Sampling Considerations for Equilibrium Dissolved Oxygen Sensors

Albert S. Tenney III Industrial Applications Manager Honeywell, Inc. Fort Washington, PA Presented to NUS 1996 International Chemistry On-line Process Instrumentation Seminar Clearwater, Florida, November 20-22, 1996

Industrial Automation and Control


Honeywell Inc.
U.S.A.: Honeywell Industrial Automation and Control, 16404 North Black Canyon Hwy., Phoenix, AZ 85023 Canada: The Honeywell Centre, 155 Gordon Baker Rd., North York, Ontario M2H 3N7 Latin America: Honeywell Inc., 480 Sawgrass Corporate Parkway, Suite 200, Sunrise, Florida 33325 Japan: Industrial Operations Tokyo, 4-28-1 Nishi-Rokugo Ohtu-ku, Tokyo 144, Japan Asia: Honeywell Asia Pacific Inc., Room 3213-3225, Sun Hung Kai Centre, No. 30 Harbour Road, Wanchai, Hong Kong Pacific Division: Honeywell Pty Ltd., 5 Thomas Holt Drive, North Ryde NSW Australia 2113 Northern Europe and Southern Africa: Honeywell Ltd., Honeywell House, Arlington Business Park, Bracknell, RG 12 1EB, U.K. Central Europe: Honeywell A.G., Kaiserleistrae 39, 63067 Offenbach, Germany Western and Southern Europe: Honeywell S.A., Avenue de Schiphol 3, 1140 Brussels, Belgium Eastern Europe: Honeywell Praha, s.r.o., Budejovicka 1, 140 00 Prague 4, Czech Republic Middle East: Honeywell Middle East Ltd., Khalifa Street, Sheikh Faisal Building, Abu Dhabi, U.A.E.

Helping You Control Your World

T talPlant
Recycled Paper

70-82-58-47 (JIT) 4/97

Honeywell Inc.

Printed in U.S.A.

Sampling Considerations for Equilibrium Dissolved Oxygen [DO] Sensors


Albert S. Tenney III Industrial Applications Manager Honeywell, Inc. Fort Washington, PA

Introduction
At standard temperature [25C] and pressure [1 atmosphere], clean water in equilibrium with air contains 8.24 milligrams of oxygen per liter of water. This is called the air saturation value at 25C and is usually reported as parts per million, ppm. The control of corrosion in modern high pressure boilers commonly demands that the dissolved oxygen concentration in the water be less than 4 parts per billion, ppb, or 1/2000th of the air saturation value. The 8.24 micrograms of oxygen contained in one milliliter [ml] of air saturated water is enough to raise the dissolved oxygen concentration in 2000 ml of water by 4 ppb. Thus, if 2000ml of water containing 4 ppb dissolved oxygen were exposed to just one ml of air saturated water, the dissolved oxygen concentration in the 2000 ml would be doubled. The concentration of oxygen in air is 20.9% by volume. The milligrams of oxygen per ml is calculated as follows: From the ideal gas law, PV = nRT where P is pressure, V is volume, n is the number of moles [gram molecular weights], R is the gas constant and T is the absolute temperature, the volume of one mole of gas at atmospheric pressure and 25C [298K] is
V = 1 mole x [0.08205 Liter-atm/mole K] x 298K/1 atm = 24.45 L The mg of oxygen per L is given by 1 L x [1 mole gas/24.45 Lgas] x [0.209 mole O2 / mole gas] x [32,000mg O2 /mole O2 ] = 274 mgO2 per liter of air.

Notes

Abstract
Boiler water in the low parts per billion dissolved oxygen concentration range can be significantly contaminated by oxygen intrusion from leaks and trapped air. Careful flow design is required to minimize errors resulting from external sources of oxygen. In many cases the flow dependence of the dissolved oxygen reading reveals the presence of contamination. The expected response for equilibrium sensors differs from that for diffusion rate sensors.

This is about 33 times as concentrated as oxygen in air saturated water. The amount of oxygen in 1 ml of air is that which is required to double the concentration of [33 x 2000] 66,000 ml of water originally containing 4 ppb of dissolved oxygen.

Distinguishing between a leak and trapped air involves comparing results over time. With a [constant] leak, the relationship between flow rate and the error in reported dissolved oxygen should be constant. With trapped air, the contamination and the positive flow rate dependence should decrease with time. Often this effect is reported as a sensor with slow response at low ppb concentrations. Whats really happening is the slow dissolution of trapped oxygen into the flowing sample. Once the trapped oxygen source is depleted, the sensor responds correctly.

