8. Protect your investment with bandwidth upgrades 9. Serial Data Analyzer and Disk Drive Analyzer
models are tailored for advanced serial data analysis and for the most complete disk drive test solution
multiplexed from the front panel or by remote control. The result its easy to hook up a passive probe even on 4 or 6 GHz modelsno more frustration and hassle of trying to find a 1 M input adapter. Plus, any existing investment in LeCroy probes, such as current probes, single-ended or differential active probes, or high voltage probes, is fully leveraged. Perfect.
Single-ended active probes, current probes, high-voltage, mixed signals, and high frequency differential probes all connect to the WavePro 7 Zi-A oscilloscope and give you a total system view.
SPI
FlexRay
LIN
A B
Instantaneous Responsiveness
With WavePro 7 Zi-A oscilloscopes you will experience remarkable responsiveness. Acquiring and manipulating the longest record lengths and performing the most complex WaveShape Analysis are all easily
X-Stream II Architecture
Optimized for Fast Throughput
X-Stream II architecture enables high throughput of dataeven when the oscilloscope is performing multiple 100 Mpts (or larger) waveforms. X-Stream II uses variable waveform segment lengths to enable all processing intensive calculations to take place in fast CPU cache memory, thus improving calculation speed and efficiency. The result10100x faster processing compared to other oscilloscopes.
Learn More
http://www.lecroy.com/dl/5213 8
Learn More
http://www.lecroy.com/dl/5214
Customized Tools
Only LeCroy completely integrates third party programs into the scopes processing stream by allowing you to create and deploy a new measurement or math algorithm directly into the oscilloscope environment and display the result on the oscilloscope in real-time! There is no need to run a separate program, or ever leave the oscilloscope window. Use C/C++, MATLAB, Excel, Jscript (JAVA), and Visual Basic to create your own customized math functions, measurement parameters, or other control algorithms.
X-Stream II fast throughput streaming architecture makes difficult analysis and deep insight possible. Above, an FFT is applied to a 50 Mpts waveform to determine root cause failure. The high frequency resolution this provides enables deep insight into signal pathologies.
XDEV Customization software package being used to implement a 1 MHz Butterworth filter using MATLAB.
PROBES
High-performance probes are an essential tool for accurate signal capture. Consequently LeCroy offers an extensive range of probes to meet virtually every application need. Optimized for use with LeCroy oscilloscopes, these probes set new standards for responsiveness and signal detection. ZS Series High Impedance Active Probes
1 GHz (ZS1000),1.5 GHz (ZS1500)
bandwidths
Wide dynamic range, low noise LeCroy ProBus System
high-voltage measurements
1.2 kV to 20 kV Works with any 1 M input
oscilloscope
(ZS1500)
LeCroy
ProBus system
ADP305, ADP300
20 MHz and 100 MHz bandwidth 1,000 Vrms common mode voltage 1,400 Vpeak differential voltage EN 61010 CAT III 80 dB CMRR at 50/60 Hz LeCroy ProBus
ZD200
200 MHz bandwidth 3.5 pF, 1 M Wide dynamic and common
mode range
LeCroy ProBus System Ideal for serial data and automotive
system
system
applications
10
WAVELINK PROBES
D600A-AT/D500PT Browser
WaveLink browser solutions offer adjustable tip widths and varying form factors and a hand held x-y-z positioner for accurate probe placement.
11
ProtoSync Solutions
ProtoSync links physical layer waveforms, data link layer decode annotation and table information, and full transaction layer protocol analysis together. By simply touching a
decode table entry in the oscilloscope software or a packet in the protocol analysis software, all views are automatically synchronized and aligned for quick and easy debug.
13
Key Features
LeCroys unique summary
view displays the Eye Pattern, TIE, Bathtub Curve and Jitter Histogram all on the screen at the same time
IEEE 1394b FireWire Rapid I/O LP-LVDS Fibre Channel IOF Ethernet Rapid I/O LP-LVDS Rapid I/O LP-LVDS MIPI D-PHY SAS SerialATA IEEE 1394b FireWire HDMI 1.2a / DVI Rapid I/O LP-LVDS Fibre Channel InfiniBand PCI Express Rapid I/O LP-LVDS
15
Jitter Decomposition
Tj
Total Jitter
SDA IIs toolset allows you to drill down for the deepest levels of insight
Rj
Random Jitter
Dj
Deterministic Jitter
BUj
Bounded Uncorrelated Jitter
DDj
Data Dependent Jitter
Pj
Periodic Jitter
OBUj
Other Bounded Uncorrelated Jitter
DCD
Duty Cycle Distortion
ISI
Intersymbol Interference
does not appear as a peak in the spectrum is Rj. This is not always the case, and in these cases the spectral method will return incorrect results. The NQ-Scale method consistenly yields correct results even in the cases where the spectral method fails to do so.
