MSA - ALTAIR 4X

Gas Meter
Instrument Overview
1 – MSA “Link” Communicator
2 – Self LED (green) and Fault
LED (yellow)
3 – Sensor Inlet
4 – Horn
5- Button
6- Button
7- Button
8 – Display
9 – Alarm LEDs (4)
Instrument Overview
10 – Clip
11 – Charging Connection
12 – Screws (4)
13 – Charge LED (red/green)
LED Descriptions
Green – flashes every 15 seconds
unit is on and operating
Red – Alarm condition
Yellow – fault condition
Red/Green – charging status
Red – is charging
Green – charge complete
On-Screen Indicators
1 – Graphic Symbols By row starting at the top left.
2 – Gas Types 4-no gas cylinder applied – expose to fresh air
3 – Gas Concentration 5-indicates “Peak” reading or high alarm
4 5 6 7 8 6-indicates STEL alarm
7-hourglass indicates user should wait
9 10 11 12 13 14 8-battery condition
9-cal gas must be applied
10-indicated minimum value or low alarm
11-indicates TWA alarm
12-indicates required interaction
13-indicates “Motion Alert” is active
14-Heart indicates sensor life warning
15-check mark indicates successful bump or
calibration
16-Sensor Labels
Battery Life Indicator
 The battery condition icon continuously displays in the
upper right-hand corner of the display. As the battery is
depleted, the battery icon segments blank until only the
battery icon outline remains.
Warning
 If the battery warning alarm activates while using the
instrument, leave the area immediately as the end of
the battery life is approaching.
 The alarm indicates the there is 30 minutes of
operating time left.
 Normal run time for the unit is 24 hours.
Operation of Unit

1. Turn “ON” in fresh air for the “Fresh Air Setup.”


2. The instrument does a self test and then goes into
“Fresh Air Setup”
1. All display segments are activated
2. Audible alarm sounds.
3. Alarm LED lights
4. Vibrating alarm activates.
Operation of Unit
 After the instrument does its self check it will display
FAS? Asking if a fresh air sample is to be done. Press
enter. If you forget to zero the meter it will revert to
its last fresh air setup.
 The “Fresh Air Setup” (FAS) is for automatic ZERO
calibration of the instrument.
 The FAS has limits. The zero of any sensor that is
outside of these limits will not be adjusted by the
FAS command.
Operation of Unit

WARNING
Do NOT activate the Fresh Air Setup unless
you are certain you are in fresh air,
uncontaminated; otherwise, inaccurate
readings can occur which can falsely
indicate that a hazardous atmosphere is
safe.
Operation of Unit
Measurement Mode
•The meter has preset measures for all four gases to alert
you when the concentration has become dangerous.

•The combustible gas measures the % of the lower


explosive limit of 5. Meaning when the meter reaches
100% it is at its LEL. When this is reached the meter will
alarm and flash XXX.
Maintenance

 Wash with a cloth of mild soap and water


 Make a note on the board if
 -there is an issue with a meter
 -the meter needs bump tested/calibrated
 -the meter needs charged