Team Breath Alert | 1

Design Context Review: Apnea Monitors in the Developing World


Rachel Alexander, Rachel Gilbert, Jordan Schermerhorn, Bridget Ugoh and Andrea Ulrich

Physiology and Prevalence


Apnea, a cessation of breathing usually occurring when a patient is asleep, is a common affliction that can have detrimental health effects.1 An apneic event is defined as a period of 20 seconds or greater without a respiratory cycle, though when accompanied by bradychardia (a heart rate under 80 beats per minute) or oxygen desaturation (O2 < 80-85%) an episode may be classified as apneic in as short as 10 seconds.2 Apnea can be characterized as one of three types: obstructive, central, or mixed. Central apnea involves both cessation of airflow and respiratory effort, resulting from a weak or underdeveloped central nervous system. Obstructive apnea (Fig. 1) is the cessation of breath despite respiratory effort, often due to muscle weakness in the diaphragm in infants, the trachea in adults, or the way the trachea is positioned while the patient is supine. Mixed apnea (also known as complex apnea) displays signs of both central and obstructive apnea.3 Roughly 0.4% of all cases of apnea are central, 84% are obstructive, and 15% are mixed. Apnea is estimated to affect nearly 40 million people in the United States alone. Many of those afflicted can cease breathing up to 100 times in a single hour.4 Physicians usually treat apnea by supplying the patient with air with greater-than-normal oxygen content,5 assisting the patient using physical or mechanical breathing Figure 1: Obstructive apnea may result from poor assist devices (such as a constant trachea positioning [1] positive airway pressure device), or
Stedmans Medical Dictionary 28th edition. Lippincott Williams & Wilkins, 2006. http://dictionary.webmd.com/terms/apnea. Accessed 28 Sep 2011. 2 Nimavat, Dharmendra, Michael Sherman, Rene Stantin. Apnea of Prematurity. Medscape Reference. Ed: Ted Rosenkrantz. 6 Apr 2011. http://emedicine.medscape.com/article/974971-overview#aw2aab6b2b2 3 Morgenthaler TI, Kagramanov V, Hanak V, Decker PA (September 2006). "Complex sleep apnea syndrome: is it a unique clinical syndrome?". Sleep 29 (9): 12039. PMID 17040008. Lay summary Science Daily (September 4, 2006). 4 Sleep Apnea. Apneos. Apneos Corporation. 06 Sep 2003. http://www.apneos.com/sleepapnea.html. Accessed 28 Sep 2011 5 Apnea of Prematurity. A.D.A.M. Medical Encylopedia. Rev: 2 Nov 2009. Rev: Neil K Keneshiro, David Zieve. Accessed 28 Sep 2011. http://www.ncbi.nlm.nih.gov/pubmedhealth/PMH0004488/
1

Team Breath Alert | 2 administrating caffeine citrate a treatment method especially effective when used in children.6

Risk Factors and Geographic Distribution


While sleep apnea in adults is generally a mild and non-life-threatening condition, apnea in children can have much more severe consequences. It is estimated that sleep apnea occurs in 2% of all children7, 7% of all infants (apnea of infancy, AOI)8, and in 50% of all premature infants born prior to 36 weeks gestation (apnea of prematurity, AOP).9 Children who are obese and those with enlarged tonsils are also at higher risk for sleep apnea.10 Conditions such as anemia, malnutrition, heart or lung problems, infection, low oxygen levels, overstimulation, and temperature problems can all trigger or exacerbate apnea in children.11 If left untreated, apnea can lead to failure to thrive, diminished growth, hypertension, cor pulmonale (failure of the right side of the heart), developmental problems such as loss of IQ, mental retardation, hyperactive behavior, acid reflux, development of a pectus excavatum deformation (sunken chest), or, in severe cases, death.12,13,14 Among children, apnea is most prevalent in premature infants. Nearly half of all premature babies suffer from AOP, while nearly 100% of premature babies born <28 weeks or at a birth weight <1000 g experience regular apneic episodes.15 Typically, apnea in neonates results from their underdeveloped central nervous systems and weak trachea muscles both of which have not yet had the time to fully develop. At 45 weeks postconceptional age, symptoms tend to disappear as these organs mature.16 Some research indicates that AoP serves as a risk factor for sudden infant death syndrome (SIDS); Moon et. al. found that infants with AoP were four times more likely
6

