MKR-Offs. 2 R149 10k0 +VREF -6V2 C136 0.1uF x100V -12V R10b 8k25 T102 BC850C -6V2 C138 3n3 R119 22k1 C121 220p R154 22k1 * R118 4k75 R155 47k5 -6V2 R10a 10k0 14 IC101D TL074 R115 10k0 R10c 6M8 R10d 6M8

T108 BC850C

+20V 60mA

C137 0.1u

C141 0.1u

R116 100k

12

C142 22u /35V

IC108 LM317

I N ADJ O U 2

1 C123 22u /35V

13

+VREF

D102 BAS16

R164 221R P111 +VREF (+13V)

R163 2k05

R128

14k3

TO FC PCB W305 S.5

C108 1nF

C122 22u C112 1nF 1 2 3 1 2 3 MK GND XF /35V

RV110

P108

MKR

10K 2

/10turns

TO XY PCB W801 S.3

X1 MPOS MOFF GND UREF CF FINE

R129 6k19

C118 22u /35V

1 2 3 4 5 6

VREF

C135 0.1u

R152

1k00

RV112 1K0

-5V +5V R121 221R R122 68k1 IC101B TL074 6 7 5 0.1u 0.1u 0.1u 0.1u C102 C104 C111 C133

-6V2 C101 C103

MK-MIN

C117 22u /35V

C119 0.1u

R151 18k2

0.1u

0.1u

-6V2

(PS) 110...210ns 15 12 %10 R222 51R1 9 IC206C 74HC02 1 0 IC203 HEF4020 12 HEF4011 8 10 9 R267 1k00 13 R262 51R1 IC204C R249 51R1 1 1 1 0 IC206D 74HC02 R S T C L K 13 11 1 1 1 1 3 2 1 5 4 2 3 6 4 5 7 9 R233 51R1 11 13 9 3 1 IC204A HEF4011 %10k 1 R257 51R1 11 10 15 12 14 CKA CKB CLR IC202B 74HC390 J201 COAX R224 51R1 TO TG J402 S.6 8 MHz OUT %4 2 D225B %2 BAW56 3 5 R255 51R1 D225A 3 IC205C 74HCU04 6 R266 51R1 R225 2k21 %8 R223 51R1 UNC D265 BAS16 5 IC204B HEF4011 6 IC206B 74HC02 R258 51R1 C265 0.1u PE QA QB QC QD 13 11 10 9 C262 4n7 LM 317 R254 2k21 IC205D 74HCU04 8 IC204D HEF4011 2 %2k %1k 6 4 5 4 R264 51R1 LE D263 BAS16 R263 100k Q Q Q Q Q 1 1 1 1 1 Q Q Q Q Q Q Q 4 3 2 1 0 9 8 7 6 5 4 1 12 8 1 R221 51R1 CKA CKB CLR QA QB QC QD 14 13 11 10 9 R250 22k1 2

+5V

TOP VIEW
3 R259 51R1 BAV70 2 1 3 IC206A 74HC02 R261 51R1 IC205A 74HCU04 1 4 2 BAW56 CKA CKB CLR IC202A 74HC390 R256 51R1 %4 1 2 QA QB QC QD 3 5 6 7

C201

R211 51R1

0.1u

%2

C202 0.1u

1 4

CKA CKB

IC201B 74HC390

K R212 51R1 BAS16

CLR

QA QB QC QD

3 5 6 7

C203 0.1u

IC201A 74HC390

C L k

C204 0.1u R251 51R1 R252 51R1 12 4

IC205F 74HCU04

C BC850 +5V B E BC860

C205 0.1u

C216 10nF

R216 51R1

C206 0.1u

IC205B 74HCU04

R169 3

LM 7805

1k00

R199 1k00

1 2

2 1

XT254 8MHz

C199 0.1u

C169 0.1u C255 1nF 1 2 IC205E 74HCU04

P201 MOLEX6

1 2 3 4 5 6

R253 10k0

MMS0204 / 0.25W / 5%

C253 27p

C254 56p

TO AL PCB W1203 S.1

+ V T G G P 2 T V N N S 0 C D D V

R265 4k75 1 2 3 4 5 6

P202 MOLEX6

HAMEG INSTRUMENTS GERMANY Title MB PCB HM5010/11 C L P U G G L E E N N N K C D D TO FC PCB W303 S.5 Size C Date: November 6, 1995 Document Number MB50101B.SCH Sheet 4 of REV

