Вы находитесь на странице: 1из 33

Resident Physics Lectures

Radiographic Grids
Prof. J.K Tonui, PhD

School of Medicine,
Department of Radiology & Imaging
Learning Outcomes
 At the end of this lecture, the student is expected to:

 Explain various grid errors or cut-offs and their causes.

 Describe how moving grids work, and why are they preferred

over stationary types.


 State the factors used in the selection of grids.

 Describe air gap technique for reducing scatter.

2 RIB 810, 811 & 812 Imaging Physics 8/19/2019


Introduction
 A Grid

 A device placed between the pt. and film for the purpose of

absorbing scatter radiation before it can interact with the


imaging receptor, e.g. X-ray film and
 First grid was made by Dr. Gustav Bucky in 1913 (used cross-

hatched), and
 Was modified by Dr. Hollis E. Potter in 1920 (used linear)

3 RIB 810, 811 & 812 Imaging Physics 8/19/2019


Creating the Image
 Transmission
 Responsible for dark areas
 Absorption
 Responsible for light areas
 Scatter
 Creates fog
 Lowers contrast
 Increases as
 kV increases
 Field size increases
 Thickness of part increases
 Z# decreases

4 RIB 810, 811 & 812 Imaging Physics 8/19/2019


Position of the Grid
 The grid

 Is placed between pt. (behind table or

upright bucky) & cassette, as shown


in Fig.,and
 If placed BACWARDS CAN CAUSE

GRID ERRORS, and


 Is recommended for use with body

parts >10 cm.

5 RIB 810, 811 & 812 Imaging Physics 8/19/2019


Grids
 Grids

 Allow primary radiation to reach

the image receptor (IR)


 Absorb most scattered radiation

 Primary disadvantage of grid use

o Grid lines on film

6 RIB 810, 811 & 812 Imaging Physics 8/19/2019


Digital Imaging Systems
 Digital imaging systems

 Use very high-frequency grids

o 103-200 lines/in

o 41-80 lines/cm

 Recommended for use with digital systems

o Minimizes grid line appearance

7 RIB 810, 811 & 812 Imaging Physics 8/19/2019


Grid Disadvantages
 Two disadvantages

 Increased patient dose

o i.e. increase amount of radiation pt. receives because it requires

exposure factors to be increased

 Positioning critical

o Require careful centering because poor positioning results in grid

cutoffs or errors, and


o Loss of primary radiation because images of lead strips projected wider

8 RIB 810, 811 & 812 Imaging Physics 8/19/2019


Grid Errors
 Proper alignment between x-ray tube and grid

 Very important

 Improper alignment will result in cut-off

 Off-level

 Off-center

 Off-focus

 Upside-down

 Moire effect

9 RIB 810, 811 & 812 Imaging Physics 8/19/2019


Grid Cutoff
 Grid cutoff

 Is loss of primary radiation that occurs when images of lead strips

are projected than would be with ordinary magnification.

 Four types of grid cutoff

1. Focused grids used upside down;


2. Lateral decentering (or angulation);
3. Focus-grid distance decentering, and
4. Combined lateral & focus-grid distance decentering
10 RIB 810, 811 & 812 Imaging Physics 8/19/2019
Upside Down Focused Grid
 Use inverted focus grid

 It produces severe peripheral cutoff with

 Dark exposed band in center of film and

 No exposure at film periphery (edges)

 When grid ratio is high, area exposed is

narrow

 When crossed grid is used upside down,

small area in the middle is only exposed.


11 RIB 810, 811 & 812 Imaging Physics
8/19/2019
Lateral Decentering
Focal Displacement
 Lateral decentering

 Is when x-ray tube is positioned lateral to

convergent line but at correct focal


distance

 Uniform loss of radiation over entire film

 Produce uniformly light radiograph, but

 Dangerous because there is no

recognizable characteristic in the film

12 RIB 810, 811 & 812 Imaging Physics 8/19/2019


Lateral Decentering
 Occurs also when grid at correct

position but tilted, and


 It also produce uniform loss of intensity

and no other clinical clues, and


 May be mistaken for technique

problems, but
 Can be compensated for by over-

exposing pt.

13 RIB 810, 811 & 812 Imaging Physics 8/19/2019


Lateral Decentering
 Magnitude of cutoff increases with

 Higher grid ratio,

 Greater decentering distance, and

 Smaller focal distances

 Loss of primary radiation is calculated by the equation:


rb
L   100
f0
L = loss of primary radiation (%)
r = grid ratio
b = lateral decentering distance (inches)
fo = focal distance of grid (inches)
14 RIB 810, 811 & 812 Imaging Physics 8/19/2019
Lateral Decentering
 Significant problem in portable radiography,

 Because it can be compensate by over-exposing the patient

 Note that

 The exact centering not possible

 To minimize the lateral decentering, use

 Low ratio grids, and

 Long focal distances

15 RIB 810, 811 & 812 Imaging Physics 8/19/2019


Distance Decentering
 Distance decentering

 Occurs when the grid is either too close or too far from focal

spot, and
 Produces darker center of the film.