Finer, Neil N.,Rosemary Higgins, John Kattwinkel, Richard Martin. Summary Proceedings from the Apnea-ofPrematurity Group. Pediatrics. Vol. 117 No. Supplement 1; 1 March 2006. pp. S47-S51. http://pediatrics.aappublications.org/content/117/Supplement_1/S47.long 7 Sleep Apnea. Apneos. 8 Rocker, Joshua, Jeffrey Israel. Pediatric Apnea. Medscape Reference. Ed: Richard G. Bachur. 25 Aug 2010. http://emedicine.medscape.com/article/800032-overview#a0199 9 Finer. 10 Sleep Apnea. Apneos. 11 Apnea of Prematurity. A.D.A.M. Medical Encylopedia. Rev: 2 Nov 2009. Rev: Neil K Keneshiro, David Zieve. Accessed 28 Sep 2011. http://www.ncbi.nlm.nih.gov/pubmedhealth/PMH0004488/ 12 Sleep Apnea. Apneos. 13 Joshua Rocker, et. al. 14 Cataletto, Mary E., Andrew J Lipton, Timothy D Murphy. Childhood Sleep Apnea. Medscape Reference. Ed: Michael R Bye. 29 Mar 2011. http://emedicine.medscape.com/article/1004104-overview#a0156 15 Finer. 16 Tauman, Riva, and Yakov Sivan, 2000. Duration of Home Monitoring for Infants Discharged with Apnea of Prematurity. Biology of the Neonate, Vol. 78, No. 3. pp 168-173. http://content.karger.com/ProdukteDB/produkte.asp?Aktion=ShowAbstract&ProduktNr=224215&Ausgabe=22 5302&ArtikelNr=14266

Team Breath Alert | 3 to die of SIDS than those without.17 Therefore, it is vital that infants at risk for AoP or AoI are carefully monitored to prevent death and other developmental complications. A recent survey18 identified the global distribution of pre-term births. The highest incidence occurred in Southern Africa, where 17.5% of all births are considered pre-term. In fact, nearly 85% of all pre-term births in 2005 took place in Africa and Asia (Fig. 2). In many of the worlds least developed areas such as sub-Saharan Africa, where premature birth rates are among the highest in the world resources for monitoring apnea in children are extremely limited, and often visual observation serves as the only indication of whether or not a child is breathing19. Unmonitored infants in these locations are at higher risk of dying or developing behavioral, physical, and mental disorders when compared to their American and European peers.

Figure 2: Distribution of preterm births around the world [2]

Measuring Breathing
Major methods of detecting apneic episodes in developed countries tend to rely on both physical and chemical indicators; these include measurement of airflow, motion, and blood oxygenation. Although none of these systems perfectly detect apneic episodes, they each possess different advantages and unique design challenges. In analyzing how these sensors could potentially be used in our device, we will examine how sensors receive measurements as well as the advantages and disadvantages of the systems, including sensor cost, complexity, and accuracy. Airflow Sensors Airflow sensors analyze patterns of breathing by measuring one or more of three parameters: temperature, pressure, and acoustics. Although some of the systems are relatively cheap, a facemask, mouth piece, or nosepiece is required to collect the
17

Cataletto, Mary E., Andrew J Lipton, Timothy D Murphy. Childhood Sleep Apnea. Medscape Reference. Ed: Michael R Bye. 29 Mar 2011. 18 Beck, Stacy, Daniel Wojdyla, Lale Say, Ana Pilar Betran, Maria Merialdi, Jennifer Harris Requejo, Craig Rubens, Ramkumar Menon, and Paul FA Van Look. 2009. The worldwide incidence of preterm birth: a systematic review of maternal mortality and morbidity. Bulletin of the World Health 19 Moons, Peter. Re: Team Breath Alert Introduction. E-mail to Team Breath Alert. 4 Oct. 2011