Tj Analysis
The SDA II analysis package has the deepest toolset for total jitter analysis. Unique tools such as IsoBER, Mask Violation Locator and PLL Track allow you to gain unparalleled insight into your serial data signal.
RjBUj Analysis
The SDA II analysis package first finds and removes Data Dependent Jitter (DDJ) from the serial data signal. This allows dedicated tools for RjBUj analysis (RjBUj Track, RjBUj Spectrum, RjBUj Histogram) to provide a view into the causes of jitter that are not distorted by the effects of DDj. These tools allow you to drill down directly to the source of your jitter problems that are caused by either random jitter (Rj) or Bounded Uncorrelated Jitter (BUj).
DDj Analysis
By first finding and removing the Data Dependent Jitter (DDj) from the serial data signal, SDA II enables DDj analysis to be performed on your serial data signal. The DDj Plot (with Digital Pattern Overlay), DDj Histogram and ISI Plot are dedicated tools for DDj Analysis that allow you to get to the root cause of jitter problems caused by Data Dependent Jitter.
Pj Analysis
A unique feature of the SDA II analysis software is the Pj Inverse FFT function. The tools gives you a new view into your periodic jitter by performing the inverse FFT of only the peaks in the spectrum. This allows you to view your periodic jitter in the time domain which can add additional insight into your jitter problems.
17
Key Features
3.5 or 6 GHz Zoom on multi-zoom
on sectors
One button access to
18
Analog-to-digital converters running at speeds up to 40 GS/s ensure the right sensitivity to measure todays high-speed read channels. In every DDA, you can run your customer-developed scripts to view the captured signal with the filters matched to your channel and media. Custom user scripts can be created in MATLAB, Visual Basic, Excel or other formats.
sample time resolution. Up to 15,000 sectors of data can be gathered with the DDA 7 Zi-A analyzers.
The DDA 7 Zi-A provides one button access to all the tools needed to accurately debug and analyze disk drive operation. The DDA 7 Zi Features:
Cascade Triggering
Triggering allows up to two events to qualify a third event (arm on A event, then qualify on B event, then trigger on C event) for precise trigger control. For instance, this could be used to Arm when the Index signal goes high, qualify when the Read Gate signal goes high, then trigger on a Head signal.
Simultaneously connecting low-speed signals, like index and servo gate, and high-speed signals, like read channels has never been easier. With integrated 50 and 1 M inputs on all models, there is no longer a need for expensive adapters.
19
SPECIFICATIONS
Vertical System
Analog (ProLink Input) Bandwidth @ 50 (-3 dB) ( 10 mV/div) Analog (ProBus Input) Bandwidth @ 50 (-3 dB) Analog (ProBus Input) Bandwidth @ 1 M (-3 dB) Rise Time (1090%, Flatness 50 ) Rise Time (Typical, 2080%, Flatness 50 ) Input Channels Bandwidth Limiters Input Impedance Input Coupling Maximum Input Voltage
WavePro 715Zi-A
Not Applicable 1.5 GHz ( 10 mV/div) 500 MHz (Typical) 235 ps 176 ps 4
WavePro 725Zi-A WavePro 735Zi-A WavePro 740Zi-A WavePro 760Zi-A (SDA) (SDA, DDA) (SDA) (SDA, DDA)
Not Applicable 2.5 GHz ( 10 mV/div) 500 MHz (Typical) 150 ps 113 ps Not Applicable 3.5 GHz ( 10 mV/div) 500 MHz (Typical) 120 ps 90 ps 4 GHz ( 10 mV/div) 3.5 GHz ( 10 mV/div) 500 MHz (Typical) 105 ps 79 ps 6 GHz ( 10 mV/div) 3.5 GHz ( 10 mV/div) 500 MHz (Typical) 70 ps 53 ps