 All parallel grids have some degree of distance decentering,

because
 Focused to infinity

16 RIB 810, 811 & 812 Imaging Physics 8/19/2019


 Far focus-grid •Near focus-grid
decentering decentering
 target below convergent line
 Target above convergent line
 cutoff more severe than far
decentering

X
 Far focus-grid
•Near focus-grid
decentering decentering

•Cutoff at periphery
 dark center

•cutoff proportional
to
 grid ratio

 decentering
distance
Minimizing Distance Decentering Cutoff

 To minimize distance decentering

cut-off, use
 low grid ratio, and

 small fields

19 RIB 810, 811 & 812 Imaging Physics 8/19/2019


Combined lateral and focus-grid distance
decentering

 Easy to recognize because

 It produces uneven exposure,

and
 The film is lighter on one end,

and darker on the other end as


seen in this Fig.

20 RIB 810, 811 & 812 Imaging Physics


Combined lateral and focus-grid distance
decentering
 Cutoff proportional to

 Grid ratio, and

 Decentering distance

 Cutoff inversely proportional to grid

focal distance, hence


 Less cutoff for longer focus grids, and

 Greater for shorter focus grids.

21 RIB 810, 811 & 812 Imaging Physics 8/19/2019


Moving Grids
 The moving grids

 Were invented by Dr. Hollis Potter in 1920, and

 As such are also referred as Potter-Bucky diaphragm.

 All stationary grids

 Form lines on the radiograph by absorbing primary x-rays, but

 Thinner Pb strips are less noticeable lines, and

 Since thinner strips have less Pb content, result in poor

“cleaning up” of scatter x-rays as well.


22 RIB 810, 811 & 812 Imaging Physics 8/19/2019
Moving Grids
 Moving grids

 Are used to both reduce appearance of

grid lines on the radiograph as well as


reducing scatter x-rays so as to realize
high quality images.

 Motion

 Starts with second trigger and blurs out

lead strip shadows.


23 RIB 810, 811 & 812 Imaging Physics 8/19/2019
Moving Grids
 The moving grid

 Uses focused grid, and

 Is placed in a holding mechanism that

begins moving just before the x-ray


exposure and continues moving after
the exposure ends, and
 Uses 2 types of movement:

o Reciprocating &

o Oscillating.

24 RIB 810, 811 & 812 Imaging Physics 8/19/2019


Moving Grids
 Grid motion is such that

 Reciprocating – motor drives the grid back and forth during

exposure.
 Oscillating - electromagnet pulls grid to one side and releases it

during exposure

 Note that

 Most grids are moving, except for portable imaging, and

 For 1 generators grid motion must not synchronize with pulses

25 RIB 810, 811 & 812 Imaging Physics 8/19/2019


Grid Tradeoff
 Advantage

 cleanup/scatter rejection

 Disadvantage

 Increases patient dose;

 Increases exposure time;

 Increases tube loading;

 Positioning & centering more critical, and

 $$$ (i.e. adds the cost)


26 RIB 810, 811 & 812 Imaging Physics 8/19/2019
Grid Selection
 Factors considered are:

 Patient Dose (mAs changes)

 Exam

 Detail required

 Part thickness

 Desired technique (kVp)

o Use low R for low kVp and high R of high kVp (books recommend 8:1

below 90 kVp and 12:1 above 90 kVp

 Equipment availability
27 RIB 810, 811 & 812 Imaging Physics 8/19/2019
Air Gap Technique
 Air gap technique

 Is also called air filtration

 Increase OID by 10 to 15 cm

 This reduces the amount of scatter reaching the IR because

some scatter will miss the IR.


 It is about the same as using an 8:1 grid

 mAs is increased 10% for every cm of air gap

 Increases magnification and reduction in detail. Has some

selective uses with chest imaging and cerebral angiography

28 RIB 810, 811 & 812 Imaging Physics 8/19/2019


Air Gap Technique
Angles of
 Principle escape

 radiation scatters uniformly

 decrease in scatter (most scatter misses

film)
o air gap decreases angle of capture; increases

angle of escape as shown in this Fig.

 Negligible attenuation in air gap

29 RIB 810, 811 & 812 Imaging Physics 8/19/2019


Air Gap Technique
 Air gap
 Is very effective in removing scatter originating closest to film
because much of scatter nearest tube doesn’t reach film

Much attenuation Air gap decreases


of scatter in the body capture angle

30 RIB 810, 811 & 812 Imaging Physics 8/19/2019


Air Gap Technique
 Problems:

 Increased OID = increase

in blur
 Must increase SID

 Motion due to lack of

contact to IR

31 RIB 810, 811 & 812 Imaging Physics 8/19/2019


Air Gap Applications
 Magnification radiography including mammography

 geometry causes air gap

 Air Gap Chest Radiography

 air gap used as alternative to grid, and

 SID increased from 6 feet to 10 feet to maintain geometric

unsharpness

 Grid not used with air gap

32 RIB 810, 811 & 812 Imaging Physics 8/19/2019


Air Gap Optimization
 Air gap more effective for thicker body parts

 First inch of air gap most effective in contrast improvement

 Image sharpness deteriorates with increasing gap

(magnification)
 compensate with

o greater SID

o smaller focal spot

33 RIB 810, 811 & 812 Imaging Physics 8/19/2019

Вам также может понравиться