Webinar Content
Differences between 8-bit and
12-bit Oscilloscopes
12 Bit Application Examples WaveRunner HRO 6 Zi Facts Complete Probing Solution Summary
WaveRunner HRO 6 Zi
5/1/2012
Analog Signal
ADC
INPUT AMPLIFIER TRACK & HOLD ANALOG TO DIGITAL CONVERTER ACQUISITION MEMORY
TDC
TIME TO DIGITAL CONVERTER
AMPLIFIER
ADC
ANALOG TO DIGITAL CONVERTER
ACQUISITION MEMORY
TRIGGER PROCESSOR
WaveRunner HRO 6 Zi
5/1/2012
ADC Resolution
No of Steps
In ppm (Quality)
6 bits
64
15.60%
15600
36.12 dB
8 bits
256
3.90%
3900
48.1 dB
12 bits
4096
0.02%
240
72.24 dB
WaveRunner HRO 6 Zi
5/1/2012
WaveRunner HRO 6 Zi
5/1/2012
WaveRunner HRO 6 Zi
5/1/2012
Highest Resolution 12-bits provides 16 times resolution compared to 8-bits Resolution = The number of available levels
= 2 bits of Resolution ADC Resolution Number of Steps Dynamic Range
8 12
256 4096
48 dB 72 dB
Quantization Error
WaveRunner HRO 6 Zi
5/1/2012
As one can see the bit resolution is no longer sufficient to show the
right half of the table. Respective screen shots of the measurements will prove this.
WaveRunner HRO 6 Zi
5/1/2012
10
Highest Resolution 12-bit allows detection of smaller signal variations The higher number of bits means the lower measurable voltage
Smallest Voltage Step Full Scale 8 bits 12 bits 80 V 312.5 mV 19.5 mV 40 V 156.2 mV 9.76 mV 20 V 78.1 mV 4.88 mV 8V 31.3 mV 1.95 mV 4V 15.6 mV 976 V 1.6 V 6.3 mV 390 V 800 mV 3.1 mV 195 V 400 mV 1.56 mV 97.6 V 160 mV 625 V 39 V 80 mV 313 V 19.5 V 40 mV 156 V 9.76 V 16 mV 62.5 V 3.9 V 8 mV 31.2 V 1.95 V
WaveRunner HRO 6 Zi
5/1/2012
11
WaveRunner HRO 6 Zi
5/1/2012
12
WaveRunner HRO 6 Zi
5/1/2012
13
the oscillations can be displayed with the 8bit instrument, comparing the measurement with that on the 12-bit instrument shows a substantial deviation.
12 Bit Scope
WaveRunner HRO 6 Zi 5/1/2012 14
Comparing Measurements of 8 bit and 12 bit Scopes The WaveStream display shows
the separate levels of the 8-bit instrument and that the last 34 levels are practical identical.
8 Bit Scope
12 Bit Scope
WaveRunner HRO 6 Zi 5/1/2012 15
Multi-tone
The FFT of a two tone signal shows how much more dynamic range
the 12-bit High Resolution Oscilloscope can detect.
8 Bit Scope
12 Bit Scope
WaveRunner HRO 6 Zi
5/1/2012
16
8-bit
12-bit
See how details of some components are completely missing on the 8bit screenshot. Also notice how low the floor is on the 12-bit trace compared to the 8-bit
Company Confidential 17
55 dB Signal to Noise Ratio 3-4 Times Better than Typical 8-bit Oscilloscopes Signal to noise ration (SNR):
The ratio, expressed in dB, of the RMS signal amplitude to the RMS sum
of all other spectral components excluding the distortion and offset error A typical SNR could be 35 dB for an 8 bit scope and 55 dB for a 12 bit Oscilloscope
~ 44 dB SNR
> 55 dB SNR
WaveRunner HRO 6 Zi
5/1/2012
20
Examples:
Revolutions per minute (RPM) Altitude difference measurement Discharge flow-through measurements ...
WaveRunner HRO 6 Zi
5/1/2012
21
8 Bit Scope
12 Bit Scope
WaveRunner HRO 6 Zi
With 8 bit Oscilloscopes the signal is comparably very noisy To get good results, a noise filter (ERES Enhanced Resolution) has to
be used
With ERES it can sometimes be the case that some fast oscillations at
edges are filtered out
WaveRunner HRO 6 Zi
5/1/2012
23
Medical Measurement
Medical measuring intruments must detect signals with low amplitudes
8 bit Scopes have problems to show these signals because the signal has
insufficient dynamic range.
