Both instruments in Oklahoma test Breath using INFRARED SPECTROSCOPY

INFRARED SPECTROSCOPY

The ethanol molecule has two ends, the hydrogen end and the hydroxide (carbonoxygen bond) end.

NOW HOLD THAT THOUGHT . . . .

In the visual spectrum:


We can tell how much fog or smoke is in the air by how much light is blocked from a light source.

BEER-LAMBERT LAW
(Provides the theoretical basis for IR breath testing.)

Each molecule absorbs infrared light waves in a unique way. Thus, each molecule has its own infrared fingerprint.
Ethanol Infrared Spectra with notation for I8000 IR filters
100% 90%

Percent Absorbtion

80% 70% 60% 50% 40% 30% 20% 10% 0% 3 4 5 6 7 8 9 10 11


3.4 micron filter 9.4 micron filter

Wavelength(microns)

Ethanol molecules absorb infrared light at wavelengths of approximately 3.00, 3.39, 7.25, 9.18, 9.50 and 11.5 microns.

No other compound absorbs infrared light at all of these wavelengths exclusively.

In the Infrared spectrum, the hydrogen end and the hydroxide end act like the fog to block the the light BUT ONLY in the 3.4 uM and 9.4 uM wavelengths, exclusively.
3.4hhhhhhh 9.4hhhhhhhhhhh

EVERYTHING ELSE PASSES THROUGH


3.4hhhhhh hhhhhhhhhhhhhhhhhhhhhhhhh hhhhhhhhhhhhhhhhhhhhhhhhhh 9.4hhhhhhhhhhh hhhhhhhhhhhhhhhhhhhhhhhhhh

THE INSTRUMENT MEASURES THE INTENSITY OF THE INFRARED LIGHT WITH CLEAN ROOM AIR.

hhhhhhhhhhhhhhhhhhhhhhhhhhhhhhhhhh hhhhhhhhhhhhhhhhhhhhhhhhhhhhhhhhh hhhhhhhhhhhhhhhhhhhhhhhhhhhhhhhhhh hhhhhhhhhhhhhhhhhhhhhhhhhhhhhhhhhh


hhhhhhhhhhhhhhhhhhhhhhhhhhhhhhhhhhhhhh hhhhhhhhhhhhhhhhhhhhhhhhhhhhhhhhhhhhhh hhhhhhhhhhhhhhhhhhhhhhhhhhhhhhhhhhhhh hhhhhhhhhhhhhhhhhhhhhhhhhhhhhhhhhhhhhh

The Instrument then compares the breath sample to determine how much of the 3.4 & 9.4uM are missing (blocked by ethanol)

hhhhhhhhhhhhhhhhhhhhhhhhhh hhhhhhhhhhhhhhhhhhhhhhhhhh

THE DIFFERENCE, OR MISSING IR WAVE INTENSITY CAN BE MEASURED AND IS DUE TO THE

BREATH ALCOHOL CONTENT

BEER-LAMBERT LAW
(Provides the theoretical basis for IR breath testing.)

Each molecule absorbs infrared light waves in a unique way. Thus, each molecule has its own infrared fingerprint.
Ethanol Infrared Spectra with notation for I8000 IR filters
100% 90%

Percent Absorbtion

80% 70% 60% 50% 40% 30% 20% 10% 0% 3 4 5 6 7 8 9 10 11


3.4 micron filter 9.4 micron filter

Wavelength(microns)

Ethanol molecules absorb infrared light at wavelengths of approximately 3.00, 3.39, 7.25, 9.18, 9.50 and 11.5 microns.

No other compound absorbs infrared light at all of these wavelengths exclusively.

IF they are not the same, by a 5% difference, THEN the instrument will
determine a substance other than ethanol, an interferent, is in the sample. The measurements of the 3.4uM and 9.4uM wavelength filters must match or the substance is an interferent.

SEE?

