Академический Документы
Профессиональный Документы
Культура Документы
radiologists Chest tubes, central lines, endotracheal tubes and NG tubes are common New devices are constantly being introduced
and to have an understanding of its function as well as complications associated with its use
advances in the treatment of the very ill have greatly increased the numbers of examinations performed at the patients bedside. Obtaining a daily chest radiograph is standard practice in most intensive care units
Chest radiograph shows that the tip of the endotracheal tube (black arrow) is slightly above the aortic arch and well above the carina, in good position. A right chest tube (white arrow), ECG leads (E), a gown snap (G), and oxygen tubing (O)
Medical devices
Extra thoracic devices
Pleural devices Tracheal
Esophageal
Vascular Cardiac Circulatory assist devices
Breast prostheses
other apparatus often lie on or under the patient and
ECG leads
Pleural Devices
chest tubes are commonly used for evacuating fluid or
air from the pleural space Antero superiorly for pneumothorax Poster inferiorly for fluid collection
Chest tubes are commonly used for evacuating uid or air from the pleural space. The normally positioned tube lies on the surface of the expanded lung, between the visceral and parietal pleura. The tube is usually placed anterosuperiorly to evacuate a pneumothorax and posteroinferiorly for uid collections.
expanded lung, between the visceral and parietal pleurae. Pigtail catheters may be used in place of standard thoracostomy tubes, and they are popular for empyema drainage and for installation of medication for treatment of an empyema
Frontal view of the chest shows a pigtail catheter that had been
inserted under fluoroscopic guidance into a loculated right empyema for instillation of urokinase and fluid drainage.
Complication Malposition
May enter the interlobar fissure Lung parenchyma Subcutaneous tissue Tube may be kinked
tube. Magnified anteroposterior chest radiograph shows misplaced chest tube (arrows) within right chest wall.
65-year-old woman with extrapleural placement of chest tube. A, Magnified anteroposterior chest radiograph shows left chest tube (arrow) in apparently adequate position. CT scan was requested to further investigate because of ineffective drainage of left pleural effusion.
69-year-old man with intrafissural placement of chest tube. A, Magnified anteroposterior chest radiograph shows
chest tube. A, Scout image shows chest tube (arrow) projecting over right mid lung field.
coursing through right upper lobe. There is associated small pneumothorax (asterisk) and subcutaneous emphysema (thin arrow).
posteroanterior chest radiograph performed after chest tube placement shows kinking of chest tube (arrow) precluding adequate pleural drainage.
37-year-old man with mediastinal placement of chest tube. A, Anteroposterior chest radiograph shows left chest tube (arrows)
in inappropriate position, directed medially and projecting across mediastinum. There is persistent left pleural effusion.
30-year-old male victim of motor vehicle trauma with abdominal placement of chest tube. A, Anteroposterior chest radiograph shows horizontally oriented chest tube (arrow) in left lower hemithorax. There are several left rib fractures, opacification of left hemithorax, and subcutaneous emphysema.
30-year-old male victim of motor vehicle trauma with abdominal placement of chest tube. B, CT image shows traumatic left diaphragmatic rupture with migration of abdominal content to left hemithorax. Chest tube (black arrow) is seen within mesenteric fat abutting small bowel loops. Note splenic rupture (white arrow) related to trauma.
Tracheal Device
Tracheal intubation is a life saving procedure but can
be life threatening if placed incorrectly The tip of the tube should be 5 cm above the carina
(black arrow) is slightly above the aortic arch and well above the carina, in good position. A right chest tube (white arrow), ECG leads (E), a gown snap (G), and oxygen tubing (O) are also visible.
