Академический Документы
Профессиональный Документы
Культура Документы
Test
Equipment
Line
Voltage
Multimeter
Impulse counter (invasive, common in older units)
No load
Turn unit on
Check for adequate incoming power
Load
Take exposure
Voltage drops specified amount under controlled load
(Example: 2 VAC drop at 98 kVp)
Electrical
Calibration
mA Preheat Calibration
Verifies voltage for warming up the filaments during preheat cycle
Important for safeguarding the filament coils during exposure
mA Calibration
Verifies tubehead current (mA)
Accomplished by taking an exposure
Dental units have fixed mA
kVP Calibration, load
Verifies output at all kVp settings (dental x-ray equipment is normally in
70-100 kVp range)
Accomplished by taking an exposure
PMCS explain
Yearly maintenance checklist PM
General considerations
Technical instructions required
Operating instructions
Service manual