Intoxilyzer 8000

The 8000 operation is similar to the 5000 with the following enhancements ACCESS CARD with bar code - used to scan in operator information Able to scan the bar code on DL for subject information No Specialist for the 8000 No Log Of Tests now on each Affidavit

Intoxilyzer 8000

Uses dry gas instead of simulator solution Able to run approx 325 test before dry gas cylinder needs to be changed or approx 2 years before cylinder expiration date Retains complete Affidavit information Sends Affidavit information to BOT and LEADRS

Intoxilyzer 8000

To perform a test on the Intoxilyzer 8000 an officer must have a valid permit. The first thing he will do is scan his ACCESS card and put in his PIN or manually type in the information if he does not have his ACCESS card. The 8000 will check an internal data base to verify if your permit is valid, if not you will not be able to operate the instrument. It takes about fifteen minutes to delete an operator from the database across the state if he is disqualified for any reason.

INTOXILYZER 8000
Access Card information The officers no longer have to type this information. LAST, FIRST, MIDDLE NAME BADGE NUMBER PERMIT NUMBER AGENCY AGENCY ADDRESS, ZIP, PHONE PERSONAL IDENTIFICATION NUMBER (PIN) (verify correct operator for digital signature)

INTOXILYZER 8000
The officers no longer type in the following information LAST NAME FIRST NAME MIDDLE NAME DATE OF BIRTH SEX M/F ADDRESS CITY

STATE ZIP D.L. NUMBER STATE OF ISSUE REVIEW Y/N

BREATH TEST

BREATH TEST Y/N:

MOUTH PIECE G/C: G DEPRIVATION Y/N: Y starts 16 minute clock N Allows officer to put in a start time for the period. If it is less than fifteen minutes prior, the clock will enforce a delay for the remaining time.

BREATH TEST
9 STEP SEQUENCE AIR BLANK BREATH SAMPLE AIR BLANK WAIT PERIOD AIR BLANK BREATH SAMPLE AIR BLANK CONTROL CHECK AIR BLANK

BREATH TEST SEQUENCE


IF THE INTOXILYZER DETECTS RADIO FREQUENCY IN CLOSE PROXIMITY Cell phones within 6 feet THE SCREEN WILL DISPLAY AND YOU WILL GET A PRINTOUT STATING

BREATH TEST SEQUENCE

IF

YOU PUSH THE START BUTTON AGAIN

THE

SCREEN WILL DISPLAY AND YOU WILL GET A PRINTOUT STATING SEQUENCE ABORTED

BREATH TEST SEQUENCE

IF

THE SUBJECT BLOWS DURING AN AIR BLANK

THE

SCREEN WILL DISPLAY AND YOU WILL GET A PRINTOUT STATING IMPROPER SAMPLE

BREATH TEST SEQUENCE

IF the subject has substantial mouth alcohol or

THE SUBJECT BLOWS A BREATH SAMPLE THAT IS ABOVE THE RANGE OF THE INTOXILYZER, 0.60

THE

SCREEN WILL DISPLAY AND YOU WILL GET A PRINTOUT STATING RANGE EXCEEDED

BREATH TEST SEQUENCE

IF THE SUBJECT DOESNT BLOW AN ADEQUATE SAMPLE, STOPS TOO SOON, STARTS AND STOPS, OR BLOWS AND SUCKS

THE

SCREEN WILL CONTINUE TO PROMPT PLEASE BLOW FOR 3 MINUTES AFTER WHICH YOU WILL GET A PRINTOUT STATING DEFICIENT SAMPLE

BREATH TEST SEQUENCE

IF

THE AIR BLANK IS CONTAMINATED

THE

SCREEN WILL DISPLAY AND YOU WILL GET A PRINTOUT STATING AMBIENT FAIL (PURGE FAIL)

BREATH TEST SEQUENCE

IF THE BREATH SAMPLE IS CONTAMINATED OR THE SUBJECT PUFFS RAPIDLY SCREEN WILL WILL DISPLAY SAMPLE REJECTED

THE

THE SCREEN WILL THEN PROMPT RETRY Y/N SELECTING Y WILL START A 16 MINUTE DEPRIVATION CLOCK.

