Академический Документы
Профессиональный Документы
Культура Документы
Laparoscopic Injuries
Fire Injuries
Properties of Electricity
Current Flow of electrons during a period of time, measured in amperes Pathway for the uninterrupted flow of electrons (must be complete/ closed to flow)
Circuit
Impedance Obstacle to the flow of current measured in ohms Voltage Force pushing current through the resistance, measured in volts
Frequency Spectrum
Current Concentration/Dispersal
The more concentrated the current, the greater the potential for a burn.
Monopolar
Active electrode at surgical site Return electrode at another site Current flows through the body between the electrodes High voltage Coag 3000 9000 Cut 1350 - 4000
Bipolar
Active and return electrodes in the
instrument
Current flow confined to tissue between electrodes Low Voltage (320 - 1200 volts)
Electrosurgical Technologies
Patient Burns
Interrogation current
20
Do not attempt to use return electrodes that disable the contact quality monitoring system Electrode is intact Conductive gel, if present, is moist Manufacturers expiration date has not been reach
Clean, dry skin Over a large, well-perfused muscle mass of the surgical side Close to the surgical site
Convex area
Prostheses
Bony prominences
Scar tissue
Hair
Follow manufacturers instructions for hair removal, cleaning, and drying site
Do not exceed 300 milliamps of power Do not activate active electrode longer than 30 seconds For neonates < .45 kg, use of bipolar electrosurgery
Increased temperature at the pad site Increased impedance at pad site Current concentration with resultant heat build up
Do not apply the patient return electrode to areas where pressure is applied to the patient
ECRI
Always follow manufacturers instructions when using a warming device Avoid close contact with warming blanket A towel, blanket, or folded sheet may be placed between the patient return electrode and the warming blanket to distance the heating source from the patient return electrode
Remove return electrode slowly, supporting the tissue underneath Observe skin under pad Chart appearance of site
Active Electrodes
When not in use, always place active electrodes in a nonconductive holster Electrodes that do not fit in the holster should be placed in a designated location with tips away from flammable material Active electrode tips should be securely seated into the hand piece
Active Electrodes
Do not use suction coagulator as a retractor Avoid tissue contact with shaft during activation Ensure that the outside of the suction coagulator remains free of fluids and mucus which may conduct electrical current Unintended activation may result if the suction coagulator handswitch is immersed in saline solution or other conductive fluids
Active Electrodes
Build-up of eschar increases resistance at the active electrode and may require the use of higher power setting Eschar debris may ignite Use a coated electrode to easily remove eschar buildup on electrode tip
Active Electrodes
Do not use red rubber catheters or other materials as a sheath
Red rubber or plastic materials may ignite Use manufacturer approved insulated tips
ERCI, 2006
Active electrode cords should not be wrapped around metal instruments Active electrode and other electrical device cords should not be bundled together
Ulmer, 2008
Pacemakers
Consult pacemaker manufacturer Use electrosurgery with care Use bipolar Use low power setting Avoid current flow through heart and pacemaker Keep cords away from pacemaker and leads Postoperatively have the device checked by a trained individual
Internal Cardioverter-Defibrillators
Obtain preoperative cardiology consult to evaluate correct functioning and determine risks Have defibrillator immediately available Deactivation of ICD by trained personnel before ESU use Continuous ECG and peripheral pulse monitoring Use bipolar If monopolar used ensure distance between the active and patient return electrode is as short as possible Avoid current flow through heart and ICD Postoperatively have the device checked by a trained individual
Hearing Aids
Best practice is to remove hearing aids prior to surgery RF leakage may cause interference to the hearing aid reception RF current may damage two piece wireless hearing aids, thus they should not be worn when electrosurgery is used
Cochlear Implant
Do not use monopolar electrosurgery on the head or neck of a cochlear implant patient Use bipolar electrosurgery on the head and neck of patients with implants that do not have extracochlear reference electrodes
The bipolar electrode must not contact the implant
If the implant has extracochlear reference electrodes, bipolar may be used if the electrode is kept more than 10 cm (4 inches) from the extracochlear electrodes Postoperatively have the device checked by a trained individual
Dental Appliances
Avoid direct contact with any metal dental appliances Cover braces with teeth guards (silicone or rubber) or use bipolar energy when possible Remove palate spreaders prior to surgery if electrosurgery will be used in immediate area of the device
Jewelry
Jewelry should be removed if it is within the activation range of the active electrode. When using a reusable, capacitivecoupled return electrode, all of the patients metal jewelry should be removed.
Tattoos
Avoid placing the patient return electrode over tattoos Inks (red in particular) contain metals which could serve as a heat or electrical conductor
No evidence to contraindicate the use of electrosurgery during pregnancy
Amniotic fluid protects fetus from any concentration of electrical current owing to the dispersion effect Just as high frequency of electrosurgery current above 100 kHz prevents neuromuscular stimulation in adults, children, and infants, the same is true for the fetus
If flammable prepping solution used follow manufacturers instructions Apply the patient return electrode to a site that will not be exposed to pooling of amniotic fluid Activate electrosurgery pencil only when in contact with intended tissue Place electrosurgery pencil in nonconductive safety holster when not in use
Pediatric applications and/or procedures performed on small anatomic structures may require reduced power settings Potential for unintended thermal damage to tissue with high current flow and longer activation For procedures where the current could flow though delicate parts of the body, use of bipolar technique may be desirable in order to avoid unwanted coagulation
Chemical Injury
Pressure Injury
Electrosurgery Injury
Questions ???
Thank you !
Program Summary
Potential adverse outcomes related to electrosurgery Basic principles of electrosurgery Best practices to achieve favorable patient outcomes related to electrosurgery
Prior to use, clinicians should review the Instructions for Use product inserts and/or Users Guide provided by the manufacturer to ensure safe and appropriate utilization of electrosurgical devices.