Академический Документы
Профессиональный Документы
Культура Документы
5ml:
0.1 100
10ml:
0.1 200
20ml:
0.1 400
30ml:
0.1 600
50/60ml:
0.1 1500 ml/h
BeneFusion VP5:
Infusion pump
Benefusion VP5
Syringe pump
Benefusion SP5
16 : 9
Selectable eight levels screen brightness, providing optimum brightness and
increasing comfort for the patient.
Display :Flow rate , Current infusion, VTBI, drug name, Remaining time,
accumulated volume, pressure, battery, key- lock, alarms etc.;
RS232:
AC connector Data transmit to PC/Upgrade software/
Upgrade docking system *
Nurse call;
DC input
SP5 – Side View
Lock catch:
Safer! Firm to fix pumps together;
Automatically clamps the line to avoid any risk of free-flow when the door is
opened and when the set is removed.
Performance
@25ml/h @ 5ml/h
when occlusion happens, the motor will turn some volume back, but there still some
left volume in the IV tubing or the syringe;
Reaction time: the pump detect the pressure limit and begin to alarm when occlusion
happens.
When occlusion happens, the motor re-back some volume automatically and
quickly to avoid the patient blood pressure rise , then to avoid much volume
hurt the patients and ensure safety;
Anti-bolus system reduces bolus rate significantly after occlusion alarm release.
75%
50%
25%
low battery alarm
0%
battery empty alarm
Still support last 10 min.,
No battery, alarm
Dual protection!
Pumps with free-flow protection greatly reduce the risk of patient injury or
death associated with free-flow over infusion;
The 3 modes being used most High level modes with advanced
of times! technology, make fluid therapy more
flexible!
Basic Modes Demo
11 Languages
In any parameter setting interface, there are reminder on the screen, do not need to
remember the function key how to use or which key moving to next step;
Smart!
Intimate!
Remember last infusion parameters when turn off, repeat settings are not required
when turn on the pump again .
Last infusion details will be showed directly on the screen, nurse do not need to
remember or re-check the last therapy parameters;
Save operator much time and workload, especially in some emergency cases;
History Records
First in, first out!
Quick research!
Customize the drug library to meet the clinical therapy needs and provide safer
medification delivery through the RS 232 interface.
Input full drug names and reduce medicaiton errors.
Hands-on / Hands-free Bolus
Standby: set time desired to stop, the pump will remind you time is
over, then continue to infuse automatically;
In some cases or some special cases, the patient have to stop infusing
for a while, but the pause can’t seting the time
Pause
Pause:
“Start”—continue infusion with same parameters;
“C”—set new parameters ;
Pause could deal with the accidents quite well, especially in some emergency cases;
KVO (keep vein open)
KVO operating KVO finish
0.1-5.0ml/h adjustable
The pump will keep slow syringe to guarantee the safety of the patients when
occlusion alarm happens.
Configurable KVO infusion rate to keep vein open.
Convenience – SP5 Operation
Open the door Set syringes Fix the clamp Close the door
Rotate the clamp
Pull the slider
Compared with the B.Braun automatic one , nurse must wait the door open, then
set syringes and still wait to close:
Noice disturb; Time-cost; High after-sale maintenance cost ;
Very easy to set syringes steps, effective for the operators;
Correct value,
End
6 Automatic Calibration
Calibrating when:
Syringe
calibration
Correct value,
End
11 12 Calibration complete
Insert value
Fuctions:
Password Data transmit to PC;
25931501
Upgrade software;
Upgrade docking
system;
Nurse call;
DC input
Click here to
browse software