INJECTOMAT AGILIA

SYRINGE PUMP
GUIDELINES FOR USE

Presented by,
Prathibha Saseedharan
Biomedical department
INTRODUCTION
Used for IV drug administration needs.
Flow rate range:0.1- 200 ml/hr
Flow rate accuracy: 1% on mechanism ; 2 % on syringes.
Syringe capacities:5, 10, 20, 30/35, 50/60 CC.
Infusion Modes:
ml/h (0.1 ml/h increments).
Volume / time: Volume to infuse 0.1 to 99.9 ml / Time: 0 h 01 to
96 h 00.
Volume limit: 0.1-999.9 ml (0.1 ml increments).
Bolus:Manual bolus: Rate 50-1200 ml/h (50 ml/h increments).
FRONT VIEW REAR VIEW

Click to add text


User panel
INSTALLATION.

Can be placed on table, pole or For fixing the pump push the
rail. fixing button and screw the
2 devices max can be clamp screw.
assembled together during Place the syringe with IV set in
infusion. cradleand secure using
3 devices max can be syringe barrel clasp.
assembled on a pole or for Move pusher forward to
transportation. syringe head.
OPERATIONS FOR USE
1.Switch ON/OFF - Connect power cord to mains supply. Press
<ON> to switch on and press <OFF> continuously for off.
2. Syringe selection- press OK to confirm syringe
C to change syringe
3. Prime function remove air bubbles in line using prime.
Press twice 1 short press and 1 continuous press and release
key to stop.
4. Flow rate selection/start - select flow rate (check infusion
parameters)
Press validation key to start infusion.
5. Silence - press to silence any alarms.
6. Access to menu press <menu/exit> key to access
functions like
- infused volume
- pressure
- battery life
7. Bolus Press bolus key until flow rate flashes, select flow rate
and confirm.
Press <bolus> twice to start and release key to stop.
8. Pause press <stop> to stop infusion.
Press validation to start infusion.
To program the pause duration press <stop> twice
and select duration.
Display and
Symbols
Alarm and safety
features
Menu
Options.