Alpha Spectrometer System

PAS 01
Model PAS-01 Para Alpha Spectrometer System is a versatile indigenously developed single channel alpha spectrometer system for measuring low activity samples that decay by Alpha-particle emission. Model PAS-01 is an integrated spectrometer in 3 width standard NIM module and includes Surface Barrier Detector or passivated ion-implanted silicon detector (active area 450 mm2), charge sensitive preamplifier, spectroscopy amplifier, detector bias supply and 1024 channel MCA with USB interface to standard PC or laptop. Alpha spectroscopy software for Pulse Height Analysis (Energy Calibration, ROI selection, Area Calculation, peak information, FWHM, FWTM calculations, data Storage and report generation, etc.) is provided along with the system. Model PAS-01 is supplied with mains operated 6 width Standard NIM BIN with low ripple Power Supply, 100 LPM Acmevac vacuum pump, digital vacuum gauge and other accessories. The front panel control PUMP, HOLD, VENT makes it easy to insert, analyze and remove samples.

Specifications
Vacuum Chamber Construction Internal Dimensions Sample Trays Sample Spacing Control Detector Type Active Area Resolution System Background Detector Efficiency Operating Voltage : : : : : Solid Brass with Nickel plating for ease of decontamination High-performance O-ring seal 61mm wide x 74 mm deep x 40 mm high S.S. Slide-in sample trays to accommodate samples of 13 mm to 51 mm dia. From 2 mm (min) to 40 mm (max) in steps of 4mm Three-position PUMP, HOLD, VENT

: : : : : :

Silicon Surface Barrier or Ion Implanted Planar Silicon Detector 450 mm2 20 keV in Vacuum for 5.5 MeV energy less than 1 CPH for above 3 MeV energy range 25 % for a detector source spacing of < 10mm for Am241 + 30 Volts

Low Noise Charge Sensitive Preamplifier Output : +100 mV for 5.15 MeV Rise Time : < 100 nSec Fall Time : 100 uS Spectroscopy Amplifier Input Output Shaping Gain Detector Bias Supply Output Ramp-up Time Wilkinson ADC Type Input pulse No. of channels Computer Interface USB Type B Connector

: : : :

0 to +200 mV 0 to +5 Volts 1 uS, Near Gaussian shape 1 to 10 adjustable by internal trimpot

: :

0 to +200 V, software settable 50 volts per minute

: : :

10 Bit Wilkinson 0 to +5 V 1024 channels

USB cable to connect Spectrometer to PC/Laptop for Alpha Spectrum Acquisition, display and Processing

Pu239 Am
241

Sample Spectrum with Pu239, Am241 Alpha Source at 15 mTorr Due to continuous R&D, specifications are likely to change without notice

Electronic Enterprises (I) Pvt. Ltd.


Para Electronics Manufacturing Division For further details contact : 216, Regal Industrial Estate A. Donde Marg, Sewri Mumbai 400 015 Phone : 022-24137096 Fax : 022-24133341 Mumbai 022-25639904 Calcutta 033-25770551 BRANCHES Hyderabad 040-23243352 Delhi 011-27240436 Bangalore 080-23380451 Kota 0744-2501113