Вы находитесь на странице: 1из 8

GYROZEN 416G

BenchtopCentrifuge

416G

UsersManual

765-22 Dohwa-dong, Nam-gu, Incheon 402-063, KOREA TEL +82 32 8661541 FAX +82 32 8661547 www.gyrozen.com

GYROZEN 416G

Contents

1. Precautions 2. TechnicalSpecifications 3. Installation 31.DeliveryChecklist 32.UnpackingtheInstrument 33.InstallationPlace 34.SupplythePower 35.On/offandthedooropen 36.Rotorcouplinganddisassembling 4. Operation 41.ControlPanel 42.FunctionDescription 43.OperationSequence 44.MiscellaneousFunctions 5.Maintenance 6.TroubleShooting 7.OrderingInformation

GYROZEN 416G

1. Precautions
Theinstrumentmustbeinstalledonaflat,stable,horizontalandsolidsurface. Beforeoperatingtheinstrument,checkiftherotorandtherotorlidaresecurelyfastened. Donotmovetheinstrumentwhenitisnotcompletelystopped. Operatewithastable&appropriatepowerinlet. Donotplacedangerousmaterialswithin30cmdistancearoundtheinstrument. OnlyuserotorsfromGYROZENCo.,Ltd.withappropriatetubesforcentrifugation. Donotusehazardous,inflammableorradioactivematerialsassamples. Whenitisnecessarytousetoxicorradioactivematerials orpathogenicmicroorganisms which belong to the Risk Group II of WHO: Laboratory Bio safety Manual, follows nationalregulations. Usetheemergencydooropenfunctiononlywhenthedoorkeyonthecontrolpanelisnot working. Thesamplemustbeloadedweightbalancedwaynottocauseimbalance. Thedensityofsamplematerialmustnotexceed1.2g/mlatMax.RPMorRCF. Anyimproperhandlingoranyusageofunqualifiedaccessoriesisnotabletobeprotected.

2. TechnicalSpecifications
Max.RPM Max.RCF Max.capacity Runtime Noiselevel Accelerationtime Decelerationtime ImbalanceRecognition Motor Safety AvailableRotors Power&Frequency Dimension(WxDxH)mm Weight *Rotor:(S)Swingrotor 416G 4,000rpm 2,700xg 15mlx16 99min59secorcontinuous 60dB 20sec 25sec Automatic HightorqueACinductionmotor Lidlock 15mlx16,50mlx4(S) 220V,50/60Hz 377x438x251 22Kg

GYROZEN 416G

3. Installation 31.DeliveryChecklist
311.Mainbody,1ea 312.Powercable,1ea 313.Emergencydooropentool,1ea 314.Usersmanual,1ea 315.Rotor(Optional)

32.Unpackingtheinstrument
321.Opentheboxandliftouttheinstrumentcarefullytogetherwiththesafetypadding. 322.Removethesafetypaddingandvinylwrap. 323.Placetheinstrumentontheflatspace.

33.Installationplace
331.Installthemachineatthesolidandflatfloorortable.Ifyouplacethecentrifugeattheslope, theaxisofrotationispossiblychangedbecauseoftherotorweight. 332.Installthemachineabout30cmisdepartedfromthewallbecauseoftheaircirculation. 333.Installthemachineattheplacewithappropriatetemperatureandhumidity.Italsohastobe maintainedwiththepropertemperature&humidity. 334.Installthemachineattheplacewithoutanykindsofcorrosivegases.

34.Supplythepower
341.416Gmodeluses110Vor220V.Checkpropervoltageofyourinstrumentandconnectto adequatepoweroutlet. 342.Ifthepowerinputismorethan+/10%oftherecommendedvoltageorfluctuatingfrequently, itmayaffectsomefunctionsoftheinstrument.ItisadvisedtouseAVR(AutomaticVoltageRegulator). 343.Ifyouwanttousetheinstrumentatothervoltagerange,pleasecontactus.

35.On/offandthedooropen
351.Turnontheinstrumentbypressingtheswitchontherightsideofthemachine. 352.PresstheDoorbuttontoopenthedoor. 353.Removetheprotectionmaterialsinsidethechamber.

36.Rotorcouplinganddisassembling
361.Mounttherotorfittingwiththemotorshaft. 362.Grasptherotorwithonehand,androtatetherotornutinthedirectionofclockwise,untilitis tightlyassembled. 363.Positiontheplasticlidproperlyandpressslightlydowntoclose. 364.Therotorlidisdisclosedbyliftinguptheknob. 365.Todisassembletherotor,rotatetherotornutinacounterclockwisedirection.

GYROZEN 416G

4. Operation 41.ControlPanel

411.DisplayFND:Showsthedataofeachfunction. 412.Up&downarrow:Usestochangeinputdata. 413.Door:Usestoopenthedoor. 414.RPM/RCF:UsestoswitchtheRPM/RCFdisplay 415.TIME:Usestosettesttimeupto99min59sec(00:continuous) 416.START:Usestostartoperation.Thelatestsettingvalueisrecalledifthereisnonewsetting 417.STOP:Usestostopoperation 418.SOFTSTART&STOP:Usestoaccelerateordecelerateslowly. *Ifyoupressthearrowbuttonlongerthan3seconds,thenumberschangeveryrapidlyandthesetup isachievedfaster.

42.FunctionDescription
421.RPM Speeddisplayunit:1rpm Speedsettingunit:10rpm Speedaccuracy:<2%(atmaximumspeed) PresstheRPM/RCFbutton. Pressthearrowbuttontochangeinputvalue. 422.RCF(RelativeCentrifugalForce) The conversion value from RPM to RCF is calculated automatically (based on the maximum radiusandrevolution) PresstheRPM/RCFbuttontwice. Pressthearrowbuttontochangeinputvalue.