Channel-Channel Isolation ProBus Input Vertical Resolution Sensitivity DC Gain Accuracy Offset Range
Offset Accuracy
50 (ProBus): 5 Vrms 50 (ProLink): 4 Vpeak 1 M (ProBus): 250 V max. (peak AC: 10 kHz + DC) > 200:1 up to 2 GHz, 200:1 up to 2 GHz, > 50:1 from 2 GHz > 50:1 from 2 GHz Not Applicable to 4 GHz to 4 GHz, > 20:1 from 4 GHz to 6 GHz 100:1 > 100:1 up to 2.5 GHz, > 30:1 from 2.5 GHz to 3.5 GHz 8 bits; up to 11 bits with enhanced resolution (ERES) 50 : 2 mV1 V/div, fully variable (29.99 mV/div via zoom); 1 M: 1 mV10 V/div, fully variable 1.5% of full scale 50 (ProBus Input): 50 (ProLink Input): 750 mV @ 10170 mV/div 750 mV @ 10118 mV/div 4 V @ 172 mV/div1 V/div 4 V @ 120 mV/div1 V/div 1 M: (ProBus Input): 50 (ProBus Input): 1 V @ 2128 mV/div 750 mV @ 10170 mV/div 10 V @ 130 mV1.28 V/div 4 V @ 172 mV/div1 V/div 100 V @ 1.3 V10 V/div 1 M: (ProBus Input): 1 V @ 2128 mV/div 10 V @ 130 mV1.28 V/div 100 V @ 1.3 V10 V/div (1.5% of full scale +1.0% of offset value +1 mV)
20 MHz, 200 MHz 1 GHz, 3 GHz 50 2% or 1 M || 16 pF, 10 M || 11 pF with supplied probe 1 M: AC, DC, GND; 50 : DC, GND 50 : 5 Vrms 1 M: 250 V max. (peak AC: 10 kHz + DC)
Horizontal System
Timebases Time/Division Range Clock Accuracy Time Interval Accuracy Jitter Noise Floor Trigger and Interpolator Jitter Channel-Channel Deskew Range External Timebase Reference (Input) External Timebase Reference (Output) External Clock Internal timebase common to 4 input channels; an external clock may be applied at the auxiliary input Real time: 20 ps/div1000 s/div (RIS mode: 20 ps/div10 ns/div; Roll mode: up to 1000 s/div) 1 ppm + (aging of 0.5 ppm/yr from last calibration) < 0.06 / SR + (clock accuracy* Reading) (rms) 1.5 ps (Typical) 1 ps (Typical) 800 fs (Typical) 750 fs (Typical) 560 fs (Typical) 2.5 ps rms (Typical) 2 ps rms (Typical) < 0.1 ps rms (Typical, software assisted) < 0.1 ps rms (Typical, software assisted) 9 x time/div. setting, 100 ms max., each channel 10 MHz; 50 impedance, applied at the rear input 10 MHz; 50 impedance, applied at the rear output 0.1 Hz-100 MHz, 50 or 1 M impediance, applied at the Auxiliary Input
Acquisition System
Single-Shot Sample Rate/Ch
WP715Zi-A
20 GS/s on 2 Ch 10 GS/s on 4 Ch
(Option WPZi-1.5GHZ-4X20GS doubles the sample rate)
WP725Zi-A (SDA)
WP740Zi-A (SDA)
40 GS/s on 2 Ch 20 GS/s on 4 Ch
Random Interleaved Sampling (RIS) Maximum Trigger Rate Intersegment Time Max. Acquisition Memory Points/Ch Standard Memory S-32 Memory Option M-64 Memory Option L-128 Memory Option
200 GS/s for repetitive signals (20 ps /div. to 10 ns/div) 1,000,000 waveforms/second (in Sequence Mode, up to 4 channels) 1 s (4 Ch / 2 Ch) 20 M / 40 M (Standard memory for SDA and DDA scopes are 32 M / 64 M) 32 M / 64 M 64 M / 128 M 128 M / 256 M
20
SPECIFICATIONS
Acquistion Processing
Averaging Enhanced Resolution (ERES) Envelope (Extrema) Interpolation
WavePro 715Zi-A
WavePro 725Zi-A WavePro 735Zi-A WavePro 740Zi-A WavePro 760Zi-A (SDA) (SDA, DDA) (SDA) (SDA, DDA)
Summed averaging to 1 million sweeps; continuous averaging to 1 million sweeps From 8.5 to 11 bits vertical resolution Envelope, floor, or roof for up to 1 million sweeps Linear or Sin x/x
Triggering System