Team Breath Alert | 4 patients expired air for analysis, and are considered more invasive than sensors placed elsewhere on the body22. One type of airflow sensor uses a thermistor to measure the temperature difference between ambient inspiratory air and lung-temperature expiratory air. Respiratory rate is calculated by tracking changes in temperature over time: as the temperature calculated approaches the ambient temperature, it is assumed that the patient has gone an extended period of time without breathing. The thermistor is shaped to form a bridge placed between the nostrils.20 This system, while low cost and easy to use, must be calibrated for various lung sizes and may experience some delay before detecting an apneic episode.20 Similar in principle to thermal sensors, pressure airflow sensors detect fluctuations in airway pressure caused by respiration. Airflow pressure is measured quantitatively with a pneumotachograph, a device that detects the pressure differential between inspiration and expiration.21 Pressure airflow sensors can be used in the form of a facemask or nasal cannula sensor. These are simple to use and require minimal calibration, but may be prohibitively expensive for use in the developing world, starting at $60 for the sensor alone. Existing pneumotachographs are also bulkier and perhaps better suited for use in adults rather than infants. Acoustic rhinometry measures air pressure as well as the airflow rate in the nasal airway during respiration by measuring reflected sound waves directed towards the patients pharynx.22 Many acoustic rhinometers require the user to hold the nosepiece near the nostril while breathing into it.23 The accuracy of acoustic rhinometry can be affected by sound interference from the patients heart beat. Furthermore, rhinometers do not pick up accurate measurements from infants, largely because readings are only accurate when patients breathe through their nose.24 Motion sensors Another means of detecting apneic episodes relies on tracking chest motion. Monitors relying on this technique tend to record motion artifacts while premature infants tend to be relatively still, older patients may move or roll in ways that expand their chest cavities when not breathing but the relatively sinusoidal nature of
20

Jovanov, Emil, and Dejan Raskovic. "Thermistor-based Breathing Sensor for Circadian Rhythm Evaluation." Web. <citeseerx.ist.psu.edu/viewdoc/summary?doi=10.1.1.127.5161>. 25 Sept. 2011. 21 Lee-Chiong, T. "Monitoring Respiration during Sleep." Clinics in Chest Medicine 24.2 (2003): 297-306. Print. 22 "Rhinometry and Rhinomanometry." Aetna - Health Insurance, Dental, Pharmacy, Group Life and Disability Insurance. Web. 07 Oct. 2011. <http://www.aetna.com/cpb/medical/data/700_799/0700.html>. 23 "Acoustic Rhinometer A1." GMI Home Page. Web. 07 Oct. 2011. <http://www.gminstruments.com/A11.htm>. 24 Folke, M., L. Cernerud, M. Ekstrm, and B. Hk. "Critical Review of Non-invasive Respiratory Monitoring in Medical Care." Medical & Biological Engineering & Computing 41.4 (2003): 377-83. Print.

Team Breath Alert | 5 breathing makes it easy to filter out this noise. The three main methods of monitoring apnea via chest motion include transthoracic impedance pneumography, stretch sensors, and motion sensor pads. Transthoracic impedance pneumography records inhalations and exhalations by measuring the change in resistance to flow of electrical current across the patients chest. Chest impedance changes as the pleural cavity expands and contracts with air flow.25 This impedance shift alters current sent through electrodes placed on the chest for straightforward measurements.26 Transthoracic impedance pneumography serves as the gold standard for hospitals in the developed world because multiple physiological monitors (including respiratory rate, heart rate and blood-oxygen level) can be obtained using one data collection system. Impedance monitors are extraordinarily accurate, but are also expensive and require a constant A/C power supply. Additional complications disruption of accurate readings by aortic blood flow (relevant for infants, who often have respiratory rates nearer in frequency to heart rates), and regular application of gel to ensure low skin-electrode impedance is necessary throughout monitoring.27 The second technique uses stretch sensors that change resistance with movement of the chest (Fig. 3).28 A wide variety of devices have been developed utilizing changes in external sensor resistance for example, Guardian Technologies has created a vest using this technique.29 Although most of these devices are inexpensive, there are several clinical problems that are present for example, a vest structure is not ideal if a nurse needs to examine the childs chest, and a vest could serve as an obstacle slowing care in an emergency. However, these sensors could be incorporated in a variety of designs, including less obstructive (if Figure 3: Apnea Belt Monitor [3]
25

