RS232/FLOW-BUS interface
General description
The RS232/FLOW-BUS interface is an interface between the FLOW-BUS and the RS232 V24 serial (computer) port. It will either be supplied as a separate enclosed unit with a FLOW-BUS connector and a RS232 connector or as an integral 14TE module of your E-7000 readout and control system. The converter offers communication with a baudrate up to 38,4 kBaud. Communication software support is available.
Specifications
Power supply Power consumption Operating temperature Storing temperature Housing dimensions Baudrates +15Vdc/+24Vdc +/- 10% +15Vdc : 50 mA +24Vdc : 35 mA 0+50 C -20+60 C box: 160x80x44mm module: 14TE 4800 Baud 9600 Baud 19200 Baud 38400 Baud FLOW-BUS: opto isolated Power : DC-DC converter
Galvanic Isolation
NOTE 1:
Normally installation on the FLOW-BUS has to be performed only one time. Mostly this will be done at Bronkhorst HighTech B.V. when your module has been built and tested. The address on the bus for this module will be stored when the module powers-off. At future power-on situations, it will be part of the FLOW-BUS on the same address each time. However, there can be situations you have to install your module on the bus again. For example: 1) When you want to rebuild your FLOW-BUS system. 2) When you want to add this module to another FLOW-BUS system. 3) When, at the check at start-up, the module finds out that its address to be on the FLOW-BUS is occupied already by another module. You will be notified by the light-pattern of the LEDs that you have to reinstall the module on the bus. In that case you may press the switch shortly for a few seconds until the green LED is blinking fast and than release it. The interface will search for a free address on the FLOW-BUS now.
Switch
Besides the command install node on FLOW-BUS several other commands can be given to the RS232/FLOW-BUS interface using the switch. The available commands depend on the situation the switch was pressed; at power-up or during normal operation. The following tables show the commands that can be given. Release the switch when the pattern appears on the green and red LED. LED pattern Red LED Off Off On Time switch Pressed 0-4s 1-4s 4-8s Command None Automatic installation on FLOW-BUS. This command can only be given during normal operation if the node address is occupied. Reset the converter.