Subject to change without notice

DIGITAL SECTION

W802

W801 6-Pol

YF1 NC YF2 CHECK-PTS. R817 +140V 10R +140V +73V R815 C824 0.1u/250V 47u/250V R868 1.5mA R816 1K82 +140V R828 10R0 R850 10R0 C827 C804 0.1u 0.1u R843 100R T809 BFS20 T814 BF422 X AMPL T813 BF422 2 R855 1K82 3 R857 100R 1 R819 1K82 RV803 4K7 A R856 1K82 R860 4K75 R846 22K1 R822 2k74 R805A 100R C803 0.1u C823 10u/25V 3 10u/25V BAS16 R862 1K50 R 0R R829 10R R847 10R -12VA -12V C812 0.1u R861 3K01 C828 C809 0.1u T807 BC847B D802 1 R818 1K82 R870 4K75 R844 0R R842 4K75 C811 0.1u R820 100R R848 22K1 R859 2 T808 BFS20 100R 3 R858 100R R849 100R 1 X POS RV801 10K RV804 4K7 A 1 +12VY YP-sym T802 BFS20 R805 100k C802 0.1uF 3 2 R823 100R R845 22K1 10u/25V C810 C822 10u/25V 47K5 1W * C826 C825 0.1u/250V R821 100R R854 100R +20V...+125V R867 47K5 1W P810 2.5MB6 +73V +12V TO CRT PCB P702 S.7 +140V +32V GND +5V +12V -12V 1 2 3 4 5 6 1 2 3 4 5 6

1 2 3

1 2 3 GND GND NC XF1 +140V XF2

Y AMP
1 2 3 4 5 6

X AMP

3-Pol

TO CRT PCB P701 S.7

+73V

+73V

R814 100R

R866 30K1 1W

R865 30K1 1W

R810 100R

2.0mA

Subject to change without notice


R827 2K21 R803 2k74 +5V +12VY +12V C896 0.1uF R895 100k R898 22k1 P801 3 2 1 -12V XF GND MK C873 100p D874B C889 0.1uF R873 2M2 R896 * C897 * (221K) R874 * 1 (BAV99) D874A 2 * 2 1 R889 1K00 R885 51R1 AUDIO - AMPL. R897 20k0 R883 100R 1 R877 0R R881 10k0 6 5 C818 10n VOLUME T810 BD850 P802 R864 10R R872 182k RV872 4k7 A 3 1 2 R876 0R 3 2 8 IC875B TL082 7 C882 220uF/16V 3 2 1 R887 332R P870 3 2 1 3-Pol NF GND +12V 2 FROM PA PCB W??? S.? 3 T885 BC850 VIDEO SIGNAL FROM IF-AMP C871 2n2 R871 221k R882 68k1 IC875A TL082 1 R863 100K R 0R i.r. R880 332k R875 0R C878 10u/25V C879 0.1uF FROM IF AMP KA1101 S.2 R879 1K00 +12V -12VA +32V P808 6 5 4 3 2 1 3 -12V C821 * P804 +140V C820 * P805 C893 0.1uF IC891 LM317 A IN D OU J 1 2 R 0R R892 221R C892 0.1uF 6 5 4 3 2 1 MOLEX6 FROM MB PCB W??? S.? -12V nc GND nc +5V +28V 1 2 3 4 5 6 4 R884 100R 2 1 3 T886 BC860 R889 0R VF BW +12V +5V GND -12V 1 2 3 4 5 6 6 5 4 3 2 1 C819 0.1u C816 * C817 0.1u 1 2 3 4 5 6 (P803A / 5005) P803 +12V 6 GND 5 +5V 4 GND 3 VS 2 BW 1 R 0R FROM FC PCB KA301 S.5 R886 51R1 R888 0R R 0R P807 1 2 3 4 5 6 GND GND GND +5V +12V -12V 1 2 3 4 5 6 1 2 3 4 5 6 C815 FROM TG UNIT KA501 S.6 * C814 0.1u C813 0.1u Title XYF PCB HM5010 Size C Date: November 6, 1995 Document Number XY50101B.SCH Sheet 3 of REV 6 5 4 3 2 1 6 5 4 3 2 1 +140V +32V GND +5V +12V -12V C891 0.1uF R891 4K75 FROM PS P1008 S.8 HAMEG INSTRUMENTS GERMANY

