Outline
Why water vapor? Why absorption in infrared region? Why cavity enhanced technologies ? Why ICOS ?
Beers Law
, =
Where
= =
= ()
= 1
Decteto r
Type CRDS (Cavity ring down spectroscopy) ICOS (Integrated cavity output spectroscopy)
In CRDS, a beam of light is injected into the cavity and the rate of decay of light which is transmitted is measured
Experiment on CRDS
On Axis TEM(3,0)
On Axis TEM(1,1)
On Axis TEM(1,0)
Cylindrical mode
Where L is the length of the cavity and 1 and 2 are the radii of curvature of the spherical mirrors. Output response [*] Where = 1 () =
(1)
= (1 + ) is the Single pass absorption loss and is the cavity gain factor.
= ()
[*].Barbara A. Paldus and Alexander A. Kachanov, An historical overview of cavity enhanced methods , Can. J. Physics 83: 975-999
A. Combi Controller (ITC 502) B. LED Mount (TCLDM 9) C. Iris Diaphragms D. Gas Chamber E. Cavity Mirror F. Converging Lens G. Optical Chopper H. IR Sensor (PDA 155) I. Lock in Amplifier (SR 510) J. Chopper Controller (SR 540) K. (Highly Pure N2 tanks L. Throttling valve M. Humidity Sensor (HoneyWell HIH-4000-001) N. Collimating system
Experimental Set Up
0.006
35
0.005 30 Probability 10115 10615 11115 11615 12115 25 20 15 0.002 10 5 0 9615 0.001 0.004
Cross-section (m2)/molecule
0.003
0 9615
10115
Wavenumber cm-1
11615
12115
Water vapor cross section is taken from the HITRAN database and LED Input intensity distribution over frequency is considered to be Gaussian. The expected output is at 25% RH at 1 atm pressure and 100 m path length
: 195 V
: 0.5 V
From ICOS calculation, absorption of 1.5 V of intensity with total output of 195 V signal is equivalent to Relative
Humidity of 0.66%.
Hence, Broadband LED source could be used to study water vapor concentration with sufficient accuracy.
Nitrogen!
Signal level decreases by different amount with different
Problem Solved ?
NO! Presence of Hydrocarbons in the commercially available Nitrogen?
Used Chromatographic Charcoal filter and water vapor filter!
Measured
Source
Manufacturing company also mentioned that some of the coating can be effected by Humid environment depending upon the process of coatings!
http://www.precisionphotonics.com/faq.asp
Conclusions
We can form sensitive Cavity by using 940 nm LED. Mirrors ( With dielectric coating by e-beam coating ) are very sensitive to moisture and they can be used to detect presence of water vapor (further study required). For the water vapor absorption cavity enhance spectroscopy , use of mirrors with e-beam coating are not good.
What Next ?
Calculation of output pattern by Using exact output pattern of LED instead of assuming Gaussian like output. Single mode 940 nm Infrared Laser Diode are available in Thor Labs ! We can test our system with these to further conform that the mirrors are moisture sensitive. Replace the mirror I have used by the mirrors coated by Ion Beam Sputtering (IBS). If collimation of the LED beam can be Improved, narrow beam of the Infrared light can be used so that we can lower the no of cavity mode and can improve the sensitivity of the system.
Acknowledgement
Advisor Prof Christopher Winstead Colleagues
Tyler Reese
Melanie Brady Rajendra Timilsina Ek Raj Adhikari