Complications
When advanced too far, the endotracheal tube usually
enters the right main bronchus, causing various combinations of hyperinflation and atelectasis of the two lungs, depending on the positions of the end and side holes. endotracheal tubes can also be placed in the esophagus the soft tissues of the neck.
anteroposterior chest radiograph shows that tip of endotracheal tube (thick arrow) is too high, at level of thoracic inlet. Endotracheal tube cuff (thin arrows) is overdistended. This abnormal position may cause vocal cord injury.
intubation. Anteroposterior chest radiograph shows endotracheal tube tip (arrow) in right main bronchus, resulting in complete collapse of left lung and leftward shift of mediastinum.
66-year-old man with left-sided double-lumen endotracheal tube. Magnified anteroposterior chest radiograph shows double-lumen endotracheal tube with its left tip (thick arrow) in left main bronchus. Right tip (thin arrow) is noted within trachea for ventilation of right lung. Double-lumen endotracheal tube allows control of distribution of ventilation to each lung. It is important to differentiate between double-lumen endotracheal tubes and inadvertent selective bronchial intubation with single-lumen catheters.
Esophageal devices
NG tubes
Esophageal balloons Esophageal stents
PH probe
NG tubes
Nasogastric tubes and feeding tubes are frequently
visualized passing through the mediastinum on their way to the stomach and intestines. Esophageal balloons and esophageal stents used to treat benign and malignant esophageal disease may manifest themselves at chest radiography.
Vascular devices
Routinely used for Monitoring hemodynamic function Hemodialysis Administrating fluids Medication Nutrition
percutaneously or surgically, via the subclavian, internal jugular, or femoral veins. Arterial devices usually are placed through the femoral artery central lines typically have one to three lumens A central venous catheter is ideally positioned in the superior vena cava for the monitoring of pressure or infusion of medication and nutrition.
measuring hemodynamic pressures and cardiac output.A better term to use is pulmonary artery catheter. Accurate measurement of pulmonary arterial wedge pressure
central venous catheter and a left subcutaneous port catheter, which enters via the left subclavian vein. Both catheter tips are in the superior vena cava.
catheter or to an arterial catheter and can be used for instillation of fluids, medications, chemotherapeutic agents, parenteral nutritional solutions, and blood products. It can also be used for withdrawal of blood samples.
catheter with its tip in the proximal most portion of the superior vena cava. (b) Close up view of the catheter tip.
Groshong catheter has closed, rounded tip Near the tip in the side of the catheter is a three-position valve. The valve is designed to allow fluid to flow in and out through the valve, but it remains closed when it is not in use. This catheter does not require routine clamping or heparin solution to keep open. It
should be within the mediastinal shadow. Placement more distally increases the chance of pulmonary infarction or vessel rupture.
Ganz catheter with its tip (arrow) in the right lower pulmonary artery.
(white arrow) in the left pulmonary artery via the inferior vena cava. Note also the bilateral chest tubes (black arrows) and ECG leads (E).
catheters used and the sites employed Pneumothorax is a common complication Vessel lacerations and perforations can produce hematomas, hemothorax, and infusion of fluid into the mediastinum, thorax, or other inappropriate space Nerve injury is usually a complication of improper puncture technique Looping of catheters may lead to knotting.
exceedingly common, found in approximately 25% of catheters placed. This may lead to false readings and an increased risk for complications. Complications of pulmonary artery catheter placement include pneumothorax, pulmonary infarction, cardiac arrhythmias, pulmonary artery perforation, endocarditis, and sepsis
Cardiac Devices
Cardiac pacemakers, valve prostheses, and artificial
hearts chest radiography is commonly employed in the assessment of patients with heart disease recognition of cardiac devices and the problems associated with them is important for all individuals involved in the care of these patients.
Heart valves
Mechanical Biologic Mechanical Heart valves most require life-long treatment
with anticoagulants. Biologic valves are less durable than mechanical valves, with some deterioration developing, frequently 510 years after placement, but they do not usually require anticoagulant treatment. It is not important and often impossible to know the specific name of a particular prosthetic heart valve, but it is important to recognize its presence The mitral and aortic valves are those most commonly replaced
tilting bileaflet mechanical mitral valve prosthesis. Median sternotomy wires and surgical clips are also evident.