Gupta, Amit K. "Respiration Rate Measurement Based on Impedance Pneumography." Texas Instruments, Feb. 2011. Web. <http://www.ti.com/lit/an/sbaa181/sbaa181.pdf>. 26 Sontheimer, D., C. B. Fischer, F. Scheffer, D. Kaempf, and O. Linderkamp. "Pitfalls in Respiratory Monitoring of Premature Infants during Kangaroo Care." Archives of Disease in Childhood - Fetal and Neonatal Edition 72.2 (1995): F115-117. Print. 27 Folke et al. 28 "Flexible Stretch Sensor." Images Scientific Instruments - Science Projects, Electronic Kits, Robotic Kits and Accessories, Microcontroller Compilers and Programmers, Parts. Web. 07 Oct. 2011. <http://www.imagesco.com/sensors/stretch-sensor.html>. 29 Guardian Technologies, 2009. Web. <http://biibs.sdsu.edu/vest.html>.

Team Breath Alert | 6 less physically stable) belts. Lastly, pressure sensors may be placed in the childs mattress to monitor chest movement. These mattress sensors are common for at-home apnea monitoring in developed countries. These mats can be difficult to position to achieve accurate measurements,30 and they are easily dislodged by infant motion, and are plagued by the highest rates of false alarms.31 On the other hand, these sensors are extremely noninvasive and simple to implement.32 Blood Gas Measurement Blood gas monitors measure respiration indirectly by examining blood oxygenation levels. Two types of blood oxygenation measurement techniques are pulse oximetry and O2/CO2 level sensing. The pulse oximeter monitors blood oxygen levels by utilizing infrared light generated by the device that passes through a finger, toe, wrist, or earlobe of the user.33 During an apneic episode, the pulse oximeter detects a drop in blood-oxygen saturation levels due to a lack of oxygen entering the blood stream via the lungs. Normal levels of arterial oxygen saturation range from 95% to 100% for infants and for adults; 34 for preterm babies, the arterial oxygen saturation levels range from 84% to 90%. 35 Saturation levels below these ranges may indicate that the patient has entered an apneic state and requires oxygen. Pulse oximeters are easy to use but are unstable and difficult to position correctly.36 An O2/CO2 sensor uses airflow to measure the blood gas concentration of expelled air from the patient and correlates it to arterial blood concentration. 37 An increase in CO2 or a decrease in O2 may indicate that the patient has entered an apneic state. One problem with this type of measurement lies in the fact that the concentration of CO2 and O2 do not correlate exactly to the arterial blood concentration. The monitor

30

Angelcare Baby Monitor | Babyphone and Baby Movement Sensor. Web. 07 Oct. 2011. <http://www.angelcare-monitor.com>. 31 Folke et al. 32 Folke et al. 33 "Pulse Oximetry and Sleep Apnea." Sleep Apnea Life - Living with Sleep Apnea. Web. 07 Oct. 2011. <http://sleepapnealife.com/pulse-oximetry-and-sleep-apnea-311.html>. 34 "Normal Oxygen Saturation For Infants | LIVESTRONG.COM." LIVESTRONG.COM - Lose Weight & Get Fit with Diet, Nutrition & Fitness Tools | LIVESTRONG.COM. Web. <http://www.livestrong.com/article/139666-normaloxygen-saturation-infants/>. 27 Sept. 2011. 35 "Normal Hemodynamic Parameters and Laboratory Values." Edwards LifeSciences LLC, 2011. Web.<http://ht.edwards.com/sci/edwards/sitecollectionimages/edwards/products/presep/ar04313hemodynp ocketcard.pdf>. 25 Sept. 2011. 36 Folke et al. 37 Folke et al.