Modes Sources Coupling Mode Pre-trigger Delay Post-trigger Delay Hold-off by Time or Events Internal Trigger Range Trigger Sensitivity with Edge Trigger (Ch 14) ProBus Inputs Normal, Auto, Single, and Stop Any input channel, Aux, Aux/10, or line; slope and level unique to each source (except line trigger) DC, AC, HFRej, LFRej 0100% of memory size (adjustable in 1% increments of 100 ns) 010,000 divisions in real time mode, limited at slower time/div settings or in roll mode From 2 ns up to 20 s or from 1 to 99,999,999 events 4.1 div from center 2 div @ < 1.5 GHz 2 div @ < 2.5 GHz 2 div @ < 3.5 GHz 1.5 div @ < 750 MHz 1.5 div @ < 1.25 GHz 1.5 div @ < 1.75 GHz 1.0 div @ < 200 MHz 1.0 div @ < 200 MHz 1.0 div @ < 200 MHz (for DC, AC, (for DC, AC, (for DC, AC, LFRej coupling, 10 mV/div, 50 ) LFRej coupling, LFRej coupling, 10 mV/div, 50 ) 10 mV/div, 50 ) 2 div @ < 4 GHz 2 div @ < 6 GHz 1.5 div @ < 2 GHz 1.5 div @ < 3 GHz Not Applicable 1.0 div @ < 200 MHz 1.0 div @ < 200 MHz (for DC, AC, (for DC, AC, LFRej coupling, LFRej coupling, 10 mV/div, 50 ) 10 mV/div, 50 ) 2 div @ < 1 GHz 1.5 div @ < 500 MHz 1.0 div @ < 200 MHz (for DC, AC, LFRej coupling) 1.5 GHz @ 10 mV/div 2.0 GHz @ 10 mV/div 2.0 GHz @ 10 mV/div 2.0 GHz @ 10 mV/div (minimum triggerable (minimum triggerable (minimum triggerable (minimum triggerable width 200 ps) width 500 ps) width 300 ps) width 250 ps) Aux (0.4 V); Aux/10 (4 V)
Basic Triggers
Edge TV-Composite Video Triggers when signal meets slope (positive, negative, or either) and level condition. Triggers NTSC or PAL with selectable line and field; HDTV (720p, 1080i, 1080p) with selectable frame rate (50 or 60 Hz) and Line; or CUSTOM with selectable Fields (18), Lines (up to 2000), Frame Rates (25, 30, 50, or 60 Hz), Interlacing (1:1, 2:1, 4:1, 8:1), or Synch Pulse Slope (Positive or Negative) Trigger when signal or exits a window defined by adjustable thresholds
Window
SMART Triggers
State or Edge Qualified Qualified First Dropout Pattern Triggers on any input source only if a defined state or edge occurred on another input source Delay between sources is selectable by time or events In Sequence acquisition mode, triggers repeatedly on event B only if a defined pattern, state, or edge (event A) is satisfied in the first segment of the acquisition. Delay between sources is selectable by time or events Triggers if signal drops out for longer than selected time between 1 ns and 20 s Logic combination (AND, NAND, OR, NOR) of 5 inputs (4 channels and external trigger input). Each source can be high, low, or dont care. The High and Low level can be selected independently. Triggers at start or end of the pattern
21
SPECIFICATIONS
Cascade (Sequence) Triggering
Capability Types Holdoff Reset
WavePro 715Zi-A
WavePro 725Zi-A WavePro 735Zi-A WavePro 740Zi-A WavePro 760Zi-A (SDA) (SDA, DDA) (SDA) (SDA, DDA)
Arm on A event, then Trigger on B event. Or Arm on A event, then Qualify on B event, and Trigger on C event. Or Arm on A event, then Qualify on B then C event, and Trigger on D event A or B event: Edge, Glitch, Width, Window, Dropout, Interval, Runt, Slew Rate, or Pattern (analog) C or D event: Edge or Pattern Delay between A and B, B and C, C and D, are all selectable by time or number of events Reset between A and B, B and C, C and D, are all selectable in time
Processor/CPU
Type Processor Memory Operating System Real Time Clock Intel Core 2 Quad, 2.5 GHz (or better) 8 GB standard Microsoft Windows 7 Professional Edition (64-bit) Date and time displayed with waveform and in hardcopy files SNTP support to synchronize to precision internal clocks
22
SPECIFICATIONS
Internal Waveform Memory Setup Storage
