Presented by
M. Anwarul Islam Medical Physicist SQUARE Hospitals Ltd Dhaka-Bangladesh
Supervised by
Golam Abu Zakaria Professor & Head Gummersbach Hospital, Academic Teaching Hospital of the University of Cologne Germany
Objectives
Optimal protection to the radiation worker & public Easy handling to regulatory compliance Feasibility check to IMRT up gradation Optimal RT service delivery with 3DCRT & IMRT Avoidance the unnecessary shielding cost
Comments
Similar for DCRT & IMRT. Workload is equal for both modalities for primary shielding thickness calculation
Scattering radiation Similar for DCRT & IMRT by patient Scattered radiation Similar for both modalities by primary barrier Linac head leakage 3 - 5 times more for IMRT than 3DCRT
Shielding features
Field Size: IAEA recommendation: 40cm 40cm Calculated data from SQUARE Hospitals: We analyzed 756 fields with 3DCRT /2D treatments Average field size: 13.81cm 15.55cm Maximum field size: 27.4cm 24.8cm
Shielding features
Field Size Effects:
Field Size VS Shielding Thickness
0.7 0.6 0.5 0.4 0.3 0.2 0.1 0 0 200 400 600 800 1000 1200 1400 1600 1800
Shielding features
Barrier Use Factor:
IAEA recommendation: 0.25 for primary beam of all angle (0, 90, 180, 270)
Calculated data from SQUARE Hospitals:
3DCRT IMRT
Use factor for 0 (316-45) = 0.19 0.20 Use factor for 90 (46-135) = 0.33 0.31 Use factor for 180 (136-225) = 0.15 0.29 Use factor for 270 (226-315) = 0.33 0.21 Combined use factor 0.25
226 - 315
46 - 135
136 - 225
Shielding features
Workload
Workload depends on - working hour per day - no. patients treat per day - working day per week
Shielding features
Workload
Workload usually 1000Gy/week used with 6MV for 3DCRT and 600Gy/week used for higher energy We calculated maximum workload 740Gy / week for 3DCRT
Shielding features
Workload for IMRT center
The proposed design limit for IDR is following an IAEA recommendation with 1 mSv/year for public and 20 mSv/year for occupational workers
IMRT center2 IDR<4 IDR<80 Output dose rate of the machine is 600MU/Min
1 2
IDR<10
Conclusion
Radiotherapy Data Analyzed from SQUARE Hospitals Ltd
Duration: 7 March, 2010 8 March, 2011 Total MU delivered (with all energies & modalities + IMRT QA) = 3601059 Total dose delivered to the patients & cubic phantom = 2108655 cGy
Conclusion
Radiotherapy Data Analyzed from SQUARE Hospitals Ltd
Survey (by NSRC) maxi. IDR in front of the door & linac control room = 02 S/h
(FS= 40 40 cm2, E = 10MV, DR = 600 MU/Min, Angles = 0, 90, 180, 270)
Calculated dose to the near door & control room = 0.2 mSv for 01 year.
Conclusion
Linac control room & In front of the door completely Controlled area and permissible dose limit = 20mSv/year but radiation dose level is 100 times less. Obviously it is a very safe for us but . . .