Академический Документы
Профессиональный Документы
Культура Документы
Lecture Overview
Introduction on AEC
Purpose of AEC Types of detectors
Introduction
The Automatic Exposure Control (AEC) requires a special
radiographer in producing consistent radiographic images from patient to patient, regardless of size or presence of pathology.
The advantages of this consistency are numerous and
include:
decreased repeat rate; decreased patient exposure; and increased department efficiency.
Purpose of AEC
AEC is a system used to consistently control
radiographic density by terminating the length of exposure based on the amount of radiation reaching the image receptor.
When an AEC device is used to terminate an exposure, the technologist sets the kVp and mA, but the time of the exposure is automatically determined by the machine.
The AEC device differs from a manual timer because the AEC does not stop the exposure until the film has reached an appropriate density.
Air
radiographs because based on sensitometric, specific amount of radiation to the film produce specific density.
If the x-ray exposure is terminated when the
exposure corresponding to optimal density is reached, the resultant radiograph should demonstrate optimal density.
AEC Systems
AEC devices work by the same principle of operation: radiation is transmitted through the patient converted into an electrical signal terminating the radiographic exposure. Two types of AEC have been used: Phototimers Ionization chamber (most common) Regardless of the specific type of AEC system used,
almost all systems use a set of three detectors, arrange in same specific manner.
exposure.
tube and both lie behind the Bucky grid and cassette.
considered as exit-type devices because the detectors are positioned behind the image receptor
electrical energy
The timer is tripped and the radiographic exposure is
the amount and energy of the X-rays reaching the fluorescent screen.
Phototimer AEC
X-RAY TUBE
SCREEN
LIGHT
PM TUBE
ELECTRONS
entrance-type devices because the detector are positioned in front of image receptor.
Compare to phototimers, ion chambers are less
X-RAY TABLE
BUCKY
X-rays penetrate the thin aluminum shell and strike the air molecules.
The electrons are attracted to the positively charged electrode and are carried through an electrical circuit
The timer is tripped and the radiographic exposure is terminated when a sufficient large charge has been exposed
Even though the ionization chamber and the phototimer operate differently, they both have the same function:
convert radiation into an electrical signal which will be used to automatically stop the exposure when the film has reached the proper density
Density setting
Backup timer
Collimation
Patient Positioning
AEC units use detectors to receive radiation passing
through the patient, it is important to properly position the anatomical part of interest over the detector. If the detector receives radiation coming from another anatomical area, the exposure can be either too long or too short, overexposing or underexposing the film. If the radiographer positions the anatomical part correctly, the AEC unit will produce a radiograph exhibiting proper density.
Patient Positioning
If the radiographer positions the anatomical part correctly, the
efficiency because of the increased amount of repeat radiographs that will result if used improperly.
Detector Selection
Proper detector selection must be made for an AEC unit to
produce the desired radiographic density. There are usually 3 detectors that are connected to a complex electrical circuit. Any single detector or a combination of two or three detectors, can be selected for a radiographic exposure. Failure to select the proper detectors will increase patient dosage because the exposure must be repeated to correct the underexposure or overexposure of he first radiograph.
Density Setting
The AEC devices are equipped with density controls allow the radiographer to fine tune the radiographic density that is produce by the unit.
These are generally are in the form of buttons on the
D. SETTING
% DENSITY CHANGE
+4 +3
+100 +75
+2
+1 0
+50
+25 0
-1
-2 -3 -4
-25
-50 -75 -100
Backup Timer
The backup timer is safety device that prevents
excessive exposure to the patient in cases of timer switch failure or radiographer error.
The backup timer must limit the exposure to a
maximum of 600mAs
Collimation
Collimation affects the production of scatter radiation. As collimation increases, the field size decreases and the quantity of
radiation decreases.
scattered radiation.
turn off too soon when scattered is excessive, which result in underexposure of area of interest.
The radiographer should open the collimator to the extent that the part
of being radiographed is image properly, but not so much as cause to stop the AEC device to stop the exposure before the area being image is properly exposed.
Bucky selection
The Bucky must be correctly selected.
Failure to do so results in the patient and the image
receptor being exposed to excessive radiation. The backup time is reached, the exposure terminated. The radiographer should be certain to deactive the AEC system and use manual technique when performing any radiographic study where the image receptor is located outside the Bucky.
set the exposure time. But the radiographer can influence the exposure time according to the mA and kV setting he chooses. If the mA is changed, X-ray quantity also changes, high mA, produces more X-rays so the exposure time will be shorter and vice versa. kV setting also influences exposure time when using AEC. High kV setting, higher energy X-rays are produced, more of the X-rays will also penetrate the part , resulting shorter required exposure time. Vice versa.
AEC TIME
kVp
AEC TIME
easily once an AEC is calibrated to produce specific densities. When calibration is performed, it is done for a particular film-screen speed. The AEC device cannot sense when the radiographer use different film or screen, resulting in too much or too little density.
exposure time possible. Minimum response time usually longer with AEC system. This can be problem with some segment of patient population, such as pediatric patient and uncooperative patients.
Lack of calibration
Failure to maintain regular calibration of the unit and
AEC device result in radiographs that lack consistence, reproducible, and appropriate density.