Front Panel and Instrument Status Store to the internal hard drive or to a USB-connected peripheral device
WavePro 715Zi-A
WavePro 725Zi-A WavePro 735Zi-A WavePro 740Zi-A WavePro 760Zi-A (SDA) (SDA, DDA) (SDA) (SDA, DDA)
4 active waveform memory traces (M1M4) store 16-bit/point full length waveforms Waveforms can be stored to any number of files limited only by the data storage media capacity
Interface
Remote Control Network Communication Standard GPIB Port (Optional) LSIB Port (Optional) Ethernet Port USB Ports External Monitor Port Via Windows Automation, or via LeCroy Remote Command Set LXI Class C, VXI-11 and VICP Supports IEEE 488.2 Supports PCI Express Gen1 x4 protocol with LeCroy supplied API Supports 10/100/1000BaseT Ethernet interface (RJ45 port) Minimum 6 total (Including 3 front panel) USB 2.0 ports support Windows compatible devices 15-pin D-Type WXGA compatible to support customer-supplied external monitor. DVI connector to support LeCroy Zi-EXTDISP-15 additional touch screen display accessory. Includes support for extended desktop operation with optional LeCroy or other second monitor LeCroy LBUS standard
Peripheral Bus
Auxiliary Input
Signal Types Coupling Max. Input Voltage Selected from External Trigger or External Clock input on front panel 50 : DC; 1 M: AC, DC, GND 50 : 5 Vrms; 1 M: 250 V (Peak AC < 10 kHz + DC)
Auxiliary Output
Signal Types Calibrator Signal Control Signals Select from calibrator, control signals or Off 500 Hz5 MHz square wave or DC level; 2.5 mV to 500 mV into 50 (5 mV1 V into 1 M) Trigger enabled, trigger out, pass/fail status
Automatic Setup
Auto Setup Find Vertical Scale Automatically sets timebase, trigger, and sensitivity to display a wide range of repetitive signals Automatically sets the vertical sensitivity and offset for the selected channel to display a waveform with the maximum dynamic range Ensures specified DC and timing accuracy is maintained for 1 year minimum
General
Auto Calibration
Probes
Probes Probe System Scale Factors Calibration Output Qty. (4) 10 passive probes ProBus (and ProLink on 4 and 6 GHz models). Automatically detects and supports a variety of compatible probes Automatically or manually selected depending on probe used 1 kHz square wave, 1 Vp-p (typical), output to probe hook
Power Requirements
Voltage Max. Power Consumption 100240 VAC 10% at 50/60 Hz; 100120 VAC 10% at 400 Hz; Automatic AC Voltage Selection 800 W/ 800 VA
Environmental
Temperature (Operating) Temperature (Non-Operating) Humidity (Operating) Humidity (Non-Operating) Altitude (Operating) Altitude (Non-Operating) Random Vibration (Operating) Random Vibration (Non-Operating) Functional Shock +5 C to +40 C including CD-RW/DVD-ROM drive 20 C to +60 C 5% to 80% relative humidity (non-condensing) up to +31 C Upper limit derates to 50% relative humidity (Non-condensing) at +40 C 5% to 95% relative humidity (non-condensing) as tested per MIL-PRF-28800F Up to 10,000 ft. (3,048 m) at or below +25 C Up to 40,000 ft. (12,192 m) 0.5 grms overall level, 5 Hz to 500 Hz, 10 minutes in each of three orthogonal axes, 30 minutes total 2.0 grms overall level, 5 Hz to 500 Hz, 10 minutes in each of three orthogonal axes, 30 minutes total 20 gpeak, half sine, 11 ms pulse, 3 shocks (positive and negative) in each of three orthogonal axes, 18 shocks total as tested per MIL-PRF-28800F 355 mm x 467 mm x 289 mm; 14" x 18.4" x 11.4" (height excludes feet) 18.4 kg; 40 lbs. 28.2 kg; 62 lbs.