Team Breath Alert | 7 assumes an ideal correlation and thus measuring CO2 and O2 changes can be inaccurate.38
Table 1: Summary of the advantages and disadvantages of methods of detecting respiratory rate

Method
Temperature Pressure Acoustics

Advantages
Airflow Sensors
Low cost, easy to use Easy to use Low cost

Disadvantages
Intrusive Expensive Potential noise interference, limited monitoring capabilities Expensive, high power requirement Noise interference, may require significant chest movement Expensive, motion artifacts, positioning, high power requirement Signal delay, may fall off, positioning Ease of use, inaccurate, intrusive

Motion Sensors
Transthoracic Impedance Pneumography Stretch Sensors Widely used in developed world, accurate Low cost, easy to use

Pressure sensor mats

Noninvasive

Blood Gas Measurement


Pulse oximeter CO2/O2 Sensor Low cost, easy to use Low cost

With a wide variety of airflow, motion and blood oxygenation monitoring techniques, no clear winner emerges as the best all-around technology the manner in which each technology is implement seems the key to success. Several technical problems, from motion artifacts to cost, may prove challenging to solve and might require intense processing. In order to narrow our options, we will now consider additional barriers imposed when using these devices in countries with the highest rates of premature infants.

38

Folke et al.

Team Breath Alert | 8

Low-resource Settings
To best assist patients in low-resource settings, it is important to understand additional problems that may be faced exclusively by those health care professionals. There are two primary economic categories used to describe lower-income countries: developing nations, as well as least-developed countries. A developing nation encompasses a wide range of countries including a per capita income range from $80 to $8,380 (Mozambique and Argentina, respectively).39 Least-developed countries (LDCs) refer to the most economically weak members of the developing countries, and include 45 countries such as Malawi, Bhutan, and Angola.40 The United Nations selects the LDCs by certain criteria, including a gross national income (GNI) per capita of less than $900, as well as meeting a certain threshold on human and economic vulnerability.41 Manufacturing Capabilities Lack of infrastructure and shortages of technical personnel often leave LDCs unable to manufacture and transport complex goods, including medical devices such as the apnea monitoring systems previously mentioned. When importing these goods from other nations, trade regulations may require price controls and tax regulations, both of which are frequently compromised due to political instability. Furthermore, the ability to adopt and operate high-end technologies is compromised by the lack of technical expertise in developing countries. In LDCs especially, consistently low rates of secondary education impede rapid technology adoption.42 Technology Awareness There are several technology gaps impeding the implementation of health devices in the developing world. Power is unreliable at best, and often no stable source of power can be found.43 In Malawi, for example frequently cited as one of the worst case scenarios for healthcare access, and largely representative of the areas in which we hope to implement our device it is not uncommon for hospitals to face power blackouts every day, and hospitals with better infrastructure may still experience
39

Trybout, James. Manufacturing firms in developing countries: how well do they do, and why? Journal of Economic Literature. 38 (1): 11-44, 2000. 40 Department of Economic and Social Affairs Statistics Division Office of the High Representative for Least Developed Countries, Landlocked Developing Countries and Small Island Developing States. World statistics pocketbook 2010: least developed countries. LDC. 35: 1-75, 2011. 41 The Criteria for the Identification of LDCs. United Nations. 2005. Accessed 6 Oct. 2011. <http://www.un.org/special-rep/ohrlls/ldc/ldc%20criteria.htm>. 42 Trybout. 43 Martnez, Andres, Valentn Villarroel, Joaqun Seoane, and Francisco del Pozo. Rural telemedicine for primary healthcare in developing countries. IEEE Technology and Society Magazine. 13-24, 2004.