Conclusion
Sampling systems for dissolved oxygen sensors should be devoid of leaks to the atmosphere, regions where air or stagnant water can communicate with the sample and downflowing. Leaks and traps can be detected from the flow dependence and time-dependent flow dependence of the reported dissolved oxygen concentration. Each time a trap is exposed to air, such as during air calibration, the trapped oxygen source is reestablished and the trap symptoms described above reappear.

Leaks and Dead Spots


Leaks can be either true leaks minute holes in the plumbing carrying the water usually observed at fittings or diffusion of oxygen from the surrounding air through the walls of non-metal tubing. Non-metal tubing is not recommended but is frequently used because of its flexibility. Any place where water fails to fill the plumbing or where the water flows down can become a dead spot and a source for oxygen contamination. An example of each is shown in Figure 1. Water flows up a sample line, over the oxygen-sensing portion of an equilibrium sensor, through the body of the sensor, out a side-arm outlet, down and then up a length of tubing leading to a funnel and, finally, over the top of the funnel. The funnel provides access to the water for the ampule of a wet chemical analysis of dissolved oxygen. The valve meters the sample flow and stops it for sensor calibration of servicing. Two things are wrong with this picture. First, as shown in the cutaway, the side arm outlet from the equilibrium probe is situated only part way up a large internal cavity.

Thick Permanent Membrane

Bifilar Pt Electrodes

anode:
2 H2O = O2 + 4H+ + 4 e-

cathode:
Permanent Internal Electrolyte O2 + 4H+ + 4 e- = 2 H2O

The water fills to the level of the side arm but, at least initially, an air pocket significantly larger that the 1 milliliter discussed in the introduction exists in the upper portion of the cavity. While the air in this pocket must eventually dissolve into the water flowing through the sensor, until it does, it serves as a source for oxygen contamination that renders the results gathered by the wet chemical method unreliable. Consequently, independent measurement of dissolved oxygen in water flowing out of an equilibrium probe is not recommended. The second error is allowing the water to the sampling funnel to flow down. Before flow starts, the tube will be filled with air. It is necessary to sweep the air from the tube before reliable results can be expected. Upflowing water will tend to lift air out above it. Downflowing water, however, can leave air bubbles trapped in the downflowing portion of the tubing. As long as the buoyant upward velocity of the bubbles exceeds the sample flow, the bubbles will remain in the downflowing portion of the tubing and will serve as a source of oxygen contamination until they are dissolved.

Ceramic Mandrel

O2 Transport

Stagnant Water

Ampule for Wet Chemical Method


Figure 6B. Equilibrium Sensor in Stagnant water: with no oxygen depletion, dissolved oxygen equilibrium between the sensor and the sample is maintained and the result is independent of flow rate.

Figure 6A. Equilibrium Sensor: Oxygen is determined by its rate of consumption at a cathode [-] while an equal amount of oxygen is generated at an anode [+]. No net oxygen is consumed from the sample.

DO Probe [Cutaway] Pocket, Initially an Air Pocket outlet Funnel

Anode [+]:

Pb Anode
2 Pb + 2 H2O = 2 PbO + 4H+ + 4 e-

DO Probe inlet

Replaceable Electrolyte

Inert Cathode Replaceable Membrane

Cathode [-]: O2 + 4H+ + 4 e- = 2 H2O

Stagnant Water

Valve

Oxygen Transport

Oxygen Transport Diminished by Depletion from Stagnant Water

Figure 7A. Typical Diffusion Cell: Oxygen diffuses from the sample, through a membrane, to the cathode where it is consumed. The resulting current is a measure of its concentration. An anode reaction, usually metal oxidation, completes the circuit.

Figure 7B. Typical Diffusion Cell in Stagnant Water: Because net oxygen is consumed from the sample, stagnant or low-flowing water is depleted of oxygen producing a result which depends upon flow velocity.

Flow
Figure 1. Sampling from probe outlet, [not recommended]

Less obvious are the problems associated with the sampling system shown in Figure 2. Water normally flows through the equilibrium sensor and out its sidearm outlet as in Figure 1. When a wet chemical calibration check is to be performed, the valve in the sidearm outlet is closed, the valve below the funnel is opened diverting flow from the sensor to the funnel. The valve on the common line stops flow for calibration and service. Again, two problems present themselves. When the flow to the sensor is stopped, the stagnant water in the sensor and the line to it will initially be at the sample dissolved oxygen concentration. However, unless the sensor has had water flowing through it long enough to dissolve the internal air pocket, oxygen from the pocket will immediately begin to diffuse into the water in the cavity and down the sample line. The question then will be whether this diffusion process is slow enough that the oxygen reaching the line flowing to the funnel will be negligible. The other problem is a general one concerning allowing any stagnant water to communicate with the sample flow. In the normal operation, flow is to the sensor and goes past the piping to the funnel. The water in this stagnant leg is sealed from contamination by valve below the funnel. However, should this valve or any other fitting in this leg leak, oxygen can enter this stagnant leg and eventually reach the T below the sensor and begin contaminating the sample. This leak might be small enough to be negligible when its oxygen is continuously being diluted by the high flow recommended for the wet chemical analysis. However, a leak of similar magnitude might be large enough