Physical Dimensions
Dimensions (HWD) Weight Shipping Weight
Certifications
CE Compliant, UL and cUL listed. Conforms to EN 61326-1, EN 61010-1, UL 61010-1 2nd edition, and CSA C22.2 No. 61010-1-04
23
SPECIFICATIONS
Standard
Math Tools
Display up to 8 math function traces (F1F8). The easy-to-use graphical interface simplifies setup of up to two operations on each function trace, and function traces can be chained together to perform math-on-math. absolute value average (summed) average (continuous) correlation (two waveforms) derivative deskew (resample) difference () enhanced resolution (to 11 bits vertical) envelope exp (base e) exp (base 10) fft (power spectrum, magnitude, phase, up to 128 Mpts) floor integral interpolate (cubic, quadratic, sinx/x) invert (negate) log (base e) log (base 10) product (x) ratio (/) reciprocal rescale (with units) roof (sinx)/x sparse square square root sum (+) zoom (identity)
Standard
Jitter and Timing Analysis
This package provides jitter timing and analysis using time, frequency, and statistical views for common timing parameters, and also includes other useful tools. Includes: Track graphs of all parameters, no limitation of number Cycle-Cycle Jitter N-Cycle N-Cycle with start selection Frequency@level Period@level Half Period Width@level Time Interval Error@level Setup Hold Skew Duty Cycle@level Duty Cycle Error
Edge@lv parameter (counts edges) Histograms expanded with 19 histogram parameters and up to 2 billion events Trend (datalog) of up to 1 million events Track graphs of all parameters Persistence histogram, persistence trace (mean, range, sigma)
Software Options
SDA II Serial Data Analysis Software (WPZi-SDAII) (Standard on SDA 7 Zi-A and DDA 7 Zi-A)
Total Jitter A complete toolset is provided to measure total jitter. Eye Diagrams with millions of UI are quickly calculated from up to 512 Mpts records, and advanced tools may be used on the Eye Diagram to aid analysis. Complete TIE and Total Jitter (Tj) parameters and analysis functions are provided. Time Interval Error (TIE) Measurement Parameter, Histogram, Spectrum and Jitter Track Total Jitter (Tj) Measurement Parameter, Histogram, Spectrum Eye Diagram Display (sliced) Eye Diagram IsoBER (lines of constant Bit Error Rate) Eye Diagram Mask Violation Locator Eye Diagram Measurement Parameters Eye Height One Level Zero Level Eye Amplitude Eye Width Eye Crossing Avg. Power Extinction Ratio Mask hits Mask out Bit Error Rate Slice Width (setting) Q-Fit Tail Representation Bathtub Curve Cumulative Density Function (CDF) PLL Track
Measure Tools
Display any 12 parameters together with statistics, including their average, high, low, and standard deviations. Histicons provide a fast, dynamic view of parameters and wave shape characteristics. Parameter Math allows addition, subtraction, multiplication, or division of two different parameters. amplitude area base cycles data delay delay duty cycle duration falltime (9010%, 8020%, @ level) frequency first last level @ x maximum mean median minimum narrow band phase narrow band power number of points +overshoot overshoot peak-to-peak period risetime (1090%, 2080%, @ level) rms std. deviation top width median phase time @ minimum (min.) time @ maximum (max.) time @ level time @ level from trigger x@ max. x@ min.
Pass/Fail Testing
Simultaneously test multiple parameters against selectable parameter limits or pre-defined masks. Pass or fail conditions can initiate actions including document to local or networked files, e-mail the image of the failure, save waveforms, send a pulse out at the front panel auxiliary BNC output, or (with the GPIB option) send a GPIB SRQ.