Team Breath Alert | 9 outages at least once a week.44 However, cell phone ownership and coverage are increasing rapidly in the developing world, with 85% 2G coverage reaching even the most desolate areas.45 The market for health-related mobile apps is predicted to expand with a compound annual growth rate of 24% between 2010 and 2014.45 Environment Technologically complex apnea monitors may not operate accurately in the extreme heat and humidity commonly present in open-air hospitals in LDCs. High humidity may cause devices to rust and affect sensor readings. In Malawi, again representing an extreme scenario in which our device might be implemented, humidity can reach 80% in the rainy season from October to May.46 Changes in temperature affect the operation of electrical components such as thermistors and resistors and may make the medical device less accurate. Temperatures in Malawi swing between extremes: temperatures can drop to just above freezing on winter evenings and rise to as high as 42 C during the summer.46

Health Care in Low-resource Settings


Health systems in developing countries are struggling to meet the needs of their sick. Worker shortages and budget deficits negatively impact essential measures of population health such as infant mortality rate. A meta-analysis of how various risk factors affect global disease burden and Disability-Adjusted-Life-Years (DALYs) showed that the poorest nations suffer more loss of life than any other region of the world. It becomes fairly clear that the areas of the world struggling with the most sickness are also the poorest regions of the world (Fig. 4).47 With child mortality specifically, neonatal deaths serves as a key indicator: neonatal deaths account for about one third of the global child mortality rate. For every 1000 births, 40-50 babies born in
Figure 4. A measure of Disability-Adjusted Life Years compared to diseases burdens worldwide
44 45

Sandy Chiume. Personal interview. 6 Oct. 2011. "Mobile Enterprise Applications Market 2010-2014." Technavio (2011). 46 Human Resources for Health Country Profile - Malawi. Africa Health Workforce Observatory. Oct. 2009.) 47 Ezzati, Majid, Alan Lopez, Anthony Rogers, et al. Selected major risk factors and global and regional burden of disease. The Lancet. 360 (9343): 1347 1360, 2002.

Team Breath Alert | 10 all LDCs combined die in the first 28 days of life.48 Governments in these countries are ill-equipped to address these health issues, often facing issues with political instability capable of crippling healthcare systems.49 Worker shortages also pose serious challenge (Table 2). This pattern is notably worse in sub-Saharan Africa, where 3% of the planets healthcare workers treat 25% of global disease burden.50

Table 2: Medical Staff per 100,000 People in Six Sub-Saharan Countries, 200451

Nurses are the primary party responsible for monitoring patients health. A study in the U.S. indicated that lower nurse-to-patient ratio led to worse patient outcome.52 Furthermore, patient crowding leads to significant restraints on hospitals. Struggling with minimal equipment, nurses often assign multiple patients to a bed, and it is not uncommon for multiple babies to share one ventilator.53 Conditions Faced by Nurses and Doctors In hospitals in the developing world, nurses face extreme stresses due to crowding, poor communication, hostile work environments, and lack of training (Fig. 5). Communication between doctors and nurses is often strained, as determined by an American study examining nurses and doctors in 36 emergency rooms. When questioned on the effectiveness of doctor-nurse communication, both doctors and nurses stated that tasks were often done by both parties because of lack of communication. It was found that doctors blame nurses for mistakes more than they blame other doctors.54 Breakdowns in communication serve as major sources of stress in the healthcare system.

48

Zaidi, Anita, W. Huskins, D. Thaver, et al. Hospital-acquired neonatal infections in developing countries. The Lancet. 365 (9465): 1175 1188, 2005. 49 Human Resources for Health Country Profile - Malawi. Africa Health Workforce Observatory. Oct. 2009. 50 Do most countries have enough health workers? World Health Organization, 26 Feb. 2008. Web. 25 Sept. 2011. <http://www.who.int/features/qa/37/en/index.html>. 51 Palmer. 52 Carayon, Pascale and Ayse Gurses. Nursing Workload and Patient Safety A Human Factors Engineering Perspective. Patient Safety and Quality: An Evidence-Based Handbook for Nurses, edited by R.G. Hughes. Rockville: Agency for Healthcare Research and Quality, 2008, pp. 2-203 2-216. 53 Bateman, Chris. Crowded wards, lousy admin contribute to death and suffering. SAMJ 100(7) (2010): 414418. Web. 25 Sept. 2011 54 Greenfield, Lazar J. Doctors and Nurses: A Troubled Partnership. Annals of Surgery 230(3): 279. 1999. Web. 25 Sept. 2011.