BREATH TEST SEQUENCE

IF THE BREATH SAMPLES ARE MORE THAN .03 APART, THE TEST IS NOT VALID AND

THE INTOXILYZER WILL PROMPT RETRY Y/N YOU CAN RETRY UP TO 3 TIMES WITHOUT HAVING TO RE-ENTER THE SUBJECT INFORMATION. THIS IS ANOTHER PROTECTION AGAINST TESTING MOUTH ALCOHOL.

BREATH TEST SEQUENCE

IF THE CONTROL TEST IS BEYOND THE

.01 TOLERANCE

THE

INTOXILYZER WILL DISPLAY Calibration Check Out of Tolerance THE PRINTOUT WILL SHOW THE RESULTS WITH THE CONTROL TEST RESULT FLAGGED INVALID AND

BREATH TEST SEQUENCE THE INTRUMENT WILL BE DISABLED

THE TEST MUST BE ADMINISTERED ON A DIFFERENT INTOXILYZER OR CONDUCT A BLOOD TEST

BREATH TEST SEQUENCE

IF THE TESTING SEQUENCE FAILS, ANY PRINTOUT IS EVIDENCE OF WHAT HAPPENED DURING THE TEST. THE SUBJECT SHOULD RECEIVE A COPY A COPY SHOULD GO TO EVERYONE WHO NORMALLY GETS A COPY OF THE THE AFFIDAVIT THE PRINTOUTS ARE SAVED IN THE MEMORY OF THE INSTRUMENT AND ARE IMMEDIATELY UPLOADED TO THE BOARD COMPUTER IN OKLAHOMA CITY. YOU WILL BE ABLE TO GET THEM.

BREATH TEST SEQUENCE

IF THE TESTING SEQUENCE COMPLETES SUCCESSFULLY THE OFFICERS AFFIDAVIT AND NOTICE OF REVOCATION/DISQUALIFICATION IS SAVED THE SAME WAY. THEY ARE NEVER PURGED FROM THE MAIN COMPUTER. YOU WOULD BE ABLE TO GET A PRINTOUT TEN YEARS FROM NOW. (It may cost you.)

The 8000 Intoxilyzer Temporary Driver License

OFFICERS AFFIDAVIT
DISTRIBUTION

On the Officer/DPS Affidavit the officer will complete SECTIONS 2, 3 & 4. SECTION 4 does NOT require a notary.

The 8000 Intoxilyzer Log of Tests is the Officer Affidavit

IMPLIED CONSENT
CIVIL ACTION NOT CRIMINAL BURDEN OF PROOF

Breath Test Operating Procedure Push the GREEN button and follow the prompts. If an Affidavit prints, the Officer followed the procedure approved by the Board of Test. If a test result appears, it is a valid, states test even if the result is 0.00.

ADMISSIBILITY

47 O.S. sec. 756 A (brutally edited) Upon the trial of any proceeding out of acts committed by any person driving a motor vehicle while under the influence of alcohol, evidence of the alcohol concentration in the breath of the person as shown by analysis of the breath performed in accordance with Sections 752 and 759 of [title 47] IS admissible.

COMPLIANCE WITH SS 752 & 759

47 O.S. 752 H (brutally edited)

Tests breath for determining the alcohol concentration, under the provisions of [O.S. Title 47], to be admissible in evidence under the provisions of this [O.S. Title 47], shall have been administered in accordance with the rules and regulations of the Board.

COMMON RED HERRINGS

The breath tube on the I8000 is shorter than the 5000D and the heater may have failed. Something in the sample chamber could increase the reading by blocking the infrared light. Mouth alcohol was measured in the breath sample.

COMMON RED HERRINGS

No one can review the source code. The particulate screen can clog. The driver can affect the test result by hyperventilating. RFI may have affected the result. Another gaseous substance in the breath may raise the result interferents.