Hancock porcine mitral valve prosthesis (arrow). A singlelead pacemaker, ECG leads, and median sternotomy wires are also seen.
Hancock porcine valve prosthesis in a Rastelli conduit going from the right ventricle to the pulmonary artery.
ring (black arrow) and a dual-lead cardiac pacemaker. Sternal wires, surgical clips, and ECG leads are also present. The sternal wires are used to close a sternal dehiscence. The patient has both horizontal sternal wires and vertical intercostal wires (white arrows).
Cardiac Pacemakers
common in older adults being treated for abnormal
cardiac rhythms caused by coronary artery disease Cardiac pacemakers improve cardiac function, reduce the severity of clinical symptoms, and reduce mortality and morbidity. A cardiac pacemaker is composed of two main elements: (a) a pulse generator and (b) lead wires with electrodes for contact with the endocardium or myocardium
electrodes to very complex pacemakers with multiple atrial and ventricular leads
Frontal (a) and lateral (b) views of the chest show a single electrode epicardial corkscrew subxiphoid pacemaker (arrowhead in a, black arrow in b). There are also coils (white arrow) occluding a previous right Blalock-Taussig shunt. In addition, ECG leads and sternal wires are evident.
sequential pacemaker with one electrode in the right atrial appendage (RA) and the other at the right ventricular apex (RV). Also shown are ECG leads (E) and the battery-control pack (B) for the pacemaker.
proper positioning of pacemaker leads, it is often difficult for the to know if a pacemaker is properly positioned. Pacemaker lead fracture is now rarely seen because of improvements in the flexibility of the metal alloys used in electrode construction.
coronary artery angioplasty, and coronary artery stent placement. Median sternotomy is the usual surgical approach for CABG surgery, and sternal wires the common method of fixation of the two sternal segments At present, almost 90% of coronary interventions include stent placement Complications associated with coronary artery stents are stent thrombosis and restenosis.
Lateral view of the chest shows sternal wires (arrowhead), vascular clips of
a saphenous vein bypass graft to the right coronary artery (curved arrow), and those of the left internal mammary graft to the left anterior descending coronary artery (straight arrow).
a patient with bilateral pulmonary artery stenosis and aneurysmal right ventricular outflow tract following tetralogy repair.
to spur efforts to develop mechanical support for the circulatory system Most mechanical support for patients with heart failure consists of devices that assist the heart without replacing it.
Mechanical cardiac assist devices can be divided into three groups: (a) temporary cardiac assist devices (b) permanent cardiac assist devices (c) heart replacement devices. short-term cardiac assist devices are the intraaotic balloon pump and newer left ventricular assist devices (VADs)
valves and the two coil, reinforced polyurethane tubes carrying pulses of compressed air to the two artificial ventricles.
Cropped frontal view (a) and full lateral view (b) of the chest show a CardioWest total artificial heart. Note the four prosthetic valves and the two coil, reinforced polyurethane tubes carrying pulses of compressed air to the two artificial ventricles.
chest radiograph shows that feeding tube has entered right main bronchus, traversed right lower lobe bronchus (white arrows), and has its tip overlying right upper quadrant of abdomen (black arrow), raising concern for possible perforation of right hemidiaphragm. Note associated right pneumothorax (asterisk).
(wax plombage). Extensive pleural calcification includes the surface of the wax ball (arrows).
A pacemaker is one of the common devices encountered on a chest x-ray. The usual location for a pacemaker is the anterior left upper chest wall (black arrow). Pacemakers may have either 1 or 2 leads. The wires connecting the pacemaker to the intracardiac electrodes must be intact (yellow arrow). The typical position of the cardiac electrodes is in the right ventricle (red arrow) for a single lead, and also in the right atrium for a dual-lead pacemaker. It is important to compare the electrode position to that in previous studies because an electrode may become dislodged.
arrows) that was placed to ameliorate the effects of an esophageal malignancy. There are also two chest tubes (), a peripherally inserted central catheter (white arrow), ECG leads (E), a gown snap (G), and a transjugular intrahepatic portosystemic shunt (T) in the liver.