Team Breath Alert | 11 When a general health questionnaire was administered, 40% of general ward nurses and 32% of neonatal nurses in Sydney had scores indicating possible psychological impairment, indicating conditions such as depression, anxiety, and fatigue derived from Figure 5: Kamuzu Central Hospital Children Wards A (Malawi) 55 [5] work-related stress. During high levels of stress, health officials often experience memory impairment, decreasing the ability of nurses to attend to their patients.56 One study examining the relationship between ward noise and heart rate found that any noises above the average daytime noise level of 61 dB resulted in significant increases, indicating higher stress.57 We anticipate that all of these stresses would be exacerbated in the developing world. The lack of training and support nurses receive in developing countries may also serve as a stressor. Support resources are offered by the West and East African Health Organizations, but access to these resources is limited.58 Many nurses receive minimal training, often on the job, and afterwards suffer from a lack of supervision and support from the hospital itself.59

Design Constraints for Apnea Monitors in LDCs


Frequently dire conditions in LDCs present us with several key constraints that define how we as a team opt to address the problem of apnea detection: We cannot assume that hospitals in the developing world will be able to utilize the constant source of A/C power, frequent automated calibration, or a significant amount of monitoring available in the US.

55

Oates, P.R., Oates, R.K. Stress and work relationships in the neonatal intensive care unit: are they worse than in the wards? Journal of Paediatrics and Child Health Vol 32 Issue 1 (1996): 57-59. Web. 20 Sept. 2011. 56 Newell, Robert. Anxiety, accuracy and reflection: the limits of professional development Journal of Advanced Nursing 17 (1992): 1326-1333. Web. 25 Sept. 2011 57 Morrison, W.E., Haas, E.C., Shaffner, D.H. et al. Noise, Stress and Annoyance in a Pediatric Intensive Care Unit. Critical Care Medicine 31(1) (2003): 113-119. Web. 25 Sept. 2011. 58 Koop, C. Everett, Pearson, Clarence E., Schwarz, M. Roy. Critical Issues in Global Health. San Francisco: JosseyBass, 2001. Print. 59 Palmer.

Team Breath Alert | 12 Worker shortages create chaotic, hectic pediatric wards with low nurseto-patient ratios; a monitor that does not require constant human monitoring is ideal. Given noise and stress issues in neonatal wards, an ideal monitor would be extremely easy to use and would require minimal training. Automatic logging of episodes and construction of patient medical records would further alleviate stressors. Our monitor will need to be self-contained and resistant to environmental extremes. Due to the lack of manufacturing capabilities, the device should be constructed in the United States and then shipped to the LDC of interest, complete with replacement parts. Lack of technical expertise means the device should be as rugged and self-sufficient as possible. Given the widespread availability of cellular networks, a monitor that can interface with wireless networks may significantly aid the health workers. High sensitivity and specificity remain our most important objectives, and should not be sacrificed as a function of environmental limitations.

Most importantly, that the device accurately identifies when the infant is not breathing for a designated period of time. Sensitivity and specificity must be seriously considered in designing an appropriate apnea monitor, and remain our most important objectives.

Problem Statement
In developing nations, 10.9 million premature babies are born each year. Roughly half of them suffer from apnea of prematurity (AOP), which could lead to serious complications and even death if left untreated. Current apnea monitors that are widely used in the developed world are not suitable for application in low-resource settings. Many of these monitors are not very sensitive and require an external power source or frequent calibration. Furthermore, current devices are designed with the assumption that only one baby is being monitored, while crowded neonatal wards often have low nurse-to-patient ratio. Properly monitoring apnea of prematurity requires a robust, inexpensive, user-friendly device that can be integrated into crowded neonatal wards in developing countries.

Team Breath Alert | 13 Work Cited for Figures and Tables


[1] http://www.91outcomes.com/2010/09/blogging-about-gulf-war-illnesses-sleep.html [2] http://www.physorg.com/news173880803.html [3] http://www.nationwidechildrens.org/apnea-prematurity [4] Figure 1b. Ezzati, Majid, Alan Lopez, Anthony Rogers, et al. Selected major risk factors and global and regional burden of disease. The Lancet. 360 (9343): 1347 1360, 2002. [5] http://webscript.princeton.edu/~sgac/malawi/stories.php