Вы находитесь на странице: 1из 21

WATER SYSTEM

a turnkey project by
Kinetics, USA

(room #81)
CA

Process Water
Chlorines Re-circulation PS
NaOCl
ELECTRIC/
SCADA
SVP

Carbon Drinking
and RO/EDI
Sand Water By Christ Small
Filter Tank Autoclave
15 ton
20 Fedegari
154 ton Autoclave
µm
Ground
Tank UV
p Deep Well

Purified Water
SYSTEM QC+Utility
Storage Tank
2
(PW Tank)
NDA-AUSSIE Prep.Tank

DU

LOOP
SYS
1 PSG
CA Process
Process
Steam PS
Cooling

Quality Process Electric-


Steam
Monitor Mechanical
Control
Panel
Distilation Unit HE
6000 lt/hours TUBE
Purified
By Finn Aqua
Water

Mixing 4 Washing
Quality
Monitor
Process WFI Process Ice
Waste Water

Storage Tank & Cooling Mixing 3 Mixing 3-4


(WFI Tank 12 m3) System
By NDA AUSSIE
65 – 70 ºC Mixing 2 Mixing 1-2

HE
TUBE Mixing 1 CIP Tank
CA
Process Process
Steam Cooling
PS

Quality Process
Monitor Steam System

Distillation
Unit HE
Purified 6000 lt/hours Tube
Water By Finn Aqua

Future
Quality
Monitor
Waste Water

Process WFI SVP Filling


Washing 30
Storage Tank
(WFI Tank 5 m3)
By NDA-AUSSIE Electric- Mixing 50
SVP Filing
65-70 ºC
Mechanical
Control Room 49
CIP 48
Panel
HE
TUBE

Chiller
System
WFI QUALITY Standard
 The distilled water produced meets and
exceeds the requirements of the United
States Pharmacopoeia (USP) 31/2008,
European Pharmacopoeia and Japan
Pharmacopoeia for WFI.
The distillate quality produced from the Multiple-Effect Water Still is as follows:
Measurement USP 31/2008 Typical Finn-Aqua®
Requirements Measured Values

Conductivity <1.3 µS/cm at 25°C 0.2 - 0.5 µS/cm at


(non-compensated <3.1 µS/cm at 25°C
Temperature) 100°C 1.0 µS/cm at 100°C

Resistivity >0.77 Mohm-cm at 5 - 2 Mohm-cm at


25°C 25°C
>0.33 Mohm-cm at 1.0 Mohm-cm at
Endotoxins 100°C
<0.25 EU/ml 100°C
<0.02 EU/ml
Clean Steam Quality Standard
The pure steam quality produced from the Pure
Steam Generator is as follows:
Measurement USP 31/2008 Typical FINN-AQUA®
Requirements Measured Values

Conductivity <1.3 µS/cm at 0.2 - 0.5 µS/cm at


(non-compensated 25°C 25°C
Temperature) <3.1 µS/cm at 1.0 µS/cm at 100°C
100°C

Resistivity >0.77 Mohm-cm 5 - 2 Mohm-cm at


  at 25°C 25°C
>0.33 Mohm-cm 1.0 Mohm-cm at
at 100°C 100°C
Endotoxins <0.25 EU/ml <0.02 EU/ml
Steris Fin
Aqua Pure
Steam
Generator
PS Capacity
1000 lt/h
Steris
Fin Aqua
Distilator
WFI Capacity
6000 lt/h