24
SPECIFICATIONS
Software Options
Jitter Decomposition Models Two jitter decomposition methods are provided and simultaneously calculated to provide maximum measurement confidence. Q-Scale, CDF, Bathtub Curve, and all jitter decomposition measurement parameters can be displayed using either method. Spectral Method NQ-Scale Method Random Jitter (Rj) and Non-Data Dependent Jitter (Rj+BUj) Random Jitter (Rj) Measurement Parameter Rj+BUj Histogram Rj+BUj Spectrum Rj+BUj Track Deterministic Jitter (Dj) Deterministic Jitter (Dj) Measurement Parameter Data Dependent Jitter (DDj) Data Dependent Jitter (DDj) Measurement Parameter DDj Histogram DDj Plot (by Pattern or N-bit Sequence)
Software Options
Disk Drive Measurements Package (WPZi-DDM2) (Standard on DDA 7 Zi-A)
This package provides disk drive parameter measurements and related mathematical functions for performing disk drive WaveShape Analysis. Disk Drive Parameters are as follows: amplitude assymetry local base local baseline separation local maximum local minimum local number local peak-peak local time between events local time between peaks local time between troughs local time at minimum local time at maximum local time peak-trough local time over threshold local time trough-peak local time under threshold narrow band phase narrow band power overwrite pulse width 50 pulse width 50 pulse width 50+ resolution track average amplitude track average amplitude track average amplitude+ auto-correlation s/n non-linear transition shift
Electrical Telecom Pulse Mask Test (WPZi-ET-PMT) (Standard on SDA 7 Zi-A and DDA 7 Zi-A)
Performs automated compliance mask tests on a wide range of electrical telecom standards.
25
ORDERING INFORMATION
Product Description
WavePro 7 Zi-A Series Oscilloscopes
1.5 GHz, 10 GS/s, 4 Ch, 20 Mpts/Ch (20 GS/s and 40 Mpts/Ch in interleaved mode) with 50 and 1 M Input 2.5 GHz, 20 GS/s, 4 Ch, 20 Mpts/Ch (40 GS/s and 40 Mpts/Ch in interleaved mode) with 50 and 1 M Input 3.5 GHz, 20 GS/s, 4 Ch, 20 Mpts/Ch (40 GS/s and 40 Mpts/Ch in interleaved mode) with 50 and 1 M Input 4 GHz, 20 GS/s, 4 Ch, 20 Mpts/Ch (40 GS/s and 40 Mpts/Ch in interleaved mode) with 50 and 1 M Input 6 GHz, 20 GS/s, 4 Ch, 20 Mpts/Ch (40 GS/s and 40 Mpts/Ch in interleaved mode) with 50 and 1 M Input
Product Code
WavePro 715Zi-A
Product Description
Memory and Sample Rate Options (contd)
Product Code
WavePro 725Zi-A
WavePro 735Zi-A
WavePro 740Zi-A
128 Mpts/Ch (256 Mpts/Ch Interleaved) Memory Option WPZi-L-128 for WavePro 7 Zi-A. Includes an additional 4 GB of RAM (8 GB total) 128 Mpts/Ch (256 Mpts/Ch Interleaved) Memory Option DDAZi-L-128 for DDA 7 Zi-A. Includes an additional 4 GB of RAM (8 GB total) 128 Mpts/Ch (256 Mpts/Ch Interleaved) Memory Option SDAZi-L-128 for SDA 7 Zi-A. Includes an additional 4 GB of RAM (8 GB total) 20 GS/s (40 GS/s Interleaved) Sampling Rate WPZi-1.5GHZ-4X20GS Option for 1.5 GHz WavePro 715 Zi-A
WavePro 760Zi-A
SDA 735Zi-A
SDA 740Zi-A
DDA 760Zi-A
DDAZi-M-64 SDAZi-M-64
26
ORDERING INFORMATION
Product Description
Serial Data Options and Accessories (contd)
Product Code
Product Description
Probes and Probe Accessories
Product Code
Audiobus Trigger and Decode Option WPZi-Audiobus TD for I2S, LJ, RJ, and TDM Audiobus Trigger, Decode, and Graph Option WPZi-Audiobus TDG for I2S, LJ, RJ, and TDM I2C Bus Trigger and Decode Option WPZi-I2Cbus TD SPI Bus Trigger and Decode Option WPZi-SPIbus TD LIN Trigger and Decode Option WPZi-LINbus TD UART and RS-232 Trigger and Decode Option WPZi-UART-RS232bus TD FlexRay Trigger and Decode Option WPZi-FlexRaybus TD FlexRay Trigger, Decode, and Physical Layer WPZi-FlexRaybus TDP Test Option CANbus TD Trigger and Decode Option WPZi-CANbus TD CANbus TDM Trigger, Decode and WPZi-CANbus TDM Measure/Graph Option MIL-STD-1553 Trigger and Decode Option WPZi-1553 TD ARINC 429 Symbolic Decode Option WPZi-ARINC429bus DSymbolic