Frontal view of the chest shows a left jugular Swan-Ganz catheter (arrows), which passes through a persistent left superior vena cava into the coronary sinus, through the right atrium and right ventricle, and into the right pulmonary artery. Also seen are a subcutaneous port (P), an endotracheal tube (ET), an ECG lead (E), and a nasogastric tube (not labeled).
The electrode is placed in the epidural space adjacent to the spinal cord. The wires are connected to a stimulating generator implanted subcutaneously. The electrode generates a weak electrical current that interrupts the transmission of pain at a spinal cord level.
Spinal cord stimulation (SCS) is recommended as a treatment option for adults with chronic pain of neuropathic origin. Most frequently: Brachial plexopathy, Post-laminectomy syndrome, Post Chemotherapy Neuropathy, Complex Regional Pain Syndrome (types I and II) and HIV polyneuropathy.
The implantable loop recorder (ILR) is a subcutaneous electrocardiographic monitoring device that stores ECG data automatically in response to specific rhythm anomalies or in response to patient activation. It is mainly used for diagnosis in patients with recurrent unexplained episodes of syncope or palpitations, but is also useful for long-term monitoring in patients with documented or suspected atrial fibrillation, for risk stratification in patients who have sustained a myocardial infarction and those who have certain genetic disorders.
dialysis catheter. A, Anteroposterior chest radiograph shows catheter (arrow) inserted via right subclavian vein with its tip projecting over right atrium.
dialysis catheter. B, Because there was clinical suspicion of malpositioning of catheter, IV contrast medium was injected and was seen to extravasate into pleural space (arrows).
pulmonary artery catheter placement. Magnified anteroposterior chest radiograph shows that tip of catheter (black arrow) is too distal (i.e., > 2 cm lateral to hilum). There is wedged-shaped opacity (white arrows) distal to catheter, consistent with pulmonary infarction.
lead. A, Posteroanterior chest radiograph shows dual-lead pacemaker. Tip of right ventricular lead (arrow) is projected at edge cardiac silhouette.
anteroposterior chest radiograph shows fracture (arrow) in pacemaker lead near battery-control pack. Lead fractures most commonly occur at venous access site, near tip, or near batterycontrol pack.
7-year-old man with normal positioning of intraaortic counterpulsation balloon pump. B, Magnified anteroposterior chest radiograph obtained during diastole shows inflated radiolucent balloon (thin arrows) as well as radiopaque tip (thick arrow) within upper descending thoracic aorta. Catheter is inflated during diastole to increase myocardial perfusion and is deflated during systole to decrease left ventricular afterload.
misplaced central venous catheters. 6-year-old girl. Posteroanterior chest radiograph shows two misplaced
catheters. Right internal jugular central venous catheter (black arrow) has its tip in right atrium. Left subclavian central venous catheter (white arrow) has its tip in right subclavian vein.
Conclusion
Various devices are used to monitor and treat critically
ill patients. The radiographic evaluation of these devices is important because the potentially serious complications arising from their introduction and use are often not clinically apparent. Familiarity with normal and abnormal radiographic findings is critical for the detection of these complications.
Recommendation
The American College of Radiology recommends daily
chest radiography for critically ill patients who have acute cardiopulmonary disease or are receiving mechanical ventilation, as well as immediate imaging for all patients who have undergone placement of endotracheal tubes (ETTs), feeding tubes, vascular catheters, and chest tubes. These recommendations are made because the malpositioning of these devices and the serious complications that may ensue are often not clinically apparent. Radiographic evaluation of these devices is important, albeit challenging, because of the technical limitations of portable chest radiography and the inability of patients to cooperate.
Thank You