Uncertainties

Using & Calculating


Uncertainties for Electrical
Measurement

Expressing Uncertainty of
Measurement
All measurements, even the most accurate,
have an unknown inaccuracy or doubt.
The is known as the UNCERTAINTY
As there is always an uncertainty with any
measurement we need to estimate this
amount. We also need to calculate our
confidence in the estimate of uncertainty, which is
how sure we are that the true value is within the
uncertainty we have estimated.

Expressing Uncertainty of
Measurement
As an example we may measure 1 Volt and
be 95% confident that we are within 10uV

Expressing Uncertainty of
Measurement
UNCERTAINTY vs. ACCURACY
There is no connection between these terms.
Uncertainty is purely the unknown in any
measurement.
Accuracy or Tolerance is the difference
between the desired value and the actual
measured value.

Expressing Uncertainty of
Measurement
Example
With a digital watch with 1 second resolution
this resolution will limit the best uncertainty
to which you can tell the time (ie. 1 second),
however the watch itself may only be
accurate to a few minutes.

Expressing Uncertainty of
Measurement
There are many sources of uncertainty in any
measurement which need to be combined
using statistical techniques to give a total.
Different types of uncertainty need to be
treated differently to obtain an accurate
estimation.

Expressing Uncertainty of
Measurement
To calculate uncertainty you must first
identify all the sources of error, estimate the
size of the contribution from each source and
also decide on the type of uncertainty for
each source.
There are two types of uncertainty
Type A Based on using statistics
e.g. repeated readings
Type

B Based on other factors


e.g. manufacturers specifications

Sources of Uncertainty in
Electrical Measurements.
Imported

Uncertainty
Drift of reference instrument
Temperature effects
Lead and thermal errors (DC volts)
Rounding errors due to resolution
Repeatability
Noise
Self heating of high current shunts

Imported Uncertainty
Imported

uncertainty is taken directly from the


certificate issued by the laboratory which
calibrated the reference instrument. The
probability distribution is NORMAL

Drift of reference instrument

Drift of reference instrument can be either taken form


historical data on the instrument or from the manufacture
specification for stability. If the drift with time can be
predicted it is possible to use a corrected figure for the
actual value of the reference with a reduced figure for
drift. However it is more normal to use an un corrected
figure.

The probability distribution is Rectangular.

Temperature Effects

The effect of temperature on many modern instruments is


often very small, and in many cases the instruments
specification covers a band of temperature without any
further addition.
Some reference standards for example resistors the
Temperature coefficient may be quite important. The
figure for TC can be taken from the manufactures spec or
measured.
The probability distribution is Rectangular.

Lead and thermal errors


(DC volts)
Thermal

emf can be difficult to evaluate. With a


little care and correct leads it is normal for
thermal EMF to be less than 1uV, or even 0.5uV
which is a figure often used in calculations.

The

probability distribution is Rectangular.

Resolution of Measurement

It is firstly important to understand that there is a big difference


between the resolution of a measuring instrument and that of a
reference source.

A source, such as a standard resistor may have no resolution at


all, but can still be very accurate, while for a measuring
instrument resolution is essential to achieve accuracy.

It is only necessary to include measurement resolution in the


uncertainty calculation. Note if a DMM is used to compare say
two resistors then the resolution must be entered twice.

The probability distribution is Rectangular

Combining Uncertainties

It is normal these days to use a spread sheet, in Excel, taking a


template from M3003.

Use column 1 for a description of source of uncertainty


Use column 2 for the value of uncertainty usually in ppm
Use column 3 for a description of type of Probability distribution
Use column 4 for the divisor, 2 for a normal dist, 1.732 for rect
Use column 5 for a coeficient used to convert say millvolts to microvolts
Use column 6 = column2(value) x column4(divisor) x column5(coeff)

Use the sum of square to total column 6 to get the Combined Standard
Uncertainty

To obtain the Expanded uncertainty (K=2, 95% Confidance) multiply the


result above by 2. Then round to 2 significant places.

Expressing Uncertainty of
Measurement
Example

Using ProCal to calculate


Uncertainties

ProCal use three key elements to dynamically calculate


uncertainties as the test is run

1) A table with imported uncertainties and calibrator specifications


2) A laboratory procedure incorporating additional factors such as
lead /connection errors etc.
3) The noise / flicker which can be input at the time of test

Set Up Instrument Spec and


Imported Uncertainties

Use Proset, select instrument Traceability in the file menu.


Select the Instrument required
Select the uncertainties button to access the table
Enter the data, note default table exists for 2000 Series

Set Up Procedure Template

Use ProSet, select laboratory procedures in the file menu.


Select the Instrument required
Select the procedure spreadsheet template say for DC voltage
Enter the parameters to be calculated, note the imported and
reference specification will always be added automatically.
Note the procedures for the main functions of the calibrator are
Installed as default

Set Up Calibration Procedure

Use ProEdit, edit a procedure and go to the Instruments tab.


Check the calibration instrument, lab. Procedure and
uncertainty line.
If a Transmille calibrator is in use these items will be set
automatically, without having to be selected.

Calibrating An Instrument

Use ProCal, run a calibration


Input the reading and select / enter the noise / flicker in the drop
down box displayed.
If required, click on the UNCERT button to view the
uncertainty calculation.