MS-250
1.5 GHz, 0.9 pF, 1 M High Impedance Active Probe ZS1500 Set of 4 ZS1500, 1.5 GHz, 0.9 pF, 1 M ZS1500-QUADPAK High Impedance Active Probe 200 MHz, 3.5 pF, 1 M Active Differential Probe ZD200 500 MHz, 1.0 pF, Active Differential Probe ZD500 1 GHz, 1.0 pF, Active Differential Probe ZD1000 1.5 GHz, 1.0 pF, Active Differential Probe ZD1500 WaveLink 6 GHz Differential Amplifier Module D600A-AT* with Adjustable Tip WaveLink 4 GHz 2.5 Vp-p Differential Amplifier D410* Small Tip Module WaveLink 4 GHz 5 Vp-p Differential Amplifier D420* Small Tip Module WaveLink 6 GHz 2.5 Vp-p Differential Amplifier D610* Small Tip Module WaveLink 6 GHz 5 Vp-p Differential Amplifier D620* Small Tip Module WaveLink 5 GHz Differential Amplifier Module D500PT* with Positioner Tip Differential Positioner Tip with Accessories Dx10-PT-kit (for use with D610 or D410) Differential Positioner Tip with Accessories Dx20-PT-kit (for use with D620 and D420) WaveLink ProLink Platform/Cable Assembly (4 6 GHz) WL-PLink WaveLink ProBus Platform/Cable Assembly (4 GHz) WL-PBus 7.5 GHz Low Capacitance Passive Probe PP066 (10, 1 k; 20, 500 ) 1 GHz, Active Differential Probe (1, 10, 20) AP034 Optical-to-Electrical Converter, 500870 nm ProLink OE525 BMA Connector Optical-to-Electrical Converter, 9501630 nm ProLink OE555 BMA Connector 10/100/1000Base-T Ethernet Test Fixture TF-ENET-B Telecom Adapter Kit 100 Bal., 120 Bal., 75 Unbal. TF-ET SATA 1.5 Gb/s, 3.0 Gb/s and 6.0 Gb/s Compliance TF-SATA-C Test Fixture SATA 1.5 Gb/s, 3.0 Gb/s and 6.0 Gb/s Compliance TF-SATA-C-KIT Test Fixture Measure Kit USB 2.0 Compliance Test Fixture TF-USB-B
* For a complete probe, order a WL-PLink or WL-PBus Platform/Cable Assembly with the Probe Tip Module.
Compatible on models with ProLink interface (4 GHz BW and higher). Includes ENET-2CAB-SMA018 and ENET-2ADA-BNCSMA.
A variety of other active voltage and current probes are also available. Consult LeCroy for more information.
Customer Service
LeCroy oscilloscopes and probes are designed, built, and tested to ensure high reliability. In the unlikely event you experience difficulties, our digital oscilloscopes are fully warranted for three years and our probes are warranted for one year. This warranty includes: No charge for return shipping Long-term 7-year support Upgrade to latest software at no charge
General Accessories
Top-mounted, Fully Integrated 15.3" WXGA with Zi-EXTDISP-15 Touch Screen Display, Including all Cabling and Software Accessory for Zi Oscilloscopes to Enable TTL-AUX-OUT TTL Level Output from the Aux Out Connector Keyboard, USB KYBD-1 Probe Deskew and Calibration Test Fixture TF-DSQ Hard Carrying Case WPZi-HARDCASE Soft Carrying Case WPZi-SOFTCASE Rackmount Accessory for Converting a Zi Series RACKMOUNT-1 Oscilloscope to an 8U Rack-mounted Package ProLink to SMA Adapter LPA-SMA-A Kit of ProLink to SMA Adapters LPA-SMA-KIT-A Oscilloscope Cart with Additional Shelf and Drawer OC1024
27
1-800-5-LeCroy www.lecroy.com
Local sales offices are located throughout the world. Visit our website to find the most convenient location.
2011 by LeCroy Corporation. All rights reserved. Specifications, prices, availability, and delivery subject to change without notice. Product or brand names are trademarks or requested trademarks of their respective holders. PCI Express is a registered trademark and/or service mark of PCI-SIG. MATLAB is a registered trademark of The MathWorks, Inc. All other product or brand names are trademarks or requested trademarks of their respective holders.
WPro7Zi-A-DS-23June11 PDF