to contaminate the stagnant water and the water flowing to the sensor since this flow is usually much less than that recommended for the wet chemical method. An arrangement eliminating sample communication with stagnant water is shown in Figure 3. Flow may be directed to either the sensor or the funnel and the 3-way valve isolates the sample flow from the stagnant water in both cases. Two problems with the arrangement shown in Figure 3 remain. First, the sensor and the funnel cannot be supplied with sample simultaneously. It is desirable to be monitoring the dissolved oxygen concentration with the sensor exactly when the sample for the wet chemical analysis is gathered. Simultaneous monitoring eliminates disagreements resulting from a fluctuating dissolved oxygen concentration. Simultaneous monitoring also allows a rapid return to the normal operation where the sensor only is monitoring without having to wait while air-contaminated water is purged from the line to the sample. Second, dissolved oxygen sensors are normally calibrated in air. It is desirable to be able to expose the sensor to air without opening plumbing connections. To achieve uncontaminated, simultaneous sampling and valve-controlled air calibration, the plumbing arrangement in Figure 4 is recommended. Normal operation consists of allowing flow to be metered through the 3-way valve to the sensor. The stagnant water in the line to the funnel is protected from air contamination by a second shutoff metering valve.

The sensor is exposed to air by switching the 3-way valve to drain. If this valve is solenoid-actuated, air calibration can be executed automatically by activating a relay in a sufficiently sophisticated analyzer. Furthermore, initiation of the calibration can be based on a programmed time or by an event. Simultaneous flow to the funnel and to the sensor is controlled by the two metering valves. Shown in Figure 5 is another arrangement that meets all of the requirements of uncontaminated simultaneous sampling and valve actuated calibration. The only drawback is that two connections into the main sample line flow are required.

Consequences Detection of Oxygen Contamination


In spite of all the precautions described above, dead spaces can exist in complicated sampling arrangements and leaks can always occur. Their presence can be revealed by the response of the sensor. The sensors dependence on sample flow rate is a key diagnostic. Equilibrium dissolved oxygen sensors are independent of flow rate. Within the sensor, oxygen is consumed at an electrode and its concentration determined from the direct current resulting from this process. The amount of oxygen consumed is balanced by oxygen produced electrochemically at a counter electrode. No net oxygen is consumed; there is no oxygen depletion in the sample; the sample just outside of the sensor need not be replenished so flow dependence is zero and oxygen may

be measured in still water if desired, see Figure 6. This is to be contrasted with oxygen sensors which consume oxygen electrochemically but which do not internally balance this consumption by a replenishing reaction. For these sensors, oxygen must continually diffuse from the sample, through a membrane, through some portion of an internal electrolyte solution to an electrode where it is consumed. Net oxygen is consumed from the sample. A concentration gradient develops from a point some distance from the sensor, through regions of diminishing dissolved oxygen concentration to the consuming electrode where the dissolved oxygen concentration is zero. The current from which the dissolved oxygen concentration is inferred depends on the diffusion path. If the flow rate past the sensor is diminished, replenishment of the consumed oxygen is lessened and the resulting drop in current is unrelated to the bulk dissolved oxygen concentration. Diffusion rate sensors demand a minimum flow rate below which the reported dissolved oxygen concentration will decrease with decreases in flow rate, see Figure 7. How does all of this allow inferring the presence of a leak? From this discussion of how these sensors work, we see that a decrease in flow rate should result in either [1] no change in signal for an equilibrium sensor or [2] a decrease in signal for a diffusion rate sensor. If, however, a decrease in flow leads to an increase in signal, this is almost surely related to oxygen leaking into the sample. For a given leak rate, a decrease in flow would correspond to an increase in dissolved oxygen concentration because lowering the flow rate dilutes the leaking oxygen less.

Outlet

Funnel

Outlet

Funnel

Outlet

Funnel

Outlet

Funnel

DO Probe
DO Probe Valve inlet

DO Probe

DO Probe Valve

inlet
inlet

Drain [for Air CAI]

inlet 3-way Valve Valve

Drain [for Air CAI]

3-way Valve
Valve Valve

Valve

Valve

3-way Valve

Valve

Valve

Flow

Flow

Figure 2. Parallel sample lines

Figure 3. Parallel sample lines eliminating cross contamination

Figure 4. Parallel sample lines eliminating cross contamination, allowing air calibration and simultaneous sampling. [Recommended in this report.]

Figure 5. Parallel independent sample lines. [An alternative recommended in this report.]