GYROZEN 416G

423.Time Timedisplay:99min59sec,2digitdisplay Timeisdowncountedafterstartingcentrifugation. PresstheTIMEbutton.ThedisplayshowstheprevioustimevalueandtheminuteLEDandvalue blinks. Pressthearrowbuttontochangetheminutevalue. Press the TIME button again to set the second value, then the second LED and value blinks automatically. Pressthearrowbuttontochangethesecondvalue. *Afterfinishingthesetting,yourinputvalueissavedautomaticallyin2~3seconds. 424.SoftStart&Stop For the sensitive samples not to disturb contents by fast acceleration and deceleration, use SOFT START/STOP. Centrifuge is accelerated or decelerated slowly. Press SOFT START / STOP buttoninsteadofSTARTbuttontousethisfunction. 425.Door Press the Door button to open the door.The door is not opened while the instrument is running. 426.START PressingtheSTARTbutton,theoperationisstartedtorunwithsettingvalue. 427.STOP PressingtheSTOP,theoperationisstoppedtoruninanycase.

43.OperationSequence
431.ConnecttheACcordtotheappropriatevoltage. 432.Turnonthepowerswitchlocatedattherightsideoftheunit.Thenthelatestoperatingvalues arerecalledwithbeepingsound. 433.Openthedoorandchecktherotor. 434.Inputthefunctionvaluesaccordingtoyourpreferences. 435.PresstheSTARTbuttontostarttheoperation. 436.Assoonasthesettimeisended,thecentrifugationwillbeceasedaccordingtotheset decelerationlevel.Ifyouwanttointerruptbeforethesettime,presstheSTOPbutton.

44.MiscellaneousFunctions
441.Inanycasesneededforemergencyopen,youcanusetheemergencydooropentoolaslongas themachineisidle.Inserttheemergencytoolintotheemergencydooropenholeattheleftupsideof theunit. 442.ImbalanceAlarm:Whentherotorisimbalancedandthevibrationissevere,theoperationis urgentlystoppedwiththeerrormessageonthedisplayandthealarmingsound.

GYROZEN 416G

5.Maintenance 51.Outerpartofinstrument
511.Cleantheoutsideofthemachinewithadrysoftcloth.Ifnecessary,diptheclothwithneutral detergentsandcleancontaminatedparts.Keepcompletelydryaftercleaning. 512.Donotuseanyvolatilechemicalssuchasalcohol,benzene,etc. 513.Ifanyrustappears,cleanwithneutraldetergentsanddryit.

52.Innerpartofinstrument
521.Keepdryinsidethechamberaftereveryuseofthemachine. 522.Cleantheshaftalwaysforavoidinganimbalanceduringtherotation. 523.Ifanypartiscontaminated,cleanwithneutraldetergents.

53.Rotor
531.Cleantherotorifanypartsarecontaminatedwithsamples. 532.Keepdryafterusage.

54.Movingorshippingofinstrument
541.Ifyouneedtomoveorshiptheinstrument,becautioustoprotecttheshaftfromanyphysical impact. 542.Removetherotorandfillinsidethechamberwithpropermaterialstokeeptheshaftonplace. 543.Itisrecommendedtomoveorshiprotorsseparatelyifnooriginalsafetypaddingisavailable.

6.TroubleShooting
Actions ChecktheRPMsensor. E1 Failuretoreachto200RPMwithin2sec. ChecktheconnectionofaRPMconnectororawire.* Checkthedoorlatch. E2 Whenrunning,thedoorisopened. Stoprunning,poweroffandon. E3 Overtemperatureofmotor. Checkthemotortemperature. Checksupplyvoltage. E4 Lowvoltage UseAVRtoprovidepropervoltage. Checksupplyvoltage. E5 Highvoltage UseAVRtoprovidepropervoltage. Stoprunning,poweroffandon. E6 Overspeed Tuningofthefirmware(Download)* Stoprunning,poweroffandon. E7 Failureofcontrolsystem Tuningofthefirmware(Download)* Checkamountofsamples. E8 Imbalance Checkweightbalancesofsamples. CheckattachedplaceofRPMsensor. E9 RPMsensorrecognitionerror CheckconnectionofcableforRPMsensor. ChangeRPMsensor. *Anywiredisconnectionortuningoftheinstrumentmustbeperformedonlybyaserviceengineerwho isauthorizedbyGYROZENCo.,Ltd. Error PossibleCauses

GYROZEN 416G

7.OrderingInformation

CatNo. G17514201 G17514.202 G17514.203 G17514.204 G1111130 G1111040 G1111140 G1111100 G1111180 G1111200 G1113010 G1113040 G1113060 G1113020 G1113030 ProductDescription 416GMainBody,withoutRotor 416GincludingAngleRotor,GAM1516&16SteelSleeves,GBT15 416GincludingSwingRotor,GSM504&4Buckets,GBT50 416GincludingSwingRotor,GMT962&2Microplateholders,GBT96B AngleRotor,GAM1516,includes16SteelSleeves,GBT15 SteelSleeve,GBT15 SwingRotor,GSM504,includes4Buckets,GBT50 Bucket,GBT5050 SwingRotor,GMT962,including2Microplateholders,GMT96B Microplateholder,GMT96B 3mlAdapterforSteelSleeve,GBT15 5mlAdapterforSteelSleeve,GBT15 15mlAdapterforSteelSleeve,GBT50 15mlAdapterforSteelSleeve,GBT50 50mlAdapterforSteelSleeve,GBT50 Capacity 16x15 4x50 2x96P 16x15 15 4x50 2x96P 96P 3ml 5ml 15mlRound 15mlConical 50mlConical

Rotors&Accessoriesfor416G