WaveRunner HRO 6 Zi
5/1/2012
24
Medical Measurement
8 Bit Scope
12 Bit Scope
WaveRunner HRO 6 Zi
Medical Measurement
Result:
The EKG Amplitude is very close to the noise level of 8 Bit Oscilloscopes Only with noise filters (ERES) could the signals be detected, but the
signal quality is very poor
WaveRunner HRO 6 Zi
5/1/2012
26
WaveRunner HRO 6 Zi
5/1/2012
27
Dynamic
Spurious
78dB
44dB
89dB
80dB
WaveRunner HRO 6 Zi
5/1/2012
28
12-bit resolution gives higher dynamic range HRO has >55 dB Signal to Noise Ratio (SNR) HRO has a better spurious-free dynamic range
For IQ-Signals EVM (Error Vector Magnitude) is one of the most important
parameters. For better EVM results the most important parameters are:
Dynamic range
Spurious free dynamic range Signal to Noise Ratio
WaveRunner HRO 6 Zi 5/1/2012 29
Power Measurement
There are two ways to make a power measurement
Measuring the current Measuring the voltage The measurement must be done very often in a big range but with high resolution for little changes Long time-measurments are needed (a lot of memory is necessary)
High dv/dt differential mode measuremts with a DA1855A Differential Amplifier and a AP015 Current Probe
5/1/2012 30
WaveRunner HRO 6 Zi
Power Measurement
8 Bit Voltage
Measurement
12 Bit Voltage
Measurement
WaveRunner HRO 6 Zi
5/1/2012
31
Power Measurement
Results:
Important under- or overshoots can only detected with 12 Bits Power calculation (voltage x current) will benefit from the higher resolution.
8-bit voltage x 8-bit current = 256 Points x 256 Points = 65,536 Points 12-bit voltage x 12-bit current = 4096 Points x 4096 Points = 16,777,216 Points
Small ripples can only be evaluated with high resolution With ERES it can be that some fast oscillations at edges are filtered out
WaveRunner HRO 6 Zi
5/1/2012
33
5/1/2012
34
Highest Accuracy
0.5% DC Vertical Gain Accuracy
Deep Memory
256 Mpts/ch
WaveRunner HRO 6 Zi
5/1/2012
36
WaveRunner HRO 6 Zi
5/1/2012
37
LeCroy
WaveRunner HRO 66Zi
Bandwidth Channels Resolution Sample Rate (All Ch) Standard Memory (per Ch.) Max Optional Memory (per ch)
256 Mpts/Ch
All Serial Data, Math, Measurement,
and Analysis Options for the 8-bit WaveRunner 6 Zi Models are also available for both WR HRO Models
WaveRunner HRO 6 Zi
5/1/2012
38
ADP300, ADP305
20 MHz, 200 MHz 1,000 VRMS Common Mode Voltage 1,400 Vpeak Differential Voltage
ZD200
200 MHz Low Cost Perfect for Serial Data/ Automotive
WaveRunner HRO 6 Zi
5/1/2012
39
5/1/2012
Company Confidential
40
5/1/2012
Company Confidential
41
ZD500
500 MHz 8 V (16Vpp) 10V 60 dB 50/60 Hz 40 dB 20 MHz 25 dB 500 MHz 50 kW (CM) 200 k (DM) 1.0 pF (Differential)
ZD1000
1 GHz 8 V (16Vpp) 10V 60 dB 50/60 Hz 40 dB 20 MHz 25 dB @ 1000 MHz 50 kW (CM) 200 k (DM) 1.0 pF (Differential)
ZD1500
1.5 GHz 8 V (16Vpp) 10V 60 dB 50/60 Hz 40 dB 20 MHz 25 dB @ 1500 MHz 50 kW (CM) 200 k (DM) 1.0 pF (Differential)
ZD200 specs and price make it a great fit for automotive customers
looking at differential signals like CAN or FlexRay No longer need to purchase large form factor, high voltage ADP305
Higher bandwidth ZD probes are ideal for higher speed serial data
signals like USB 2.0 ZD500 and ZD1000 replace AP033 and AP034
5/1/2012 Company Confidential 42
Yes 16 Std
Power
Serial Data Analysis Clock Jitter History Mode Measurement Trigger All instance measurements WaveScan Customization Rare event capture Speed Responsiveness
Yes
SDA II JitKit Standard Standard No Limit Std Full capability with XDEV Triggerscan Best in long memory processing Best in responsiveness
43
Decode Trigger
Auto
44
Market Specifications
WaveRunner HRO 6 Zi
General Purpose
Noise
Analysis
Good
Full
Average
Full
Company Confidential
45
High Resolution Want a display but still want high resolution Want to capture wide voltage Range
8-bit Scope
Company Confidential
46
Summary
12-bit resolution (15-bit resolution with ERES) 55 dB Signal to Noise Ratio (SNR)
Signals with a low amplitude can be measured easy
0.5% DC Vertical Gain Accuracy 1 mV vertical Sensitivity @ full bandwidth Up to 400 V offset capability
8 bit scopes only have a range of 10 V
ERES limits the bandwidth and can modify the signal waveform
Better to use 12 bit without ERES than 8 bit with ERES for authentic
signal waveform rendering
WaveRunner HRO 6 Zi
5/1/2012
47
Any Questions?