+133V

R825

R806 100R

R824 100R

100R

T803 BFS20

Y AMPL

T811 BF422

T812 BF422

R826 *

R807 100R

R812 1K50

R813 100R

R809

1K50

RV802A 4K7

R811 1K50

R804 2K21

R808

1K50

R836 10K0

R801 4K75

T801 BC847B

R802 1K0

D801 BAS16

C801 0.1u

R830 10R

+12V

R841 100R

R833 100K

R838 2K74

C806 0.1u

T806 BC860C

R853 10R

3 2 1

C807

FROM MB PCB P111 S.4

22p

T805 BC850C

R839 1K50

R834 12K1

T804 BC850C

R852

10K0

R835 12K1

R832 22K1

R840 *

R837 10K0

C808 22p

R851 6K81

R831 10R

C805 0,1uF

-12V

-12V YPOS GND nc +5V +12V

FROM RB PCB KA??? S.?

SOT23

Top View

BC850

R-Chip 1% 1206

BAS16

BC860

R-Chip 5% 1206

nc

XY Board

BFS20

SXA 0411 Met. Fi. 1% 1W

33

34
+5V IC321A 74HC74 4 D P R 6 CLK C L Q D340 BAW56 +5V 3 1 BAV70 4 1 ALPS R342 Q 2K21 1 C L Q 3 R344 221K +5V 8 1 3 +5V R359 100R 1 3 T340 BC850 2 3 RV310 10K A 9 CF FINE 2 MARKER OFF 2 1 3 2 1 0 D P R CLK 5 2 D341 TO MB PCB P108 S.4 6 3 1 SW305 R343 10K0 1 Q 5 /60mm W305 6pol 2 3

FC-Board

IC320 HEF4017

14 13 3 2 4 7 10 1 5 6 9 11 12 12 11 +5V IC321B 74HC74

CLK ENA

15

RST

Q0 Q1 Q2 Q3 Q4 Q5 Q6 Q7 Q8 Q9

1 2 3 4 5 6

X1 MPOS MOFF GND UREF CF FINE

CO

D301

+5V

BAV70

D302 BAV70

L302 100uH

IC300 ICM7217N

7 6 5 4 D303 BAS16 6 6 1 6 6 1 1 6 1 1

18 17 16 15

T301 BC850

LED304 GREEN CTR R306 10K0

10

R301

14

4K75

C301 220u /16V

C302 0.1u

24

19

10 9 8 5 4 2 3 7 10 9 8 5 4 2 3 7 10 9 8 5 4 2 3 7 10 9 8 5 4 2 3 7 10 9 8 5 4 2 3 7

