Keller Gauge
In order for the Keller Gauge Leo Record (Ei) to work on the computer the following must first be installed:
1. Keller Gauge Software
2. K-104A Cable Software
Once the software from each item has been downloaded on the computer, the Keller Gauge can be programmed to:
Log Pressure and Temperature readings at different timing intervals
Plot and Print Graph of the recorded data
http://www.keller-druck.com/english/homee/hme.html
1. Select Products
2. Select Digital Pressure Gauges / Manometers / Calibrator
3. Select Leo Record (Ei)
4. Select Software Logger DCX 4.x
5. Select and Download Logger DCX
- There are several downloadable software versions on that page, ensure that the LOGGER DCX is the software
version that you download.
1. Connect the Gauge to the computer using the K-104 A Cable (USB R485)
2. Insert CD Driver Disc: K-104 USB Converter
3. Follow the direction prompts from the CD
Once the Keller Gauge Software and K-104 A Cable software is installed, open the Writer and Reader Programs
Note: Shortcuts for the Writer and Reader can be placed onto the desktop.
Writer
Writer
1. Ensure that the Type reads Pressure: If the Writer indicates Water Level, then select the Type
Button until it reads Pressure
3. If the Keller Gauge is communicating with the computer, the Battery Capacity will change from 0% (Blue
status bar) to 100% (Green Status Bar)
Writer
Battery Capacity =
100%, therefore the
Computer is
communicating with
the Gauge
Reader
Ensure that they Viewer is contained in the
Modules to Execute box.
If an error message occurs, it means that the computer is not communicating with the gauge. Assuming that the
software has been installed correctly, adjust the COM PORT setting to the highest levels and re-try the previous
steps.
NOTE:
2 6
1
4
If the Gauge is connected to the computer during the operation, the software can go online and display exactly
what the gauge is recording
Note:
The Temperature
reading is the
value at that
location.
NOTE:
WHEN THE SYSTEM IS ONLINE, THE READER WILL NOT
DOWNLOAD THE GRAPH. GO OFFLINE TO DOWNLINE THE
GRAPH FROM THE READER IF NEEDED
The data from the Gauge can be downloaded at any time during the operations.
If the gauge is active, a message will pop-up
Un-check the Synchronize Test with PC-Clock and DONT SET A MARK
Setting a mark will split the graph / data into different event and not have a continuous graph
DO NOT SET A
MARK
The data on the graph cannot be edited, but the look of the graph can be edited
On the viewer: Options Edit Graph
Adjust the line of the Graph to Solid Lines
Use P1 Blue and TOB1 Red
NOTE:
DO NOT CHANGE THE NAME OF THE FILES. SAVE THEM AS IS IN THE RESPECTIVE FOLDER
Once save in the respective folder, the files will be displayed as:
The accepted format for presenting the graph to the client is as followed:
1. Copy and paste the Graph into the given template
2. Copy and paste a 2nd Graph into the given template and label all key event points
The client should be presented with 2 Graphs for the given operation.
1 document will have a clean graph
1 document will have a reported graph with all the event points explained