R303 150R R308 3K32

9 IC301 HP3603 1000MHz IC302 HP3603 100MHz IC303 HP3603 10MHz IC304 HP3603 1MHz IC305 HP3603 0.1MHz

20 23 27 25 28 22 26 21

12

R307 10K0 MKR LED303 GREEN T302 BC860

R302 10K0

R304 10K0

R340 475R +5V R335 39R2

R341 475R

R336 39R2

LED301 RED UNCAL

+5V

R311 182R LED305 2 LED306 IC308 LED307 IC307 MC14029 +5V 74HC42 3 R313 22K1

T303 BC860

D304 BAS16

100

IC306A MC14011 1

R310 221R

50

D305 BAV99

20

1 LED308 5 LED309 2 LED310 7 CO 1 LED311 5 15 1 9 10

R309

+5V

10

R314 33K2 A B C D QA QB QC QD A B C D 15 14 13 12 6 11 14 2 C305 0.1u 4 12 13 3

10K0

C310 0.1u

C311 0.1u

C309 0.1u

T304 BC850

C304 0.1u IC306B MC14011 6 4 5 C306 22n L301 100uH

1 2 3 4 5 6 7 9 10 11 0 1 2 3 4 5 6 7 8 9 CI CLK PE B/D U/D

C303 220u /16V

T305 BC850 LED312

0.5

R325 100R

R324 100R RV301

R315 221R LED313 +5V LED314 8 10 LED315 T306 BC860 IC306C MC14011 9 R335 4k75

0.2

R323 82K5

R312 1M00 R326 100R IC306D 13 11 12 MC14011 VF BW +5V +5V R339 * 3 1 1 2 3 4 5 6 R322 1K00 R331 33K2 2 BW C339 * R338 * 1 2 3 4 5 6 W304 6pol /70mm 5 2 B A B C D T W V Z C TO MB PCB P107 S.4 4 1 5 4 2 1 SW304 6 3 6 SW303 3 T331 BC850 TO XYF PCB P802 S.3 R321 33K2 C308 0.1u W301 6pol R327 100R

5K0 TR

P301 TRACE ROTATION COIL 1 P302

0.1

BAV70

D306

R316 221R

ZERO

2 R319 33K2 R332 1K00 VF

R336 *

BAW56

D307

/300mm

R317 100K +12V R329 *

R318 100K

R330 10K0

- G + + B V 1 N 5 1 WF 2 D V 2 V V R320 33K2 C307 0.1u

1 2 3 4 5 6

W303 6pol

/65mm

C L P U L E E N K C SW302 3 ON 2 1 OFF ALPS

G G N N D D

SW301

12.5kHz

TO MB PCB P202 S.4

250kHz

ALPS

UP

SCANWIDTH

DOWN Title

HAMEG INSTRUMENTS, GERMANY

BANDWIDTH 12.5kHz

VIDEOFILTER OFF

FC-PCB HM 5010/11 Size C Date: November 22, 1995 Document Number FC50101A.sch Sheet 5 of REV

Subject to change without notice

PA-Board

Subject to change without notice

22

23
D701 1 2 3 1 2 3 D14-364 TO MB PCB P110 S.4 13 9 R706 10K0 4 12 R710 3 2 1 5 IC701 6 CNY17 11 7 6 GND BLANC GND R701 681R KA701

CRT-Board

LL4151

D702 LL4151

2 R705 T701 BC850 10K0 C701 0.1u R704 1k0 D703 TLSG5101 1 1 4

51R1

KA702

BLV FCD G1 -1925V H1 H14

1 2 3 4 5 6

1 2 3 4 5 6

TO PS P1002 S.8

R708 47k5 VR701

R703 47k5

20k0 C703 R707 51k1 R702 68k1 C702 0.1u/250 1 2 3

P701 1 2 3 YF1 NC YF2

0.1u/100

P702

TO XYF PCB KA802 S.3

XF2 150V XF1 NC GND GND

6 5 4 3 2 1

6 5 4 3 2 1

HAMEG INSTRUMENTS GERMANY Title CRT PCB HM5005/6 Size A Date: March 14, 1994 Document Number CRT5005.SCH Sheet 7 of 8 REV

Subject to change without notice

TO XYF PCB KA801 S.3

Power Supply Board

Subject to change without notice

24

RA-Board

25

Subject to change without notice

Tracking-Generator

Subject to change without notice

26

RB-Board

27

Subject to change without notice

TO

Main Board

Subject to change without notice

28

Main board

29

Subject to change without notice

XY-Board

Subject to change without notice

30

XY-Board

31

Subject to change without notice

FC-Board

Subject to change without notice

32

FC-Board

33

Subject to change without notice

PA-Board

Subject to change without notice

34

CRT-Board

35

Subject to change without notice

PS-Board

Subject to change without notice

36

PS-Board

37

Subject to change without notice

PS-Board

Subject to change without notice

38

Block Diagram HM5010/HM5011

39

Subject to change without notice

Oscilloscopes Multimeters Counters Frequency Synthesizers Generators R- and LC-Meters Spectrum Analyzers Power Supplies Curve Tracers Time Standards

HAMEG GmbH Industriestrae 6 D-63533 Mainhausen Telefon: +49 (0) 6182 / 800-0 Telefax: +49 (0) 6182 / 800-100 E-mail: sales@hameg.de service@hameg.de

Internet:

www.hameg.de

4S-5010-0010

Printed in Germany

Вам также может понравиться